HYPODERMIC NEEDLE-PRO EDGE SAFETY DEVICE, HYPODERMIC NEEDLE-PRO EDGE SAFETY DEVICE WITH SYRINGE
K071785 · Smiths Medical Asd, Inc. · FMI · Aug 2, 2007 · General Hospital
Device Facts
| Record ID | K071785 |
| Device Name | HYPODERMIC NEEDLE-PRO EDGE SAFETY DEVICE, HYPODERMIC NEEDLE-PRO EDGE SAFETY DEVICE WITH SYRINGE |
| Applicant | Smiths Medical Asd, Inc. |
| Product Code | FMI · General Hospital |
| Decision Date | Aug 2, 2007 |
| Decision | SESE |
| Submission Type | Abbreviated |
| Regulation | 21 CFR 880.5570 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
Hypodermic Needle-Pro® EDGE" Safety Device: This device is intended for injection or aspiration of fluids using a Luer lock or Luer slip syringe. The needle protection device covers the needle after use to help prevent needle sticks. Hypodermic Needle-Pro® EDGE" Safety Device with Syringe: This device is intended for use to inject fluids into or withdraw fluids from the body. The needle protection device covers the needle after use to help prevent needle sticks.
Device Story
Hypodermic needle with integrated safety protection mechanism; used for fluid injection or aspiration. Features one-piece hub and protective sheath with living hinge. Bevel orientation indicated by arrow on sheath. Post-procedure, user performs one-handed technique to press needle into sheath; lugs on hub snap into clips on sheath (bottom snap) to contain needle. Includes retaining hooks for added security. Used in clinical settings by healthcare professionals. Benefits include reduction of accidental needle stick injuries.
Clinical Evidence
Simulated clinical use studies conducted. Results confirmed device effectiveness in shielding needles post-use. No human clinical trials performed; bench testing and simulated use only.
Technological Characteristics
Hinged-style manual protective sheath; Luer lock/slip compatible. Conforms to ISO 594/1:1986, ISO 594-2:1998, ISO 7864:1993, and ISO 7886-1:1993. Sterile, single-use device.
Indications for Use
Indicated for injection or aspiration of fluids in patients requiring hypodermic needle use. No specific age or gender restrictions provided. Contraindications not specified.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
Predicate Devices
- Hypodermic Needle-Pro® EDGE™ Needle Protection Device (K041399)
- Hypodermic Needle-Pro® EDGE™ Needle Protection Device (K063450)
- Portex® Needle-Pro® EDGE™ Safety Device with Syringe (K061194)
- SurGuard2™ Safety Hypodermic Needle (K031453)
Related Devices
- K073188 — HYPODERMIC NEEDLE-PRO EDGE SAFETY DEVICE · Smiths Medical Asd, Inc. · Feb 8, 2008
- K063450 — PORTEX HYPODERMIC NEEDLE- PRO EDGE SAFETY DEVICE, 18GX1, 401810, 19GX1, 401910, 20GX1,402010, 21GX1402110 · Smiths Medical Asd, Inc. · Jan 25, 2007
- K122249 — TERUMO SURGUARD3 SAFETY NEEDLE & HYPODERMIC SYRINGE WITH SAFETY NEEDLE · Terumo (Philippines) Corporation · Aug 28, 2012
- K061194 — PORTEX NEEDLE-PRO EDGE SAFETY WITH SYRINGE, MODELS 432310, 432210, 432510, 432558 · Smiths Medical Asd, Inc. · Aug 24, 2006
- K011925 — HYPODERMIC NEEDLE-PRO NEEDLE WITH NEEDLE PROTECTION DEVICE; HYPODERMIC NEEDLE-PRO SYRINGE & NEEDLE WITH NEEDLE PROTECTIO · Sims Portex, Inc. · Jul 12, 2001
Submission Summary (Full Text)
{0}------------------------------------------------
# smiths
## SECTION 5, 510(k) Summary
### Company Information:
Smiths Medical ASD, Inc. 10 Bowman Drive Keene, NH 03431 (603) 352-3812, prompt 4, ext 2493 Contact: Brian D. Farias Regulatory Affairs Manager
### Smiths Medical ASD, Inc.
Anesthesia and Safety Devices Division
10 Bowman Drive Keene NH 03431 USA Tel: +1 603 352 3812 Fax: +1 603 352 3703 www.smiths-medical.com
AUG - 2 2007
Summary Prepared: June 29, 2007
### Product Name:
# Trade Namc: Hypodermic Needle-Pro® EDGE" Safety Device Hypodermic Needle-Pro® EDGE" Safety Device with Syringe
Common Name: Hypodernic Needle with attached needle protection and Syringe with attached needle and needle protection
Classification Name: Hypodermic Single Lumen Needles (21 CFR 880.5570, Product Code FMI) and Piston Syringe (21 CFR 880.5860, Product Code FMF)
### Predicate Device(s):
K041399 and K063450 (Smiths Medical ASD, Inc.) Hypodermic Needlc-Pro® EDGE" Needle Protection Device
K061194 (Smiths Medical ASD, Inc.) Portex® Needle-Pro® EDGE™ Safety Device with Syringe
K031453 (Terumo Medical Corporation) SurGuard2™ Safety Hypodermic Needle
### Device Description:
This device is intended for injection or aspiration of fluids into the body. The needle protection device covers the needle after use to help prevent needle sticks. The device features a "onepiece" design of needle hub and protective sheath with a living hinge. The needle cannula is permanently affixed into the hub. The sheath has an "arrow" indicating the bevel orientation, i.e. when the sheath is oriented to the right, the bevel is in the "up position". After the procedure is completed, the needle is pressed into the sheath using a one-handed technique. As the needle enters the protective sheath, the lugs on the needle hub snap into the clips of the protective sheath resulting in a "bottom snap" and the needle is contained within the sheath. In addition to the "bottom snap", the needle protection sheath for needles equal to or less than !" has one needle retaining hook (except 30g) and the sheath for needles longer than 1" has two hooks. The device is then discarded into a sharps container.
{1}------------------------------------------------
### Indications for Use:
Hypodermic Needle-Pro® EDGE" Safety Device: This device is intended for injection or aspiration of fluids using a Luer lock or Luer slip syringe. The needle protection device covers the needle after use to help prevent needle sticks.
Hypodermic Needle-Pro® EDGE" Safety Device with Syringe: This device is intended for use to inject fluids into or withdraw fluids from the body. The needle protection device covers the needle after use to help prevent needle sticks.
### Technological Characteristics:
The proposed and predicate devices all employ the same hinged style protective sheath that is manually activated after use.
### Non-Clinical Data:
This abbreviated 510(k) submission declares conformance to the following standards:
ISO 594/1:1986(E), International Standard, Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment-Part 1: General requirements
ISO 594-2:1998(E), International Standard, Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment-Part 2: Lock Fittings
ISO 7864:1993(E), International Standard, Sterile hypodermic needles for single use
ISO 7886-1:1993, International Standard, Sterile hypodermic syringes for single use
### Clinical Data:
Simulated clinical use studies were conducted which confirmed that the device could be used effectively, with the needle shielded inside the protection device after use.
### Conclusion:
The standards compliance and simulated clinical use studies demonstrate that the proposed device is safe and effective and is substantially equivalent to the predicate devices,
Very truly yours,
SMITHS MEDICAL ASD, INC.
rian D. Farias Regulatory Affairs Manager
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, rendered in a simple, bold line drawing.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Brian D. Farias Regulatory Affairs Manager Smiths Medical ASD, Incorporated 10 Bowman Drive Keene, New Hampshire 03431
AUG - 2 2007
Re: K071785
Trade/Device Name: Hypodermic Needle-Pro® EDGE™ Safety Device Regulation Number: 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: FMI Dated: June 29, 2007 Received: July 3, 2007
Dear Mr. Farias:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 -- Mr. Farias
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Shinn-Linn, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# SECTION 4, Indications for Use Statement
# Indications for Use
510(k) Number (if known):
Device Name: Hypodermic Needle-Pro® EDGE™ Safety Device
Indications for Use:
This device is intended for injection or aspiration of fluids using a Luer lock or Luer slip syringe. The needle protection device covers the needle after use to help prevent needle sticks.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
---
( Javision Sign-Off) Division of Anesthesiology, General Hospital, Intection Control, Dental Devices