← Product Code [GDM](/submissions/SU/subpart-e%E2%80%94surgical-devices/GDM) · K071732

# MARROWMINER (K071732)

_Stemcor Systems, Inc. · GDM · Sep 24, 2007 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GDM/K071732

## Device Facts

- **Applicant:** Stemcor Systems, Inc.
- **Product Code:** [GDM](/submissions/SU/subpart-e%E2%80%94surgical-devices/GDM.md)
- **Decision Date:** Sep 24, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4800
- **Device Class:** Class 1
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

For harvesting bone marrow.

## Device Story

MarrowMiner is a battery-powered bone marrow harvesting system. Components include a motorized handle, aspiration chamber, flexible aspiration shaft, and access guide. The device operates by rotating the tip of the flexible shaft to aspirate bone marrow under controlled vacuum through a single puncture site. Used by clinicians to collect bone marrow samples while maintaining iliac crest integrity. The system allows for the collection of the desired volume of marrow through a minimally invasive approach.

## Clinical Evidence

Animal studies were conducted to evaluate the device's ability to obtain adequate bone marrow samples without compromising the integrity of the iliac crest.

## Technological Characteristics

Battery-powered motorized handle; flexible aspiration shaft; aspiration chamber; access guide. Operates via mechanical rotation and controlled vacuum aspiration. Single-puncture access.

## Regulatory Identification

A manual surgical instrument for general use is a nonpowered, hand-held, or hand manipulated device, either reusable or disposable, intended to be used in various general surgical procedures. The device includes the applicator, clip applier, biopsy brush, manual dermabrasion brush, scrub brush, cannula, ligature carrier, chisel, clamp, contractor, curette, cutter, dissector, elevator, skin graft expander, file, forceps, gouge, instrument guide, needle guide, hammer, hemostat, amputation hook, ligature passing and knot-tying instrument, knife, mallet, disposable or reusable aspiration and injection needle, disposable or reusable suturing needle, osteotome, pliers, rasp, retainer, retractor, saw, scalpel blade, scalpel handle, one-piece scalpel, snare, spatula, stapler, disposable or reusable stripper, stylet, suturing apparatus for the stomach and intestine, measuring tape, and calipers. A surgical instrument that has specialized uses in a specific medical specialty is classified in separate regulations in parts 868 through 892 of this chapter.

## Predicate Devices

- Biomedical Eng, Inc. BoneHog/Excaliber ([K961902](/device/K961902.md), [K964073](/device/K964073.md), [K971783](/device/K971783.md))
- Medical Device Technologies, Inc. InterV SnareLok Bone Marrow Biopsy Needle ([K043523](/device/K043523.md))
- Vidacare Bone Marrow Aspiration Set ([K062833](/device/K062833.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## PREMARKET NOTIFICATION [510(K)] SUMMARY

| June, 2007                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | K071732     |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Trade Name:                       | StemCor Systems, Inc. MarrowMiner                                                                                                                                                                                                                                                                                                                                                                                                                                                     |             |
| Common Name:                      | Biopsy instrument                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |             |
| Classification Name:              | Instrument, biopsy (per 21 CFR section 876.1075)                                                                                                                                                                                                                                                                                                                                                                                                                                      |             |
| Manufacturer's Name:              | StemCor Systems, Inc.<br>801 Hermosa Way<br>Menlo Park, CA 94025                                                                                                                                                                                                                                                                                                                                                                                                                      | SEP 24 2007 |
| Corresponding Official:           | Sharon Rockwell<br>Consultant<br>5582 Chalon Road<br>Yorba Linda, CA 92886<br>Phone: (714) 695-9269<br>Fax: (714) 779-1239                                                                                                                                                                                                                                                                                                                                                            |             |
| Predicate Device(s):              | Biomedical Eng, Inc. BoneHog/Excaliber, K961902, K964073, K971783.<br>Medical Device Technologies, Inc. InterV SnareLok Bone Marrow Biopsy Needle, K043523.<br>Vidacare Bone Marrow Aspiration Set, K062833                                                                                                                                                                                                                                                                           |             |
| Device Description:               | The MarrowMiner is a device intended for harvesting bone marrow. The StemCor MarrowMiner bone marrow collection system is comprised of four components; a MarrowMiner battery powered handle, an aspiration chamber, a flexible aspiration shaft, and an access guide. The tip of the flexible aspiration shaft rotates and aspirates bone marrow. The device has been shown in animal studies to obtain adequate bone marrow samples without compromising the iliac crest integrity. |             |
| Intended Use:                     | For harvesting bone marrow.                                                                                                                                                                                                                                                                                                                                                                                                                                                           |             |
| Technological<br>Characteristics: | Aspiration module and flexible shaft allow the marrow to be collected under controlled vacuum through a single puncture until the desired volume is collected.                                                                                                                                                                                                                                                                                                                        |             |

:

.

:

:

:

:

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three abstract human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

SEP 2 4 2007

Stemcor Systems, Inc. % Rockwell & Associates Ms. Sharon Rockwell Consultant, Regulatory Affairs 5582 Chalon Road Yorba Linda, California 92886

Re: K071732

Trade/Device Name: StemCor Systems, Inc. MarrowMiner Regulation Number: 21 CFR 878.4820 Regulation Name: Surgical instrument motors and accessories/attachments Regulatory Class: I Product Code: GDM, GAA Dated: August 30, 2007 Received: August 31, 2007

Dear Ms. Rockwell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA mav publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{2}------------------------------------------------

Page 2 - Ms. Sharon Rockwell

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation ventitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2141 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html .

Sincerely vours.

Par Vdo Rmo De
Mark N. Melkerson 241.03

Mark N. Melkerson Director

Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## INDICATIONS FOR USE STATEMENT

510(k) Number (if known): _ KO7l732

Device Name: StemCor Systems, Inc. MarrowMiner

Indications for Use:

The MarrowMiner is intended for harvesting bone marrow.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Mark A. Milliman

(Division Sign-Off) (Division of General, Restorative, and Neurological Devices

**510(k) Number** KO

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GDM/K071732](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GDM/K071732)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
