K071708 · Power Medical Interventions, Inc. · GDW · Sep 27, 2007 · General, Plastic Surgery
Device Facts
Record ID
K071708
Device Name
I45, I45S, I60, AND I60S
Applicant
Power Medical Interventions, Inc.
Product Code
GDW · General, Plastic Surgery
Decision Date
Sep 27, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The i45, i45s, i60, i60s have applications for general and endoscopic surgery in gastrointestinal, gynecological, general abdominal, thoracic surgical procedures for resection, transection, creation of anastomoses, and for open occlusion of the heart's left atrial appendage.
Device Story
Reusable, battery-powered linear stapler; articulates left and right; used for resection, transection, and anastomosis. Input: manual control via buttons/rocker switches on proximal housing. Transformation: 14.8Vdc battery powers two DC motors; rotational power transmitted through rigid shaft to stapling/cutting mechanism. Output: 4 rows of staples with central transection. Used in OR by surgeons for general/endoscopic procedures. Untethered design replaces previous console-tethered systems. Benefits: provides motorized stapling/cutting without external power console, improving surgical maneuverability.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on technological comparison and bench testing of the power, transmission, and control systems.
Technological Characteristics
Reusable, steam-sterilizable linear stapler. Powered by 14.8Vdc battery. Features integrated DC motors for stapling, cutting, and articulation. Rigid shaft transmission. Control via buttons/rocker switches on proximal housing. Patient contact materials adhere to industry standards.
Indications for Use
Indicated for patients undergoing general or endoscopic surgery in gastrointestinal, gynecological, general abdominal, or thoracic procedures requiring resection, transection, or anastomosis, and for open occlusion of the heart's left atrial appendage.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
Predicate Devices
SurgASSIST® System with Circular Stapler Digital Loading Units® (K003277)
SurgASSIST® Computer Mediated Linear Cutter Digital Loading Units® (K040720)
Submission Summary (Full Text)
{0}------------------------------------------------
K071708
Page ① of ②
## Section E
SEP 2 7 2007
## Traditional 510(k) - Summary
In accordance with 21 CFR Section 807.92, Power Medical Interventions, Inc., is submitting the following 510(k) Summary:
- 1) Submitter Information:
Power Medical Interventions, Inc. 2021 Cabot Blvd. West Langhorne, PA 19047 USA Ph: 267-775-8151 Fax: 267-775-8123
Applicant: Barbara J. Whitman
Date of Notification: June 19, 2007
- 2) Name of Device:
| Trade Name: | i45, i45S, i60, i60S |
|-----------------|------------------------------------------|
| Common Name: | Linear Staplers with Implantable Staples |
| Classification: | Staple, Implantable, GDW |
- 3) Predicate Devices:
- SurgASSIST® System with Circular Stapler Digital Loading a) Units® Power Medical Interventions, Inc., Langhorne, PA (K003277). REF PC100; RCW100; FS14, CS21, CS25, CS29, CS33.
- FlexShaft II, Power Medicai Interventions, Inc., Langhorne, PA. b) (K021249). REF FS214
- SurgASSIST® Computer Mediated Linear Cutter Digital Loading C) Units®. Power Medical Interventions, Inc., Langhorne, PA. (K040720) REF CMLC30, CMLC30, CMLC55, CMLC75.
- Device Description: 4)
The devices described here are reusable i45, i45S, i60, i60S used in conjunction with a variety of Reloads, which were previously cleared to market under K052415. The i45, i45S, i60, i60S are designed to be cleaned and sterilized for multi-patient use.
00015
. ..
{1}------------------------------------------------
K071708
Page ② of ②
Power Medical Interventions, Inc 145, 145s, 160, 160s 45, 1453, 100, 1000
Traditional 510(k) Premarket Notification, June 19, 2007
## Device Modification: 5)
The i45, i45S, i60, i60S have identical technological features as the predicate devices (K003277, K021249, and K040720). The i45, i45S, i60, ions is an integrated stapler that articulates left and right. In addition to motors for stapling and cutting, and articulation, the proximal housing holds the battery and electronics enabling untethered use. Device control is managed with buttons and rocker switches on the proximal housing. The device will be steam sterilized. All product patient contact materials adhere to industry standards.
The predicate devices for the power, transmission, and control aspects are the PC100, previously cleared to market via K003277 and the FS214, previously cleared to market via K021249. In the predicate device, AC power was converted to 24Vdc driving DC motors. In the subject device, a battery rated at 14.8Vdc powers two DC motors directly. In the a battery ransmission of rotational power was accomplished production of a FlexShaft. In the subject device, the rotational power is transmitted through a rigid shaft. In the predicate device, control for articulation and stapling was achieved through a Remote Control Unit plugged into the Power Console. In the subject device, control is integrated into the handpiece.
The predicate device for the stapling and cutting aspects is the Computer Mediated Linear Cutter, previously cleared to market via K040720. Both the predicate and the subject devices provided 4 rows of staples with a transection in the middle. Both are steam sterilizable, multi-patient use devices.
- Indications For Use (identical to those in the predicate CMLC DLU, 6) K040720) -
The i45, i45S, i60, i60S have applications for general and endoscopic surgery in gastrointestinal, gynecological, general abdominal, thoracic surgical procedures for resection, transection, creation of anastomoses, and for open occlusion of the heart's left atrial appendage.
7) Comparison to Predicate Devices
> The i45, i45S, i60, i60S have the same function as the previously cleared predicate devices. The subject device provides stapling and cutting, similar to that of the CMLC, but is no longer tethered to the PC100.
The Indications For Use statement is identical to that which appears in the CMLC predicate device (K040720).
00016
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Inside the circle is an abstract symbol that resembles a stylized caduceus or a representation of human figures. The symbol is composed of curved lines and shapes, giving it a flowing and dynamic appearance.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Power Medical Interventions, Inc. % Peter Shearstone Sr. VP. Regulatory/Quality Assurance 2021 Cabot Boulevard West Langhorne, Pennsylvania 19047
SEP 2 7 2007
Re: K071708
Trade/Device Name: i45/i45s/i60/i60s Regulation Number: 21 CFR 878.4750 Regulation Name: Linear Stapler with Implantable Staples Regulatory Class: II Product Code: GDW Dated: September 13, 2007 Received: September 13, 2007
Dear Mr. Shearstone:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable. the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Mr. Shearstone
marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N. Melkerson 9/26/22
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Power Medical Interventions. Inc 145. 145s. 160. 160s Traditional 510(k) Premarket Notification, June 19, 2007
## Section D
# Indications for Use
510(k) Number (if known): *K071708*
Device Name: *i45, i45S, i60, i60S*
Indications for Use:
The i45, i45s, i60, i60s have applications for general and endoscopic
surgery in gastrointestinal, gynecological, general abdominal, thoracic
surgical procedures for resection, transection, creation of anastomoses,
and for open occlusion of the heart's left atrial appendage.
(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices
510(k) Number *K071708*
| <strong>Labels</strong> | <strong>Values</strong> |
|----------------------------------------------|-------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | X |
| Over-The-Counter Use (21 CFR 801 Subpart C) | |
AND/OR
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER
PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Concurrence of CDRH, Office of Device Evaluation (ODE)
**00014**
Page | of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.