K071555 · Organogenesis, Inc. · NPL · Aug 30, 2007 · Dental
Device Facts
Record ID
K071555
Device Name
FORTAGEN ORAL MEMBRANE
Applicant
Organogenesis, Inc.
Product Code
NPL · Dental
Decision Date
Aug 30, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3930
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
FortaGen Oral Membrane is intended to be used during guided bone regeneration (GBR) and guided tissue regeneration (GTR) procedures as a biodegradable barrier membrane for: - Ridge augmentation for later implant insertions; - Simultaneous ridge augmentation and implant insertions; - Ridge augmentation around implants inserted in delayed extraction sites; - Ridge augmentation around implants inserted in immediate extraction sites; - Alveolar ridge preservation consequent to tooth (teeth) extraction(s); - Over the window in lateral window sinus elevation procedures; - In implants with vertical bone loss due to infection, only in cases where satisfactory debridement and implant surface disinfection can be achieved; - In intra bony defects around teeth; - For treatment of recession defects, together with coronally positioned flap; - In furcation defects in multi-rooted teeth.
Device Story
FortaGen Oral Membrane is a multi-laminate, resorbable barrier membrane composed primarily of Type I porcine collagen. Supplied in hydrated sheet form, it is used by dental clinicians during guided bone regeneration (GBR) and guided tissue regeneration (GTR) procedures. The membrane acts as a physical barrier to prevent soft tissue ingrowth into bone defects, facilitating bone healing and regeneration. It is designed to resorb within approximately six months post-implantation. The device is intended for use in various clinical scenarios, including ridge augmentation, sinus elevation, and treatment of periodontal defects. By maintaining space and excluding non-osteogenic cells, the membrane supports the development of new bone and tissue, potentially improving outcomes for dental implant placement and periodontal health.
Clinical Evidence
Bench testing only. Biocompatibility and resorption profile were evaluated to demonstrate equivalence to predicate devices.
Technological Characteristics
Multi-laminate sheet composed of Type I porcine collagen. Resorbable barrier membrane. Supplied hydrated in sterile peelable packaging. Dimensions range from 13 mm x 25 mm to 40 mm x 50 mm. Designed for resorption within six months.
Indications for Use
Indicated for patients undergoing guided bone or tissue regeneration procedures, including ridge augmentation, alveolar ridge preservation, sinus elevation, and treatment of periodontal defects (recession, furcation, intra-bony) around teeth or implants.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows a handwritten string of characters, which appears to be a code or identifier. The string reads 'K0715555'. The characters are written in a clear, legible style, with distinct strokes that make each character easily identifiable. A horizontal line is present beneath the string of characters.
Image /page/0/Picture/3 description: The image shows a handwritten number that appears to be "182". The numbers are written in a cursive style, with the "1" and "8" clearly visible. There is a line underneath the number, possibly indicating it is part of a calculation or a label.
# 5. 510(k) Summary
#### Submitted by:
Organogenesis, Inc. 150 Dan Road Canton, MA 02021 AUG 3 0 2007
## Contact:
Patrick R. Bilbo Telephone: 781-401-1155 Facsimile: 781-401-1288
Date: June 6, 2007
### Device:
Trade Name: Common/Usual Name: Classification Name: Regulatory Class:
FortaGen® Oral Membrane Resorbable dental barrier membrane, animal source Not applicable Not classified
## Predicate Devices:
The device is similar to predicate collagen-based surgical mesh devices and collagen dental membranes previously cleared for commercial distribution. The relevant predicate devices include FortaGen Surgical Mesh (K021105), formerly FortaFlex Surgical Mesh, manufactured by Organogenesis Inc., Ossix-Plus (K053260) manufactured by Colbar LifeScience, Ltd., and Bio-Gide Resorbable Bilayer Membrane (K050446) manufactured by Ed. Geistlich Soehne Ag für Chemische Industrie.
### Intended Use:
FortaGen Oral Membrane is intended to be used during guided bone regeneration (GBR) and guided tissue regeneration (GTR) procedures as a biodegradable barrier membrane for:
- Ridge augmentation for later implant insertions;
- . Simultaneous ridge augmentation and implant insertions;
- . Ridge augmentation around implants inserted in delayed extraction sites;
- . Ridge augmentation around implants inserted in immediate extraction sites;
- . Alveolar ridge preservation consequent to tooth (teeth) extraction(s);
- Over the window in lateral window sinus elevation procedures; .
- . In implants with vertical bone loss due to infection, only in cases where satisfactory debridement and implant surface disinfection can be achieved;
- . In intra bony defects around teeth;
# Confidential
{1}------------------------------------------------
\$\frac{K07/555}{307^{2}}\$
- . For treatment of recession defects, together with coronally positioned flap;
- . In furcation defects in multi-rooted teeth.
# Device Description:
FortaGen Oral Membrane consists of a multi-laminate sheet predominantly of Type I porcine collagen. The device is supplied hydrated in sheet form in sizes ranging from 13 mm x 25 mm to 40 mm x 50 mm in sterile double layer peelable packaging.
# Technological Characteristics and Performance Data:
The FortaGen Oral Membrane, like its predicate devices, is manufactured from porcine collagen and is designed to resorb within six months. FortaGen Oral Membrane has been shown to be biocompatible. Its handling characteristics and resorption profile are substantially equivalent to those of its predicate devices, and the FortaGen Oral Membrane should function equivalently to Ossix-Plus and Bio-Gide when used as intended for GTR and GBR procedures.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/2/Picture/2 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized eagle with outstretched wings, symbolizing protection and care. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" are arranged in a circular pattern around the eagle, indicating the department's name and national affiliation. The seal is presented in black and white, giving it a formal and official appearance.
MAY 27 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Patrick R. Bilbo Vice President, Regulatory and Clinical Affairs Organogenesis, Incorporated 150 Dan Road Canton, Massachusetts 02021
Re: K071555
Trade/Device Name: FortaGen Oral Membrane Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: II Product Code: NPL Dated: August 17, 2007 Received: August 20, 2007
Dear Mr. Bilbo:
This letter corrects our substantially equivalent letter of August 30, 2007.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Bilbo
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Suite Y. Michie Davis.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
107/
| K071555 |
|---------|
|---------|
Indications for Use
510(k) Number (if known): _
Device Name: FortaGen Oral Membrane
Indications for Use:
FortaGen Oral Membrane is intended to be used during guided bone and guided tissue regeneration procedures as a biodegradable barrier membrane for:
- Ridge augmentation for later implant insertions; ●
- Simultaneous ridge augmentation and implant insertions; .
- . Ridge augmentation around implants inserted in delayed extraction sites;
- Ridge augmentation around implants inserted in immediate extraction sites; .
- . Alveolar ridge preservation consequent to tooth (teeth) extraction(s);
- Over the window in lateral window sinus elevation procedures;
- . In implants with vertical bone loss due to infection, only in cases where satisfactory debridement and implant surface disinfection can be achieved;
- In intra bony defects around teeth;
- . For treatment of recession defects, together with coronally positioned flap;
- In furcation defects in multi-rooted teeth. .
Prescription Use __ X Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susy Runoer
(Division Sign-Jfl) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page of
510(k) Number: K071555
Confidential
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.