UNI-VENT 731 SERIES VENTILATORS, MODEL 731, 731EMV
K071526 · Impact Instrumentation, Inc. · CBK · Dec 10, 2008 · Anesthesiology
Device Facts
Record ID
K071526
Device Name
UNI-VENT 731 SERIES VENTILATORS, MODEL 731, 731EMV
Applicant
Impact Instrumentation, Inc.
Product Code
CBK · Anesthesiology
Decision Date
Dec 10, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Model 731EMV (EMV) is indicated for use in the management of adolescent and adult patients weighing ≥30 kg with acute or chronic respiratory failure or during resuscitation by providing continuous positive-pressure ventilation. It is appropriate for use in hospitals, outside the hospital, during transport, and in austere environments where it may be exposed to rain, dust, rough handling and extremes in temperature and humidity. With an appropriate third-party filter in place, it may be operated in environments where chemical and/or biological toxins are present (see External Filter Use). It is not intended to operate in explosive environments. The EMV is intended for use by skilled care providers with knowledge of mechanical ventilation, emergency medical services (EMS) personnel with a basic knowledge of mechanical ventilation and by first responders under the direction of skilled medical care providers.
Device Story
Portable, microprocessor-controlled ventilator; provides positive-pressure ventilation using oxygen source or ambient air via internal compressor. Inputs: gas source, power (AC/DC/battery), and noninvasive pulse oximetry sensors for oxygen saturation/pulse rate. Operation: point-turn-and-click interface for ventilation settings/monitoring; includes alarm system with context-sensitive help. Used in hospitals, transport, and austere environments by skilled clinicians, EMS, or first responders. Output: continuous ventilation and physiological monitoring data. Benefits: provides respiratory support in diverse, harsh environments; protects against environmental hazards via impact-resistant polycarbonate case and specialized filtration.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Portable, microprocessor-controlled ventilator; impact-resistant polycarbonate case. Power: AC, external DC, or internal battery. Gas source: 55 psig oxygen or ambient air via internal compressor. Connectivity: noninvasive pulse oximetry sensors. Filtration: particulate, bacterial/viral, or chemical/biologic filters. Interface: point-turn-and-click.
Indications for Use
Indicated for adolescent and adult patients weighing ≥30 kg with acute or chronic respiratory failure or requiring resuscitation. Contraindicated for use in explosive environments.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
Predicate Devices
Impact Uni-Vent- Eagle, Model 754 (K870861/B and K931473)
{0}------------------------------------------------
IMPACT Instrumentation, Inc.
27 Fairfield Place, West Caldwell, NJ 07006 P.O. Box 508, West Caldwell, NJ 07007-0508
Image /page/0/Picture/2 description: The image shows the word "IMPACT" in large, bold letters. The word is positioned in the center of the image, with a world map outline above and below it. There is also a heart rate line to the left of the word, and some handwritten text in the upper right corner.
DEC10 2008
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS Impact, Uni-Vent® Model 731EMV 510(k) Number K071526
Manufacturer:
Impact Instrumentation, Inc. P.O. Box 508/27 Fairfield Place West Caldwell, New Jersey 07006 Phone: 973-882-1212 Fax: 973-882-4993
Contact Person:
December 10, 2008 Date Summary Prepared:
Uni-Vent 10 Model 731EMV Trade Name:
Continuous Ventilator (per 21 CFR 868.5895) Classification Name:
Mr. Leslie H. Sherman
Classification: Class II
Product Code: CBK, DQA
Device Description:
The Uni-Vent @ Model 731EMV is a portable, microprocessor controlled, electrically or pneumatically powered intensive care ventilator designed to use either oxygen (02) from a 55 psig source or ambient air using an internal compressor power to deliver a positive pressure breaths. The unit can be electrically powered from an external alternating current source, external direct current (DC) source or the internal DC battery. An intuitive point-turn-and click interface allows the operator to set and monitor ventilation in all operating environments. A series of alarms alert the user operator to all conditions that affect the ventilator's operation and/or performance and provide context sensitive help relevant to the alarm condition. Ambient air is filtered using a particulate filter or when the operating environment requires either a bacterial/viral or chemical/biologic (NATO No: 4240-01-361-1319) filter. The unit is contained in an impact resistant polycarbonate case which protects of the controls from damage and inadvertent manipulation.
Page 1 of 2
electro-medica
{1}------------------------------------------------
K071526
p. 2 of 2
Description of Noninvasive Pulse Oximeter
The Uni-Vent @ Model 731EMV internal pulse oximeter connects to the patient using noninvasive sensors to monitor oxygen saturation and pulse rate. Pulse oximeter specific alarms and instructions are presented to the operator through the user interface. Isolated DC power is provided to the pulse oximeter.
Intended Use:
The Model 731EMV (EMV) is indicated for use in the management of adolescent and adult patients weighing ≥30 kg with acute or chronic respiratory failure or during resuscitation by providing continuous positivepressure ventilation. It is appropriate for use in hospitals, outside the hospital, during transport, and in austere environments where it may be exposed to rain, dust, rough handling and extremes in temperature and humidity. With an appropriate third-party filter in place, it may be operated in environments where chemical and/or biological toxins are present (see External Filter Use). It is not intended to operate in explosive environments. The EMV is intended for use by skilled care providers with knowledge of mechanical ventilation, emergency medical services (EMS) personnel with a basic knowledge of mechanical ventilation and by first responders under the direction of skilled medical care providers.
Substantial Equivalence:
The Impact, Model 731EMV, is substantially equivalent to the predicate devices listed below:
Predicate Devices:
- 1. Impact Uni-Vent- Eagle, Model 754. 510(k) #K870861/B and K931473.
- 2. Impact Uni-Vent Model 730. 510(k) #K032386
- 3. Impact Uni-Vent Model 73X. 510(k) #K051476.
- 4. Masimo SET RAD 5 MS-11 PCB 510(k) #K033296, Masimo Corp, Irvine, CA
- 5. Versamed SmartVent™ 201, 510(k) #K061627, Versamed Medical Systems, Inc., Pearl River, NY
- 6. Vela Bird Products, Palm Springs, CA, 510(k) # K032451
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular pattern around the eagle. The eagle is black, and the text is also black. The logo is simple and recognizable.
## Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 1 0 2008
Mr. Leslie H. Sherman President PACT Instrumentation, Incorporated 27 Fairfield Place West Caldwell, New Jersey 07006
Re: K071526
Trade/Device Name: Uni-Vent ® Model 731 EMV Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: CBK, DQA Dated: December 3, 2008 Received: December 3, 2008
Dear Mr. Sherman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Sherman
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Sybil y. Mckain PhD.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## INDICATIONS FOR USE
510(k) Number (if known): K071526 Device Name: Uni-Vent @ Model 731EMV
## Indications for Use:
The Model 731EMV (EMV) is indicated for use in the management of adolescent and adult patients weighing ≥30 kg with acute or chronic respiratory failure or during resuscitation by providing continuous positive-pressure ventilation. It is appropriate for use in hospitals, outside the hospital, during transport, and in austere environments where it may be exposed to rain, dust, rough handling and extremes in temperature and humidity. With an appropriate third-party filter in place, it may be operated in environments where chemical and/or biological toxins are present (see External Filter Use). It is not intended to operate in explosive environments. The EMV is intended for use by skilled care providers with knowledge of mechanical ventilation, emergency medical services (EMS) personnel with a basic knowledge of mechanical ventilation and by first responders under the direction of skilled medical care providers.
X Prescription Use (Per 21 CFR 801 Subpart D) (Per 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Suage y. Michie Ous
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number:_大07 15 こん
Page 1 of 1
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.