PSS SYSTEM

K071438 · Spinal USA · MNI · Aug 21, 2007 · Orthopedic

Device Facts

Record IDK071438
Device NamePSS SYSTEM
ApplicantSpinal USA
Product CodeMNI · Orthopedic
Decision DateAug 21, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3070
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The PSS System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudoarthrosis). The PSS System is also intended for non-cervical pedicle screw fixation for the following indications: severe spondylolisthesis (grades 3 and 4 of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion. It is also intended for the following indications: trauma (i.e. fracture or dislocation); spinal stenosis; curvatures (i.e. scoliosis; and/or lordosis); spinal tumor; pseudoarthrosis; and failed previous fusion.

Device Story

PSS System is a top-loading, posterior spinal fixation system; components include pedicle screws, rods, and locking cap screws. Used by surgeons in clinical settings to provide immobilization and stabilization of spinal segments as an adjunct to fusion. Components are available in various sizes to match patient anatomy. System is supplied non-sterile. Device provides mechanical support to the spine to facilitate fusion; benefits include stabilization of acute/chronic instabilities or deformities.

Clinical Evidence

Bench testing only. Testing performed in accordance with ASTM F1717.

Technological Characteristics

Top-loading posterior spinal fixation system. Components: pedicle screws, rods, locking cap screws. Material: medical grade titanium alloy (ASTM F136 or ISO 5832-3). Supplied non-sterile.

Indications for Use

Indicated for skeletally mature patients requiring spinal immobilization/stabilization as an adjunct to fusion for thoracic, lumbar, and sacral spine instabilities or deformities including degenerative spondylolisthesis, fracture, dislocation, scoliosis, kyphosis, spinal tumor, pseudoarthrosis, failed fusion, and spinal stenosis. Severe spondylolisthesis (grades 3-4, L5-S1) requires autogenous bone graft and L3-sacrum fixation.

Regulatory Classification

Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ #### 1438 510 (k) Summary of Safety and Effectiveness | Date Summary Prepared: | August 17, 2007 | | |--------------------------------------------|-----------------------------------------------------------------------------------------------|-------------| | Submitter Information: | Spinal USA<br>644 Lakeland East Drive Suite A<br>Flowood, MS 39232 | AUG 21 2007 | | Contact Name:<br>Phone:<br>Fax:<br>E-mail: | Jeffrey Johnson<br>601-420-4244<br>601-420-5501<br>jeff@spinalusa.com | | | Device Trade Name: | PSS System | | | Common Name: | Pedicle Screw Spinal System<br>MNH (Class II) 888.3070(b)(1)<br>MNI (Class II) 888.3070(b)(1) | | ## INTENDED USE: The PSS System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudoarthrosis). The PSS System is also intended for non-cervical pedicle screw fixation for the following indications: severe spondylolisthesis (grades 3 and 4 of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion. It is also intended for the following indications: trauma (i.e. fracture or dislocation); spinal stenosis; curvatures (i.e. scoliosis; and/or lordosis); spinal tumor; pseudoarthrosis; and failed previous fusion. ## DEVICE DESCRIPTION: The PSS System is a top-loading, multiple component, posterior spinal fixation system which consists of pedicle screws, rods, locking cap screws. All of the components are available in a variety of sizes to match more closely the patient's anatomy. All components are made from medical grade titanium alloy described by such standards as ASTM F136 or ISO5832-3. The products are supplied clean and ''NON-STERILE". ## EQUIVALENT DEVICE: Testing in accordance with ASTM F171 7 "Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model" of the PSS System demonstrates that the device is substantially equivalent to the Moss Miami Pedicle Fixation System (K980447), Optima Spine System (K031585), and the Xia Spinal System (K984251). {2}------------------------------------------------ Image /page/1/Picture/1 description: The image shows a partial view of a document with a logo and some text. On the left side, there's a stylized logo featuring an abstract bird-like figure, possibly an eagle, with its wings spread. To the right of the logo, there's a portion of a word, with the letters 'DE' visible, suggesting it's the beginning of a longer word or title. The overall image appears to be a fragment of a larger document, possibly a letterhead or official paper. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 #### AUG 2 1 2007 Spinal USA % Mr. Jeffrey Johnson Manager, Regulatory Affairs 644 Lakeland East Drive, Suite A Flowood, Mississippi 39232 Re: K071438 . Trade/Device Name: PSS System Regulation Number: 21 CFR 888.3070(b)(1) Regulation Name: Pedicle Screw Spinal System Regulatory Class: II Product Codes: MNI, MNH Dated: May 22, 2007 Received: May 23, 2007 Dear Mr. Johnson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This {3}------------------------------------------------ Page 2 – Mr. Jeffrey Johnson letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally premainer noticated wice results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address <http://www.fda.gov/cdrh/industry/support/index.html>>. Sincerely vours, Souhare Bvenp Mark N. Melkerson Director Division of General, Restorative and Neurological Devicesd Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): ٻڳڙ 38 Device Name: PSS System Indications for Use: The PSS System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairroent, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudoarthrosis). The PSS System is also intended for non-cervical pedicle screw fixation for the following indications: severe spondylolisthesis (grades 3 and 4 of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion. It is also intended for the following indications: trauma (i.e. fracture or dislocation); spinal stenosis; curvatures (i.e. scoliosis; and/or lordosis); spinal tumor; pseudoarthrosis; and failed previous fusion. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Barbara Brown Division Sign off Division of General, Restorativ and Neurological Devices **510(k) Number** k071438
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...