K071436 · Konica Minolta Medical & Graphic, Inc. · LLZ · Jun 27, 2007 · Radiology
Device Facts
Record ID
K071436
Device Name
REGIUS UNITEA
Applicant
Konica Minolta Medical & Graphic, Inc.
Product Code
LLZ · Radiology
Decision Date
Jun 27, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The REGIUS Unitea software is intended for installation on an off-the-shelf PC meeting or exceeding minimum specifications. The REGIUS Unitea software primarily facilitates processing and presentation of medical images on display monitors suitable for the medical task being performed. The REGIUS Unitea software can process and display images from the following modality types: Plain X-ray Radiography, X-ray Computed Tomography, Magnetic Resonance imaging, Ultrasound, Nuclear Medicine and other DICOM compliant modalities. The REGIUS Unitea must not be used for primary image diagnosis in mammography.
Device Story
REGIUS Unitea is software installed on off-the-shelf PCs; manages/controls cassette-based CR systems (REGIUS 170, 190, 110) via network; receives/displays DICOM images from modalities and digital media (DVD/CD-R, USB). Features include: demographic retrieval from HIS; reading condition specification; image processing (contrast/density, frequency, equalization, hybrid, integral, masking, rotation/flipping, re-sampling/resizing, stitching, grid suppression); annotation/marker addition; output to printers/DICOM devices. Used by clinicians to facilitate image presentation on monitors; aids diagnostic workflow by enhancing image quality and managing patient data. Benefits include improved visualization of medical images and streamlined data handling.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Software-only device for off-the-shelf PC. Features DICOM connectivity, image processing algorithms (frequency, equalization, hybrid, integral), and grid suppression. Operates on standard PC hardware. No specific materials or energy sources defined as it is software.
Indications for Use
Indicated for processing and display of medical images from X-ray, CT, MRI, Ultrasound, Nuclear Medicine, and other DICOM modalities. Contraindicated for primary image diagnosis in mammography.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
K071436
# 510(k) Summary
#### 1. Company Identification
JUN 2 7 2007
Konica Minolta Medical & Graphic, Inc. 2970 Ishikawa-machi, Hachioji-shi, Tokyo 192-8505, Japan Phone: 81-42-660-9607 Fax: 81-42-660-9588
### 2. Official Correspondent
Koji Matsushima (Mr.) Manager Standards & Regulations Department Quality Assurance Center
### 3. Date of Submission
May 21, 2007
### 4. Establishment Registration No. 3003769120
- 5. Device Trade Name REGIUS Unitea
### 6. Common Name
Medical Image Processing Work Station
### 7. Classification
Class LI, 90 LLZ, 21 CFR 892. 2050, Picture archiving and communications system
#### 8. Predicate Device
REGIUS CS-2000 and CS-3000, 510(k) number: K051523 cleared on Jul.20,2005
### 9. Description of Device
REGIUS Unitea is a software which is intended for installation on an off-the-shelf PC meeting or exceeding minimum specifications. REGIUS unitea software controls and manages the cassette type CR (Computed Radiography) such as REGIUS MODEL 170,190 and 110 that is connected via the network. REGIUS Unitea receives and displays images from other DICOM compliant modalities connected via the network and whole use digital media such as DVD/CD-R, DSC and USB memory cards connected as disk drives
REGIUS Unitea software has the following set of features.
{2}------------------------------------------------
- 1. Feature to automatically obtain patient demographic information (Name, Age, Sex, Date of Birth, etc) from Hospital Information Systems.
- 2. Feature to specify the reading condition (Sampling Sensitivity of the sensor and so on) of the connected CR device.
- 3. Receive and store image data from REGIUS MODEL 170, 190 and 100 CR, other DICOM compliant modalities and digital media (such as DVD/CD-R, DSC and USB memory cards).
- 4. Display image data received from the REGIUS MODEL 170, 190 and 100 CR, other DICOM compliant modalities and digital media (such as DVD/CD-R, DSC and USB memory cards).
- 5. Feature to apply image processing to images received from REGIUS MODEL 170. 190 and 100 CR.
Image processing applied includes:
- 1) Contrast and Density Adiustments.
- 2) F-processing: Frequency processing that highlights fine details in the image, or enhances detail that has been blurred, without affecting density.
- 3) E-processing: Equalization processing that improves the image quality that cannot be fully expressed by the film latitude due to wide distribution of the subject.
- 4) H-processing: Hybrid processing which combines frequency enhancement processing and equalization processing based on multi-resolution analysis.
- 5) 1- processing: Integral processing that automatically adjusts luminance (film density) and contrast of an image.
- 6) Masking: Filis black on the area of the frame where X-ray is not irradiated.
- 7) Rotating/Flipping: Rotate/Reverse an image.
- 8) Re-sampling and Resizing: Function that re-samples and resizes the image data.
- 9) Stitching: Function that manually or automatically recognizes long body part and stitch multiple images to create a composite image. The stitched image can be divided into several small images before outputting to a storage device or film printer.
- 10) Grid Suppression Function to suppress the grid and moire patterns within the image exposed with the grid
- 11) Function to add digital marker for patient information and to add annotations to the image data
- 12) Function that outputs image data to a printer, other DICOM devices and digital media.
#### 10. Intended Use
The REGIUS Unitea software is intended for installation on an off-the-shelf PC meeting or exceeding minimum specifications. The REGIUS Unitea software primarily facilitates processing and presentation of medical images on display monitors suitable for the medical task being performed. The REGIUS Unitea
{3}------------------------------------------------
software can process and display images from the following modality types: Plain X-ray Radiography, X-ray Computed Tomography, Magnetic Resonance imaging, Ultrasound, Nuclear Medicine and other DICOM compliant modalities. The REGIUS Unitea must not be used for primary image diagnosis in mammography.
#### 11. Substantial Equivalence to Predicate Device
REGIUS Unitea is substantially equivalent to our REGIUS CS-2000 and CS-3000, 510(k) number: K051523. Comparison of the principal characteristics of these devices is shown in the Section 3.
#### 12. Safety Information
REGIUS Unitea introduces no new safety and efficacy issues other than those already identified with the predicate device. The results of a hazard analysis, combined with the appropriate preventive measure taken indicate that the device is of minor level of concern as per the May 11, 2005 issue of the "Guidance for the Content of Premark Submissions for Software Contained in Medical Devices".
{4}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
## UN 2 7 2007
Konica Minolta Medical & Graphic, Inc. % Mr. Russell Munves US Agent and Attorney Storch, Amini, & Munves, P.C. 140 East 45th St., 25th Floor Two Grand Central Tower NEW YORK NY 10017
Re: K071436
Trade/Device Name: REGIUS Unitea Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: May 21, 2007 Received: May 23, 2007
#### Dear Mr. Munves:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/3/Picture/9 description: The image is a circular logo for the FDA Centennial, celebrating 1906-2006. The letters "FDA" are prominently displayed in a bold, stylized font. Below the letters, the word "Centennial" is written in a cursive font. Three stars are arranged in a horizontal line beneath the word "Centennial".
roloting and Promoting Public Stoa
{5}------------------------------------------------
Page 2 -
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (2) CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely yours,
Nancy C Hodgon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name REGIUS Unitea
Indications for Use:
The REGIUS Unitea software is intended for installation on an off-the-shelf PC meeting or exceeding minimum specifications. The REGIUS Unitea software primarily facilitates processing and presentation of medical images on display monitors suitable for the medical task being performed. The REGIUS Unitea software can process and display medical images from the following modality types: Plain X-ray Radiography, X-ray Computed Tomography, Magnetic Resonance imaging, Ultrasound, Nuclear Medicine and other DICOM compliant modalities. The REGIUS Unitea must not be used for primary image diagnosis in mammography.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Herbert Lewis
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K071436
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.