← Product Code [KPR](/productcode/KPR) · K071402

# X3C 1600 PLUS DIGITAL RADIOGRAPHIC SYSTEM (K071402)

_Imaging Dynamics Company , Ltd. · KPR · Jun 8, 2007 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K071402

## Device Facts

- **Applicant:** Imaging Dynamics Company , Ltd.
- **Product Code:** [KPR](/productcode/KPR.md)
- **Decision Date:** Jun 8, 2007
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.1680
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

The 1600 Plus is intended for use by a qualified/trained doctor or technologist on both adult and pediatric patients for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts but not mammography. Applications can be performed with patient sitting, standing, or lying in the prone or supine positions.

## Device Story

X3C 1600 Plus is a digital radiographic system; modification of Xplorer 1600 Plus. System replaces Xplorer 1000 detector with X3C digital radiographic detector. Device captures X-ray signals to produce diagnostic images of skeletal and soft tissue structures. Operated by qualified physicians or technologists in clinical settings. Patient positioning includes sitting, standing, prone, or supine. Output provides clinicians with radiographic images for diagnostic assessment. System requires operation with specified medical X-ray equipment by trained personnel following radiation safety standards.

## Clinical Evidence

No clinical tests conducted. Bench testing only.

## Technological Characteristics

Digital radiographic system; utilizes X3C digital radiographic detector. Conforms to 21 CFR part 820, ISO 9001, and ISO 13485 quality systems. Stationary X-ray system architecture.

## Regulatory Identification

A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Xplorer 1600 Plus Digital Radiographic System ([K062586](/device/K062586.md))

## Reference Devices

- Xplorer 1000 digital radiographic detector ([K992955](/device/K992955.md))
- X3C digital radiographic detector ([K070079](/device/K070079.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## Special 510k Submission X3C 1600 Plus Digital Radiographic System

## 510k Summary

| 1. | Submitter:             | Imaging Dynamics Company Ltd<br>Suite 151, Pegasus Way NE<br>Calgary, AB,<br>Canada T2E 8M5                                                                                                                                                                                                                                                                           | JUN - 8 2007 |
|----|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
|    | Contact person:        | Shirantha Samarappuli<br>Manager - Regulatory Affairs<br>Tel: 403 251 9939; Fax: 403 251 1771                                                                                                                                                                                                                                                                         |              |
|    | Date Prepared:         | May 14, 2007                                                                                                                                                                                                                                                                                                                                                          |              |
| 2. | Device Name:           | X3C 1600 Plus Digital Radiographic System,                                                                                                                                                                                                                                                                                                                            |              |
| 3. | Device Classification: | Class II, 892.1680 (KPR), 892.1630 (MQB),                                                                                                                                                                                                                                                                                                                             |              |
| 4. | Predicate Device:      | Xplorer 1600 Plus Digital Radiographic System (K062586)                                                                                                                                                                                                                                                                                                               |              |
| 5. | Device Description:    | The X3C 1600 Plus is a modification to Xplorer 1600 Plus where the<br>Xplorer 1000 digital radiographic detector (a previously marketed device covered by 510k K992955) in<br>Xplorer 1600 Plus system is replaced with X3C digital radiographic detector, previously marketed<br>device under K070079. The X3C 1600 Plus system is manufactured by Imaging Dynamics. |              |

6. The X3C 1600 PLUS is intended for use by a qualified/trained doctor or Indications for Use: technologist on both adult and pediatric patients for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts on both adult and pediatric patients. Applications can be performed with patient sitting, standing or lying in the prone or supine positions. The X3C 1600 Plus (510k submission device) is not intended for mammography.

- 7. Comparison with predicate device: The X3C 1600 PLUS is substantially equivalent to the currently marketed Xplorer 1600 Plus. X3C 1600 Plus device does not alter the fundamental scientific technology from Xplorer 1600 Plus predicate device. The replacement of Xplorer 100 digital radiographic detector (K992955) with X3C digital radiographic detector (K070079) is the only significant change between the 2 devices. X3C 1600 Plus has the same intended use as the predicate device.
	- a. Non-clinical tests: The device has been evaluated for performance, biocompatibility and effectiveness as well as thermal, electrical and mechanical safety and has been found to substantially equivalent to predicate device. The design and development process of the manufacturer conforms to 21 CFR part 820, ISO 9001 and ISO 13485 quality systems.
	- b. Clinical tests: No clinical tests conducted.
	- C. Conclusion: The device was evaluated against the predicate device (Xplorer 1600 Plus -K062586) for all performance, safety & effectiveness requirements and found as substantially equivalent to the predicate device.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for the Department of Health & Human Services. The logo consists of two parts: a symbol on the left and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" on the right. The symbol is a stylized representation of a human figure, and the text is written in a clear, sans-serif font.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

Ms. Shirantha Samarappuli Manager-Regulatory Affairs Imaging Dynamics Company, Ltd. Suite 151, 2340 Pegasus Way NE CALGARY ALBERTA T2E 8M5 CANADA

AUG - 9 2013

Re: K071402

Trade/Device Name: X3C 1600 Plus Digital Radiographic System Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: KPR and MQB Dated: May 18, 2007 Received: May 21, 2007

Dear Ms. Samarappuli:

This letter corrects our substantially equivalent letter of June 8, 2007.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21. Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Parts 801 and 809): medical device reporting (reporting of

{2}------------------------------------------------

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours,

Janine M. Morris

Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## · Indications for Use 1 - 2 ·

## 1.1 Indications for Use

The U.S. Food and Drug Administration (FDA) requires the following statement to appear in this manual:

"The 1600 Plus is intended for use by a qualified/trained doctor or technologist on both adult and pediatric patients for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts but not mammography. Applications can be performed with patient sitting, standing, or lying in the prone or supine positions."

FDA 510(k) for the X4C 1600 Plus: K062586

FDA 510(k) for the X3C 1600 Plus: to be assigned by FDA

FDA 510(k) for the X3C Detector Head: K070079

- · The 1600 Plus digital radiography system is not to be used for mammography.
- · The 1600 Plus digital radiography system should only be operated in conjunction with specified medical x-ray equipment and by trained and licensed personnel.
- WARNING! Only certified and properly trained, authorized personnel should be permitted to take x-ray exposures. No practical design can incorporate complete protection for personnel who do not follow proper safety precautions.
- WARNING! The appropriate x-ray imaging exposure parameters should be consistent with established Federal and Provincial/State radiation protection practices and medical imaging standards in your institution.

*Prescription Use*

Herbert Hever

vision Sion-Off) ivision of Reproductive. Abdomina Radiological Devices 510(k) Number

Imaging Dynamics Company · PN: 902-0001-015 v1.0 · May 29, 2007

---

**Source:** [https://fda.innolitics.com/device/K071402](https://fda.innolitics.com/device/K071402)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
