K071398 · Applied Spectral Imaging , Ltd. · JOY · Oct 4, 2007 · Hematology
Device Facts
Record ID
K071398
Device Name
SCANVIEW, MODEL SC-300
Applicant
Applied Spectral Imaging , Ltd.
Product Code
JOY · Hematology
Decision Date
Oct 4, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.5260
Device Class
Class 2
Indications for Use
The ScanView System is an automated scanning microscope and image analysis system. It is intended for in-vitro diagnostic use as an aiding tool to the pathologist or cytogeneticist in the detection, classification and enumeration of cells of interest based on color, intensity, size, pattern, and shape. The ScanView is indicated to detect the following cell types: CEP® X Spectrum Orange™/CEP® Y Spectrum Green™ DNA Probe Kit (Vysis, Inc., Downer's Grove, IL) and is limited to the analysis of CEP XY probes via high magnification capture and analysis of interphase nuclei. CEP XY is indicated for use to assess the effectiveness of bone marrow transplantation in opposite-sex transplants
Device Story
ScanView is an automated scanning microscope and image analysis system; integrates external microscope, motorized multi-slide stage, camera, and workstation. Operates via brightfield and fluorescent illumination; motorized Z-axis, X-Y stage, and filter turret enable automated slide scanning. Captures high-magnification images of interphase nuclei; software enables image enhancement, archiving, retrieval, and printing. Used in clinical laboratory settings by pathologists or cytogeneticists. System aids in detection, classification, and enumeration of cells based on color, intensity, size, pattern, and shape. Output allows clinicians to review digital images to assess bone marrow transplantation effectiveness; facilitates diagnostic decision-making by providing visual evidence of cell characteristics.
Clinical Evidence
Bench and clinical data demonstrate device meets required specifications. No adverse effects detected. Testing confirms performance in detection, classification, and enumeration of cells using specified DNA probe kits.
Technological Characteristics
Integrated digital imaging system; motorized microscope with Z-motion, X-Y stage, and motorized fluorescence filter turret. Includes camera and workstation. Software-based image analysis for cell detection/enumeration. Complies with 21 CFR 820.30.
Indications for Use
Indicated for use by pathologists or cytogeneticists to detect, classify, and enumerate cells using CEP X Spectrum Orange/CEP Y Spectrum Green DNA Probe Kit. Limited to analysis of CEP XY probes in interphase nuclei to assess bone marrow transplantation effectiveness in opposite-sex transplants.
Regulatory Classification
Identification
An automated cell-locating device is a device used to locate blood cells on a peripheral blood smear, allowing the operator to identify and classify each cell according to type. (Peripheral blood is blood circulating in one of the body's extremities, such as the arm.)
{1}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY INSTRUMENT ONLY TEMPLATE
A. 510(k) Number:
K071398
B. Purpose for Submission:
Traditional 510(k)
C. Manufacturer and Instrument Name:
Applied Spectral Imaging Ltd. (ASI)
D. Type of Test or Tests Performed:
The ScanView is indicated to detect the following cell types: CEP X Spectrum Orange/CEP Y Spectrum Green DNA Probe Kit (Vysis, Inc. Downer's Grove, IL) and is limited to the analysis of CEP XY probes via high magnification capture and analysis of interphase nuclei.
E. System Descriptions:
1. Device Description:
The ScanView System is an integrated digital imaging system comprising an external microscope, motorized multi-slide stage, camera, and a workstation. It is designed to acquire images of cells and enables identification and examination of cells of interest. Cytogenetics experts can view and scan cells and record the images using both bright field and fluorescent illumination. The acquired images can be enhanced, archived, retrieved, and printed. The automated microscope also includes a motorized filter turret containing fluorescence filters.
2. Principles of Operation:
The ScanView system works with bright field and fluorescent samples with all currently applied staining techniques. The system transfers images of cells from the microscope to a computer where it is analyzed and processed. The operator with the support of the ScanView software views the images. The results are interpreted by a competent cytogeneticist who exercises his judgment in the use of the provided information when formulating the diagnosis. The results are documented in hard copy and archive.
{2}
2
3. **Modes of Operation:**
a. Fully Automated Scan and Relocation; b. Image Acquisition and Semi-automatic Relocation; c. Automatic Scanning and Semi-automatic Relocation on Remote Stations; d. Finding Metaphases in a SKY Slide; e. Finding Metaphases in a Multi-color FISH Sample; f. Fluorescence Multi-color FISH Samples-Dual Scan.
4. **Specimen Identification:**
Manual keyboard entry into a Case Data Manager.
5. **Specimen Sampling and Handling:**
ScanView scans patient specimen slides stained with the CEP X Spectrum Orange / CEP Y Spectrum Green DNA Probe Kit (Vysis, Inc.).
6. **Calibration:**
Factory calibrated
7. **Quality Control:**
ProbeCheck slides, supplied by Vysis, are designed for use as controls for interphase FISH and laboratory quality control, specifically, for the CEP X/Y IVD Kit. ProbeCheck Low Level Male Control Slides (95% XX, 5% XY) and ProbeCheck Low Level Female Control Slides (95% XY, 5% XX) are tested and optimized for use in FISH assays. These slides are manufactured from a mixture of cultured normal male and female lymphoblast cell lines.
8. **Software:**
FDA has reviewed applicant’s Hazard Analysis and Software Development processes for this line of product types:
Yes ☐ X or No ☐
F. **Regulatory Information:**
1. **Regulation section:**
21 CFR 866.4700, Automated fluorescence in situ hybridization (FISH) enumeration systems.
{3}
2. Classification:
Class II
3 Product code:
NTH
4. Panel:
Immunology (82)
# G. Intended Use:
1. Indication(s) for Use:
The ScanView System is an automated scanning microscope and image analysis system. It is intended for in-vitro diagnostic use as an aiding tool to the pathologist or cytogeneticist in the detection, classification and enumeration of cells of interest based on color, intensity, size, pattern, and shape. The ScanView is indicated to detect the following cell types: CEP® X Spectrum Orange™/CEP® Y Spectrum Green™ DNA Probe Kit (Vysis, Inc., Downer's Grove, IL) and is limited to the analysis of CEP XY probes via high magnification capture and analysis of interphase nuclei. CEP XY is indicated for use to assess the effectiveness of bone marrow transplantation in opposite-sex transplants
2. Special Conditions for Use Statement(s):
NA
# H. Substantial Equivalence Information:
1. Predicate Device Name(s) and 510(k) numbers:
a. Applied Spectral Imaging, FISHView, K050236
b. Applied Imaging, CytoVision, K042542
2. Comparison with Predicate Device:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Automated Scanning | Yes | CytoVision: Yes |
| Device Components | Automated Microscope. | CytoVision: Same |
{4}
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Constraints | PC with windows based-based operating system. Keyboard and control panel. Color monitor for display of information. CCD Camera. Motorized Stage.
Reliance on automatic scanning and on the expertise and judgment of the cytogenetics technician for cell selection, editing, examination, and correction. All final diagnoses must be made by qualified medical personnel using all information available from the clinical evaluation and other diagnostic procedures.
The system does not suggest an interpretation, diagnosis, or treatment.
The operator should adhere to the standard safety procedures employed when handling biological substances, glass microscope slides, and electronic equipment. | CytoVision: Same |
| Software Application | Control of the man-machine-interface.
Scan, Review, Capture, and Analysis of the images.
Handling the images | CytoVision: Same |
4
{5}
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Spatial resolution | display, storage, and communication.1280 X 1024 | FISHview: Same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Indication for Use | The ScanView System is an automated microscope and image analysis system. It is intended for in-vitro diagnostic use as aiding tool to the pathologist or cytogeneticist in the detection, classification and enumeration of cells of interest based on color, intensity, size, pattern, and shape. The ScanView is indicated to detect the following cell types: CEP® X Spectrum Orange™/CEP® Y Spectrum Green™ DNA Probe Kit (Vysis, Inc., Downer's Grove, IL) and is limited to the analysis of CEP XY probes via high magnification capture and analysis of interphase nuclei. CEP XY is indicated for use to assess the effectiveness of bone marrow transplantation in opposite-sex transplants. | The FISHView system is intended to be used for karyotyping with real-time microscope images from cultured and stained cell specimens in their metaphase. The system works with bright field and fluorescent samples. Specimens suitable for banding analysis are: amniotic fluid, chorionic villus, bone marrow and solid tumor. Karyotyping is normally applied for the pre and postnatal diagnosis of birth defects, chromosome abnormalities, genetic diseases (such as Down's Syndrome), cancer and for the follow up of cancer treatment. The FISHView system does not locate metaphase spreads; it does not rank the given according to quality; nor does it automatically classify chromosomes. In addition, the FISHView is intended as an aiding tool to the pathologist or cytologist for digital visualizing, processing, |
{6}
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | | counting, and
classification of stained cells and for storage of FISH multi-dye images of the following specimens: amniotic fluid, peripheral blood, chorionic villus, bone marrow and solid tumor.
The FISHView does require and relies on the operator to analyze the digitized microscope images.
The CytoVision system is an automated scanning microscope and image analysis system. It is intended for in vitro diagnostic use as an aid in chromosomal analysis. CytoVision assists in the location of interphase and metaphase nuclei on standard microscope slides using both brightfield and fluorescent microscopy. This particular CytoVision software application (CEP XY_ENG) is an accessory to the CEP® X Spectrum Orange™/CEP® Y Spectrum Green™ DNA Probe Kit (Vysis, Inc., Downer’s Grove, IL) and is limited to the analysis of CEP XY probes via high magnification capture and analysis of interphase nuclei. CEP |
6
{7}
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Device components | Motorized stage | XY is indicated for use to assess the effectiveness of bone marrow transplantation in opposite-sex transplants. |
| Constraints | Reliance on automatic scanning and on the expertise and judgment of the cytogenetics technician for cell selection, editing, examination, and correction. All final diagnoses must be made by qualified medical personnel using all information available from the clinical evaluation and other diagnostic procedures. | FISHView: no motorized stage.FISHView: Reliance on the expertise and judgment of the cytogenetics technician for cell selection, editing, examination, and correction. All final diagnoses must be made by qualified medical personnel using all information available from the clinical evaluation and other diagnostic procedures. |
| | The system does not suggest an interpretation, diagnosis, or treatment. | The system does not suggest an interpretation, diagnosis, or treatment. |
| | The operator should adhere to the standard safety procedures employed when handling biological substances, glass microscope slides, and electronic equipment. | The operator should adhere to the standard safety procedures employed when handling biological substances, glass microscope slides, and electronic equipment. |
| | Yes | FISHView: No |
| Automated scanning | 1280 x 1024 | CytoVision: Not known. |
| Spatial resolution | | |
{8}
I. Special Control/Guidance Document Referenced (if applicable):
Class II Special controls Guidance document: Automated Fluorescence in situ Hybridization (FISH) Enumeration Systems. May 23, 2005.
J. Performance Characteristics:
1. Analytical Performance:
a. Accuracy:
Measurements of 61 slides (bone marrow samples from leukemic patients at different post-transplant periods, starting from 7 days up to 2 years and more) using the manual gold standard were compared to results obtained by the ScanView XY System. Regression of manual results on Scan results indicated that:
- Scan and Manual are highly related with R-squares of 0.997, 0.997, and 0.980 for XY-percent, XX-percent, and their ratio respectively.
- The slopes of the Manual-Scan regressions did not differ significantly from 1.0 for all three parameters.
- The intercepts of the Manual-Scan regressions did not differ significantly from 0 for all three parameters.
b. Precision/Reproducibility:
For the reproducibility study, 5 slides from 5 different patients were chosen randomly from the 61 slides mentioned in the accuracy section above. In addition, 5 Vysis ProbeCheck control slides were also used to test reproducibility. ProbeCheck Low Level Male Control Slides (95% XX, 5% XY) and ProbeCheck Low Level Female Control Slides (95% XY, 5% XX) were tested. Reproducibility standard deviations for XY-percent, XX-percent and their Ratio were all 3.3% and below. Using these data a statistical model was constructed from which population standard deviations were estimated. The modeled standard deviations for XY-percent, XY-percent, and their Ratio ranged between 0.02% and 0.99%.
d. Carryover:
N/A
{9}
e. Interfering Substances:
N/A
2. Other Supportive Instrument Performance Data Not Covered Above:
N/A
K. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
L. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
9
{10}
10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.