JAZZ BLOOD GLUCOSE MONITORING SYSTEM

K071393 · Agamatrix · NBW · Jul 20, 2007 · Clinical Chemistry

Device Facts

Record IDK071393
Device NameJAZZ BLOOD GLUCOSE MONITORING SYSTEM
ApplicantAgamatrix
Product CodeNBW · Clinical Chemistry
Decision DateJul 20, 2007
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Intended Use

AgaMatrix JAZZ™ Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood from the finger stick, palm, and/or forearm. Testing is done outside the body (in vitro diagnostic use). It is indicated for use at home (over the counter (OTC) ) by persons with diabetes, or in a clinical setting by healthcare professionals, as an aid to monitor the effectiveness of diabetes control. AgaMatrix JAZZ™ Blood Glucose Meter is intended for use with AgaMatrix JAZZTM Blood Glucose Test Strips for the quantitative measurement of glucose in fresh capillary whole blood from the finger stick, palm, and/or forearm. Testing is done outside the body (in vitro diagnostic use). It is indicated for use at home (over the counter (OTC) ) by persons with diabetes, or in a clinical setting by healthcare professionals, as an aid to monitor the effectiveness of diabetes control. AgaMatrix JAZZ™ Blood Glucose Test Strips are intended for use with AgaMatrix JAZZ™ Blood Glucose Meter for the quantitative measurement of glucose in fresh capillary whole blood from the finger stick, palm and/or forearm. Testing is done outside the body (in vitro diagnostic use). It is indicated for use at home (over the counter (OTC) ) by persons with diabetes, or in a clinical setting by healthcare professionals, as an aid to monitor the effectiveness of diabetes control. AgaMatrix JAZZ™ Control Solutions are intended for use with the AgaMatrix JAZZ™ Meter and AgaMatrix JAZZ™ Test Strips as a quality control check to verify the accuracy of blood glucose test results.

Device Story

Portable, battery-operated blood glucose meter; uses test strips to measure glucose in fresh capillary whole blood samples; intended for OTC home use by diabetics or clinical use by professionals; provides quantitative glucose concentration output; aids in monitoring diabetes control effectiveness; system includes meter, lancing device, lancets, and control solutions; brand name change from Liberty Blood Glucose Monitoring System with minor user interface improvements.

Clinical Evidence

Bench testing only; no clinical data provided. Verification and validation activities performed per design control requirements to confirm that modifications met predetermined acceptance criteria.

Technological Characteristics

Portable, battery-operated glucose test system; in vitro diagnostic; includes meter, test strips, and control solutions; complies with ISO 15197:2003, ISO 14971:2000, IEC 61010-1, IEC 61010-2-101, and IEC 61000-4-3.

Indications for Use

Indicated for persons with diabetes for quantitative blood glucose monitoring using fresh capillary whole blood (finger, palm, forearm). Not for use with neonates.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ AgaMatrix 10 Manor Parkway Salem, NH 03079 Office: +1 (603) 328-6000 Fax: +1 (603) 893-4191 #### 510 (k) SUMMARY This summary of 510(k) safety and effectiveness information is being submitted in JUL 2 0 2007 accordance with the requirements of SMDA 1990 and 21 CFR 807.92. ### 1. Submitter's Name, Address, Telephone Number, Contact Person, and date the summary was prepared. | Submitter's Name: | AgaMatrix, Inc.<br>10 Manor Parkway<br>Salem, NH 03079 | |-------------------|--------------------------------------------------------| | Contact Person: | Connie Hertel<br>Director Quality & Regulatory Affairs | | Telephone: | (603) 328-6000 | | Fax: | (603) 893-4191 | Date the summary prepared: May 17, 2007 #### 2. Device Name | Trade/Proprietary Name: | Jazz™ Blood Glucose Monitoring System | |-------------------------|-------------------------------------------| | Common/Usual Name: | Blood Glucose Monitoring System | | Classification Name: | Glocuse test system (per 21 CFR 862.1345) | | Class: | II | | Panel: | Chemistry | #### 3. Modification Device: The Jazz 104 Blood Glucose Monitoring System is a brand name change to the Liberty ™ Blood Glucose Monitoring System. Some user interface modifications were improved. #### 4. Description of the Device The AgaMatrix JazzTM Blood Glucose Monitoring System includes a meter with batteries, compact carrying case, Lancing device, lancets, control solution and owner's booklet. Test Strips are sold separately. It is intended for over-the-counter home use by diabetics to monitor their blood glucose levels, or for use in a clinical setting by healthcare professionals. The system tests fresh capillary whole blood. The meter is a portable, battery-operated instrument. {1}------------------------------------------------ AgaMatrix #### 510(k) Summary (Continued) #### 5. Intended use of the device The Jazz™ Blood Glucose Monitoring System is intended to quantitatively measure blood glucose levels, also known as blood sugar, from fresh capillary whole blood samples taken from the fingertips or the palm. The Jazz™ Test Strips are for in vitro diagnostic (outside of the body) use only. The Jazz™ System is not intended for use with neonates. #### 4. Testing The manufacturer of the Jazz™ Blood Glucose Monitoring System certifies that the device complies with the following: ISO 15197:2003 In vitro diagnostic test systems - Requirements for bloodglucose monitoring systems for self-testing in managing diabetes mellitus ISO 14971:2000 Medical devices - Application of risk management to Medical devices IEC 61010-1 Medical electrical equipment -- General requirements for safety IEC 61010-2-101 Safety requirements for electrical equipment for measurement, control and laboratory use - particular requirements for in vitro diagnostic (IVD) medical equipment IEC 61000-4-3 Electromagnetic compatibility (EMC) #### 8. Conclusions Based upon the testing and comparison to the predicate device, the Jazz™ Blood Glucose Monitoring System has the same intended uses, with similar technological characteristics as the predicate device. The system performs as intended and raises no new safety of effectiveness issues. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three overlapping wings, symbolizing service to the nation. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 JUL 2 0 2007 AgaMatrix, Inc. c/o Connie Hertel Director Quality & Regulatory Affairs 10 Manor Parkway Salem, NH 03079 Re: k071393 Trade/Device Name: Jazz Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: Class II Product Code: NBW, CGA Dated: June 28, 2007 Received: June 29, 2007 #### Dear Ms. Hertel: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {3}------------------------------------------------ Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device. or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. Jean M. Cooper, M.S., D.V.M. Yéan M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indication for Use 510(k) Number (if known): k071393 Device Name: JAZZ™ Indication For Use: ### AgaMatrix JAZZ™ Blood Glucose Monitoring System: AgaMatrix JAZZ™ Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood from the finger stick, palm, and/or forearm. Testing is done outside the body (in vitro diagnostic use). It is indicated for use at home (over the counter (OTC) ) by persons with diabetes, or in a clinical setting by healthcare professionals, as an aid to monitor the effectiveness of diabetes control. ## AgaMatrix JAZZ™ Blood Glucose Meter: AgaMatrix JAZZ™ Blood Glucose Meter is intended for use with AgaMatrix JAZZTM Blood Glucose Test Strips for the quantitative measurement of glucose in fresh capillary whole blood from the finger stick, palm, and/or forearm. Testing is done outside the body (in vitro diagnostic use). It is indicated for use at home (over the counter (OTC) ) by persons with diabetes, or in a clinical setting by healthcare professionals, as an aid to monitor the effectiveness of diabetes control. ### AgaMatrix JAZZ™ Blood Glucose Test Strips: AgaMatrix JAZZ™ Blood Glucose Test Strips are intended for use with AgaMatrix JAZZ™ Blood Glucose Meter for the quantitative measurement of glucose in fresh capillary whole blood from the finger stick, palm and/or forearm. Testing is done outside the body (in vitro diagnostic use). It is indicated for use at home (over the counter (OTC) ) by persons with diabetes, or in a clinical setting by healthcare professionals, as an aid to monitor the effectiveness of diabetes control. ### AgaMatrix JAZZ™ Control Solutions (Normal level and High level): AgaMatrix JAZZ™ Control Solutions are intended for use with the AgaMatrix JAZZ™ Meter and AgaMatrix JAZZ™ Test Strips as a quality control check to verify the accuracy of blood glucose test results. Prescription Use X (21 CFR Part 801 Subpart D) And/Or Over the Counter Use X . (21 CFR Part 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OVD) Caro Senson Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) K071393
Innolitics
510(k) Summary
Decision Summary
Classification Order