K071352 · Gebr. Brasseler GmbH & Co. KG · EJL · Jun 29, 2007 · Dental
Device Facts
Record ID
K071352
Device Name
CERADRILL
Applicant
Gebr. Brasseler GmbH & Co. KG
Product Code
EJL · Dental
Decision Date
Jun 29, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3240
Device Class
Class 1
Attributes
3rd-Party Reviewed
Indications for Use
The drill is used to pre-drill into the maxilla or mandible to create an initial osteotomy for the later endosseous dental implant placement. The drills are used with a handpiece. CeraDrill is a ceramic drill for the dental implantology made of a modern high performance ceramic. The drills are intended to be used with a handpiece. The intended use for this device is to cut into the maxilla or mandible to create an osteotomy for endosseous dental implant placement.
Device Story
CeraDrill is a ceramic pilot drill used in dental implantology to create an initial osteotomy in the maxilla or mandible. The device is operated by a dentist using an electrical powered handpiece. It features a high-performance ceramic (ZrO2/Al2O3/Y2O3) construction with specific twist drill blade geometry and lasered depth markings for precise penetration control. Unlike metal drills, it is metal-free, biocompatible, and corrosion-resistant. The device provides high cutting efficiency and improved thermal safety compared to traditional steel burs. It is designed for repeated use, capable of being cleaned, disinfected, and sterilized for at least 10 cycles without significant loss of mechanical stability. The output is a prepared bone site ready for subsequent implant placement, facilitating positive osseointegration.
Clinical Evidence
No clinical data. Substantial equivalence is based on non-clinical performance testing, including physical and biological characterization, thermal safety analysis, and mechanical stability testing after repeated reprocessing cycles.
Technological Characteristics
Material: ZrO2/Al2O3/Y2O3 ceramic. Design: Twist drill with 2 or 3 cutting flutes; external irrigation. Dimensions: Various diameters and lengths. Energy: Powered by electrical handpiece. Biocompatibility: ISO 7405 compliant. Reprocessing: Sterilizable, validated for at least 10 cycles. Features: Lasered depth markings at 2 mm intervals.
Indications for Use
Indicated for patients requiring endosseous dental implants. Used to create an initial osteotomy in the maxilla or mandible for implant placement. Intended for use by dentists with a powered handpiece.
Regulatory Classification
Identification
A dental bur is a rotary cutting device made from carbon steel or tungsten carbide intended to cut hard structures in the mouth, such as teeth or bone. It is also intended to cut hard metals, plastics, porcelains, and similar materials intended for use in the fabrication of dental devices.
{0}------------------------------------------------
Premarket notification 510(k) submission Gebr. Brasseler Gmbh & Co KG CeraDrill K:
JUN 2 9 100
# 10(k) Summarv
Gebr. Brasseler GmbH & Co. KG owner's name: address: Trophagener Weg 25 32631 Lemgo Germany phone: + 49 5261 701 467 fax numbers: + 49 5261 701 289 Mr. Olaf Brand name of contact person: date the summary was prepared: February 2007 Establishment Registration number: 8010468 name of the device: CeraDrill trade or proprietary name: CeraDrill the classification name: Bone cutting instrument and accessories (21 CFR 872.4120 Product Code DZI)
> Date: 2007-02-01 Revision: 3 Chapter 05 1/8 Page:
{1}------------------------------------------------
Nobel Biocare USA, Inc.
Device:
Amorphous Diamond Coated Drill
510(k) No.:
K990846
#### Description of the device:
CeraDrills within the new "CeraLine" is intended for use in implantology. These are made of high efficiency ceramics featuring an excellent cutting performance. The new CeraDrill offers the possibility of working without any metal, thus guaranteeing a biocompatible operation. Damages, as they occur in particular when disinfecting and cleaning steel burs with inappropriate agents, are now a thing of the past. The multifunctional ceramic drills are suited for initial preparation of the implant site axis and depth. The drills feature high initial sharpness and optimal cutting efficiency, thus achieving an effective material reduction. Due to their high efficiency toothing and the distinct tip-transversing blade, the drill has very good axial drilling properties for precise cutting without the need for prior use of a centering bur. The special twist drill blade geometry ensures smooth and precise operation. Moreover, the depth markings, which are lasered on to the working part in 2 mm intervals guarantee precise and safe control of the penetration depth.
The CeraDrill family is available in diverse dimensions and shapes for the individual preparation on the osteotomy. The following table shows the REF numbers, diameters, penetration depth, shank types, number of cutting flutes:
05
{2}------------------------------------------------
Remark:
further explanations concerning the columns of the following
table can be found in chapter 11 of this submission
| Model | REF /<br>"Bezeichnung" | ID Nr. /<br>Ident<br>Nr. | Description | Diameter /<br>SØK Kopf-<br>durch-<br>messer" | penetration<br>depth / KL<br>"Kopflänge" | Total<br>length/<br>Gesamt-<br>länge | Shank<br>type<br>ISO<br>1797 | Shank<br>type<br>US | Cutting<br>flutes/<br>Schnei-<br>denzahl |
|------------------|------------------------|--------------------------|-------------------------------------|----------------------------------------------|------------------------------------------|--------------------------------------|------------------------------|---------------------|------------------------------------------|
| K210L16.RA.020 | K210L16.204.020 | 033792 | CeraDrill<br>pilot drill<br>2,0 L16 | 2,0 mm | 16,0 mm | 30,5 mm | 204 | RA | 2 |
| K210L19.RA.020 | K210L19.204.020 | 033793 | CeraDrill<br>pilot drill<br>2,0 L19 | 2,0 mm | 19,0 mm | 33,5 mm | 204 | RA | 2 |
| K210L16.RA.028 | K210L16.204.028 | 033794 | CeraDrill<br>pilot drill<br>2,8 L16 | 2,8 mm | 16,0 mm | 32,0 mm | 204 | RA | 3 |
| K210L19.RA.028 | K210L19.204.028 | 033795 | CeraDrill<br>pilot drill<br>2,8 L19 | 2,8 mm | 19,0 mm | 35,0 mm | 204 | RA | 3 |
| K210L20.RASL.020 | K210L20.205.020 | 034899 | CeraDrill<br>pilot drill<br>2,0 L20 | 2,0 mm | 20,6 mm | 41,0 mm | 205 | RASL | 2 |
| K210L20.RASL.028 | K210L20.205.028 | 034900 | CeraDrill<br>pilot drill<br>2,8 L20 | 2,8 mm | 20,84 mm | 41,0 mm | 205 | RASL | 3 |
| K210L20.RASL.035 | K210L20.205.035 | 035211 | CeraDrill<br>pilot drill<br>3,5 L20 | 3,5 mm | 21,05 mm | 41,0 mm | 205 | RASL | 3 |
| K210L20.RASL.042 | K210L20.205.042 | 035212 | CeraDrill<br>pilot drill<br>4,2 L20 | 4,2 mm | 21,26 mm | 41,0 mm | 205 | RASL | 3 |
Date: 2007-02-01 Revision: 3 Chapter
{3}------------------------------------------------
## Intended use:
CeraDrill is a ceramic pilot drill for the dental implantology made of a modern high performance ceramic. The drills are intended to be used with a handpiece.
- Indications for use .
The drill is used to pre-drill into the maxilla or mandible to create an initial osteotomy for the later endosseous dental implant placement. The drills are used with a handpiece.
- Target population
The target population for the CeraDrill are Patients applicable for dental treatment with endosseous implants CeraDrill is used in the following anatomical sites: maxilla or mandible
CeraDrill is intended to be used at the Dentist
The device is an integral part of any implantological preparation.
Due to the fact that the device is intended to be used in the framework of the dental implantology the patients must be applicable for dental treatment with endosseous implants.
05
{4}------------------------------------------------
# Summary of technological Characteristics
# (compared to the predicate device)
The following table shows the technological characteristics of both
devices:
.
| Aspect | New device | Predicate device |
|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------|
| | CeraDrill | Amorpheus Diamond<br>coated drill |
| energy used and/or<br>delivered | The device is intended to<br>be used in electrical<br>powered hand pieces | same |
| design | two / three cutting flutes<br>external irrigation | two / three cutting flutes<br>internal /external<br>irrigation |
| performance | Ordinary concentricity | Ordinary concentricity |
| corrosion | Corrosion resistive<br>Advantage:<br>not leading to decolouri-<br>sation while reprocessing | Limited corrosion<br>resistance for stainless<br>steel |
| materials | The CeraDrill is made of<br>ZrO2/Al2O3/Y2O3<br><br>Advantage:<br>- just one material<br>- no coating<br>- limited probability of<br>abrasion due to<br>inadequate bonding | stainless steel<br>coated with diamond |
| Biocompatibility | According ISO 7405 | same |
Date: 2007-02-01 Revision: 3 Chapter 05 Page: 5/8
:
:
:
:
: ﺗﺒ ......... :
:
{5}------------------------------------------------
#### Premarket notification 510(k) submission Gebr. Brasseler Gmbh & Co KG K: ---------------------------------------------------------------------------------------------------------------------------------------------------------------------------CeraDrill ..
:
| Aspect | New device | Predicate device |
|-----------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------|
| | | |
| Ability to be<br>reprocessed | CeraDrill | Amorpheus Diamond<br>coated drill |
| | The CeraDrill can be<br>used, cleaned,<br>disinfected and sterilised<br>according to the<br>specifications as given in<br>the instructions for use<br>for at least 10 cycles.<br>(see test report# 06m1166 in<br>chapter 14A of this<br>submission) | No limitation for<br>reprocessing indicated |
| Mechanical stability<br>after reprocessing | CeraDrill may be<br>reprocessed in 100<br>cycles without a<br>significant loss of<br>mechanical stability<br>(details see chapter 18 d) | No further explanation /<br>information given |
| | thermal safety | Improved thermal safety |
| | | |
| Advantage: | | |
| It was observed that the | | |
| temperature rise when | | |
| using ceramic burs was<br>lower than that observed<br>with steel burs | | |
| (details see chapter 18 a) | | |
:
:
: .
:
: 上一篇:
:
. : :
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Page: ୧ାଃ
{6}------------------------------------------------
# Substantial equivalence determination
#### Non-clinical performance data:
The determination of substantial equivalence is based on several nonclinical tests, most of which are repeated in the body of this 510(k) submission.
This includes especially those tests showing the physical and biological characteristics. The test results can be summarised as follows:
- General technological equivalence between predicate and current device
- improved thermal safety -
- improved safety concerning the risk of inadequate bonding of the coating
- the assumable lower resistive ness against bending forces during operation is negligible, since those bending force being necessary for fracture would lead to unacceptable osteotomies (=misuse)
(Details on non-clinical test can be found in chapter 18 of this submission)
### Clinical performance data
The determination of substantial equivalence is not based on clinical data.
{7}------------------------------------------------
# Conclusion:
The main intention of the CeraDrill is to pre-drill into the maxilla or mandible to create an initial osteotomy for the later endosseous dental implant placement. As such the CeraDrill can be concluded to be substantial equivalent to the predicate device:
Amorphous Diamond Coated Drill, Nobel Biocare, K990846
The only technical difference is the material used: whereas the predicate device is made of steel and coated with amorphous diamond the CeraDril is made of a single material which is Zirconium oxide.
This technological difference leads to the following advantages of the CeraDrill device:
+ Metalfree preparation
+ positve Osteointgration
+ non allgergenic compared to CrNi Steal
There is no known disadvantage; the known difference in the bending strength of the materials does not have an impact in the intended use as identified in the chapter 18 of the submission.
The safety and performance data submitted in this file support a finding of substantial equivalence between the CeraDrill and the predicate device as specified above.
#### Any other information:
none
{8}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Gebr. Brasseler GmbH & Company KG C/O Mr. Olaf Brand TUV Rheinland of North America, Incorporated 12 Commerce Road Newtown, Connecticut 06470
JUN 2 9 2007
Re: K071352
Trade/Device Name: CeraDrill Regulation Number: 21 CFR 872.4130 Regulation Name: Intraoral Dental Drill Regulatory Class: I Product Code: EJL Dated: June 25, 2007 Received: June 27, 2007
Dear Mr. Brand:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{9}------------------------------------------------
#### Page 2 - Mr. Brand
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Susan Runore
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{10}------------------------------------------------
K0713352
# Indications for Use
510(k) Number (if known): 1007 12 50
Device Name: CeraDrill
- . Indications for use
The drill is used to pre-drill into the maxilla or mandible to create an initial osteotomy for the later endosseous dental implant placement. The drills are used with a handpiece.
- specific indications:
CeraDrill is a ceramic drill for the dental implantology made of a modern high performance ceramic. The drills are intended to be used with a handpiece. The intended use for this device is to cut into the maxilla or mandible to create an osteotomy for endosseous dental implant placement.
- clinical settings / target population: · .
Patients must be applicable for dental treatment with endosseous implants.
- anatomical sites: .
The products are used in the framework of dental implantology for the preparation of the upper jaw (Maxilla) or the lower jaw
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susa Roose
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page I of 1
510(k) Number: K071352
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.