← Product Code [ELL](/productcode/ELL) · K071182

# PROCERA BRIDGE ZIRCONIA 9 UNITS-14 UNITS, MODELS 30-4004 TO 30-4013 (K071182)

_Nobel Biocare AB · ELL · Jul 11, 2007 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K071182

## Device Facts

- **Applicant:** Nobel Biocare AB
- **Product Code:** [ELL](/productcode/ELL.md)
- **Decision Date:** Jul 11, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3920
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

Nobel Biocare's Procera Bridge Zirconia is indicated for use as the core structure of an artificial prosthesis for partially edentulous patients in the need of prosthetic oral reconstruction in order to restore chewing function. The Procera Bridge Zirconia may be cemented or bonded to either natural or artificial tooth abutments.

## Device Story

Procera Bridge Zirconia is a prefabricated dental prosthesis core structure; precision milled from a single solid piece of densely sintered Zirconia. Used in oral cavity to restore chewing function; cemented or bonded to natural or artificial tooth abutments. Operated by dental professionals in clinical settings. Provides structural support for artificial teeth; enables prosthetic reconstruction for partially edentulous patients.

## Clinical Evidence

No clinical data provided; substantial equivalence based on technological characteristics and intended use.

## Technological Characteristics

Material: Densely sintered Zirconia. Form factor: Prefabricated, precision-milled core structure for artificial prosthesis. Connectivity: None. Energy source: None. Sterilization: Not specified.

## Regulatory Identification

A porcelain tooth is a prefabricated device made of porcelain powder for clinical use (§ 872.6660) intended for use in construction of fixed or removable prostheses, such as crowns and partial dentures.

## Predicate Devices

- Procera Bridge Zirconia ([K053125](/device/K053125.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## 1.4 510(k) Summary of Safety and Effectiveness

| Submitted by:                          | Ms. Phuong Nguyen Son<br>Regulatory Affairs Specialist                     |
|----------------------------------------|----------------------------------------------------------------------------|
| Address:                               | Nobel Biocare USA LLC<br>22715 Savi Ranch Parkway<br>Yorba Linda, CA 92887 |
| Telephone:                             | (714) 282-4800, ext. 7830                                                  |
| Facsimile:                             | (714) 282-9023                                                             |
| Date of Submission:                    | April 26, 2007                                                             |
| Classification Name:                   | Porcelain Tooth (21 CFR 872.3920)                                          |
| Trade or Proprietary<br>or Model Name: | Procera Bridge Zirconia                                                    |
| Legally Marketed Device(s):            | Procera Bridge Zirconia (K053125)                                          |

## Device Description:

Nobel Biocare's Procera Bridge Zirconia is a prefabricated device intended for use as the core structure of an artificial prosthesis for placement in the oral cavity in order to restore chewing function. The Procera Bridge Zirconia is precision milled from one solid piece of densely sintered Zirconia, and can be cemented or bonded to either natural or artificial tooth abutments.

Indications for Use:

Nobel Biocare's Procera Bridge Zirconia is indicated for use as the core structure of an artificial prosthesis for partially edentulous patients in the need of prosthetic oral reconstruction in order to restore chewing function. The Procera Bridge Zirconia may be cemented or bonded to either natural or artificial tooth abutments.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the eagle. The eagle is black, and the text is in a simple, sans-serif font.

Public Health Service

JUL 1 1 2007

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Nobel Biocare AB C/O Phuong Nguyen Son Regulatory Affairs Specialist Nobel Biocare USA LLC 22715 Savi Ranch Parkway Yorba Linda, California 92887

Re: K071182

Trade/Device Name: Procera Bridge Zirconia Regulation Number: 21 CFR 872.3920 Regulation Name: Porcelain Tooth Regulatory Class: II Product Code: ELL Dated: April 26, 2007 Received: April 30, 2007

Dear Ms. Son:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

Page 2 - Ms. Son

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Clubs

Chiu Lin. Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use

510(k) Number (if known):

Device Name: Procera Bridge Zirconia

Indications For Use:

Nobel Biocare's Procera Bridge Zirconia is indicated for use as the core structure of an artificial prosthesis for partially edentulous patients in the need of prosthetic oral reconstruction in order to restore chewing function. The Procera Bridge Zirconia may cemented or bonded to either natural or artificial tooth abutments.

X Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

| Concurrence of CDRH, Office of Device Evaluation (ODE) |
|--------------------------------------------------------|
| Kein Muly for MSR                                      |

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: Page 1 of 1 ___ KOTI

## 1.3

---

**Source:** [https://fda.innolitics.com/device/K071182](https://fda.innolitics.com/device/K071182)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
