OHMEDA MEDICAL GIRAFFE OMNIBED

K071175 · GE Healthcare · FMZ · Jul 17, 2007 · General Hospital

Device Facts

Record IDK071175
Device NameOHMEDA MEDICAL GIRAFFE OMNIBED
ApplicantGE Healthcare
Product CodeFMZ · General Hospital
Decision DateJul 17, 2007
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 880.5400
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The OmniBed is a combination of an infant incubator and an infant warmer. The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user's demand. It cannot be operated in both modes at the same time. Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Incubators provide an enclosed, temperature-controlled environment and warmers provide infrared heat in an open environment. They may also be used for short periods of time to facilitate the neonate's transition from the uterus to the external environment. This device may incorporate a Servo Controlled Oxygen Delivery System. This is indicated to provide stable oxygen concentration within the infant compartment at the value set by the operator (21-65%).

Device Story

Giraffe OmniBed functions as hybrid infant incubator and radiant warmer; provides thermal support for neonates unable to thermo-regulate. Device transitions between enclosed incubator mode and open radiant warmer mode. Optional Servo Controlled Oxygen Delivery System maintains operator-set oxygen concentrations (21-65%). New feature: Uninterruptible Power Supply (UPS) provides short-term battery power (15-minute life) to support intra-hospital mobility. Operated by clinical staff in hospital settings. UPS consists of medical-grade battery and shelf; does not alter core control mechanisms or operating principles. Device provides stable thermal environment to facilitate transition from uterus to external environment.

Clinical Evidence

Bench testing only. Performance of the OmniBed with the addition of the UPS was verified against product specifications and recognized consensus standards, including IEC 601-2-21 and IEC 601-2-19 for alarm decibel levels.

Technological Characteristics

Hybrid incubator/warmer; infrared heat source for open mode; servo-controlled oxygen delivery. UPS includes medical-grade battery (15-minute capacity, 6-hour recharge). Standards: IEC 601-2-21 Amend 1, IEC 601-2-19 Amend 1. Connectivity: Standalone with battery backup.

Indications for Use

Indicated for neonates unable to thermo-regulate; provides controlled thermal environment (incubator or radiant warmer) and optional servo-controlled oxygen delivery (21-65%).

Regulatory Classification

Identification

A neonatal incubator is a device consisting of a rigid boxlike enclosure in which an infant may be kept in a controlled environment for medical care. The device may include an AC-powered heater, a fan to circulate the warmed air, a container for water to add humidity, a control valve through which oxygen may be added, and access ports for nursing care.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## K071175 JUL 1-7 2007 ### 510(k) Summary Submitter Information and Date Prepared Agata Smieja GE Healthcare 8880 Gorman Road Laurel, MD 20723 USA Phone: 410 888 5218 Mobile: 410 456 0329 Prepared July 16, 2007 Device Identification | Proprietary Name: | |-------------------| | Common Name: | Ohmeda Medical Giraffe OmniBed Giraffe OmniBed Predicate Device Information | Predicate Device | 510(k) Number | |-------------------------------------------------------------------------------------------|---------------| | Ohmeda Medical Giraffe OmniBed | K993407 | | Ohmeda Medical Giraffe OmniBed (with optional<br>Servo Controlled Oxygen Delivery System) | K020543 | Intended Use Statement The OmniBed is a combination of an infant incubator and an infant warmer. The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user's demand. It cannot be operated in both modes at the same {1}------------------------------------------------ time. Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Incubators provide an enclosed, temperature-controlled environment and warmers provide infrared heat in an open environment. They may also be used for short periods of time to facilitate the neonate's transition from the uterus to the external environment. This device may incorporate a Servo Controlled Oxygen Delivery System. This is indicated to provide stable oxygen concentration within the infant compartment at the value set by the operator (21-65%). #### Functional Description and Technological Characteristics Giraffe OmniBed is an infant bed, which provides thermal support to infants who are unable to thermo-regulate based on their own physiology. The bed has two modes of operation, enclosed bed operation and open bed operation. In an open mode, the device functions as an incubator and in an enclosed mode it functions as a conventional, infant radiant warmer. #### Functional description of the new features added to the device The Giraffe Uninterruptible Power Supply (UPS) is intended to provide a short term source of electrical power Giraffe OmniBed, thus aiding its intra hospital mobility. The Giraffe UPS does not change the indications for use, control mechanisms, operating principles, performance specifications, or other features of the Giraffe Omnibed. The UPS serves as an extension to the Giraffe OmniBed by providing uninterrupted electric power to the device. The UPS comprises a medical grade battery and a shelf. #### Performance Data The battery has a life of 15 minutes at 37° C and 70% RH at steady state. The recharge time is 6 hours. The addition of the UPS to the Giraffe OmniBed reduces the decibel level of the alarm at 3 meters in front of the equipment by less than 1 dB, within the minimum lower level of 50dB as specified by IEC 601-2-21 Amend 1 clause 102.1 and IEC-601-2-19 Amend 1 clause 102.3. {2}------------------------------------------------ 3 8+3 Performance of the OmniBed with the addition of UPS has been established by bench testing against product specifications and recognized consensus standards. | Prepared by: | | |--------------|--| | Date | | {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines representing its body and wings. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 ## JUL 1 7 2007 Ms. Agata Smieja Global Compliance Leader Ohmeda Medical, a Division of Datex Ohmeda, a GE Healthcare Company 8880 Gorman Road Laurel, Maryland 20723 Re: K071175 Trade/Device Name: Giraffe OmniBed Regulation Number: 21 CFR 880.5400 Regulation Name: Neonatal Incubator Regulatory Class: II Product Code: FMZ Dated: June 28, 2007 Received: June 29, 2007 Dear Ms. Smieja: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Ms. Smieja Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Chiu Lin, Ph.D. Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use 510(k) Number (if known): K071175 Device Name: Indications for Use: The OmniBed is a combination of an infant incubator and an infant warmer. The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user's demand. It cannot be operated in both modes at the same time. Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Incubators provide an enclosed, temperature-controlled environment and warmers provide infrared heat in an open environment. They may also be used for short periods of time to facilitate the neonate's transition from the uterus to the external environment. This device may incorporate a Servo Controlled Oxygen Delivery System. This is indicated to provide stable oxygen concentration within the infant compartment at the value set by the operator (21-65%). Prescription Use _ X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) S = 006 (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: (
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...