The Stryker Intraline Anchor is a soft tissue anchor which will be used for tissue fixation to bone in the shoulder, foot/ankle, knee, hand/wrist, elbow, and pelvis. The anchor is intended for use in the following procedures: Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift/Capsulolabral Reconstruction. Foot and Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Midfoot Reconstruction, Metatarsal Ligament Repair. Knee: Anterior Cruciate Ligament Repair, Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Illiotibial Band Tenodesis. Hand and Wrist: Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction. Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction. Pelvis: Bladder Neck Suspension Procedures. The Stryker Soft Eyelet RC Anchor is intended for single-use only.
Device Story
Stryker Intraline Anchor is a screw-in soft tissue anchor system; pre-threaded with nonabsorbable USP braided UHMWPE suture; pre-assembled on disposable inserter. Used by surgeons in orthopedic procedures (shoulder, foot/ankle, knee, hand/wrist, elbow, pelvis) for tissue-to-bone fixation. Device provides mechanical attachment point for sutures to secure soft tissue to bone. Single-use device.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
Materials: PEEK-OPTIMA or Titanium alloy (Ti 6Al 4V ELI). Design: Screw-in anchor with pre-threaded UHMWPE suture on disposable inserter. Sterilization: Ethylene oxide (SAL 10^-6). Standards: ISO 10993-1, ISO 10993-7, EN 550-1, EN 11607-1/2, EN 980, EN 1041, ISO 14971.
Indications for Use
Indicated for patients requiring soft tissue-to-bone fixation in shoulder, foot/ankle, knee, hand/wrist, elbow, and pelvic surgical procedures.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
K071157 page 1 of 2
Image /page/0/Picture/1 description: The image shows the word "stryker" in a bold, sans-serif font. The word is all lowercase, and the letters are closely spaced together. There is a registered trademark symbol to the upper right of the letter "r". The word is black and the background is white.
5900 Optical Ct San Jose, CA 95138 t: 408 754 2124 f: 408 754 2914 jeff.semone@stryker.com
### Endoscopy
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
SEP 1 7 2007
Date: April 2, 2007
# Contact Person:
K Jeffrey Semone Director, Regulatory Affairs 408-754-2124(phone) 408-754-2521 (fax) jeff.semone@stryker.com
## Device Name:
| Proprietary Names: | Stryker PEEK Intraline Anchor |
|-------------------------|-------------------------------------------------------------|
| | Stryker Titanium Intraline Anchor |
| Common and Usual Names: | Soft Eyelet Anchor |
| | PEEK Soft Eyelet Anchor |
| | Titanium Soft Eyelet Anchor |
| Classification Name: | Screw, Fastener, Fixation, Nonabsorbable, Bone, Soft Tissue |
| | (Class II, 21 CFR 888.3040, Product Code MBI, Orthopedics |
| | Review Panel) |
## Predicate Devices:
Arthrex 5.5mm PEEK Corkscrew FT: #K061665
## Device Description and Intended Use:
.
The Stryker Intraline Anchors are soft tissue anchors which will be used for tissue fixation to bone in the shoulder, foot/ankle, knee, hand/wrist, elbow, and pelvis. The anchors are intended for use in the following procedures:
This space is intentionally left blank
{1}------------------------------------------------
### Shoulder:
- Rotator Cuff Repair
- Bankart Repair
- SLAP Lesion Repair
- Biceps Tenodesis ■
- י Acromio-Clavicular Separation Repair
- Deltoid Repair
- Capsular Shift/Capsulolabral Reconstruction.
### Foot and Ankle:
- Lateral Stabilization
- . Medial Stabilization
- Achilles Tendon Repair
- Hallux Valgus Reconstruction
- Midfoot Reconstruction
- Metatarsal Ligament Repair.
### Knee:
- Anterior Cruciate Ligament Repair
- Medial Collateral Ligament Repair
- Lateral Collateral Ligament Repair
- . Patellar Tendon Repair
- Posterior Oblique Ligament Repair
- Illiotibial Band Tenodesis
### Hand and Wrist:
- Scapholunate Ligament Reconstruction
- Ulnar Collateral Ligament Reconstruction .
- Radial Collateral Ligament Reconstruction
#### Elbow
- Biceps Tendon Reattachment
- Ulnar or Radial Collateral Ligament Reconstruction
#### Pelvis:
- Bladder Neck Suspension Procedures
Each configuration of the Stryker Intraline Anchor Family is a screw-in anchor that is pre-threaded with nonabsorbable USP braided ultra high molecular weight polyethylene (UHMWPE) suture (K033654, K040472 and K063778) and pre-assembled on a disposable inserter. The Stryker PEEK Intraline Anchor will be manufactured from PEEK-OPTIMA® (polyetheretherketone). The Stryker Titanium Intraline Anchor will be manufactured from titanium alloy Ti 6Al 4V EL1 The Stryker Intraline Anchor Family will be validated to a SAL of 10th using ethylene oxide. The EtO residuals will be tested according to ISO 10993-7.
Prior to introducing the Stryker Intraline Anchor Family to market, the devices will conform to the following voluntary safety and performance standards: ISO 10993-1, Blue Book Memorandum G95-1, EN 550-1, EN 11607-1, EN 11607-2, EN 980, EN 1041, and EN ISO 14971.
The Stryker PEEK and Intraline Anchor is considered substantially equivalent in performance, material composition, intended use, safety, and efficacy to the Arthrex PEEK Corkscrew FT.
The Stryker Titanium Intraline Anchor is considered substantially equivalent in performance, intended use, safety, and efficacy to the Arthrex PEEK Corkscrew FT.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular fashion around the symbol. The text is in all caps and is evenly spaced around the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Stryker Endoscopy c/o Mr. K. Jeffrey Semone Director, Regulatory Affairs 5900 Optical Court San Jose, CA 95138
SEP 17 2007
Re: K071157
Trade/Device Name: Stryker PEEK and Titanium Intraline Anchor Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: HWC, JDR, MBI Dated: August 20, 2007 Received: August 22, 2007
Dear Mr. Semone:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Mr. K. Jeffrey Semone
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Barbara Buchr
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## INDICATIONS FOR USE
Device Name: Stryker Intraline Anchor
510(k) Number if known: K071157
The Stryker Intraline Anchor is a soft tissue anchor which will be used for tissue fixation to bone in the shoulder, foot/ankle, knee, hand/wrist, elbow, and pelvis. The anchor is intended for use in the following procedures:
Hand and Wrist:
Elbow
Pelvis:
Scapholunate Ligament Reconstruction
Bladder Neck Suspension Procedures.
Biceps Tendon Reattachment
Ulnar Collateral Ligament Reconstruction Radial Collateral Ligament Reconstruction
Ulnar or Radial Collateral Ligament Reconstruction
#### Shoulder:
- Rotator Cuff Repair
- Bankari Repair
- SLAP Lesion Repair
- Biceps Tenodesis
- Acromio-Clavicular Separation Repair
- Deltoid Repair
- Capsular Shift/Capsulolabral Reconstruction.
#### Foot and Ankle:
- Lateral Stabilization
- Medial Stabilization
- Achilles Tendon Repair
- Hallux Valgus Reconstruction
- Midfoot Reconstruction
- Metatarsal Ligament Repair.
#### Knee:
- Anterior Cruciate Ligament Repair
- Medial Collateral Ligament Repair
- Lateral Collateral Ligament Repair
- Patellar Tendon Repair
- Posterior Oblique Ligament Repair
- Illiotibial Band Tenodesis
The Stryker Soft Eyelet RC Anchor is intended for single-use only.
Prescription Use X (Part 21 CFR 801 Subpart D)
Over-The-Counter AND/OR (21 CFR 801 Subpart C)
\$\mu_o\$
Use
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Houban Buchnd
Division of General. Restorative. and Neurological Devices
510(k) Number k071157
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.