← Product Code [DZE](/productcode/DZE) · K071070

# OLYMPUS DENTAL IMPLANT SYSTEM (K071070)

_Keystone Dental, Inc. · DZE · Oct 23, 2007 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K071070

## Device Facts

- **Applicant:** Keystone Dental, Inc.
- **Product Code:** [DZE](/productcode/DZE.md)
- **Decision Date:** Oct 23, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3640
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

The Olympus Dental Implant System is intended to be surgically placed, either immediately after extraction or following healing, in the bone of the maxillary and/or mandibular arch to provide support for crowns, bridges or overdentures. Olympus Dental Implants with an insertion depth of 8 mm or more can be loaded immediately if they have achieved adequate primary stability.

## Device Story

Olympus Dental Implant System consists of parallel-sided, threaded, root-form dental implants and straight abutments. Implants are surgically placed into maxillary or mandibular bone by dental professionals to support prosthetic restorations (crowns, bridges, overdentures). System includes laboratory copings, impression copings, and analogs to facilitate prosthetic preparation. Device provides mechanical foundation for dental prosthetics; immediate loading is an option for implants ≥8 mm achieving primary stability. Benefits include restoration of oral function and aesthetics for edentulous or partially edentulous patients.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Parallel-sided, threaded, root-form dental implants and straight abutments. Materials and specific ASTM standards not disclosed in provided text. Mechanical fixation via surgical placement in bone.

## Regulatory Identification

An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.

## Special Controls

*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Olympus Dental Implant System

# 510(k) Summary

# 0CT 2 3 2007

# Olympus Dental Implant System

Linda Jalbert

#### ADMINISTRATIVE INFORMATION

| Manufacturer Name: | Keystone Dental, Inc.      |
|--------------------|----------------------------|
|                    | 3 Burlington Woods Drive   |
|                    | Burlington, MA 01803       |
|                    | Telephone 1 (781) 272-9272 |
|                    | Fax 1 (781) 272-9972       |

Official Contact:

Representative/Consultant:

Floyd G. Larson PaxMed International, LLC 11234 El Camino Real, Suite 200 San Diego, CA 92130 Telephone 1 (858) 792-1235 Fax 1 (858) 792-1236 email: flarson@paxmed.com

#### DEVICE NAME AND CLASSIFICATION

| Trade/Proprietary Name:     | Olympus Dental Implant System                                                                                                 |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| Common Name:                | dental implants and abutments                                                                                                 |
| Classification Regulations: | Endosseous dental implant<br>(21 CFR 872.3640), Class II<br>Endosseous dental implant abutment<br>(21 CFR 872.3630), Class II |
| Product Codes               | DZE, NHA                                                                                                                      |

#### DEVICE CLASSIFICATION PANEL

The Classification Panel for these devices is the Dental Products Panel, and they are reviewed by the Dental Devices Branch.

#### INTENDED USE

The Olympus Dental Implant System is intended to be surgically placed, either immediately after extraction or following healing, in the bone of the maxillary and/or mandibular arch to provide support for crowns, bridges or overdentures. Olympus Dental Implants with an insertion depth of 8 mm or more can be loaded immediately if they have achieved adequate primary stability.

{1}------------------------------------------------

#### DEVICE DESCRIPTION

The Olympus Dental Implant System includes various sizes of parallel sided, threaded, root-form dental implants and straight abutments intended to support prosthetic restorations in edentulous or partially edentulous patients. The implants can be placed immediately following extraction or after a healing period. If good primary stobility is reached according to the Instructions for Use, Olympus Dental Implants with an insertion depth of 8 mm or more may be immediately loaded. The system includes a variety of laboratory (burnout) copings, impression copings, analogs and other components (Class 1 exempt, not a subject of this submission) intended to facilitate the prevaration of prosthetic restorations.

### EQUIVALENCE TO MARKETED PRODUCT

Keystone Dental, Inc. demonstrated that, for the purposes of FDA's regulation of medical devices, the Olympus Dental Implant System is substantially equivalent in indications and design principles to predicate devices, each of which has been determined by FDA to be substantially equivalent to preamendment devices.

{2}------------------------------------------------

:

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 2 3 2007

Keystone Dental, Incorporated C/O Mr. Floyd G. Larson President PaxMed International, LLC 11234 EL Camino Real Suite 200 San Diego, California 92130

Re: K071070

Trade/Device Name: Olympus Dental Implant System Regulation Number: 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE Dated: October 1, 2007 Received: October 2, 2007

Dear Mr. Larson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

#### Page 2 -Mr. Larson

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Neil R.P. Oslin
Chiu Lin, Ph.D.

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known): K071070

Device Name:

Olympus Dental Implant System

Indications for Use:

The Olympus Dental Implant System is intended to be surgically placed, either immediately after extraction or following healing, in the bone of the maxillary and/or mandibular arch to provide support for crowns, bridges or overdentures. Olympus Dental Implants with an insertion depth of 8 mm or more can be loaded immediately if they have achieved adequate primary stability.

Prescription Use X (Part 21 CFR 801 Subpart D)

Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)

### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

|  | <img alt="Signature" src="signature.png"/> |
|--|--------------------------------------------|
|--|--------------------------------------------|

(ision Sign-Off)
Division of Anesthesiology, General Hospital,
Infection Control, Dental Devices

| 510(k) Number: | K071070 |
|----------------|---------|
|----------------|---------|

---

**Source:** [https://fda.innolitics.com/device/K071070](https://fda.innolitics.com/device/K071070)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
