Volume Interactions Pte Ltd's Image Processing System is a medical device for the display and visualization of 3D medical image data derived from tomographic radiology images, excluding mammography images. It is intended to be used by qualified and trained medical professionals, after proper installation.
Device Story
System processes DICOM 3.0 and other tomographic image data (CT, MRI) to generate 3D reconstructions; enables spatial manipulation, structure highlighting, and distance/volume measurements. Comprises three modules: RadioDexter (processing software), Dextroscope (interactive console with stereoscopic display for two-handed manipulation), and Dextrobeam (collaborative stereoscopic projection system). Used by neurosurgeons and specialists in clinical settings for surgical planning and collaborative discussion. Provides visual reference of white matter fiber tracks via DTI module. Output allows clinicians to interact with virtual 3D anatomy, facilitating precise hand-eye coordinated planning; benefits include improved surgical visualization and collaborative decision-making.
Clinical Evidence
No clinical data provided. Device safety and effectiveness were established through bench testing and verification/validation procedures for product design and development.
Technological Characteristics
PC-based system running Windows NT/2000. Uses industry-standard rendering methods and algorithms for 3D reconstruction. Hardware includes stereoscopic displays, mirrors, and projection systems for 3D interaction. Connectivity via DICOM 3.0. Software-based processing of tomographic data.
Indications for Use
Indicated for qualified and trained medical professionals for the display and visualization of 3D medical image data derived from tomographic radiology images (excluding mammography). Includes an add-on module for visualizing white matter anatomy via fiber tracks for neurosurgery planning.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Volume Interactions Bracco Group. The logo is black and white. The word "Volume" is in large, bold letters, with a hand pointing to the "o" in Volume, which is a sphere made of dots. Below Volume is the word "INTERACTIONS" in smaller, bold letters, and below that is "Bracco Group" in a smaller, non-bold font.
Tuvalu
K07/1054
| 510(k) Summary Of Safety and Effectiveness | | MAY 22 2007 |
|----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| RadioDexter™™, Dextroscope™ and Dextrobeam™™ | | |
| Submitter Information | | |
| Manufacturer: | Volume Interactions Pte Ltd<br>1 Kim Seng Promenade #12-01<br>Great World City East Tower<br>Singapore 237994<br>Phone: + 65 62226962 ext 232<br>Facsimile: + 65 62226215 | |
| Contact Person: | Michael Sim, Director, Worldwide Operations | |
| Summary Date: | 9th April 2007 | |
| Device Name<br>Common Name: | Image Processing System | |
| Trade Name:<br>& Model No. | RadioDexter™™ 1.0, Dextroscope™™ MK10, Dextrobeam™ MK3 &<br>MK4 | |
| Classification Name: | Picture Archiving and Communications System (per 21 CFR<br>892.2050) | |
| Classification Number: | LLZ | |
# Intended Use
Volume Interactions Pte Ltd's Image Processing System is a medical device for the display and visualization of 3D medical image data derived from tomographic radiology images, excluding mammography images. It is intended to be used by qualified and trained medical professionals, after proper installation.
#### Predicate device
| Trade Name & Model No.: | RadioDexter™ 1.0, Dextroscope™ MK10,<br>Dextrobeam™ MK3 & MK4 |
|-------------------------|---------------------------------------------------------------|
| 510(k) number: | K063730 |
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Volume Interactions, a Bracco Group company. The logo features the word "Volume" in a bold, sans-serif font, with a graphic of a hand touching a dotted sphere in place of the "o". Below "Volume", the word "INTERACTIONS" is written in a smaller, sans-serif font. Underneath "INTERACTIONS" is the text "Bracco Group".
#### Device Description
Volume Interactions Image Processing System reads DICOM 3.0 format medical image data sets (and other formats) and displays 3D image reconstructions of these data sets through various user selectable industry standard rendering methods and algorithms. The clinical users can spatially manipulate, process to highlight structures and volumes of interest, and measure distances and volumes in the 3D image reconstructions. The processed data can be stored either as 3D image data in a proprietary format, or as 2D picture projections of the 3D image data in TIFF image format. The system runs on commercially available PC compatible computers and hardware components with the Microsoft Windows NT and 2000 operating systems.
The system consists of three product modules namely, RadioDexter™, Dextroscope™ and Dextrobeam™. The modules are described as follows:
RadioDexter™ is software that processes tomographic (e.g.: Computer Tomography, Magnetic Resonance Imaging) data and produces stereoscopic 3D renderings for surgery planning and visualization purposes. The software uses various user selectable industry standard rendering methods and algorithms.
# DTI (Diffusion Tensor Imaging)
DTI (Diffusion Tensor Imaging) is an add-on module to RadioDexter™. This module allows the user to visualize white matter anatomy in the form of fiber tracks. The intended use of this module is to generate and provide a visual reference of white matter fiber tracks in a 3D virtual reality environment during the process of neurosurgery planning using the Dextroscope™ or during a discussion/collaboration using the Dextrobeam™. It is not intended to be used otherwise.
Dextroscope™ is an interactive console and display system that allows the user to interact with two hands with the 3D images generated by the RadioDexter™ software. The Dextroscope™ user works seated, with both forearms positioned on armrests. Wearing stereoscopic glasses, the user looks into a mirror and perceives the virtual image within comfortable reach of both hands for precise hand-eye coordinated manipulation. The hardware uses various industry standard components.
Dextrobeam™ is an interactive console intended for group collaborative discussions with 3D images using a stereoscopic projection system. The Dextrobeam™ system uses the base of the Dextroscope™ as the 3D interaction interface with the virtual objects. The monitor of the Dextroscope™ is replaced by a screen projection system, so instead of looking into the mirror of the Dextroscope™, the user looks at large stereoscopic screen projections while working with the virtual data in reach of his hands. This enables the discussion of 3D data sets with other specialists in stereoscopic 3D. The hardware uses various industry standard components.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a logo for Volume Interactions, a Bracco Group company. The logo features the word "Volume" in bold, with a hand pointing towards a sphere made of interconnected lines. Below "Volume" is the word "INTERACTIONS" in a smaller, sans-serif font, and below that is "Bracco Group" in a similar font.
# Substantial Equivalence
The Image Processing System in this submission remains unchanged as previously received 510(k)063730, in the following aspects:
- Use the same operating principle .
- Have the same technological characteristics .
- Incorporate similar basic software and hardware design .
- Have the same fundamental scientific technology .
The only differences are as follows:
#### Intended Use
The following statement has been removed from the intended use:
"Volume Interactions Pte Ltd's Image Processing System is not intended to be used in direct contact with the patient nor is it intended to be connected to equipment that is used in direct contact with the patient."
The device's input is derived from tomographic radiology images and the data format of the input is in either DICOM 3.0 or other common imaging formats (classic SGI formats. TIFF, Raw volume chunk or slices). It is clear that the input is not derived from patients directly which is why our company finds it redundant to highlight that there is no direct patient contact and connection to equipment that is used in direct contact with the patient. Removal of the statement does not affect the safety and effectiveness of the device as this does not change the system in terms of performance, design and technological characteristics.
#### DTI (Diffusion Tensor Imaging)
DTI (Diffusion Tensor Imaging) is an add-on module to RadioDexter™. This module allows the user to visualize white matter anatomy in the form of fiber tracks. The intended use of this module is to generate and provide a visual reference of white matter fiber tracks in a 3D virtual reality environment during the process of neurosurgery planning using the Dextroscope™ or during a discussion/collaboration using the Dextrobeam™. It is not intended to be used otherwise.
DTI has been verified and validated according to Volume Interactions' procedures for product design and development. The validation proves the safety and effectiveness of the module.
#### Conclusion
In summary, the Image Processing System (Dextroscope™, Dextrobeam™ with RadioDexter™ ) described in this submission are, in our opinion, substantially equivalent to the predicate device.
{3}------------------------------------------------
Image /page/3/Picture/13 description: The image is a seal for the Department of Health & Human Services (HHS). The seal features the HHS logo, which is a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the logo.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
Mr. Michael Sim Director, Worldwide Operations Volume Interactions Pte Ltd 1 Kim Seng Promenade #12-01 Great World City East Tower SINGAPORE 237994
MAY 2 2 2007
Re: K071054
Trade/Device Name: Image Processing System (Dextroscope™ MK10. Dextrobeam™ MK3 and MK4, and RadioDexter™ 1.0) Regulation Number: 21 CFR §892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: April 9, 2007 Received: April 16, 2007
Dear Mr. Sim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/3/Picture/10 description: The image is a circular logo or emblem. The text "1906-2006" is at the top of the logo. The letters "FDA" are in the center of the logo, with the word "Centennial" underneath. The text "FDA Centennial" is in a stylized font. The text "Health is Confidence Protected" is around the perimeter of the logo.
*Protecting and Promoting Public Health*
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address
http://www.fda.gov/cdrh/industry/support/index.html
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### INDICATION FOR USE STATEMENT
510(k) Number
# KD71054
#### Image Processing (Dextroscope™ MK10, Device Name System . . Dextrobeam™ MK3 and MK4, RadioDexter™ 1.0)
#### Indication for Use
Volume Interactions Pte Ltd's Image Processing System is a medical device for the display and visualization of 3D medical image data derived from tomographic radiology images, excluding mammography images. It is intended to be used by qualified and trained medical professionals, after proper installation.
Over-The-Counter Use Prescription Use AND/OR (21 CFR 801 Subpart C) (Per 21CFR 801 Subpart D) .
IPLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancyc brogdon
(Division Sign Off)
(Division Sign-Off Division of Reproductive, Abdominal, a Radiological Devices 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.