IFS BONE FIXATION DEVICES

K071035 · Internal Fixation Systems, Inc. · HWC · Aug 13, 2007 · Orthopedic

Device Facts

Record IDK071035
Device NameIFS BONE FIXATION DEVICES
ApplicantInternal Fixation Systems, Inc.
Product CodeHWC · Orthopedic
Decision DateAug 13, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The IFS Bone Fixation Devices are indicated for use in fixation of bone fractures, for bone reconstructions, as guide pins for insertion of other implants or implantation through the skin so that traction may be applied to the skeletal system.

Device Story

IFS Bone Fixation Devices consist of metallic cancellous, malleolar, and cortical screws (lengths 6-160mm; diameters 1.5-7.3mm) and guide wires (threaded, spaded, or blunt; diameters 0.028-0.188 inches; lengths 4-12 inches). Used by surgeons for orthopedic bone fixation and skeletal traction. Devices are distributed non-sterile and require end-user steam sterilization. No software or electronic components.

Clinical Evidence

Bench testing only.

Technological Characteristics

Materials: 316L stainless steel. Form factor: Screws (cancellous, malleolar, cortical) and guide wires (threaded, spaded, blunt). Sterilization: Non-sterile, intended for end-user steam sterilization per AAMI/ANSI/ISO 11737 guidelines to achieve SAL 10-6.

Indications for Use

Indicated for fixation of bone fractures, bone reconstructions, use as guide pins for other implants, or percutaneous implantation for skeletal traction. Not indicated for spinal fixation.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for Internal Fixation Systems (IFS). The logo consists of a circular graphic to the left of the letters "IFS". The words "Internal Fixation Systems" are printed below the graphic and letters. KO71035 pg 1/2 ## Internal Fixation Systems, Inc. 510(k) Summary AUG 1 3 2007 | Company Name: | Internal Fixation Systems, Inc.<br>10100 N.W. 116th Way<br>Miami, Florida 33178 | |--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Name: | Steve Hernandez<br>10100 N.W. 116th Way<br>Miami, Florida 33178<br>(305) 216-4766 | | Trade Name: | IFS Bone Fixation Devices | | Common Name: | Bone Fixation Fasteners | | Regulation Name: | Smooth or Threaded Metallic Bone Fixation Fastener | | Regulation<br>Number: | 21 CFR 888.3040 | | Regulatory Class: | II | | Device Product<br>Code: | HWC | | Substantially<br>Equivalent Devices: | Depuy Kirschner Wires and Steinmann Pins (K960385)<br>Plus Cancellous Bone Screws (K011719)<br>Biodynamic Technologies, Inc. (K972403)<br>GSO Bone Fixation Fasteners (K063589)<br>Nexa Bone Screw System (K053394)<br>Synthes 3.5 Cortex Screws (K043185)<br>Synthes 7.0/7.3 Cannulated Screws (K962011)<br>OsteoMed 1.2mm Auto-Drive Screw System (K023260) | | Device Description: | The IFS Bone Fixation Devices consist of cancellous,<br>malleolar, or cortical screws ranging from 6 to 160mm in<br>length and 1.5mm to 7.3mm in diameter. Threaded, spaded or<br>blunt guide wires range from 0.028 to 0.188 of an inch in<br>diameter and 4 to 12 inches in length. All IFS Bone Fixation | May 27, 2007 Page 1 of 2 {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for Internal Fixation Systems (IFS). The logo consists of a circular graphic to the left of the letters IFS. The graphic contains three arrows pointing in a circular direction. The text "Internal Fixation Systems" is located below the graphic and letters. Devices are manufactured using 316L stainless steel. Intended Use: The IFS Bone Fixation Devices are indicated for use in fixation of bone fractures, for bone reconstructions, as guide pins for insertion of other implants or implantation through the skin so that traction may be applied to the skeletal system. Technological The IFS Bone Fixation Devices are substantially equivalent to Characteristics the predicate devices with respect to design and material. Comparison: Sterilization IFS Bone Fixation Devices will be distributed non-sterile. The Information: devices are sterilized by the end user per the AAMI Guidelines "Good Hospital Practice: Steam Sterilization and Sterility Assurance" and ANSI/AAMI/ISO 11737 guidelines to achieve the Sterility Assurance Level (SAL) of 10-6. Conclusion: There are no significant differences between bone fixation devices and the other devices as listed in the Substantially Equivalent Devices. The IFS Bone Fixation Devices and the predicate devices have similar design attributes, material, and intended use thus is substantially equivalent. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo features a stylized eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle. AUG 1 3 2007 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Public Health Service Internal Fixation Systems, Inc. % Mr. Steve Hernandez President 10100 NW 116th Way, Ste. 18 Miami, Florida 33178 Re: K071035 Trade/Device Name: IFS Bone Fixation Devices Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: II Product Code: HWC Dated: June 19, 2007 Received: June 21, 2007 Dear Mr. Hernandez: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 - Mr. Steve Hernandez This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Barbara Brelup Mark N. Melker Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## 40710 ## Indications for Use 510(k) Number (if known): Device Name: IFS Bone Fixation Devices Indications for Use: The IFS Bone Fixation Devices are indicated for use in fixation of bone fractures, for bone reconstructions, as guide pins for insertion of other implants or implantation through the skin so that traction may be applied to the skeletal system. ***Warning: IFS Bone Fixation Devices are not indicated for spinal fixation.*** Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) 510(k) Number____<
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...