INION BIORESTORE

K070998 · Inion OY · MQV · Oct 11, 2007 · Orthopedic

Device Facts

Record IDK070998
Device NameINION BIORESTORE
ApplicantInion OY
Product CodeMQV · Orthopedic
Decision DateOct 11, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3045
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The INION BIORESTORE™ implants are bone graft substitutes indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. INION BIORESTORE™ is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., the extremities, spine and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process.

Device Story

Inion BioRestore™ is a synthetic, non-collagenous, radiopaque, degradable bioactive glass bone graft substitute. Available as cylinders, blocks, and morsels; implanted by surgeons into non-structural bony voids or gaps in the extremities, spine, or pelvis. Principle of operation: material undergoes surface kinetic modification upon implantation, forming a calcium phosphate layer similar to bone hydroxyapatite; this layer acts as a scaffold for new bone growth. Material is osteoconductive and osteostimulative, releasing ionic dissolution products that actively stimulate osteoblast proliferation and differentiation. Device resorbs in vivo, typically within six months, and is replaced by host bone. Intended for single-use; provided sterile.

Clinical Evidence

No clinical data on humans regarding rate and extent of bone formation. Evidence limited to in vitro cell culture tests (human adipose mesenchymal stem cells) demonstrating osteostimulative effects (alkaline phosphatase activity) and in vivo animal tests demonstrating bone formation compared to hydroxyapatite.

Technological Characteristics

Synthetic, non-collagenous, degradable bioactive glass. Form factors: cylinders, blocks, morsels. Radiopaque. Sterile, single-use. Osteoconductive and osteostimulative properties via ionic dissolution products.

Indications for Use

Indicated for patients requiring bone void filling in the skeletal system (extremities, spine, pelvis) for non-structural bony voids or gaps resulting from surgery or trauma.

Regulatory Classification

Identification

A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ ## 510(k) SUMMARY # Inion BioRestore™ 0CT 1 1 2007 #### Manufacturer and submitter Inion Oy, Lääkärinkatu 2, FIN-33520 Tampere, FINLAND #### Contact Person Kati Marttinen, Regulatory Affairs Specialist Phone: +358 10 830 6600 Fax: +358 10 830 6691 kati.marttınen@inion.com # Establishment registration number 9710629 Trade name of the device Inion BioRestore™ ## Device classification and product code Class II (Special controls) Classification Panel: Orthopedic Product Code: MQV Common name: Osteoconductive and osteostimulative bone void filler; Synthetic resorbable bone graft material Regulation number: 888.3045 #### Predicate devices NovaBone ® (K000149) NovaBone ® Resorbable Bone Graft Substitute (K021336) NovaBone ® Porous Bone Graft Scaffold (K060432) NovaBone ® Resorbable bone graft substitute (K052494) PerioGlass ® Bioglass ® Bone Graft Particulate (K040278) #### Conformance with performance standards No applicable mandatory performance standards exist for this device. Compliance to voluntary consensus standards is listed in the application. {2}------------------------------------------------ # 510(k) SUMMARY Inion BioRestore™ ## Device description and principles of operation Inion BioRestore™ is an osteoconductive and osteostimulative bioactive bone graft device. In vivo tests have demonstrated more bone formation at each post-implantation timepoint and more total bone formation compared to other osteoconductive devices such as hydroxyapatite. In vitro cell culture tests with human adipose mesenchymal stem cells have demonstrated an osteostimulative effect, defined as the active stimulation of osteoblast proliferation and differentiation as evidenced by alkaline phosphatase activity. This stimulation has been attributed as being the result of the interaction between osteoblasts and the ionic dissolution products released from Inion BioRestore™ during its absorption. Clinical data on humans on rate and extent of bone formation observed in cell culture and animal models has not been established. Bioactive materials are those materials that elicit a specific biological response at the interface of the material that results in the formation of a bond between the tissues and the material. Osteostimulation is a "... property of some bioactive materials to enhance, actively stimulate both the proliferation and differentiation of progenitor cells (e.g. mesenchymal stem cells) ... "1.2 Inion BioRestore™ system consists of different size cylinders, blocks and morsels made of degradable bioactive glass. When implanted, a kinetic modification of the surface occurs, resulting in the formation of a calcium phosphate layer that is essentially similar in composition and structure to the hydroxyapatite found in bone mineral. This apatite layer provides scaffolding onto which the patient's new bone will grow allowing complete repair of the defect. Based on pre-clinical testing, most of the material degrades in vivo in six months. The material is radiopaque. Inion BioRestore™ implants are intended for single use and are provided sterile to the user. They are completely synthetic and noncollagenous. # Indications for use The Inion BioRestore™ implants are bone graft substitutes indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. Inion BioRestore™ is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., the extremities, spine and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process. <sup>1</sup> Society for Biomaterials presentation, 24th Annual Meeting, 1998:511-518 <sup>2</sup> J Appl Biomat 1992, 3:123-129 {3}------------------------------------------------ # 510(k) SUMMARY # Inion BioRestore™ ## Substantial equivalence to marketed products Based on the performance data and specifications presented, it can be concluded that the intended use, material composition and scientific technology, degradation properties, bioactive, osteoconductive and osteostimulative properties of Inion BioRestore™ implants are substantially equivalent with the predicate Class II devices, when used in the indications for use described above, because the differences between Inion BioRestore™ and the predicate devices do not raise new questions of safety and effectiveness. {4}------------------------------------------------ Image /page/3/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, clutching a staff with a snake wrapped around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 OCT 1 1 2007 Inion Oy % Ms. Kati Marttinen Regulatory Affairs Specialist Lääkärinkatu 2 33520Tampere, Finland Re: K070998 Trade/Device Name: BioRestore™ Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler Regulatory Class: Class II Product Code: MQV Dated: August 6, 2007 Received: August 9, 2007 Dear Ms. Marttinen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {5}------------------------------------------------ #### Page 2 - Ms. Kati Marttinen This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Sarbay Buehum Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ # Statement of Indications for Use 510(k) Number: K070998 Device Name: Inion BioRestore™ The INION BIORESTORE™ implants are bone graft substitutes indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. INION BIORESTORE™ is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., the extremities, spine and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) currence of CDRH, Office of Device Evaluation (ODE) Arbera Knerp Division of General, Restorative, and Neurological Devices Page 1 of 1 510(k) Number K070998
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...