← Product Code [CAF](/productcode/CAF) · K070948

# HSINER NEBULIZER BOTTLE (K070948)

_Hsiner · CAF · Apr 25, 2007 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K070948

## Device Facts

- **Applicant:** Hsiner
- **Product Code:** [CAF](/productcode/CAF.md)
- **Decision Date:** Apr 25, 2007
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 868.5630
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The Hsiner Nebulizer Bottle is used to administer various aerosol treatments to adult and pediatric patients in both the homecare and hospital settings. This device is intended for use only with FDA-approved drugs upon the specific direction by a physician. Its use is indicated whenever a physician or healthcare professional administers or prescribes medical aerosol products to a patient using a Small Volume Nebulizer.

## Device Story

The Hsiner Nebulizer Bottle is a small volume nebulizer used in homecare and hospital settings to deliver aerosolized medication. It operates via a compressed gas source; the venturi principle draws liquid medication from a refillable jar, which is then aerosolized into respirable particles through impaction and baffling. The device is operated by healthcare professionals or patients under physician direction. It does not contain medication and is intended for use with FDA-approved drugs. The output is a mist of aerosolized liquid delivered directly to the patient for inhalation, facilitating the administration of prescribed respiratory treatments.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Small volume nebulizer; operates via venturi principle using compressed gas source; utilizes impaction and baffling for aerosolization; refillable jar design; intended for adult and pediatric use.

## Regulatory Identification

A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.

## Predicate Devices

- Hsiner Jet Nebulizer ([K052811](/device/K052811.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Hsiner Company Modified Jet Nebulizer 510(k) Submission

<070948

# 6. 510(k) Summary

In accordance with 21 CFR section 807.92 Hsiner is submitting the following 510(k) summary.

APR 2 5 2007

#### 6.1. Submitter Information

Hsiner Company, LTD No.312, Jhongshan Rd., Shengang Township, Taichung County 429 Taiwan, ROC

Phone: +86-4-25152480

Registration No .: 3003862188 Owner/Operator No .: 9053474

#### 6.2. Name of Device

| Proprietary Name:    | Nebulizer Bottle                     |
|----------------------|--------------------------------------|
| Common Name:         | Nebulizer                            |
| Classification Name: | Nebulizer (direct patient interface) |
| Product Code:        | CAF                                  |

Regulation Number: 868.5630 Device Class 2

#### 6.3. Substantially equivalent to:

- 0 Hsiner Jet Nebulizer (K052811)
## 6.4. Description of the device

The Hsiner Nebulizer Bottle is used to administer various aerosol treatments in both the homecare and hospital settings. This device is intended to only be use with FDA-approved drugs upon the specific direction of a physician. This device is not used specific drug nor is it distributed with such drugs.

The nebulizer sprays respiratory size aerosolized liquids into gasses that are delivered directly to the patient for breathing. The nebulizer operates on a compressed gas source which draws liguids from a refillable Jar by the venturi principle and aerosolizes it into respirable particles by impaction and baffling.

## 6.5. Intended Use of the Device

The Hsiner Nebulizer Bottle is used to administer various aerosol treatments to adult and pediatric patients in both the homecare and hospital settings. This device is intended for use only with FDAapproved drugs upon the specific direction by a physician. Its use is indicated whenever a physician or healthcare professional administers or prescribes medical aerosol products to a patient using a Small Volume Nebulizer.

## 6.6. Comparison to Predicate Devices

The Hsiner Nebulizer Bottle is equivalent in design, materials and performance to the Hsiner Jet Nebulizer and utilizes the same principles of operation and has the same intended use.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features the department's name in a circular arrangement around the perimeter. In the center of the seal is the HHS logo, which consists of a stylized caduceus symbol. The caduceus is a staff with two snakes coiled around it, and it is a symbol of medicine and health care. The seal is a symbol of the department's mission to protect the health of all Americans and provide essential human services.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Hsiner Company C/O Mr. Tom Shanks Principal MDVentures 29201 Via Norte Temecula, California 92591

APR 2 5 2007

Re: K070948

Trade/Device Name: Hsiner Nebulizer Bottle Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: II Product Code: CAF Dated: April 1, 2007 Received: April 4, 2007

Dear Mr. Shanks:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

# Page 2 - Mr. Shanks

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Susa Rusone

Chiu Lin. Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Hsiner Company Modified Jet Nebulizer 510(k) Submission

# Indications for Use

510(k) Number (**K**^070948):

Hsiner Nebulizer Bottle Device Name:

Indications for Use:

The Hsiner Nebulizer Bottle is used to administer various aerosol treatments to adult and pediatric patients in both the homecare and hospital settings. This device is intended for use only with FDA-approved drugs upon the specific direction by a physician. Its use is indicated whenever a physician or healthcare professional administers or prescribes medical aerosol products to a patient using a Small Volume Nebulizer.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Chl Udo

rion of Anactheriology, General Hospital Jon Control, Central Devices

'(k) Number

Page 1 of 1

Page 10

---

**Source:** [https://fda.innolitics.com/device/K070948](https://fda.innolitics.com/device/K070948)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
