K070931 · Promedtek, Inc. · ILX · May 24, 2007 · Physical Medicine
Device Facts
Record ID
K070931
Device Name
MODEL PMT850
Applicant
Promedtek, Inc.
Product Code
ILX · Physical Medicine
Decision Date
May 24, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5290
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Model PMT850 is indicated for adjunctive use in the palliative treatment of postoperative edema and pain in superficial soft tissue.
Device Story
Model PMT850 is a pulsed shortwave diathermy device; consists of a generating unit connected via cable to a treatment applicator. Operates by delivering pulsed shortwave radiofrequency (27.12 MHz) energy to superficial soft tissue. Used in clinical settings; operated by healthcare providers. Provider places applicator over standard dressing on treatment area; selects one of 36 combinations of 6 pulse frequencies and 6 power settings via generating unit dials. Device provides non-thermal energy to assist in pain and edema management. Benefits include adjunctive palliative care for postoperative patients. Performance verified via applicator field measurements and output energy comparisons against predicate.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including applicator field measurements and output energy comparisons against the predicate device to demonstrate equivalent performance. Biocompatibility testing (ISO 10993-5) performed on applicator foam material.
Indicated for adjunctive palliative treatment of postoperative edema and pain in superficial soft tissue.
Regulatory Classification
Identification
A nonthermal shortwave therapy is a prescription device that applies to the body pulsed electromagnetic energy in the RF bands of 13.56 MHz or 27.12 MHz and that is intended for adjunctive use in the palliative treatment of postoperative pain and edema of soft tissue by means other than the generation of deep heat within body tissues as described in paragraph (a) of this section.
Special Controls
*Classification: Class II (special controls).* The device is classified as class II. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible.
(ii) Appropriate analysis/testing must demonstrate that the device is electrically safe and electromagnetically compatible in its intended use environment.
(iii) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Non-clinical performance testing must characterize the output waveform of the device and demonstrate that the device meets appropriate output performance specifications. The output characteristics and the methods used to determine these characteristics, including the following, must be determined:
(A) Peak output power;
(B) Pulse width;
(C) Pulse frequency;
(D) Duty cycle;
(E) Characteristics of other types of modulation that may be used;
(F) Average measured output powered into the RF antenna/applicator;
(G) Specific absorption rates in saline gel test load or other appropriate model;
(H) Characterization of the electrical and magnetic fields in saline gel test load or other appropriate model for each RF antenna and prescribed RF antenna orientation/position; and
(I) Characterization of the deposited energy density in saline gel test load or other appropriate model.
(iv) A detailed summary of the clinical testing pertinent to use of the device to demonstrate the effectiveness of the device in its intended use.
(v) Labeling must include the following:
(A) Output characteristics of the device;
(B) Recommended treatment regimes, including duration of use; and
(C) A detailed summary of the clinical testing pertinent to the use of the device and a summary of the adverse events and complications.
(vi) Nonthermal shortwave therapy devices marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with these special controls.
{0}------------------------------------------------
A final order reclassifying shortwave diathermy (SWD) intended for adjunctive use in the palliative treatment of postoperative pain and edema of soft tissue by means other than the generation of deep heat within body tissues, a preamendments Class III device, into class II, and renaming the device "nonthermal shortwave therapy" (SWT), was published on October 13, 2015. See here: https://www.federalregister.gov/documents/2015/10/13/2015-25923/physical-medicine-devices-reclassification-of-shortwave-
diathermy-for-all-other-uses-henceforth-to
While the device submitted and cleared through K070931 may serve as a valid predicate device for a new SWT device, please refer to the aforementioned final order for current regulatory requirements for this device type.
{1}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
ProMedTek, Inc. % The Weinberg Group, Inc. Diane Mandell, Ph.D., RAC Senior Consultant 1220 Nineteenth St NW, Suite 300 Washington, District of Columbia 20036
MAY 2 4 2007
Re: K070931
Trade/Device Name: Model PMT850 Regulation Number: 21 CFR 890.5290 Regulation Name: Shortwave diathermy Regulatory Class: Class III Product Code: ILX Dated: April 3, 2007 Received: April 3, 2007
Dear Dr. Mandell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
## Page 2 - Dr. Diane Mandell
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally prematicated predicated. The PDF intembly sification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please n you uchie specific acriviance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general informational and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark A. Milhem
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## INDICATIONS FOR USE
510(K) Number (if known): K070 931
Device Name: Model PMT850
Indications for Use:
The Model PMT850 is indicated for adjunctive use in the palliative treatment of postoperative edema and pain in superficial soft tissue.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mark N. Millhusan
(Division Sign-Off) Division of General, Restorative. and Neurological Devices
**510(k) Number** K07 093
Page 1 of 1
ProMedTek Premarket Notification Model PMT850
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the text "K070931" above the word "PROMEDTEK". The word "PROMEDTEK" is in bold, and there is a graphic of a person with a globe for a head above the letters "MEDTEK". The text "K070931" appears to be handwritten.
# 510(k) Summary
# Model PMT850
MAY 2 4 2007
| Name of Device: | Model PMT850 |
|-----------------------|------------------------------------------------------------------------------------|
| Common Name: | Shortwave Diathermy |
| Classification Name: | 21 CFR 890.5290(b), Class III |
| Product Code: | ILX |
| Classification Panel: | Physical Medicines Devices Panel |
| Sponsor: | ProMedTek, Inc.<br>7272 East Indian School Road, Suite 102<br>Scottsdale, AZ 85251 |
| Contact: | Daniel R. Puchek |
| Tele: | 480-385-2468 |
| Fax: | 480-385-2499 |
| Date Prepared: | May 14, 2007 |
#### LEGALLY MARKETED PREDICATE DEVICE A.
K903675. MRT sofPulse shortwave diathermy device, ElectroPharmacology, Inc.
#### DEVICE DESCRIPTION B.
The Model PMT850 device is a pulsed shortwave diathermy device. The Model PMT850 system consists of a Generating Unit connected by a cable to the Treatment Applicator. The Treatment Applicator and cable are a single unit which is connected or disconnected easily from the Generating Unit. The Generating Unit connects to a standard AC electrical outlet by a grounded 3-prong UL/CSA approved, medically rated electric cord.
Treatment with the Model PMT850 requires placing the Treatment Applicator on top of any standard dressing over the area to be treated, setting the dials on the Generating Unit, and turning the device on. There are 6 possible pulse frequency and 6 possible power control settings on the generating unit of the Model PMT850, providing 36 setting options for treatment. Suggested treatment settings are discussed in the Operating Manual for the Model PMT850.
#### INTENDED USE ்.
The Model PMT850 is indicated for adjunctive use in the palliative treatment of postoperative edema and pain in superficial soft tissue.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "PROMEDTEK" in a bold, sans-serif font. The "O" in "PROMEDTEK" is replaced with a globe graphic. Above the globe is a curved line with a dot above it.
#### SUBSTANTIAL EQUIVALENCE SUMMARY D.
The Model PMT850 has the same intended use and indications for use statement as the predicate device. The Model PMT850 has the same technological characteristics as the predicate device, with the exception of the lower power required from the Generating Unit to provide an equivalent amount of Peak Power from the Treatment Applicator, which does not affect safety or effectiveness. In addition to a side-by-side comparison of the descriptive characteristics of the Model PMT850 and the predicate presented below, performance studies were conducted to ensure equivalent performance compared with the predicate: Applicator Field Measurements, and Output Energy Comparison. The decision algorithm brings us to a determination of device substantial equivalence with the predicate.
#### TECHNOLOGICAL CHARACTERISTICS ប៉
The general technological characteristics of the Model PMT850 and the predicate are equivalent. This includes: power supply, operating principle, energy transfer mechanism, treatment applicator, device physical conformation (weight and size), and packaging.
The following technical specifications are also found equivalent between the Model PMT850 and the predicate: power required, fuse protection, operating frequency, RF Output Impedance, RF Pulse Width, RF Pulse waveform, pulse frequency, pulse duty cycle, pulse control accuracy, peak RF applicator power, average RF applicator output power, exposure timer, and timer accuracy (see Table 1 below). The only technical specification that differs is the lower peak RF generator power required for the Model PMT850 Generating Unit to provide an equivalent amount of Peak Power as the predicate from the Treatment Applicator for each of the six power settings.
| | ProMedTek Inc.<br>MODEL PMT850 | Predicate Device<br>MRT sofPulse |
|------------------------|------------------------------------|------------------------------------|
| Power required | 110 VAC, 60 Hz | 110 VAC, 60 Hz |
| Fuse protection | AC: 1 Amp, 250 V | AC: 1 Amp, 250 V |
| Operating frequency | Short Wave RF<br>27.12 MHz ± 0.05% | Short Wave RF<br>27.12 MHz ± 0.05% |
| RF Output Impedance | 50 Ohms | 50 Ohms |
| RF Pulse width | 65 µsec | 65 µsec |
| RF Pulse waveform | 27 MHz Sine Wave Envelope | 27 MHz Sine Wave Envelope |
| Pulse control accuracy | 0.0025% | Not reported |
| Exposure timer | 0 - 30 minutes | 0 - 30 minutes |
| Timer accuracy | 0.001% | 1% |
## Table 1. COMPARISON OF TECHNICAL SPECIFICTIONS
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "PROMEDTEK" in a bold, sans-serif font. The "O" in "PROMEDTEK" is replaced by a globe with lines of latitude and longitude. Above the globe is a curved line with a dot above it, resembling a person with their arm raised.
#### г. TESTING
## Substantial Equivalence
Applicator Field Measurements and Output Energy Comparison testing was conducted by two independent laboratories to determine equivalence in device performance of the Model PMT850 and the predicate device. The results of these tests support the substantial equivalence of the Model PMT850 with the predicate device.
### Conformance to Standards
Prior to marketing, the Model PMT850 will comply with all standard requirements listed by an NRTL and also will conform to electrical safety (UL 60601-1), electromagnetic compatibility (IEC 60601-1-2) and packaging guidelines (ASTM D4169-05).
Biocompatibility testing was determined not to be required because the device does not contact patient skin. However, it should be noted that the treatment applicator covering is made of a high-grade polyethylene foam, which has passed biocompatibility tests for cytotoxicity/irritancy (ISO 10993-5).
#### CONCLUSIONS G.
ProMedTek, Inc. has demonstrated through the comparison of characteristics and comparison of performance testing, that the Model PMT850 is substantially equivalent to the predicate, MRT sofPulse shortwave diathermy device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.