K070910 · Sysmex America, Inc. · LKM · May 25, 2007 · Hematology
Device Facts
Record ID
K070910
Device Name
SYSMEX MODEL UF1000I
Applicant
Sysmex America, Inc.
Product Code
LKM · Hematology
Decision Date
May 25, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.5200
Device Class
Class 2
Attributes
AI/ML
Indications for Use
The Sysmex® UF-1000i is an automated urine particle analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories. The UF-1000i analyzes the following parameters in urine samples: RBC, WBC, Epithelial Cells, Cast, and Bacteria and flags the presence of the following: Pathologic Cast, Crystal, Sperm, Small Round Cell, Yeast like cell and Mucus.
Device Story
The Sysmex UF-1000i is an automated urine particle analyzer used in clinical laboratories to analyze microscopic formed elements in urine. The system comprises a main unit for aspiration, dilution, mixing, and analysis; an auto-sampler for up to 50 tubes; and an Information Processing Unit (IPU) for data processing and operator interface. The device utilizes flow cytometry with a red semiconductor laser to analyze urine elements. Inputs include urine samples; the system transforms these via forward scatter, fluorescence, and side-scattered light signals (using a dedicated bacteria channel) into classified particle counts. Adaptive cluster analysis identifies and classifies elements. Results are displayed on the IPU, printed, or transmitted to a host computer. By automating the identification of RBCs, WBCs, bacteria, and other elements, the device assists clinicians in screening for urinary tract conditions, improving laboratory efficiency and diagnostic throughput.
Clinical Evidence
Bench testing only. Performance was established through comparison to the predicate UF-100. Data provided includes carryover, linearity, accuracy, and reproducibility studies, which demonstrated excellent correlation with the predicate device and met manufacturer specifications.
Technological Characteristics
Flow cytometry; red semiconductor laser; automated aspiration/dilution/mixing. Parameters: RBC, WBC, Epithelial Cells, Casts, Bacteria. Connectivity: IPU for data processing/operator interface. Software: Moderate Level of Concern. Calibration: Factory-set (UFII Calibrator). QC: UFII Control (2 levels). Throughput: 100 samples/hour.
Indications for Use
Indicated for in vitro diagnostic screening of urine samples in clinical laboratory settings to analyze RBC, WBC, Epithelial Cells, Casts, and Bacteria, and to flag Pathologic Casts, Crystals, Sperm, Small Round Cells, Yeast-like cells, and Mucus.
Regulatory Classification
Identification
An automated cell counter is a fully-automated or semi-automated device used to count red blood cells, white blood cells, or blood platelets using a sample of the patient's peripheral blood (blood circulating in one of the body's extremities, such as the arm). These devices may also measure hemoglobin or hematocrit and may also calculate or measure one or more of the red cell indices (the erythrocyte mean corpuscular volume, the mean corpuscular hemoglobin, or the mean corpuscular hemoglobin concentration). These devices may use either an electronic particle counting method or an optical counting method.
{1}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY INSTRUMENT ONLY TEMPLATE
A. 510(k) Number:
K070910
B. Purpose for Submission:
This traditional 510(k) is for a modification of a previously cleared device.
C. Manufacturer and Instrument Name:
Sysmex America, Inc., Sysmex® UF-1000i, Automated Urine Particle Analyzer
D. Type of Test or Tests Performed:
The device is a dedicated system for the analysis of microscopic formed elements in urine specimens. The parameters are: RBC, WBC, Epithelial Cells, Casts, Bacteria, and flags the presence of Casts, Crystals, Sperm, Small Round Cells, Yeast like cells, and Mucus.
E. System Descriptions:
1. Device Description:
The Sysmex UF-1000i, an automated urine particle analyzer, is a dedicated system for the analysis of microscopic formed elements in urine specimens. The instrument consists of three principal units; (1) Main unit which aspirates, dilutes, mixes analyzes urine samples; (2) Auto Sampler Unit supplies samples to the Main Unit automatically; (3) IPU (Information Processing Unit) which processes data from the Main Unit and provides the operator interface with the system. The UF-1000i is equipped with a Sampler that provides continuous automated sampling for up to 50 tubes.
2. Principles of Operation:
The instrument utilizes Sysmex flow cytometry using a red semiconductor laser for analyzing organized elements of urine particle characterization and identification is based on detection of forward scatter, fluorescence, and adaptive cluster analysis. Using its own reagents, the UF1000i automatically classifies organized elements of urine and carries out all processes automatically from aspiration of the sample to outputting the results.
{2}
3. **Modes of Operation:**
Manual Mode Measurement and Sampler Mode Measurement
4. **Specimen Identification:**
Bar Code
5. **Specimen Sampling and Handling:**
Automated sampling and mixing with a throughput of 100 samples/hour.
6. **Calibration:**
UFII Calibrator. Calibration is done at the factory. There is no user calibration.
7. **Quality Control:**
UFII Control (2 levels), QC programs includes L-J management and X-bar management.
8. **Software:**
FDA has reviewed applicant’s Hazard Analysis and Software Development processes for this line of product types:
☑ Yes ☐ No Software documentation at a Moderate Level of Concern conforms to the FDA software guidance.
F. **Regulatory Information:**
1. **Regulation Section:**
21 CFR 864.5200 Automated cell counter
2. **Classification:**
Class II
3. **Product code:**
LKM
4. **Panel:**
Hematology (81)
{3}
G. Intended Use:
1. Indication(s) for Use:
Sysmex UF-1000i is an automated urine particle analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories. The UF-1000i analyzes the following parameters in urine samples: RBC, WBC, Epithelial Cells, Cast, and Bacteria and flags the presence of the following: Pathologic Cast, Crystal, Sperm, Small Round Cell, Yeast-like cell and Mucus.
2. Special Conditions for Use Statement(s):
N/A
H. Substantial Equivalence Information:
1. Predicate Device Name(s) and 510(k) numbers:
Sysmex UF-100 (K961054)
2. Comparison with Predicate Device:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Specimen Type | Random urine sample | Same |
| Throughput | 100 samples/ hour | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Parameters | RBC, WBC, Epithelial Cells, Cast, and Bacteria and flags the presence of the following: Pathologic Cast, Crystal, Sperm, Small Round Cell, Yeast-like cell. | RBC, WBC, Epithelial Cells, Cast, and Bacteria and flags the presence of the following: Pathologic Cast, Crystal, Sperm, Small Round Cell, Yeast-like cell. |
| Methodology | The device utilizes Sysmex flow cytometry using a red semiconductor laser for analyzing organized elements or urine. Particle characterization and identification is | The UF100 utilizes Sysmex flow cytometry using an argon laser for analyzing organized elements of urine. In combination with flow cytometry the UF-100 uses an impedance |
{4}
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Reagents | UFII SHEATH
UFII SEARCH-SED
UFII PACK-SED
UFII SEARCH-BAC
UFII PACK-BAC | measurement using a double stain with two fluorescent dyes. Particle characterization and identification is based on detection of forward scatter, fluorescence and impedance signals, and on adaptive cluster analysis. |
| Quality Control/ Calibrator | UFII CONTROLS-2 levels
UFII CALIBRATOR | UF-CHECK-3 levels
UF-CAL |
| Software/Hardware Differences | Two channels for bacteria and sediment | One channel for sediment only |
I. Special Control/Guidance Document Referenced (if applicable):
Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA, December 4, 2001
J. Performance Characteristics:
1. Analytical Performance:
a. Accuracy: The UF1000i was compared with the UF100. Specimens included 310 urine samples with various abnormal conditions and normal samples at three outside clinical sites. Linear regression analysis was used for the correlation for RBC, WBC, EC, and CAST parameters. Percent agreement was performed for the Bacteria parameter. A new reference interval was established for the bacteria parameter n the UF-1000i since there is a separate channel and additional stain in the UF-1000i for bacteria detection. Therefore, a comparison between the UF-1000i and UF-100 using each instrument's bacteria reference interval as the cutoff
{5}
exhibited the percent agreement between the two instruments.
Table 1: UF-1000i vs. UF100 Enumerated Parameter Comparison
| Parameter | Correlation (r) | Regression (r²) | Regression Equation | Range |
| --- | --- | --- | --- | --- |
| RBC | 0.9921 | 0.9842 | y=0.9544x-3.009 | 0.0-4628.1 |
| WBC | 0.9669 | 0.9348 | y=0.8622x-1.6818 | 0.0-2557.5 |
| EC | 0.9777 | 0.9558 | y=0.8646x-0.1134 | 0.0-176.5 |
| CAST | 0.9558 | 0.9136 | y=0.6125x-0.0629 | 0.00-28.04 |
Table 2: UF-100i vs. UF100 Bacteria Parameter Comparison
| % Agreement | 83.22% |
| --- | --- |
| Predictive Value Positive | 91.66% |
| Predictive Value Negative | 84.30% |
Table 3: Overall Flagging Parameter Comparison: A flagging comparison of the UF-1000i to the UF-100 was performed on 310 samples. The flagging of the UF-1000i was also compared to the manual microscopic samples on 71 samples.
| | UF-1000i vs. UF-100 | UF-1000i vs. Manual |
| --- | --- | --- |
| % Agreement= (TP + TP)
N | 91.29% | 81.69% |
Conclusion: the UF-000i results compared favorably with UF-100 results.
b. Precision/Reproducibility: Within run precision data was collected for the quality control material (Lot # A6006) for the five enumerated parameters (RBC, WBC, Epithelial Cells, Cast, and Bacteria). Ten consecutive replicates of QC levels L and H were tested on the UF-1000i. The mean, standard deviation and coefficient of variation were determined and compared to performance claims on the assay sheet.
Table 4: Within Run Precision using QC Material (Lot # A6006)
| UFII-L Lot # A6006 | | | | | |
| --- | --- | --- | --- | --- | --- |
| | RBC | WBC | EC | Cast | Bacteria |
| Mean | 37.9 | 40.5 | 11.0 | 5.4 | 202.7 |
| SD | 4.7 | 5.0 | 1.4 | 0.9 | 17.3 |
| CV% | 12.3 | 12.4 | 12.8 | 17.2 | 8.5 |
| Acceptable | Yes | Yes | Yes | Yes | Yes |
{6}
6
| UFII-H Lot # A6006 | | | | | |
| --- | --- | --- | --- | --- | --- |
| | RBC | WBC | EC | Cast | Bacteria |
| Mean | 200.7 | 814.2 | 75.6 | 19.2 | 796.1 |
| SD | 4.7 | 9.2 | 4.0 | 1.4 | 34.6 |
| CV% | 2.3 | 1.1 | 5.3 | 7.1 | 4.4 |
| Acceptable | Yes | Yes | Yes | Yes | Yes |
Within Run Urine Specimens: Within Run precision was determined on select urine specimens each having a minimum volume of 50 mL which exceeded the established Normal population Reference Range for the five enumerated parameters (RBC, WBC, EC, CAST, Bacteria). The specimens were analyzed 10 times consecutively on the UF-1000i in the manual mode.
Conclusion: The within run precision on the five enumerated parameters were acceptable.
d. Carryover: Carryover data was collected by analyzing a high sample three consecutive times (H1, H2, H3) then analyzing a low sample three times (L1, L2, L3). The formula (L1-L3)/(H3-L3) x 100% was used where L1 is the first low sample analyzed; L3 is the last low sample analyzed; and H3 is the last high sample analyzed.
Table 5: UF-1000i Carryover
| Parameter | Carryover | Carryover | Carryover | Carryover | Manufacturer Specification |
| --- | --- | --- | --- | --- | --- |
| RBC | 0.05% | 0.08% | 0.09% | 0.05% | ≤0.1% or ≤5.0/μL |
| Bacteria | 0.00% | 0.0012/μL | 0.00% | 0.0035/μ | ≤0.05% or ≤5.0/μL |
Conclusion: Carryover on the testing parameters met manufacturer specifications.
e. Interfering Substances: The following conditions can volume settings to be difficult: High density samples within pyuria; macroscopic hematuria samples; samples that include fluorescent matter due to the inclusion of crystals or chemicals; samples that include preservatives.
2. Other Supportive Instrument Performance Data Not Covered Above:
Linearity: Linearity was performed on RBC, WBC, EC, Cast, and Bacteria by diluting samples with instrument diluent.
{7}
Table 6: Linearity Results
| Parameter | Range Tested | L | R² | r | Slope | Intercept |
| --- | --- | --- | --- | --- | --- | --- |
| WBC | 0.0-3603.0 | /μL | 0.99 | 1.00 | 1.02 | 48.24 |
| | 0.6-1623.3 | /μL | 1.00 | 1.00 | 0.99 | -3.61 |
| | 1.9-5259.7 | /μL | 1.00 | 1.00 | 0.99 | 138.15 |
| RBC | 0.0-5030.9 | /μL | 1.00 | 1.00 | 1.00 | 15.34 |
| | 1.8-2568.3 | /μL | 1.00 | 1.00 | 1.00 | 4.58 |
| | 2.8-5249.1 | /μL | 1.00 | 1.00 | 1.00 | -115.30 |
| EC | 0.3-171.3 | /μL | 0.90 | 1.00 | 1.07 | 12.67 |
| | 3.2-250.9 | /μL | 1.00 | 0.95 | 1.00 | 5.06 |
| CAST | 0.06-5.57 | /μL | 1.00 | 1.00 | 1.00 | -061 |
| | 0.0-15.20 | /μL | 0.94 | 0.97 | 0.95 | -0.73 |
| Bacteria | 0.0-320.2 | /μL | 0.99 | 1.00 | 0.97 | -1.19 |
| | 144.1-6263.75 | /μL | 1.00 | 1.00 | 0.99 | 125.62 |
Conclusion: RBC, WBC, EC, Cast, and Bacteria parameters were linear.
Verification of Reference Intervals: Since the UF-1000Ii is similar to the UF-100, 169 urine specimens fulfilling sample requirement conditions or the laboratory were tested and compared to existing reference values.
Table 7: Reference Interval Data Collected for UF-1000i
| Parameter | Female Reference Range | Male Reference Range | Combined Reference Range |
| --- | --- | --- | --- |
| RBC | 0-23.6 | 0-7.3 | 0-23.6 |
| WBC | 0-19.0 | 0-10.1 | 0-19.0 |
| EC | 0-33.9 | 0-3.0 | 0-31.2 |
| CAST | 0-0.86 | 0-0.61 | 0-11.1 |
| BACTERIA | 0-993.9 | 0-14.6 | 0-1175.8 |
Table 8: Reference Intervals for UF-1000i
| RBC | WBC | EC | CASTS | BACTERIA |
| --- | --- | --- | --- | --- |
| < 23/μL | < 25/μL | < 31/μL* | < 1/μL | < 1200/μl |
*Epithelial cell data is considered for information only given these cells are urinary contaminate are dependent on how the urine was collected.
Conclusion: The existing UF100 reference intervals for RBC, WBC, EC, and Casts can be used for the UF-1000i. The reference interval for Bacteria on the UF-1000i has been reestablished due to better bacteria detection with the addition of the bacteria channel and specific bacteria stain in the UF-1000i instrument.
{8}
K. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
L. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
8
{9}
9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.