← Product Code [NHB](/productcode/NHB) · K070886

# MEROGEL INJECTABLE BIORESORBABLE STENT (K070886)

_Medtronic, Inc. · NHB · Apr 26, 2007 · Ear, Nose, Throat · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K070886

## Device Facts

- **Applicant:** Medtronic, Inc.
- **Product Code:** [NHB](/productcode/NHB.md)
- **Decision Date:** Apr 26, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 874.3620
- **Device Class:** Class 2
- **Review Panel:** Ear, Nose, Throat
- **Attributes:** Therapeutic

## Indications for Use

MeroGel Injectable Bioresorbable Stent is a space occupying gel stent intended to separate and prevent adhesions between mucosal surfaces, help control minimal bleeding following surgery or nasal trauma, and act as an adjunct to aid in the natural healing process. MeroGel Injectable Bioresorbable Stent is indicated for use in the middle ear and external ear canal following canalplasty, myringoplasty, tympanoplasty, and stapes and mastoid surgery. The device is indicated following nasal/sinus surgery or trauma to prevent lateralization of the middle turbinate and nasal adhesions during the post operative period.

## Device Story

Sterile, transparent, viscoelastic, bioresorbable gel composed of cross-linked hyaluronic acid polymers; supplied in syringe. Used in ENT surgery/trauma to fill cavities, separate mucosal surfaces, and provide tamponade for minimal bleeding. Applied by physician in clinical setting. Acts as physical barrier during healing; naturally eliminated or aspirated. Benefits include adhesion prevention and support of natural healing process.

## Clinical Evidence

No clinical data provided; substantial equivalence based on technological characteristics and material similarity to legally marketed predicate devices.

## Technological Characteristics

Cross-linked hyaluronic acid polymer gel; sterile; viscoelastic; bioresorbable; syringe-based delivery; non-electronic; no software.

## Regulatory Identification

Ear, nose, and throat synthetic polymer material is a device material that is intended to be implanted for use as a space-occupying substance in the reconstructive surgery of the head and neck. The device is used, for example, in augmentation rhinoplasty and in tissue defect closures in the esophagus. The device is shaped and formed by the suregon to conform to the patient's needs. This generic type of device is made of material such as polyamide mesh or foil and porous polyethylene.

## Predicate Devices

- Sepragel 104 ENT Nasal/Sinus and Otologic Dressing ([K043035](/device/K043035.md))
- MeroGel Otologic Pack ([K001148](/device/K001148.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K070886

APR 2 6 2007

# 510(k) SUMMARY SUMMARY OF SAFETY AND EFFECTIVENESS FOR MeroGel Injectable Bioresorbable Dressing

| 510(k) Owner          | Medtronic Xomed, Inc<br>6743 Southpoint Drive North<br>Jacksonville, Florida 32216-0980 USA<br>904-296-9600<br>904-296-2386 (FAX) |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| Contact Name          | Jayme Wilson<br>Senior Regulatory Affairs Specialist<br>Medtronic Xomed, Inc                                                      |
| Date Summary Prepared | April 24, 2006                                                                                                                    |
| Proprietary Name      | MeroGel Injectable Bioresorbable Dressing (Final<br>name TBD)                                                                     |
| Common Name           | Polymer, Ear, Nose and Throat, Synthetic,<br>Absorbable                                                                           |
| Classification Name   | Ear, nose and throat synthetic polymer material<br>(21 CFR 874.3620, Product Code NHB, Class II)                                  |

#### Marketed device claiming equivalence to

MeroGel Injectable Bioresorbable Stent is equivalent to Genzyme Corporation's Sepragel 104 ENT Nasal/Sinus and Otologic Dressing, K043035 and Medtronic Xomed's MeroGei™ Otologic Pack, K001148.

#### Device Description

MeroGel Injectable Bioresorbable Stent is a sterile, transparent, viscoelactic, bioresorbable gel composed of cross-linked polymers of hyaluronic acid. The MeroGel Injectable Bioresorbable Stent fills ENT cavities following surgery or trauma to keep to tissues or structures separate during the healing process. During this time, the tamponade effect helps control minimal bleeding normally associated with routine ENT surgery. MeroGel Injectable leaves the site of placement by natural elimination, or it may be aspirated from the cavity earlier at the discretion of the physician.

{1}------------------------------------------------

# Intended Use

MeroGel Injectable Bioresorbable Stent is a space occupying gel stent intended to separate and prevent adhesions between mucosal surfaces, help control minimal bleeding following surgery or nasal trauma, and act as an adjunct to aid in the natural healing process.

## Indications for Use

MeroGel Injectable Bioresorbable Stent is indicated for use in the middle ear and external ear canal following canalplasty, myringoplasty, tympanoplasty, and stapes and mastoid surgery. The device is indicated following nasal/sinus surgery or trauma to prevent lateralization of the middle turbinate and nasal adhesions during the post operative period.

{2}------------------------------------------------

| Device Name                | Comparison to Marketed Devices<br>MeroGel™ Injectable Bioresorbable Stent<br>Medtronic Xomed<br>PROPOSED                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Sepragel™ ENT Nasal/Sinus and<br>Otologic Dressing<br>Genzyme Corporation<br>K043035                                                                                                                                                                                                                                                                                                                                                                                                                                                         | MeroGel™ Otologic Pack<br>Medtronic Xomed<br>K001148                                                                                                                                                                                                                                                                                                                                                         |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code               | ENT synthetic polymer material<br>77NHB                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | ENT synthetic polymer material<br>77KHJ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | ENT synthetic polymer material<br>77KHJ                                                                                                                                                                                                                                                                                                                                                                      |
| Intended<br>Use/Indication | MeroGel Injectable Bioresorbable Stent is<br>a space occupying gel stent intended to<br>separate and prevent adhesions between<br>mucosal surfaces, help control minimal<br>bleeding following surgery or nasal trauma,<br>and act as an adjunct to aid in the natural<br>healing process. MeroGel Injectable<br>Bioresorbable Stent is indicated for use in<br>the middle ear and external ear canal<br>following canalplasty, myringoplasty,<br>tympanoplasty, and stapes and mastoid<br>surgery. The device is indicated following<br>nasal/sinus surgery or trauma to prevent<br>lateralization of the middle turbinate and<br>nasal adhesions during the post operative<br>period. | For use in patients undergoing<br>nasal/sinus surgery as a space-<br>occupying gel stent to separate and<br>prevent adhesions between<br>mucosal surfaces in the nasal<br>cavity, to help control minimal<br>bleeding following surgery or<br>nasal trauma, and to prevent<br>lateralization of the middle<br>turbinate during the postoperative<br>period. The device is also<br>indicated for use in the middle ear<br>and external ear canal following<br>canalplasty, myringoplasty,<br>tympanoplasty and, stapes and<br>mastoid surgery | MeroGel Otologic Pack is a space<br>occupying dressing and/or stent<br>intended to separate mucosal<br>surfaces, help control minimal<br>bleeding and act as an adjunct to<br>aid in the natural healing process.<br>MeroGel Otologic Pack is<br>indicated for use in the middle ear<br>and external ear canal following<br>canalplasty, myringoplasty,<br>tympanoplasty, and stapes and<br>mastoid surgery. |
| Material<br>Bioresorbable  | Derivative hyaluronic acid<br>YES                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Derivative hyaluronic acid<br>YES                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Derivative hyaluronic acid<br>YES                                                                                                                                                                                                                                                                                                                                                                            |
| Sterile                    | YES                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | YES                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | YES                                                                                                                                                                                                                                                                                                                                                                                                          |
| Product<br>Matrix          | Gel in a syringe                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Gel in a syringe                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Non-woven pad in a protective<br>folded sheet                                                                                                                                                                                                                                                                                                                                                                |

.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird-like figure with outstretched wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird. The logo is black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Medtronic Xomed c/o Jayme Wilson 6743 Southpoint Drive North Jacksonville, Florida 32216-0980

APR 2 6 2007

Re: K070886

Trade/Device Name: MeroGel™ Injectable Bioresorbable Stent Regulation Number: 21 CFR 874.3620 Regulation Name: Ear Nose & Throat Synthetic Polymer Material Regulatory Class: II Product Code: NHB Dated: March 29, 2007 Received: March 30, 2007

Dear Ms. Wilson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 - Jayme Wilson

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

M.B. Eichelman SimD

Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure .

{5}------------------------------------------------

### INDICATIONS FOR USE

510(k) Number (if known): _ Kon 08862

MeroGel Injectable Bioresorbable Stent (Final name to be Device Name: determined)

Indications for Use:

MeroGel Injectable Bioresorbable Stent is a space occupying gel stent intended to separate and prevent adhesions between mucosal surfaces, help control minimal bleeding following surgery or nasal trauma, and act as an adjunct to aid in the natural healing process.

MeroGel Injectable Bioresorbable Stent is indicated for use in the middle ear and external ear canal following canalplasty, myringoplasty, tympanoplasty, and stapes and mastoid surgery. The device is indicated following nasal/sinus surgery or trauma to prevent lateralization of the middle turbinate and nasal adhesions during the post operative period.

Prescription Use X AND/OR

Over-The-Counter Use

(Part 21 CFR 801 Subpart D) C)

(21 CFR 801 Subpart

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Keun Bahn

Division Sign-Off) ivision of Opnthalmic Ear, Nose and Throat Devises

**510(k) Number** K070886

| Prescription Use     | <span style="text-decoration: line-through;"></span> |
|----------------------|------------------------------------------------------|
| (Per 21 CFR 801.109) | <span style="text-decoration: line-through;"></span> |

9 of 56

---

**Source:** [https://fda.innolitics.com/device/K070886](https://fda.innolitics.com/device/K070886)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
