← Product Code [MBI](/productcode/MBI) · K070882

# STRYKER PEEK TWINLOOP TAC (K070882)

_Stryker Endoscopy · MBI · Jul 20, 2007 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K070882

## Device Facts

- **Applicant:** Stryker Endoscopy
- **Product Code:** [MBI](/productcode/MBI.md)
- **Decision Date:** Jul 20, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Stryker PEEK TwinLoop Tac is intended to be used for suture or tissue fixation in the foot, ankle, knee, hip, hand, wrist, elbow, and shoulder. Specific indications are listed below and are size appropriate per patient needs: Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction. Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction. Scaphulolunate Ligament Reconstruction, Carpal Ligament Reconstruction, Hand/Wrist: Repair/Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP and MCP joints for all digits, digital tendon transfers. Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Foot/Ankle: Ligament Repair, Hallux Valgus Reconstruction, digital tendon transfers, Mid-foot reconstruction. Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis. Pelvis: Bladder Neck Suspension for Female urinary incontinence due to urethral hypermobility or intrinsic sphincter deficiency. Hip: Capsular repair, acetabular labral repair. The Stryker PEEK TwinLoop Tac is intended for single-use only.

## Device Story

Push-in type suture anchor; PEEK-OPTIMA LT3 polymer body; two USP#1 high-strength eyelets; pre-threaded with USP#2 non-absorbable braided UHMWPE sutures; pre-assembled on disposable inserter. Used by orthopedic surgeons in clinical/OR settings for soft tissue-to-bone fixation. Device provides mechanical anchor point for sutures to reattach tendons/ligaments to bone. Benefits include secure fixation during healing. Single-use, sterile, ethylene oxide sterilized.

## Clinical Evidence

No clinical data. Bench testing only.

## Technological Characteristics

Material: PEEK-OPTIMA LT3 (polyetheretherketone). Suture: Braided UHMWPE. Form factor: Push-in anchor with two eyelets, pre-assembled on disposable inserter. Sterilization: Ethylene oxide (SAL 10^-6). Standards: ISO 10993-1, ISO 10993-7, EN 550, EN 556-1, EN 11607, EN 980, EN 1041, EN ISO 14971.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Arthrex PEEK SutureTak ([K971723](/device/K971723.md), [K000506](/device/K000506.md), [K050749](/device/K050749.md), [K061863](/device/K061863.md))
- Arthrex Bio-Corkscrew FT ([K043337](/device/K043337.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K070882 page '92

5900 Optical Court San Jose, CA 95138 t: 408 754 2000 f: 408 754 2521

stryker®

Endoscopy

JUL 2 0 2007

# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

Date: March 30, 2007

Contact Person: Erica A. Walters, RAC Sr. Regulatory Representative 408-754-2078(phone) 408-754-2521 (fax) erica.walters@stryker.com

### Device Name:

| Proprietary Name:      | Stryker PEEK TwinLoop Tac                            |
|------------------------|------------------------------------------------------|
| Common and Usual Name: | PEEK Suture Anchor                                   |
| Classification Name:   | Screw, Fastener, Fixation, Nonabsorbable, Bone, Soft |
|                        | Tissue (Class II, 21 CFR 888.3040, Product Code MBI, |
|                        | Orthopedics Review Panel)                            |

### Predicate Device(s):

Arthrex PEEK SutureTak: K971723, K000506, K050749, K061863 Arthrex Bio-Corkscrew FT: K043337

## Device Description and Intended Use:

The Stryker PEEK TwinLoop Tac is intended to be used for suture or tissue fixation in the foot, ankle, knee hip, hand wrist, elbow, and shoulder. Specific indications are listed below and are size appropriate per patient needs:

Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction.

Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction.

{1}------------------------------------------------

page of

Scaphulolunate Ligament Reconstruction, Carpal Ligament Reconstruction, Hand/Wrist: Repair/Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendoros and PIP, DIP and MCP joints for all digits, digital tendon transfers.

Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Foot/Ankle: Ligament Repair, Hallux Valgus Reconstruction, digital tendon transfers, Mid-foot reconstruction.

Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis.

Pelvis: Bladder Neck Suspension for Female urinary incontinence due to urethral hypermobility or intrinsic sphincter deficiency.

Hip: Capsular repair, acetabular labral repair.

The Stryker PEEK TwinLoop Tac is intended for single-use only.

The Stryker PEEK TwinLoop Tac is a push-in type implant with two USP#1 high strength suture eyelets pre-threaded with USP#2 non-absorbable braided surgical sutures and pre-assembled on a disposable inserter. The Stryker PEEK TwinLoop Tac will be manufactured from PEEK-OPTIMA® LT3 (polyetheretherketone), a biocompatible polymer manufactured by Invibio Inc. The sutures will be USP braided ultra high molecular weight polyethylene (UHMWPE) suture (K033654, K040472 and K063778). The Stryker PEEK TwinLoop Tac will be validated to a SAL of 10th using ethylene oxide. The EtO residuals will be tested according to ISO 10993-7.1995.

Prior to introducing the Stryker PEEK TwinLoop Tac to market, the device will conform to the following voluntary safety and performance standards: ISO 10993-1, Blue Book Memorandum G95-1, EN 550, EN 556-1, EN 11607-1, EN 11607-2, EN 980, EN 1041, and EN ISO 14971.

The technological and material differences between the Stryker PEEK TwinLoop Tac, and the Arthrex PEEK SutureTak (K971723, k000506, k050749, K061863), do not affect the safety and efficacy of the product; therefore, the Stryker PEEK TwinLoop Tac is considered substantially equivalent in performance, intended use, safety, and efficacy to the Arthrex PEEK SutureTac.

The Stryker PEEK TwinLoop Tac is considered substantially equivalent in material composition, intended use, safety and efficacy to the Arthrex PEEK SutureTak (K971723, k000506, k050749, K061863).

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)". Inside the circle is a stylized symbol that resembles an abstract human figure or a bird in flight, composed of three curved lines.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 2 0 2007

Stryker Endoscopy % Mr. K. Jeffrev Semone Director, Regulatory Affairs 5900 Optical Ct. San Jose, CA 95138

Re: K070882

> Trade/Device Name: Stryker PEEK TwinLoop Tac Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI, HWC, JDR Dated: July 6, 2007 Received: July 9, 2007

Dear Mr. Semone:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 – Mr. K. Jeffrey Semone

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or 240-276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Soubayo Spichum

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

page 1 of 1

#### INDICATIONS FOR USE

Device Name: Stryker PEEK TwinLoop Tac

510(k) Number if known: K070882

The Stryker PEEK TwinLoop Tac is intended to be used for suture or tissue fixation in the foot, ankle, knee, hip, hand, wrist, elbow, and shoulder. Specific indications are listed below and are size appropriate per patient needs:

Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction.

Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction.

Scaphulolunate Ligament Reconstruction, Carpal Ligament Reconstruction, Hand/Wrist: Repair/Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP and MCP joints for all digits, digital tendon transfers.

Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Foot/Ankle: Ligament Repair, Hallux Valgus Reconstruction, digital tendon transfers, Mid-foot reconstruction.

Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis.

Pelvis: Bladder Neck Suspension for Female urinary incontinence due to urethral hypermobility or intrinsic sphincter deficiency.

Hip: Capsular repair, acetabular labral repair,

The Stryker PEEK TwinLoop Tac is intended for single-use only.

Prescription Use X Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE

K070882

and Neurological Devices

STRYKER PEEK TWINLOOP TAC 510(K) SUBMISSION

**510(k) Number** k070882

---

**Source:** [https://fda.innolitics.com/device/K070882](https://fda.innolitics.com/device/K070882)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
