K070859 · Sterilmed, Inc. · NLL · Sep 12, 2007 · General, Plastic Surgery
Device Facts
Record ID
K070859
Device Name
REPROCESSED ETHICON ETS ENDOSCOPIC LINEAR CUTTERS
Applicant
Sterilmed, Inc.
Product Code
NLL · General, Plastic Surgery
Decision Date
Sep 12, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The ETS 45, ETS-Flex45, and Compact-Flex 45 reprocessed, reloadable Endoscopic Linear Cutters are intended for transection, resection and/or anastomosis. These instruments have applications in open and minimally invasive general, gynecologic, urologic (including radical prostatectomy), thoracic and pediatric surgery. The ETS Flex 45, ETS 45 and ETS Compact Flex 45 device can be used with staple line or tissue buttressing materials such as bovine pericardium. The ETS and ETS-Flex (articulating) reprocessed, reloadable Endoscopic Linear Cutters (35mm) have applications in general, gynecologic, urologic, and thoracic surgery for transection, and/or creation of anastomoses.
Device Story
Reprocessed, sterile, single-patient-use endoscopic linear cutter; delivers two double-staggered or triple-staggered rows of staples while simultaneously dividing tissue; reloadable design allows maximum 8 firings per procedure; articulation lever on ETS-Flex models enables bilateral jaw movement; used in OR by surgeons for general, gynecologic, urologic, thoracic, and pediatric procedures; device facilitates tissue transection and anastomosis; implantable staples and cartridges are not reprocessed; device provides mechanical surgical assistance.
Clinical Evidence
Bench testing only. Functional testing performed on representative samples to verify performance characteristics. Process validation conducted for cleaning, sterilization, and packaging. Manufacturing process includes visual and functional testing of all units.
Technological Characteristics
Reloadable linear cutter; mechanical actuation; manual operation; stainless steel/plastic construction; sterile; single-patient-use; compatible with staple line or tissue buttressing materials (e.g., bovine pericardium).
Indications for Use
Indicated for patients undergoing open or minimally invasive general, gynecologic, urologic (including radical prostatectomy), thoracic, or pediatric surgery requiring transection, resection, or anastomosis.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
Predicate Devices
Ethicon Endopath ETS Endoscopic Linear Cutter and ETS-Flex Endoscopic Articulating Linear Cutter (K961390)
{0}------------------------------------------------
Ko70.859
## I. SUMMARY AND CERTIFICATION
- A. 510(k) Summary
:
| Submitter: | SterilMed, Inc. | SEP 12 2007 |
|------------------------|-------------------------------------------------------------------------------------------------------------------|-------------|
| Contact Person: | Caroline Butterfield<br>11400 73rd Avenue North<br>Maple Grove, MN 55369<br>Ph: 888-856-4870<br>Fax: 763-488-3350 | |
| Date Prepared: | March 27, 2007 | |
| Trade Name: | Reprocessed Ethicon ETS Endoscopic Linear Cutters | |
| Classification Name: | Staple, Implantable | |
| Classification Number: | Class II, 21 CFR 878.4750 | |
| Product Code: | NLL | |
| Predicate<br>Devices: | The reprocessed Ethicon ETS Linear Cutters are substantially equivalent to the Ethicon<br>Endopath ETS Endoscopic Linear Cutter and ETS-Flex Endoscopic Articulating Linear<br>Cutter (K961390). |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device<br>Description: | SterilMed's reprocessed Ethicon ETS Linear Cutter is a sterile, single patient use<br>device. It delivers two double-staggered or two triple-staggered rows of staples while<br>simultaneously dividing the tissue between the rows of staples. An articulation lever<br>on the ETS-Flex Linear Cutter enables bilateral movement of the instrument jaws.<br>These devices are reloadable linear cutters that allow for a maximum of 8 firings in a<br>single surgical procedure. |
| | Note: Only the linear cutter is the subject of this submission, the implantable staple and<br>the staple cartridge are not reprocessed and therefore are not included. |
| Intended Use: | The ETS 45, ETS-Flex 45, and ETS Compact-Flex 45 reprocessed, reloadable Ethicon<br>Linear Cutters are intended for transection, resection and/or anastomosis. These<br>instruments have applications in open and minimally invasive general, gynecologic,<br>urologic (including radical prostatectomy), thoracic and pediatric surgery. The ETS-<br>Flex 45, ETS 45 and ETS Compact-Flex 45 device can be used with staple line or tissue<br>buttressing materials such as bovine pericardium. |
| | The ETS, ETS-Flex (articulating) Endoscopic reprocessed, reloadable Linear Cutters<br>(35mm) have applications in general, gynecological, and thoracic surgery for<br>transection, resection, and/or creation of anastomoses. |
| Functional and<br>Safety Testing: | Representative samples of reprocessed Linear Cutters are tested to demonstrate<br>appropriate functional characteristics. Process validation testing is performed to<br>validate the cleaning and sterilization procedures as well as device packaging. In<br>addition, the manufacturing process includes visual and validated functional testing of<br>all products produced. |
| Conclusion: | The reprocessed ETS Linear Cutters are substantially equivalent to the Endopath ETS<br>Endoscopic Linear Cutter and ETS-Flex Endoscopic Articulating Linear Cutter<br>(K961390, K980023, K002398 and K020079) manufactured by Ethicon. |
| | This conclusion is based upon the devices' similarities in functional design (principle of<br>operation), materials, indications for use and methods of construction. |
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of a stylized caduceus, which is a symbol often associated with healthcare, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" arranged in a circular fashion around the symbol. The caduceus in the logo is a modern interpretation, featuring a series of curved lines that suggest the wings and staff of the traditional symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 1 2 2007
SterilMed, Inc. % Mr. Dennis Toussaint Director of Regulatory Affairs 11400 731d Avenue North, Suite 100 Maple Grove, Minnesota 55369
Re: K070859
Trade/Device Name: Reprocessed Ethicon ETS Linear Cutters Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable staple Regulatory Class: II Product Code: NLL Dated: August 27, 2007 Received: August 28, 2007
Dear Mr. Toussaint:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
Page 2 -- Mr. Dennis Toussaint
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
K070859
Page (8-
## Indications for Use
510(k) Number (if known): K070859
Device Name: Reprocessed Ethicon ETS Linear Cutters
Indications for Use:
The ETS 45, ETS-Flex45, and Compact-Flex 45 reprocessed, reloadable Endoscopic Linear Cutters are intended for transection, resection and/or anastomosis. These instruments have applications in open and minimally invasive general, gynecologic, urologic (including radical prostatectomy), thoracic and pediatric surgery. The ETS Flex 45, ETS 45 and ETS Compact Flex 45 device can be used with staple line or tissue buttressing materials such as bovine pericardium.
The ETS and ETS-Flex (articulating) reprocessed, reloadable Endoscopic Linear Cutters (35mm) have applications in general, gynecologic, urologic, and thoracic surgery for transection, and/or creation of anastomoses.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices
Page 1 of 1
510(k) Number_ L670859
{4}------------------------------------------------
Ko 70859 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 -
Page 2 of 2
| Manufacturer | Base Device | Model # | Description |
|--------------|------------------------|---------------|----------------------------------------------------------|
| Ethicon | ETS Flex 45 | ATW45 | ETS-Flex 45mm, 34cm long (no cartridge included) |
| | | ATB45 | ETS-Flex 45mm, 34cm long (no cartridge included) |
| | | ATG45 | ETS-Flex 45mm, 34cm long (no cartridge included) |
| | | 6TB45 | ETS-Flex 45mm, 34cm long (no cartridge included) |
| | ETS Flex 35 | ATW35 | ETS-Flex 35mm, 34cm long (no cartridge included) |
| | | ATB35 | ETS-Flex 35mm, 34cm long (no cartridge included) |
| | ETS 45 | TSW45 | ETS Straight 45mm, 34cm long (no cartridge included) |
| | | TSB45 | ETS Straight 45mm, 34cm long (no cartridge included) |
| | | TSG45 | ETS Straight 45mm, 34cm long (no cartridge included) |
| | ETS 35 | TSW35 | ETS Straight 35mm, 34cm long (no cartridge included) |
| | | TSB35 | ETS Straight 35mm, 34cm long (no cartridge included) |
| | ETS Flex<br>Compact 45 | SCW45 Compact | ETS-Flex Compact 45mm, 24cm long (no cartridge included) |
| | | SCB45 Compact | ETS-Flex Compact 45mm, 24cm long (no cartridge included) |
| | | 6CB45 Compact | ETS-Flex Compact 45mm, 24cm long (no cartridge included) |
| | | SCG45 Compact | ETS-Flex Compact 45mm, 24cm long (no cartridge included) |
| | | 6SB45 Compact | ETS-Flex Compact 45mm, 24cm long (no cartridge included) |
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
| | | | Table 1 - List of Devices included in this Premarket Notification Submission |
|--|--|--|------------------------------------------------------------------------------|
|--|--|--|------------------------------------------------------------------------------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.