← Product Code [MNR](/productcode/MNR) · K070855

# SLEEP QUALITY (K070855)

_Dynadx Corporation · MNR · Mar 20, 2008 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K070855

## Device Facts

- **Applicant:** Dynadx Corporation
- **Product Code:** [MNR](/productcode/MNR.md)
- **Decision Date:** Mar 20, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.2375
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Software as a Medical Device

## Indications for Use

DynaDx Sleep Quality is intended for use on adult patients only as a screening device to determine the need for clinical diagnosis and evaluation by polysomnography based on the patient's score. The ECG recording may be obtained at any location specified by a physician including home, hospital or clinic. Subjects screened for sleep apnea should have periods of sleep of at least 4 hours duration during which the ECG is predominantly sinus rhythm in nature.

## Device Story

Sleep Quality is a software system for screening sleep-disordered breathing (SDB). It processes single-lead ECG recordings exported in ASCII format from FDA-cleared Holter monitors. The system analyzes the ECG to derive RR intervals and ECG-Derived Respiratory (EDR) information. By identifying SDB episodes, the software calculates an Apnea-Hypopnea Index (AHI). Used in clinical or home settings, the software output assists physicians in determining the necessity for formal polysomnography. The device provides a non-invasive screening method to identify patients requiring further diagnostic evaluation, potentially improving access to sleep apnea care.

## Clinical Evidence

Clinical study performed on 26,100 minutes of sleep data. Results: 95.0% per-subject sensitivity, 90.0% per-subject specificity, and 92.9% per-subject accuracy. Per-minute metrics: 84.5% sensitivity, 83.0% specificity, and 83.8% accuracy. Comparison against Lifescreen Apnea predicate showed improved performance metrics across all categories.

## Technological Characteristics

Software-based system; processes ASCII-formatted ECG data from external Holter monitors. Uses proprietary algorithm to extract RR intervals and ECG-Derived Respiratory (EDR) signals. Operates on standard computing hardware. No specific materials or energy sources as it is a software-only analysis tool.

## Regulatory Identification

A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.

## Predicate Devices

- Lifescreen Apnea ([K042745](/device/K042745.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K070855

# Section 5: 510(k) Summary

MAR 2 0 2008

DynaDx Corporation 213 Houghton Street Mountain View, CA 94041 Phone: 650-386-6369 Fax: 330-734-0004

Contact: Yanhui Liu

Summary Prepared: March 20, 2007

Trade Name: Sleep Ouclity

Common Name: Apnea Examination System

Classification Name: Ventilatory Effort Recorder

Predicate Devices Identification:

CFR21: 868.2375 Product Code: MNR Device Class: II Legally Marketed Device: Lifescreen Apnea Manufacturer: Del Mar Reynolds Medical, Inc. K#: K042745

#### Description:

Sleep Quality is a software system, which can probe the dynamical interactions between heart rate and respiratory variability during sleep by using a single lead Electrocardiogram (ECG) recording. It is designed to support FDA approved Holter Monitor devices on the market. Sleep Quality system takes text format ECG recording as input. It requires that supported Holter devices are able to export their recording into ASCII data format. Most of the commencially available Holter Monitors devices have this capability. ECG recording data is first collected by a standard Holter devices na subject's sleep, is converted into ASCII data format, and is then analyzed by the Sleep Quality system. By recognizing episodes of Sleep Disordered Breathing (SDB), the software indicates that apneic events have occurred and estimated the Apnea-Hypopnea Index (AH). To detect SDB, the algorithm derives both RR Interval and ECG-Derived Respiratory (EDR) information from the ECG recording.

{1}------------------------------------------------

#### Intended Use:

DynaDx Sleep Quality is intended for use on adult patients only as a screening device to determine the need for clinical diagnosis and evaluation by polysomnography based on the patient's score. The ECG recording may be obtained at any location specifical by a physician including home, hospital or clinic. Subjects screened for sleep apneas should have periods of sleep of at least 4 hours duration during which the ECG is predominantly sinus rhythm in nature.

| Product<br>Parameter                                | Sleep Quality<br>(Submission Device)<br>DynaDx Corporation<br>K070855                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Lifescreen Apnea<br>(Predicate Device)<br>Del Mar Reynolds Inc.<br>K042745                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|-----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Class                                        | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Product Code                                        | MNR                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | MNR                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Device Type                                         | Ventilator Effort Recorder                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Ventilator Effort Recorder                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Common Name                                         | Apnea Examination Software                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Apnea Examination Software                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Regulation Number                                   | 868.2375                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | 886.2375                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Intended Use                                        | Sleep Quality is intended for<br>use on adult patients only as<br>a screening device to<br>determine the need for<br>clinical diagnosis and<br>evaluation by<br>polysomnography based on<br>the patient's score. The ECG<br>recording may be obtained at<br>any location specified by a<br>physician including home,<br>hospital or clinic. Subjects<br>screened for sleep apnea<br>should have periods of sleep<br>of at least 4 hours duration<br>during which the ECG is<br>predominantly sinus rhythm<br>in nature. | Lifescreen Apnea is intended<br>for use on adult patients only<br>as a screening device to<br>determine the need for<br>clinical diagnosis and<br>evaluation by<br>polysomnography based on<br>the patient's score. The ECG<br>recording may be obtained at<br>any location specified by a<br>physician including home,<br>hospital or clinic. Subjects<br>screened for sleep apnea<br>should have periods of sleep<br>of at least 4 hours duration<br>during which the ECG is<br>predominantly sinus rhythm<br>in nature. |
| Target Population                                   | Adults                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Adults                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Apnea-Hypopnea<br>Index (AHI)<br>determination      | Apnea-Hypopnea Index<br>(AHI) derived from ECG<br>signal through proprietary<br>algorithm                                                                                                                                                                                                                                                                                                                                                                                                                               | Apnea-Hypopnea Index<br>(AHI) derived from ECG<br>signal through proprietary<br>algorithm                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Detection of Sleep<br>Disordered<br>Breathing (SDB) | To detect SDB, the algorithm<br>derives both RR Interval and<br>ECG-Derived Respiratory<br>(EDR) information from the<br>ECG recording.                                                                                                                                                                                                                                                                                                                                                                                 | To detect SDB, the algorithm<br>derives both RR Interval and<br>ECG-Derived Respiratory<br>(EDR) information from the<br>ECG recording.                                                                                                                                                                                                                                                                                                                                                                                    |
| Device for<br>collecting signal                     | FDA approved Holter<br>Monitor with text format<br>ECG signal output                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Del Mar Reynolds Lifescreen<br>digital Holter recorder                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Data Collection                                     | Home, hospital, clinic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Home, hospital, clinic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Location                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Environment of Use                                  | Hospital and clinic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Hospital and clinic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Software Validation                                 | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Clinical Study                                      | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Clinical Trial                                      | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Per-subject<br>Sensitivity                          | 95.0%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 85.0%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Per-subject<br>Specificity                          | 90.0%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 83.3%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Per Subject<br>Accuracy                             | 92.9%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 84.3%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Per Subject Positive<br>Predictivity                | 92.7%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 87.1%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Sleep Length for<br>Test (minutes)                  | 26100                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 26100                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Per-minute<br>sensitivity                           | 84.5%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 80.9%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Per-minute<br>specificity                           | 83.0%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 78.1%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Per-minute<br>Accuracy                              | 83.8%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 79.5%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Per-minute Positive<br>Predictivity                 | 83.3%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 78.7%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |

### Predicate Product Comparison Chart:

{2}------------------------------------------------

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring a symbol that resembles a caduceus or a staff entwined with a serpent.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 2 0 2008

DynaDx Corporation C/O Mr. EJ Smith Regulatory Consultant Smith Associates 1676 Village Green, Suite A Crofton, Maryland 21114

Re: K070855

Trade/Device Name: Sleep Quality Regulation Number: 868.2375 Regulation Name: Breathing Frequency Monitor Regulatory Class: II Product Code: MNR Dated: February 19, 2008 Received: March 10, 2008

Dear Mr. Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Smith

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Sutte Y. Michie Ons.

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# Indications for Use

510(k) Number (if known):

Device Name: Sleep Quality

Indications for Use:

DynaDx Sleep Quality is intended for use on adult patients only as a screening device to determine the need for clinical diagnosis and evaluation by polysomnography based on the patient's score. The ECG recording may be obtained at any location specified by a physician including home, hospital or clinic. Subjects screened for sleep apnea should have periods of sleep of at least 4 hours duration during which the ECG is predominantly sinus rhythm in nature.

Prescription Use イ (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C) .

## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Min Thulal

Page ___of____________________________________________________________________________________________________________________________________________________________________

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number:________________________________________________________________________________________________________________________________________________________________

---

**Source:** [https://fda.innolitics.com/device/K070855](https://fda.innolitics.com/device/K070855)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
