K070818 · Tom Medical Entwicklungs GmbH · DSH · Apr 6, 2007 · Cardiovascular
Device Facts
Record ID
K070818
Device Name
MEDILOGAR4, MEDILOGAR12
Applicant
Tom Medical Entwicklungs GmbH
Product Code
DSH · Cardiovascular
Decision Date
Apr 6, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2800
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
The MedilogAR4 and MedilogAR12 Digital Holter Recorders are used to record a 3 channel ECG and to measure the time intervals between consecutive R peaks in the ECG. The system is designed for a measuring duration of more than 24h and is therefore worn by the patient throughout the whole day. The preparation for the recording (attaching electrodes, etc.) is undertaken by the technician or doctor. During the recording, the instrument is carried in a special protective pouch which can be attached to a belt. The instrument is not designed for emergency monitoring purposes (intensive medical monitor); the instrument is only designed to record the ECG and/or information derived from ECG. The recorder is not protected against the effects of defibrillation as outlined in EN60601-2-25. The instruments also record other information: HF component of the ECG, time point of the P-wave, time point of the T-wave, and time point of the pacemaker pulse. In addition the MedilogAR12 records the ECG amplitude.
Device Story
Ambulatory Holter recorder (AR4/AR12) captures 3-channel ECG via body surface electrodes; worn by patient in protective pouch for 24+ hours. Technician/doctor performs electrode attachment. Device digitizes analog signals via 16-bit A/D converter; microprocessor performs signal processing including pacemaker spike detection, R-peak detection, PT-wave detection, and R-S amplitude measurement (EDR). Data stored on removable CompactFlash card. Output transferred via infrared interface to PC/PDA for clinician review. Facilitates diagnosis of arrhythmias, ischemia, and pacemaker function; aids in assessing patient recovery and heart rate variability. Benefits include long-term cardiac rhythm monitoring in ambulatory settings.
Clinical Evidence
Bench testing only. R-peak detection validated against MIT, AHA, and NST databases (sensitivity/positive predictivity >99%). Pacemaker detection validated using phantom artificial ECG signal generator (100% detection). PT-wave detection validated against QT database (sensitivity/positive predictivity >80%). R-S amplitude measurement validated against recordings with airflow channel (mean deviation <1 c/min).
Technological Characteristics
3-channel digital Holter recorder. 16-bit A/D conversion. CompactFlash storage. Infrared interface for data transfer. Microprocessor-based signal processing. Powered by batteries. Not defibrillation-protected per EN60601-2-25. Software performs digital signal processing including R-peak, PT-wave, and pacemaker spike detection.
Indications for Use
Indicated for patients requiring 24+ hour 3-channel ECG monitoring. Used for evaluation of symptoms suggesting arrhythmia or myocardial ischemia, documenting therapeutic interventions, ST segment changes, patient response after resuming activities (e.g., post-MI or cardiac surgery), clinical/epidemiological research, pacemaker evaluation, heart rate variability, and QT interval reporting. Not for emergency/intensive monitoring. Not suitable for infants <10 kg.
Regulatory Classification
Identification
A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.
{0}------------------------------------------------
sstrengrubergas A-8010 Graz, AUSTRIA al: + 43 (316) 261609
# MK3A 510(k) Summary
Version: 1
Status: RELEASE
K070818
p1/3
| Proprietary name: | MK3A | APR - 6 2007 |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Trade name: | medilogAR4, medilogAR12 | |
| Common name: | Ambulatory ECG Monitor, Holter Recorder | |
| Classification name: | Ambulatory ECG Monitor (21 CFR 870.2800, Product Code DSH) | |
| 510(k) submitter: | TOM Medical Entwicklungs GmbH<br>Anzengrubergasse 8, A-8010 Graz, Austria<br>Tel. + 43 (316) 261609-0<br>Fax. + 43 (316) 261609-10 | |
| Contact person: | Thomas Hasiba, General Manager Graz,<br>Tel. + 43 (316) 261609-14 | |
| Manufacturer: | Huntleight Healthcare Cardiology Products Division<br>35 Portmanmoor Road, Cardiff, Wales CF24 5HN, UK | |
| Predicate devices: | Medilog FD4 [510 (k) Number K970902]<br>DXP1000 [510 (k) Number K993618] (marketed as Medilog FD5) | |
### 1 Concise description of the device
#### 1.1 General
The recorder (type reference AR4 or AR12) is used to record a 3 channel ECG and to measure the time intervals between consecutive R peaks in the ECG. The system is designed for a measuring duration of more than 24h and is therefore worn by the patient throughout the whole day. The preparation for the recording (attaching electrodes, etc.) is undertaken by the technician or doctor. During the recording, the instrument is carried in a special protective pouch which can be attached to a belt. After inserting a memory card and batteries, the instrument can be switched on by pressing and holding the button on the top of the instrument. The recording is stopped by pressing and holding the same button on the top of the instrument or the recorder stops the recording automatically after a pre-set time. The instrument is not designed for emergency monitoring purposes (intensive medical monitor); the instrument is only designed to record the ECG and/or information derived from ECG. The recorder is not protected against the effects of defibrillation as outlined in EN60601-2-25.
As well as measuring the ECG and the beat to beat heart rate, the instrument also records other information: HF component of the ECG (quantifies muscle artefacts or disturbances), time point of the P-wave, time point of the T-wave, and time point of the pacemaker pulse.
In addition the MedilogAR12 records the ECG amplitude - Due to the heart's mechanical connection to the ribcage, breathing causes the electrical heart vector to turn and therefore changes the ECG amplitude. This reveals information about the breathing frequency and QPA (pulse /breathing quotient).
The data is stored digitally on a memory card. After removing the memory card, the data can be read by an appropriate card reader (e.g. PCMCIA slot and adapter).
The content of this document is intellectual property of TOM Medical Entwicklungs GmbH and has confidential. You are not allowed to print, copy, store, modify, publish or use any material for any purpose at all without the written consent of TOM Medical Entwicklongs GmbH.
{1}------------------------------------------------
Version: 1
Status: RELEASE
The recorder is not designed for emergency monitoring purposes (intensive medical monitor only designed for recording the ECG and/or heart rate.
The recorder is not protected against the effects of defibrillation as outlined in EN60601-2-25.
The recorder is not suitable for infants weighing less than 10 kg (22 lbs).
### 1.2 Technology
The MedilogAR4 and MedilogAR12 Digital Holter Recorders measure the voltage difference to derive the electrical ECG from body surface electrodes. A ground electrode is used to keep ground potential of the measured patient and of the recorder at the same level. After amplification the analogue signal is digitised using conventional 16 bit A/D converter. A microprocessor gathers the so sampled data. The microprocessor processes the data before it stores them onto a removable flash memory card (CompactFlash technology). Processing includes:
- Pacemaker spike detection .
- R- peak detection .
- . R-S amplitude measurement
- . PT- wave detection
- . Resample to storage sample rate
- Loss-less delta coding (Huffmann) .
The data stored on the flash memory card is organised in one Windows FAT16 file.
### 1.3 Indications for use
The MedilogAR4 and MedilogAR12 Digital Holter Recorders are used to record a 3 channel ECG and to measure the time intervals between consecutive R peaks in the ECG. The system is designed for a measuring duration of more than 24h and is therefore worn by the patient throughout the whole day. The preparation for the recording (attaching electrodes, etc.) is undertaken by the technician or doctor. During the recording, the instrument is carried in a special protective pouch which can be attached to a belt.
The instrument is not designed for emergency monitoring purposes (intensive medical monitor); the instrument is only designed to record the ECG and/or information derived from ECG. The recorder is not protected against the effects of defibrillation as outlined in EN60601-2-25.
The instruments also record other information: HF component of the ECG, time point of the Pwave, time point of the T-wave, and time point of the pacemaker pulse.
In addition the MedilogAR12 records the ECG amplitude.
Such monitoring is most frequent used for the indications below:
- . Evaluation of symptoms suggesting arrhythmia or myocardial ischemia.
- Evaluation of ECG documenting therapeutic interventions in individual patients or groups of . patients.
- . Evaluation of patients for ST segment changes (no online ST measurement implemented in recorder)
- Evaluation of a patient's response after resuming occupational or recreational activities (e.g., . after M.I. or cardiac surgery)
- Clinical and epidemiological research studies. .
- . Evaluation of patients with pacemakers
- Reporting of time and frequency domain heart rate variability (no online calculation . implemented in recorder)
- Reporting of QT interval (no online QT interval measurement implemented in recorder) .
The content of this document is intellectual property of TOM Medial Entwicklungs GmbH and has contidential. You are not allowed to print, copy, store, modily, publish or use any material for any purpose at all without the written consent of TOM Medical Entyvicklungs GrtbH
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three heads facing to the right. The eagle is surrounded by a circular border with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" written around it.
MAY 2 4 2007
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Tom Medical Entwicklungs Gmbh c/o Mr. Stefan Preiss TUV America, Inc. 1175 Old Highway 8 NW, Suite 104 New Brighton, MN 55112
Re: K070818
Trade/Device Name: Medilog AR4, MedilogAR12 Regulation Number: 21 CFR 870.2800 Regulation Name: Medical magnetic tape recorder Regulatory Class: Class II Product Code: DSH Dated: March 22, 2007 Received: March 26, 2007
Dear Mr. Preiss:
This letter corrects our substantially equivalent letter of April 6, 2007.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other
{3}------------------------------------------------
Page 2 - Mr. Preiss
requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120_ Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
Version: 1 Status: RELEASE
Image /page/4/Picture/1 description: The image shows the text "K070818" on the top line and "P 3/3" on the second line. The text is written in a handwritten style with a black ink or marker. The numbers and letters are clearly visible, and the image appears to be a close-up shot of the text.
# 2 Device comparison table
Tel: + 43 (316) 261GOS
The candidate devices Medilog AR4 and Medilog AR12, subject of this 510(k), have been compared to the following predicate devices: Medilog FD4 [510 (k) Number K970902], DXP1000 [510 (k) Number K993618] (marketed as Medilog FD5).
The significant differences to the Medilog FD4 are found as:
- Infrared interface Medilog AR4 and Medilog AR12 recorders provide a infrared interface . instead of a fibre optical cable. This improves the usability, as fibre optic interfaces to PCs became obsolete resently
- Enhanced input voltage range more contemporary electronic provides better ECG quality for . the sake of better ECG quality
- Digital pacemaker detection The Medilog AR4 and Medilog AR12 are equipped with an . internal spike detection algorithm offering a method for pacemaker functional analysis in Holter recordings
- Digital R peak detection The R- peak detection feature is a preprocessing tool for off line . beat detection algorithms to realign off line detected beats at a higher resolution, and, hence, provide a higher accuracy for e.g. HRV (heart rate variability) analysis.
- R-S amplitude measurement -- The Medilog AR12 also records the R-S amplitude, which . represents, according to several publications, an indicator for the heart position and movement in the chest, which is mainly induced by breathing and body position.
The significant differences to the Medilog FD5 (DXP 1000) are found as:
- No display Instead of showing ECG traces directly on the recording device, the Medilog AR4 and Medilog AR12 recorders provide an infrared interface, compatible with PDAs and PCs. The ECG traces can then reviewed on a remote PDA or PC.
- Enhanced input voltage range see discussion above .
- . Digital R peak detection - see discussion above
- R-S amplitude measurement see discussion above .
## 3 Summary for performance testing
Only bench tests were required to prove the device's performance.
The following tests have been performed:
| Test | Benchmark | Criteria | Result |
|--------------------------------------|-----------------------------------------------------|-----------------------------------------------------------------------------|--------|
| R peak detection | against MIT, AHA, and NST<br>database | sensitivity and positive<br>predictivity higher than 99%<br>for MIT and AHA | PASSED |
| Pacemaker<br>detection | Phantom artificial ECG signal<br>generator | 100% detection of<br>pacemaker spikes in all<br>programms | PASSED |
| PT wave<br>detection | against QT database | sensitivity and positive<br>predictivity higher than 80%<br>for QT database | PASSED |
| RS amplitude<br>measurement<br>(EDR) | against a set of recordings<br>with airflow channel | differences of mean<br>deviation lower than 1 c/min | PASSED |
#### Conclusion ব
The AR4/AR12 were found to meet all requirements. The devices are save and at least as effective, and perform at least as well as the predicate devices.
The content of this document is intellectual property of TOM Medial Entwickungs GmbH and has contidential. You are not allowed to prin, copy, store, modily, publish or use any material from this document for any purpose at all without the written consent of TOM Medical Entyvishings GmbH.
{5}------------------------------------------------
'5818
### Indications for Use
510(k) Number (if known):
not assigned yet
Device Name:
MedilogAR4, MedilogAR12
Indications for Use:
The MedilogAR4 and MedilogAR12 Digital Holter Recorders are used to record a 3 channel ECG and to measure the time intervals between consecutive R peaks in the ECG. The system is designed for a measuring duration of more than 24h and is therefore worn by the patient throughout the whole day. The preparation for the recording (artaching electrodes, etc.) is undertaken by the technician or doctor. During the instrument is carried in a special protective pouch which can be attached to a belt.
The instrument is not designed for emergency monitoring purposes lintensive medical monitor); the instrument is only designed to record the ECG and/or information derived from ECG, The recorder is not protected against the effects of defibrillation as outlined in ENRO601-2-25.
The instruments also record other information: HF component of the ECG, time point of the Pwave, time point of the T-wave, and time point of the pacemaker pulse.
In addition the MedilogAR12 records the ECG amplitude.
Such monitoring is most frequent used for the indications below:
- Evaluation of symptoms suggesting archythmia or myocardial ischomia.
- Evaluation of ECG documenting therapeutic in individual patients or groups of patients.
- Evaluation of patients for ST segment changes ino online ST measurement in promotive in recorder! 3.
- Evaluation of a patiant's responsa after resuming occupational activities in recentional actually of র্ব cardiac surgary)
- 5 Clinical and epidemiological revearch studies.
- 6. Evaluation of pationts with pacemakers
- Reporting of time and frequency the variebility the online calculation implemented in racorded) 7 য়ে
- Reporting of OT interval (no online, OT interval messurement implemented in recorder)
Prescription Use yES (Part 21 GFR 801 Subpart D) AND/OR
Over-The Gounter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number K07818
Page 1 of 1
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.