The NeoCoil 3.0T 6-Channel Carotid Array Coil is a receive only phased array RF coil used to produce diagnostic images of the carotid arteries that can be interpreted by a trained physician. The coil provides coverage of the carotid arteries and associated vasculature from the sternal notch through the internal carotid arteries at the level of cervical vertebrae C1 in Magnetic Resonance Imaging systems. The NeoCoil 3.0T 6-Channel Carotid Array Coil is designed for use with the HD series (3.0Tesla) MRI scanners manufactured by General Electric Healthcare (GEHC). Anatomic Regions: Head and neck vasculature. Nuclei Excited: Hydrogen. The indications for use are the same as for standard imaging: The GE scanner is indicated for use as an NMR device that produces images that: (1) correspond to the distribution of protons exhibiting NMR signal, (2) depend upon NMR parameters (proton density, spin lattice relaxation time T1, spin-spin relaxation time T2, and (3) display the vasculature of the head and neck regions specifically the carotid arteries and associated soft tissue. When interpreted by a trained physician, these images yield information that can be useful in the determination of a diagnosis.
Device Story
The NeoCoil 3.0T 6-Channel Carotid Array Coil is a receive-only phased array RF coil designed for high-resolution imaging of carotid arteries. It consists of two foam-covered paddles, each containing three antennas, held in place by a headband with a ball-and-socket joint for patient positioning. The device captures hydrogen proton signals during MRI procedures. It utilizes pre-amplifier decoupling and antenna overlap to minimize mutual coupling. Safety features include active PIN diode decoupling and passive crossed-diode detuning during RF transmit. Used in clinical settings by trained physicians, the coil provides high SNR and a large field-of-view for visualizing carotid lumen, vessel walls, and atherosclerotic plaques. The output is diagnostic MR images interpreted by a physician to assist in clinical diagnosis. The device is compatible with ASSET parallel imaging techniques.
Clinical Evidence
Bench testing only. Performance was evaluated through SNR and image uniformity testing to demonstrate substantial equivalence to the predicate devices.
Technological Characteristics
Phased-array receive-only RF coil; 6 elements; foam-covered paddle form factor; headband positioning system with ball-and-socket joints; active PIN diode decoupling and passive crossed-diode detuning; compatible with GE 3.0T Signa EXCITE HD series MRI scanners; supports ASSET parallel imaging.
Indications for Use
Indicated for patients requiring diagnostic MRI imaging of the head and neck vasculature, specifically the carotid arteries and associated soft tissue, from the sternal notch through the internal carotid arteries at the level of cervical vertebrae C1. Compatible with GE 3.0T HD series MRI scanners.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
USAI Millennium III 3T-8-Channel Neurovascular Coil (K042342)
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(k) Summary of Safety and Effectiveness
| Device Name: | NeoCoil 3.0T 6-Channel Carotid Array Coil |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Proprietary Name: | NeoCoil 3.0T 6-Channel Carotid Array Coil |
| | APR 11 2007 |
| Common/Usual Name: | Magnetic Resonance Specialty Coil |
| Classification Name: | Magnetic Resonance Specialty Coil |
| Classification Number: | 892.1000 |
| Classification Panel: | Radiology Device Panel |
| CDRH Product Code: | MOS |
| Regulatory Class: | II |
| Reason for 510(k): | New device |
| Applicant: | Brian Brown<br>Executive Director<br>NeoCoil<br>N27 W23910A Paul Rd<br>Pewaukee, WI 53072<br>262-347-1250 x 12 (office)<br>261-347-1251 (fax)<br>brian.brown@neocoil.com |
| Preparation date: | 12/12/2006 |
| Est. Registration No: | |
| Intended Use: | The NeoCoil 3.0T 6-Channel Carotid Array Coil is a receive only phased array RF coil used to produce diagnostic images of the carotid arteries that can be interpreted by a trained physician. The coil provides coverage of the carotid arteries and associated vasculature from the sternal notch through the internal carotid arteries at the level of cervical vertebrae C1 in Magnetic Resonance Imaging systems. The NeoCoil 3.0T 6-Channel Carotid Array Coil is designed for use with the HD series (3.0Tesla) MRI scanners manufactured by General Electric Healthcare (GEHC). Anatomic Regions: Head and neck vasculature. Nuclei Excited: Hydrogen. |
| | The indications for use are the same as for standard imaging:<br>The GE scanner is indicated for use as an NMR device that produces images that: (1) correspond to the distribution of protons exhibiting NMR signal, (2) depend upon NMR parameters (proton density, spin lattice relaxation time T1, spin-spin relaxation time T2, and (3) display the vasculature of the head and neck regions specifically the carotid arteries and associated soft tissue. When interpreted by a trained physician, these images yield information that can be useful in the determination of a diagnosis. |
| Standards:<br>Performance: | No applicable performance standards have been issued under Section 514 of the Food, Drug and Cosmetic Act. |
| Device Description: | The NeoCoil 3.0T 6-Channel Carotid Array Coil is a multi-element phased array receive only coil used for obtaining diagnostic images of the carotid arteries and associated vasculature from the sternal notch through the internal carotid arteries at the level of cervical vertebrae C1 in Magnetic Resonance Imaging Systems. Compared to predicate devices, the submitted device offers greater SNR due to its operating field strength of 3.0T, and a larger field-of-view due to the antenna layout. |
{1}------------------------------------------------
| | The submitted device consists of a dual set of foam covered "paddles",<br>consisting of three antennas each. The three antennas in each paddle are<br>uniquely positioned with the appropriate overlap to cancel out mutual coupling<br>effects from adjacent antennas. Pre-amplifier decoupling reduces any remaining<br>decoupling between the antennas. |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The paddles are held in place over the imaging area via a headband. A system<br>interface cable connects to the coil at the top of the headband. The foam<br>paddles connect to the headband using a ball and socket joint that enable proper<br>positioning of the paddles over the imaging area. |
| | To ensure safety, each antenna is equipped with two transmit decoupling<br>circuits; one active and the other passive. Active decoupling is achieved by PIN<br>diodes that receive signals from the scanner to turn the coil to a high impedance<br>state during system RF transmit. Crossed diodes are installed on each antenna<br>acting as passive switches. These passive switched detune the antennas further<br>during RF transmit. |
| Predicate Devices: | Machnet Carotids Coil Array Assembly (K012491)<br>USAI 3T HD Breast Array (K052585)<br>USAI Millennium III 3T-8-Channel Neurovascular Coil (K042342) |
| Comparison to Predicate: | It is our opinion that the NeoCoil 3.0T 6-Channel Carotid Array Coil in this<br>submission is substantially equivalent to the previously cleared Machnet Carotids<br>Coil Array Assembly (K012491), the USAI 3T HD Breast Array (K052585), and<br>the USAI Millennium III 3T-8-Channel Neurovascular Coil (K042342). Remaining<br>differences do not impact indications for use or have an impact on safety. |
| Summary of Studies: | In all material respects, the NeoCoil 3.0T 6-Channel Carotid Array is substantially<br>equivalent to the Machnet Carotids coil Assembly. SNR and image uniformity<br>testing was performed which support the conclusion that the submitted device<br>satisfies design objectives. |
| Conclusion: | The NeoCoil 3.0T 6-Channel Carotid Array is substantially equivalent to the<br>predicate device. Use of the NeoCoil 3.0T 6-Channel Carotid Array does not<br>result in any new potential hazards and does not alter the safety of the MRI<br>scanner. |
:
.
.
:
:
{2}------------------------------------------------
# Appendix F: Preliminary Product Data Sheet
:
.
.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a dense collection of text, arranged in a vertical format. The text appears to be of a small font size, making it difficult to read individual words or sentences without closer inspection. The overall impression is that of a document or page filled with a large amount of written information.
- Neocoll
# 3.0T 6-Channel Carotid Array Coll
The 3.0T &Channel Carolld Array Cost enables by Isteral diagnoctic maging of the caroed arrary bifurcation using a QE Signa® 3.0T EXCITE® HD series MR system.
#### Primary Applications
The 3.0T 6-Channel Carolid Array Coll has been designed and tested to provide high quality imaging of the carond arteries. The unique array configuration offers complete coverage head and neck vasculature from the sternal notch through the internal carolid arteries at the level of cervical vertebrae C1. The coll allows for sub-millimeter resolution of the carolids lumen, vessel walls, and atheroscierolic plaques,
Image /page/3/Picture/6 description: The image shows a pair of headphones with a white rectangular device attached to the headband. The headphones are dark in color, and the white device appears to be some sort of control or interface. A cable extends from the headphones, ending in a white connector. The headphones and cable are resting on a textured surface.
#### Technical Features
This phased-array coll has 6 elements. The element configuration is optimized to provide daep penetrabon SNR and improved Sil coverage.
The unique headband and positioning system facilitate patient seeip and fit, providing greater patient confort and imaging periormance.
The coll is companible with ASSET® parallel imaging iechniques
NeoCoil, LLC N27 W23810A Paul Rd. Penaukee, WI 53072 @ 2006 NeoColl All Rights Reserved
Coil Coverage (FOV) Supence inferor 15om Anterior / Posterior 15 cm Right | Lett 24 mm
| Coil Dimensions | | |
|-----------------|---------|---------|
| Length | 17.5 cm | 6.9 in |
| Width | 28.0 cm | 11.0 in |
| Height | 48.0 cm | 18.9 in |
| Weight | 1.5 kg | 3.4 lbs |
Image /page/3/Picture/14 description: This is a black and white image of a medical scan. The scan appears to be of the neck and upper chest area, showing the blood vessels. The vessels are highlighted in white against the dark background, allowing for a clear view of their structure and potential abnormalities.
System Requirements and Ordering Requires a GE 3.0T Signa EXCITE HD zeries MR system. Compatible with all such models. Offered by GE through catalog MTBD. or NeoCoil catalog number NC001
#### Warranty
The published Company warranty in effect on the date of shipment shall apply. The Company reserves the right to make changes
Manufactured by NecCoil, LLC.
{4}------------------------------------------------
## Third Party Review Quality Assessment
#### Section 1 - Submission Information
| 510(k) No.: | K070778 |
|---------------------------|------------------------------------|
| Third Party Organization: | Intertek Testing Services NA, Inc. |
Third Party's Primary Reviewer(s): Daniel Lehtonen
ODE/OIVD Division: DRARD Branch/Team: Radiological Devices Branch
### Section 2 - 510(k) Decision
Third party recommendation: SE √ ____ NSE _______ Other (specify): ___________________________________________________________________________________________________________
ODE/OIVD final decision: SE _ V___ NSE _____ Other (specify): _________________________________________________________________________________________________________________________________________
### Section 3 -- Assessment of Third Party Review
| Review Element | Rating (check one) | | |
|---------------------------------------------------------------------------------------------------------------------------------------------|--------------------|----------------|----------------|
| | Adequate | Minor Issue(s) | Major Issue(s) |
| a. Determination of device eligibility for third party review | √ | | |
| b. Extent of pre-submission consultation with ODE/OIVD division | | | |
| c. Organization and format of review documentation | √ | | |
| d. Determination of 510(k) administrative completeness (screening review) | √ | | |
| e. Summary of device characteristics, intended use, and performance (including accessories, if applicable) and reason for 510(k) submission | √ | | |
| f. Comparison to legally marketed devices—identification and analysis of key similarities and differences | | √ | |
| g. Rationale for conclusions and recommendation | √ | | |
| h. Use of guidance documents and standards | √ | | |
| i. Resolution of 510(k) deficiencies and FDA requests for additional information | | | |
| j. Scope of reviewer expertise and use of consulting reviewers | | | |
| k. Other (specify): | | | |
Comments (explanation of ratings/issues): The performance of the coil did not match the labeling, IFU and 510(k) summary. This was corrected by the Sponsor after I requested additional information.
### Section 4 - ODE/OIVD Assessor Information
Sunder Rajan - Date: April 5, 2007 Tel. No.: 240 276 3968 Assessed by:
Routing: Division--Clip completed assessment (this page only) to inside front cover of 510(k). DMC -- Forward this page only to Eric Rechen, POS/ODE, Rm. 120J, Corp. Blvd. (HFZ-402).
{5}------------------------------------------------
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
APR 1 2007
c/o Mr. Daniel W. Lehtonen Staff Engineer - Medical Devices and Responsible Third Party Official Intertek Testing Services NA, Inc. 2307 East Aurora Rd., Unit B7 TWINSBURG OH 44087
Re: K070778
NeoCoil
Trade/Device Name: NeoCoil 3.OT 6-Channel Carotid Array Coil Regulation Number: 21 CFR §892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: MOS Dated: March 20, 2007 Received: March 21, 2007
Dear Mr. Brown:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivelon (for the indiver use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to device that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of the evideral manufacturing practice, labeling, and prohibitions against misbranding and adultersting.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/5/Picture/9 description: The image shows the FDA Centennial logo, which is a circular emblem with the letters "FDA" in the center. The logo also includes the years "1906-2006" at the top and the word "Centennial" below the letters. Below the logo is the text "Protecting and Promoting Public Health" in a decorative font.
{6}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely yours.
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## Indications for Use
长070778 510(k) Number (if known):
Device Name: NeoCoil 3.0T 6-Channel Carotid Array Coil
#### Indications For Use:
The NeoCoil 3.0T 6-Channel Carotid Array Coil is a receive only phased array RF coil used to produce diagnostic images of the carotid arteries that can be interpreted by a trained physician. The coil provides coverage of the carotid arteries and associated vasculature from the sternal notch through the internal carotid arteries at the level of cervical vertebrae C1 in Magnetic Resonance Imaging systems. The NeoCoil 3.0T 6-Channel Carotid Array Coil is designed for use with the HD series (3.0Testa) MRI scanners manufactured by General Electric Healthcare (GEHC). Anatomic Regions: Head and neck vasculature. Nuclei Excited: Hydrogen.
The indications for use are the same as for standard imaging:
The GE scanner is indicated for use as an NMR device that produces images that: (1) correspond to the distribution of protons exhibiting NMR signal, (2) depend upon NMR parameters (proton density, spin lattice relaxation time T1, spin-spin relaxation time T2, and (3) display the vasculature of the head and neck regions specifically the carotid arteries and associated soft tissue. When interpreted by a trained physician, these images yield information that can be useful in the determination of a diagnosis.
David A. Seymour
(Division Sign-Off) Division of Reproductive, Abdomina and Radiological Devices 510(k) Number
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page
Page 17 of 65
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.