TRABECULAR METAL VERTEBRAL BODY REPLACEMENT (VBR), MODEL# 06-115, 06-155

K070754 · Zimmer Trabecular Metal Technology · MQP · Aug 31, 2007 · Orthopedic

Device Facts

Record IDK070754
Device NameTRABECULAR METAL VERTEBRAL BODY REPLACEMENT (VBR), MODEL# 06-115, 06-155
ApplicantZimmer Trabecular Metal Technology
Product CodeMQP · Orthopedic
Decision DateAug 31, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3060
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Trabecular Metal Vertebral Body Replacement System is comprised of vertebral body replacement devices intended for use in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged or unstable vertebral body due to tumor or trauma (i.e., fracture). The Trabecular Metal Vertebral Body Replacement is intended for use with supplemental internal spinal fixation systems. The Trabecular Metal Vertebral Body Replacement may be used with bone graft.

Device Story

Trabecular Metal Vertebral Body Replacement (VBR) System; porous tantalum deposited on vitreous carbon skeleton; replaces diseased/damaged vertebral bodies and adjacent discs in thoracolumbar spine (T1-L5). Used in spinal surgery by surgeons; requires supplemental internal spinal fixation. Provides structural support to collapsed/unstable vertebrae; facilitates fusion when used with bone graft. Available in multiple configurations to match patient anatomy.

Clinical Evidence

Bench testing only; mechanical performance testing and analyses conducted to demonstrate that the device meets predetermined requirements for structural integrity.

Technological Characteristics

Material: Trabecular Metal (porous tantalum on vitreous carbon skeleton). Form factor: Vertebral body replacement device in various configurations. Energy source: None (passive implant). Sterilization: Not specified.

Indications for Use

Indicated for patients requiring replacement of a collapsed, damaged, or unstable vertebral body in the thoracolumbar spine (T1-L5) resulting from tumor or trauma (fracture). Intended for use with supplemental internal spinal fixation and optional bone graft.

Regulatory Classification

Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

Submission Summary (Full Text)

{1}------------------------------------------------ K070754 Page 1 of 2. # 510(k) Summary AUG 3 1 2007 | Submitter: | Zimmer Trabecular Metal Technology, Inc.<br>10 Pomeroy Road<br>Parsippany, New Jersey 07054 | |---------------------------------------|------------------------------------------------------------------------------------------------------------------| | Contact Person: | Jennifer P. Harakal<br>Senior Specialist, Regulatory Affairs<br>Telephone: (973) 576-0133<br>Fax: (973) 884-8795 | | Date: | March 16, 2007 | | Trade Name: | Trabecular Metal™ Vertebral Body Replacement System | | Common Name: | Vertebral Body Replacement Device | | Classification Name and<br>Reference: | Spinal Intervertebral Body Fixation Orthosis<br>21 CFR § 888.3060, MOP | ## DEVICE DESCRIPTION The Trabecular Metal Vertebral Body Replacement (VBR) System is designed to be used as a replacement for a diseased or damaged vertebral body and the adjacent disc when spinal surgery is indicated. The Trabecular Metal VBR System is wholly comprised of Trabecular Metal Porous Tantalum deposited on a vitreous carbon skeleton) and is available in a variety of configurations to accommodate the anatomical requirements of different patients. #### INDICATIONS FOR USE The Trabecular Metal Vertebral Body Replacement System is comprised of vertebral body replacement devices intended for use in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged or unstable vertebral body due to tumor or trauma (i.e., fracture). The Trabecular Metal Vertebral Body Replacement is intended for use with supplemental internal spinal fixation systems. The Trabecular Metal Vertebral Body Replacement may be used with bone graft. ## DEVICE TECHNOLOGICAL CHARACTERISTICS AND COMPARISON TO PREDICATE DEVICE(S) Zimmer Trabecular Metal Technology, Inc. has submitted documentation demonstrating the substantial equivalence of the proposed System to its predicate devices. The subject {2}------------------------------------------------ K070754 Page 2 of 2 System is similar to its predicate devices with respect to intended use/indications for use, materials, and basic principles of operation. ## PERFORMANCE DATA The results of testing and analyses conducted demonstrate that the worst cases of the proposed System adequately meet the predetermined requirements established for its mechanical performance. # SUBSTANTIAL EQUIVALENCE The Trabecular Metal™ Vertebral Body Replacement System is substantially equivalent to its predicate devices with respect to intended use/indications for use, technological characteristics and basic principles of operation. As demonstrated by supporting performance data, these technological differences do not present any new issues of safety or effectiveness. {3}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 AUG 3 1 2007 Zimmer Trabecular Metal Technology % Ms. Jennifer P. Harakal Senior Regulatory Affairs Specialist 10 Pomeroy Road Parsippany, NJ 07054 Re: K070754 Trade/Device Name: Trabecular Metal™ Vertebral Body Replacement System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: MOP Dated: July 18, 2007 Received: July 19, 2007 Dear Ms. Harakal: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ Page 2 - Ms. Jennifer P. Harakal This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Mark A. Milken Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ #### Indications for Use 510(k) Number (if known): _ K0 70754 Device Name: Trabecular Metal" Vertebral Body Replacement System Indications for Use: The Trabecular Metal Vertebral Body Replacement System is comprised of vertebral body replacement devices intended for use in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged or unstable vertebral body due to tumor or trauma (i.e., fracture). The Trabecular Metal Vertebral Body Replacement is intended for use with supplemental internal spinal fixation systems. The Trabecular Metal Vertebral Body Replacement may be used with bone graft. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) # Concurrence of CDRH, Office of Device Evaluation (ODE) Page _1_ of _1_ Marle A. Miller (Division Sign-Off) (Division Sign of General, Restorative, and Neurological Devi **510(k) Number**
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...