K070726 · Barcoview, Medical Imaging Systems · LLZ · May 2, 2007 · Radiology
Device Facts
Record ID
K070726
Device Name
NIO COLOR 3MP AND MDNC 3120
Applicant
Barcoview, Medical Imaging Systems
Product Code
LLZ · Radiology
Decision Date
May 2, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Indications for Use
The MDNC 3120 is intended to be used in displaying and viewing digital images for review by trained medical practitioners. These devices must not be used in primary image diagnosis in mammography. The MDNC 3120, which is part of the Nio Color 3MP system, will be marketed separately.
Device Story
Nio Color 3MP is a medical display system comprising a 20.8" color LCD (MDNC 3120), a high-resolution display controller board (e.g., MXRT 2100, 5100, or 7100), and MediCal QAWeb Agent software. The system connects to a PACS workstation to display digital medical images for review by trained practitioners. The display controller board utilizes a 32-bit lookup table to provide up to 64-bit color. The MediCal QAWeb Agent software enables local calibration, quality assurance, and setting of display test patterns to ensure DICOM-compliant viewing. The system supports both landscape and portrait orientations. It is intended for clinical environments where medical images are reviewed; it does not contact patients and is not used for life-sustaining functions. The output allows clinicians to visualize diagnostic images for review purposes, aiding in clinical decision-making.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
20.8" color LCD flat panel display; 2048 x 1536 pixel resolution; 32-bit display controller board with 32-bit in/out lookup table; supports up to 64-bit color. Software: MediCal QAWeb Agent for calibration and QA. Connectivity: Plugs into PACS workstation. Orientation: Landscape/Portrait. Non-patient contacting.
Indications for Use
Indicated for trained medical practitioners to display and view digital images for review. Contraindicated for primary image diagnosis in mammography.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
# 510(K) SUMMARY
In accordance with 21 CFR 807.92
### 1. Date of preparation
March, 8 2007
#### 2. Company information
BarcoView 35 President Kennedypark B-8500 Kortrijk, Belgium Tel. +32-(0)56-233-211 Fax +32-(0)56-233-457
### 3. Contact person
Lieven De Wandel Official correspondent
#### 4. Device information
- . Trade name: MDNC 3120
- . Common name: Display system, medical image workstation, and others
- . Classification name: System, Image Processing
- . Classification number: 21 CFR 892.2050 / Procode 90LLZ
#### 5. Predicate device
- . Name: E-2320C
- . 510(k) number: K052958
- . Manufacturer: Barco NV
### 6. Device description
MDNC 3120 is a 20.8" color LCD display for medical viewing. It is combined with MediCal QAWeb Agent, a user-friendly software that allows to optimize the display for DICOM-compliant viewing.
### 7. Intended use
"The MDNC 3120 is intended to be used in displaying and viewing digital images for review by trained medical practitioners. These devices must not be used in primary image diagnosis in mammography. The MDNC 3120, which is part of the Nio Color 3MP system, will be marketed separately.
### 8. Summary of technological characteristics
K070726
MAY - 2 2007
{2}------------------------------------------------
The flat panel display has a resolution of 2048 x 1536 pixels. It can be used in landscape or portrait mode.
The MediCal QAWeb Aqent software allows to set the display function, display test patterns, calibrate the display and view additional display controller information.
Compared to the predicate device, the MDNC 3120 display has a different LCD panel, other electronic and mechanical parts. The basic specifications and functions, however, are the same.
The device does not come into contact with the patient. It does not control any life sustaining devices either.
#### 9. Conclusion:
The Barco MDNC 3120 is substantially equivalent to the predicate device, E-2320C. The new and predicate devices are substantially equivalent in the areas of technical characteristics, general function, application and intended use.
Any difference between both devices does not affect safety or efficacy.
The 510(k) Pre-Market Notification for the Barco MDNC 3120 contains adequate information and data to enable FDA - CDRH to determine substantial equivalence to the predicate device.
{3}------------------------------------------------
# 510(K) SUMMARY
In accordance with 21 CFR 807.92
# 1. Date of preparation
March, 8 2007
### 2. Company information
BarcoView 35 President Kennedypark B-8500 Kortrijk, Belgium Tel. +32-(0)56-233-211 Fax +32-(0)56-233-457
# 3. Contact person
Lieven De Wandel Official correspondent
### 4. Device information
- . Trade name: Nio Color 3MP
- Common name: Display system, medical image workstation, and others .
- Classification name: System, Image Processing .
- . Classification number: 21 CFR 892.2050 / Procode 90LLZ
### 5. Predicate device
- . Name: Color Nio 2MP
- . 510(k) number: K052958
- . Manufacturer: Barco NV
# 6. Device description
Nio Color 3MP is a display system for medical viewing. It consists of 3 components:
MDNC 3120 is a 20.8" color LCD display. MediCal QAWeb Agent is a softcopy QA software application for local calibration and QA control. The system also contains a fast high-resolution display controller board that plugs into a PACS workstation computer. Standard the system is delivered with the MXRT 2100 board, but the user can opt for another type of display controller board (MXRT 5100 or MXRT 7100)
् द्व
The display system can be a single-head system or multi-head system. In the last case it contains multiple displays and display controller boards.
# 7. Intended use
"The Nio Color 3MP is intended to be used in displaying and viewing digital images for review by
{4}------------------------------------------------
trained medical practitioners. These devices must not be used in primary image diagnosis in mammography. The Nio Color 3MP, containing the display MDNC 3120, the software MediCal QAWeb Agent and a graphic board, will be marketed as separate device.
#### 8. Summary of technological characteristics
The device consists of three components:
- One 3-megapixel flat panel display (MDNC 3120) .
- . One 32-bit display controller
- MediCal QAWeb Agent software .
The flat panel display has a resolution of 2048 x 1536 pixels. It can be used in landscape and portrait mode.
The display controller board is an ultra-high speed board with a 32-bit in, 32-bit out lookup table, providing up to 64 bit color.
The MediCal QAWeb Aqent software allows to set the display test patterns. calibrate the display and view additional display controller information.
Compared to the predicate device, the display from the Nio Color 3MP system has a different LCD panel, other electronic boards and other mechanical parts. The software is a new version. However, the basic specifications and functions of all the parts are the same.
The device does not come into contact with the patient. It does not control any life sustaining devices either.
#### 9. Conclusion:
The Barco Nio Color 3MP is substantially equivalent to the predicate device, Color Nio 2MP.
i, i.
The new and predicate devices are substantially equivalent in the areas of technical characteristics, general function, application and intended use.
Any difference between both devices does not affect safety or efficacy.
The 510(k) Pre-Market Notification for the Barco Nio Color 3MP contains adequate information and data to enable FDA - CDRH to determine substantial equivalence to the predicate device.
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in all caps and is in a sans-serif font. The text is black and is on a white background. The text is centered in the image.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
Mr. Lieven De Wandel Official Correspondent BarcoView Medical Imaging Systems President Kennedypark 35 B-8500 Kortrijk BELGIUM
MAY - 2 2007
K070726 Re:
Trade/Device Name: Nio Color 3MP and MDNC 3120 Regulation Number: 21 CFR §892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: March 8, 2007 Received: March 15, 2007
Dear Mr. De Wandel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition FDA may publish further announcements concerning your device in the Federal Register.
Image /page/4/Picture/9 description: The image is a circular logo with the text "1906-2006" at the top and "Centennial" at the bottom. The letters "FDA" are prominently displayed in the center of the logo. There are three stars at the bottom of the logo, and the entire design is surrounded by a dotted border.
Protecting and Promoting Public Health
{6}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all . the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address
http://www.fda.gov/cdrh/industry/support/index.html
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director. Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### INDICATIONS FOR USE
510(k) Number (if known): K010226
Device Name: Nio Color 3MP
Indications for Use:
"The Nio Color 3MP is intended to be used in displaying and viewing digital images for review by trained medical practitioners. These devices must not be used in primary image diagnosis in tramou mounty. The Nio Color 3MP, containing the display MDNC 3120, the software MediCal QAWeb Agent and a graphic board, will be marketed as separate device.
Prescription Use __XX_
(Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy Brogdon
(Division Sign-Off Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number _
{8}------------------------------------------------
# INDICATIONS FOR USE
510(k) Number (if known): K076226
Device Name: MDNC 3120
Indications for Use:
"The MDNC 3120 is intended to be used in displaying and viewing digital images for review by trained medical practitioners. These devices must not be used in primary image diagnosis in mammography. The MDNC 3120, which is part of the Nio Color 3MP system, will be marketed separately.
Prescription Use XX
(Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy Chrogdon
Dinsion Sian-Off ann of Reproductive, Abdo Anlogical Devices neiki Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.