K070695 · Zimmer, Inc. · KRR · Jun 7, 2007 · Orthopedic
Device Facts
Record ID
K070695
Device Name
ZIMMER PATELLOFEMORAL JOINT PROSTHESIS
Applicant
Zimmer, Inc.
Product Code
KRR · Orthopedic
Decision Date
Jun 7, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3540
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
-Osteoarthritis, traumatic arthritis, polyarthritis, and/or severe chondrocalcinosis of the patellofemoral joint; -The salvage of previously failed surgical attempts (e.g., arthroscopy, lateral release, cartilage transplantation); -History of patellar dislocation or patella fracture; -Dysplasia-induced degeneration This device is intended for cemented use only.
Device Story
Zimmer Patellofemoral Joint (PFJ) Prosthesis; knee prosthesis component designed to replicate femoral patellar groove anatomy. Incorporates NexGen knee system frontal profile geometry for patella tracking; accommodates resurfaced and unresurfaced patellas. Available in five sizes; left and right configurations. Implanted by orthopedic surgeons in clinical settings to address patellofemoral joint degeneration or failure of previous surgeries. Device provides mechanical replacement of articulating surface; intended for cemented fixation. Benefits include restoration of joint function and stability for patients with specified arthritic or traumatic conditions.
Clinical Evidence
Bench testing only. No clinical data provided. Non-clinical performance testing demonstrated the device meets requirements and is as safe and effective as the predicate devices.
Technological Characteristics
Knee joint patellofemoral polymer/metal semi-constrained cemented prosthesis. Design replicates femoral patellar groove anatomy using NexGen knee system frontal profile geometry. Five sizes available in left/right configurations. Cemented fixation.
Indications for Use
Indicated for patients with osteoarthritis, traumatic arthritis, polyarthritis, severe chondrocalcinosis of the patellofemoral joint, dysplasia-induced degeneration, history of patellar dislocation or fracture, or salvage of failed prior surgical attempts (e.g., arthroscopy, lateral release, cartilage transplantation). Intended for cemented use only.
Regulatory Classification
Identification
A knee joint patellofemoral polymer/metal semi-constrained cemented prosthesis is a two-part device intended to be implanted to replace part of a knee joint in the treatment of primary patellofemoral arthritis or chondromalacia. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes a component made of alloys, such as cobalt-chromium-molybdenum or austenitic steel, for resurfacing the intercondylar groove (femoral sulcus) on the anterior aspect of the distal femur, and a patellar component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those devices intended for use with bone cement (§ 888.3027). The patellar component is designed to be implanted only with its femoral component.
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Applications for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,” and
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-Aluminum 4-Vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-Chromium-Molybdenum Casting Alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-Chromium-Molybdenum Alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,”
(vii) ISO 7207-2:1998 “Implants for Surgery—Components for Partial and Total Knee Joint Prostheses—Part 2: Articulating Surfaces Made of Metal, Ceramic and Plastic Materials,” and
(viii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal Coatings,”
(vii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants,” and
(viii) F 1672-95 “Specification for Resurfacing Patellar Prosthesis.”
{0}------------------------------------------------
K070695
# JUN - 7 2007
# Summary of Safety and Effectiveness
| Submitter: | Zimmer, Inc.<br>P.O. Box 708<br>Warsaw, IN 46581-0708 |
|----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Dalene T. Binkley<br>Senior Associate, Regulatory Affairs<br>Telephone: (574) 372-4907<br>Fax: (574) 372-4605 |
| Date: | June 4, 2007 |
| Trade Name: | Zimmer® Patellofemoral Joint Prosthesis |
| Common Name: | Knee Prosthesis Component |
| Classification Name<br>and Reference: | Knee joint patellofemoral polymer/metal semi-<br>constrained cemented prosthesis<br>21 CFR § 888.3540 |
| Predicate Device: | Stryker Compartmental Knee System, manufactured<br>by Howmedica Osteonics Corp., K052917, cleared<br>December 27, 2005<br><br>Natural-Knee® II Patellofemoral Joint Prosthesis,<br>manufactured by Zimmer, Inc., K002356, cleared<br>October 30, 2000<br><br>NexGen® Knee Gender Solutions Female Femoral<br>Components, manufactured by Zimmer, Inc.,<br>K060370, cleared April 28, 2006 |
| Device Description: | The Zimmer Patellofemoral Joint (PFJ) Prosthesis is<br>designed to closely replicate the anatomic features<br>of the patellar groove on the femur. The implant's<br>articulating surface incorporates the NexGen® knee<br>system's frontal profile geometry for optimal patella<br>tracking for both resurfaced and unresurfaced<br>patellas. The construct of the PFJ prosthesis<br>accommodates a wide range of patients, both female<br>and male. Five sizes in left and right configurations |
| Intended Use: | -Osteoarthritis, traumatic arthritis, polyarthritis,<br>and/or severe chondrocalcinosis of the<br>patellofemoral joint;<br>-The salvage of previously failed surgical attempts<br>(e.g., arthroscopy, lateral release, cartilage<br>transplantation);<br>-History of patellar dislocation or patella fracture;<br>-Dysplasia-induced degeneration<br>This device is intended for cemented use only. |
| Comparison to Predicate Device: | The Zimmer Patellofemoral Joint Prosthesis is<br>packaged, manufactured, and sterilized using the<br>same materials and processes as the predicate<br>devices. The subject device also has the same<br>intended use and similar fixation methods as the<br>predicate devices. |
| Performance Data (Nonclinical<br>and/or Clinical): | Non-Clinical Performance and Conclusions:<br>Non-clinical testing demonstrated that the Zimmer<br>Patellofemoral Joint Prosthesis met performance<br>requirements and is as safe and effective as the<br>predicate devices. |
:
{1}------------------------------------------------
are being offered for this implant.
:
.
{2}------------------------------------------------
Public Health Service
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular fashion around the caduceus symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
#### JUN - 7 2007
Zimmer, Inc. % Ms. Dalene T. Binkley M.S., RAC Senior Associate, Corporate Regulatory Affairs P.O. Box 708 Warsaw, Indiana 46581-0708
Re: K070695 Trade/Device Name: Zimmer® Patellofemoral Joint Prosthesis Regulation Number: 21 CFR 888.3540 Regulation Name: Knee joint patellofemoral polymer/metal semi-constrained Cemented prosthesis Regulatory Class: Class II Product Code: KRR Dated: March 12, 2007 Received: March 13, 2007
Dear Ms. Binkley:
We have reviewed your Section 510(k) premarket notification of intent to market the devices referenced above and have determined the devices are substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the devices, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your devices are classified (see above) into either class II (Special Controls) or class III (PMA), they may be subject to such additional controls. Existing major regulations affecting your devices can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your devices in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your devices comply with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 – Ms. Dalene T. Binkley M.S., RAC
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic forth in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your devices as described in your Section 510(k) rms lotter will and my Jorve FDA finding of substantial equivalence of your devices to a legally premation notificate device results in a classification for your devices and thus, permits your devices to proceed to the market.
If you desire specific advice for your devices on our labeling regulation (21 CFR Part 801), please If you desire specific acreed at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small general informational and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or on the Internet address http://www.fda.gov/odrh/industry/support/index.html.
Sincerely yours,
Carbone Buehr
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
K070695
## 510(k) Number (if known):
Device Name:
Zimmer® Patellofemoral Joint Prosthesis
Indications for Use:
-Osteoarthritis, traumatic arthritis, polyarthritis, and/or severe
chondrocalcinosis of the patellofemoral joint;
chondrocalchiosis of the patchoremoral joints,
-The salvage of previously failed surgical attempts (e.g., arthroscopy, lateral release, cartilage transplantation),
-History of patellar dislocation or patella fracture;
-Dysplasia-induced degeneration
This device is intended for cemented use only.
Prescription Use _X_ (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(Please do not write below this line - Continue on another page if needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
tenbas breens
Division of General, Restorative, and Neurological Devices
Page 1 of 1
510(k) Number k070695
009
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.