FLUOROPERM 92, 60, 30 AND 151 AND PARAGON HDS, HDS 100 AND THIN RIGID GAS PERMEABLE CONTACT LENSES
Applicant
Paragon Vision Sciences
Product Code
HQD · Ophthalmic
Decision Date
May 1, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5916
Device Class
Class 2
Indications for Use
FluoroPerm® 92 rigid gas permeable spherical or aspheric contact lenses are indicated for daily wear and extended wear from 1 to 7 days between removals for cleaning and disinfection as recommended by the eye care practitioner. FluoroPerm® 92 rigid gas permeable bifocal, or toric contact lenses are indicated for daily wear only. FluoroPerm® 92 rigid gas permeable spheric and bifocal contact lenses are indicated for the correction of visual acuity in not-aphakic persons with nondiseased eyes that are nearsighted (myopic), farsighted (hyperopic), and may exhibit corneal astigmatism up to 4.00 diopters that does not interfere with visual acuity. FluoroPerm® 92 toric contact lenses are indicated to correct astigmatism of up to 6.00 diopters. FluoroPerm® 92 bifocal lenses are indicated to treat presbyopia up to +4.00 D add power. FluoroPerm® 92 contact lenses are indicated for management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty, radial keratotomy, or LASIK surgery in otherwise nondiseased eyes. [Note: Similar statements provided for FluoroPerm® 60, Paragon HDS®, FluoroPerm® 30, Paragon Thin™, FluoroPerm® 151, and Paragon HDS® 100, with varying wear schedules.]
Device Story
Rigid gas permeable (RGP) contact lenses; lathe-cut fluorosilicone acrylate polymers; daily or extended wear (1-7 days) depending on specific lens model. Corrects refractive errors and manages irregular corneal conditions. Practitioner-prescribed; patient-worn. Lenses feature spherical, aspheric, bifocal, or toric designs. May include UV-blocking additives (Uvinul D-49) and color tints (D&C dyes) for handling. Provides visual acuity correction and corneal surface regularization.
Clinical Evidence
No clinical data provided; substantial equivalence based on material safety and performance characteristics of previously approved RGP lens materials.
Technological Characteristics
Lathe-cut rigid gas permeable contact lenses; materials: paflufocon A, B, C, D (fluorosilicone acrylate polymers). Features: spherical, aspheric, bifocal, or toric designs; optional UV absorber (Uvinul D-49) blocking up to 97% of light <380 nm; optional color tints (D&C Green No. 6, Yellow No. 9, Violet No. 2, Red No. 17).
Indications for Use
Indicated for non-aphakic patients with nondiseased eyes for correction of myopia, hyperopia, and corneal astigmatism (up to 4.00D, or up to 6.00D for toric). Bifocal lenses indicated for presbyopia (up to +4.00D add). Also indicated for irregular corneal conditions including keratoconus, pellucid marginal degeneration, or post-surgical (penetrating keratoplasty, radial keratotomy, LASIK).
Regulatory Classification
Identification
A rigid gas permeable contact lens is a device intended to be worn directly against the cornea of the eye to correct vision conditions. The device is made of various materials, such as cellulose acetate butyrate, polyacrylate-silicone, or silicone elastomers, whose main polymer molecules generally do not absorb or attract water.
Predicate Devices
Boston® XO (hexafocon A)
Boston EO® (enflufocon B)
Boston ES® (enflufocon A)
Submission Summary (Full Text)
{0}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/0/Picture/2 description: The image shows a logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its wings, and the text "U.S. Department of Health & Human Services" is arranged in a circle around the eagle. The logo is black and white.
MAY - 1 2007
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Paragon Vision Sciences William E. Mevers, Ph.D. Vice President, Science & Technology 947 East Impala Mesa, AZ 85204-6619
Re: K070637
FluoroPerm® 30 and Paragon Thin™ (paflufocon C); Trade/Device Name: FluoroPerm® 60 and Paragon HDS® (paflufocon B): FluoroPerm® 92 (paflufocon A); and, FluoroPerm® 151 and Paragon HDS® 100 (paflufocon D) RGP Contact Lenses Regulation Number: 21 CFR 886.5916 Regulation Name: Rigid gas permeable contact lenses Regulatory Class: Class II Product Code: HQD Dated: March 5. 2007 Received: March 7, 2007
Dear Dr. Meyers:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Page 2 - Dr. William Meyers
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address (car) supply industr. Support inationali.
Sincerely yours,
M.B. Egclehner SiwD
Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications For Use
K070637
510(k) Number (if known):
Device Name:
FluoroPerm® 92 (paflufocon A) rigid gas permeable contact lenses FluoroPerm® 60 (paflufocon B) rigid gas permeable contact lenses Paragon HDS® (paflufocon B) rigid gas permeable contact lenses FluoroPerm 30 (paflufocon C) rigid gas permeable contact lenses FluoroPerm® 151 (paflufocou D) rigid gas permeable contact lenses Paragon HDS® 100 (paflufocon D) rigid gas permeable contact lenses
#### Indications For Use:
#### FluoroPerm® 92
FluoroPerm® 92 rigid gas permeable spherical or aspheric contact lenses are indicated for daily wear I horor chil. 92 ngla ga permission between removals for cleaning and disinfection as recommended by the eye care practitioner. FluoroPerm® 92 rigid gas permeable bifocal, or toric contact lenses are indicated for daily wear only.
FluoroPerm® 92 rigid gas permeable spheric and bifocal contact lenses are indicated for the correction of visual acuity in not-aphabic persons with nondiseased eyes that are nearsighted (myopic), farsighted (byperopic), and may exhibit corneal astigmatism up to 4.00 diopters that does not interfere with visual acuity. FluoroPerm® 92 toric contact lenses are indicated to correct astigmatism of up to 6.00 diopters. FluoroPerm® 92 bifocal leases are indicated to treat presbyopia up to +4.00 D add power.
FluoroPerm® 92 contact lenses are indicated for management of irregular corneal conditions such as kerator ones, pellucid marginal degeneration, or following penctrating keratoplasty, radial keratotomy, or LASIK surgery in otherwise nondiseased eyes.
# FluoroPerm® 60
FluoroPerm® 60 rigid gas permeable spherical or aspheric contact lenses are indicated for daily wear I tobior and out from 1 to 7 days between removals for cleaning and disinfection as recommonded by the eye care practitioner. FluoroPerm® 60 rigid gas permeable bifocal, or toric contact leases are indicated for daily wear only.
FluoroPerm® 60 rigid gas permeable spheric and bifocal contact leases are indicated for the correction of visual acuity in not-aphakic persons with nondiseased eyes that are nearsighted (myopic), farsighted (byperopic), and may exhibit corneal astigmatism up to 4.00 diopters that does not interfece with visual acuity. FivoroPerm® 60 toric contact lenses are indicated to correct astigmatism of up to 6.00 diopters. FluoroPerm® 60 bifocal lenses are indicated to treat presbyopia up to +4.00 D add power.
FluoroPerm® 60 contact leases are indicated for management of irregular comcal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoriasty, radial keratotomy, or LASIK surgery in otherwise nondiseased eyes.
Page 1 of 3
{3}------------------------------------------------
# Paragon HDS40
Paragon HDS® rigid gas permeable spheric contact lenses are indicated for daily wear and extended wear from 1 to 7 days between removals for cleaning and disinfection as recommended by the cye care practitioner. Paragon HDS® rigid gas permeable bifocal, or toric contact lenses are indicated for daily wear only.
Paragon HDS rigid gas permeable spherical, asphenc and bifocal contact lenses are indicated for the correction of visual aculty in not-aphakic persons with nondiseased eyes that are nearsighted (myopic), farsighted (hyperopic), and may exhibit corneal astigmatism up to 4.00 diopters that does not interfere with visual acuity. Paragon HDS tone contact leases are indicated to correct astigmatism of up to 6.00 diopters. Faragon HDS® bifocal lenses are indicated to treat presbyopia up to +4.00 D add power.
Paragon HDS contact lenses are indicated for management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty, radial keratotorny, or LASIK surgeryin otherwise nondiseased eyes.
# FluoroPerm® 30
FluoroPerm® 30 rigid gas permeable contact leases are indicated for daily wear as recommended by the eye care practitioner.
FluoroPerm* 30 rigid gas permeable spheric and bifocal contact lenses are indicated for the correction of visual acuity in not-aphakic persons with nondiseased eyes that are nearsighted (myopic), farsighted (hyperopic), and may exhibit corneal astigmatism up 10 4.00 diopters that does not interfere with visual acuity. FluoroPerm® 30 toric contact lenses are indicated to correct astigmatism of up 10 6.00 diopters. FluoroPerm* 30 bifocal lenses are indicated to treat presbyopia up to +4.00 D add power.
FluoroPerm® 30 contact lenses are indicated for management of irregular comeal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty, radial keratotomy, or LASIK surgery in otherwise nondiseased eyes.
## Paragon Thin"
Paragon Thin " rigid gas permeable contact lenses are indicated for daily wear as recommended by the eve care practitioner.
Paragon Thin rigid gas permeable spheric and bifocal contact lenses are indicated for the correction of visual acuity in not-aphakic persons with nondiseased tyes that are nearsighted (myopic), farsighted (hyperopic), and may exhibit comes astigmatism up to 4.00 diopters that does not interfere with visual acuity. Paragon Thin" witc contact lenses are indicated to correct astigmatism of up to 6.00 diopters. Paragon Thin bifocal lenses are indicated to treat presbyopia up 10 +4.00 D add power.
Paragon Thin" contact lenses are indicated for management of iregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty, radial keratotomy, or LASIK surgery in otherwise nondiseased eyes.
Page 2 of 3
{4}------------------------------------------------
# FluoroPerm® 151
FluoroPerm® 151 rigid gas permeable spheric contact lenses are indicated for daily wear and extended wear from 1 to 7 days between removals for cleaning and disinfection as recommended by the eye care practitioner. FluoroPerme 151 rigid gas permeable bifocal, or toric contact lenses are indicated for daily wear only.
FluoroPerm 151 rigid gas permeable spherical, aspheric and bifocal contact lenses are indicated for the correction of visual acuity in not-aphakic persons with nondiseased cycs that are nearsighted (myopic), farsighted (hyperopic), and may exhibit comeal astigmatism up to 4.00 diopters that does not interfere with visual acuity. FluoroPenn® 151 toric contact lenses are indicated to correct astigmatism of up to 6.00 diopers. FluoroPerm 151 bifocal lenses are indicated to treat presbyopia up to +4.00 D add power.
FluoroPerm 151 contact lenses are indicated for management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty, radial keratotomy, or LASIK surgery in otherwise nondiseased eyes.
#### Paragon HDS 100
Paragon HDS 100 rigid gas permeable spheric contact lenses are indicated for daily wear and extended wear from 1 to 7 days between removals for cleaning and disinfection as recommended by the eye care practitioner. Paragon HDS 100 rigid gas permeable bifocal, or toric contact lenses are indicated for daily wear only.
Paragon HDS 100 rigid gas permeable spheric and bifocal contact lenses are indicated for the correction of visual scuity in not-aphakic persons with nondiseased eves that are nearsighted (myopic), farsighted (hyperopic), and may exhibit comeal astigmatism up to 4.00 diovers that docs not interfere with visual acuity. Paragon HDS® 100 toric contact lenses are indicated to correct astigmatism of up to 6.00 diopters. Paragon HDS 100 bifocal lenses are indicated to treat presbyopia up to +4.00 D add power.
Paragon HDS 100 contact leases are indicated for management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following peactrating keratoplasty, radial keratotomy, or LASIK surgery in otherwise nondiseased eyes.
Prescription Use X (Part 21 CFR 801 Sub) bart D)
Over -
(2) CFR
Over - the - Counter Use (2) CFR 801 Subpan C).
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IS NEEDED)
AND/OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
Marc Roblez
0
(Division Sign-Off)
Division of Ophthalmic Ear,
Nose and Throat Devises
Page 3 of 3
umbal
{5}------------------------------------------------
# 510(k) Summary of Safetv and Effectiveness
MAY - 1 2007
# Submitter Information
Paragon Vision Sciences 947 E. Impala Ave. Mesa, AZ 85204
Contact Person:
William Meyers PhD Vice President, Science & Technology
Email: wemevers@MSN.com Phone: 480 507 7606
## Device Name
FluoroPerm® 92 (patlufocon A) rigid gas permeable contact lenses FluoroPerm® 60 (paflufocon B) rigid gas permeable contact lenses Paragon HDS (paflufocon B) rigid gas permeable contact lenses FluoroPerm 30 {paflufocon C) rigid gas permeable contact lenses " (paflufocon C) rigid gas permeable contact lenses Paragon Thin FluoroPerm 151 (paflufocon D) rigid gas permeable contact lenses Paragon HDS* 100 (paflufocon D) rigid gas permeable contact lenses
## Reason for Submission
Expanded Indication
## Predicate Devices
Classification Name: rigid gas permeable (hydrophobic) contact lens
Proprietary Names: Boston® XO* (chexafocon A),
Boston EO® (enflufocon EO® (enflufocon B)
Boston ES* (enflufocon A)
### Description of the device:
FluoroPerm® 92, FluoroPerm® 60, Paragon HDS*, FluoroPerm* 30, Paragon Thin"*, and FluoroPerm® 151, and Paragon HDS® 100 lenses are available as lath cut rigid gas permeable contact lenses for daily wear only. The lenses are manufactured from these FDA approved contact lens materials; pastufocon A (FluoroPerm" 02), pallufocon B (FluoroPerm) 60, Paragon HDS®), patlufocon C (Fluorosco (Film The Paragon Time), and Poly San Poly San Parised Bron fluorosilicone acrylate monomers.
The lenses may be tinted to offer a handling and for locating the lens. The tinted lenses contain one or more of the following approved color additives: D&C Green No. 6, Peroxide Yellow No. 9, D&C Violet No. 2 and D&C Red No. 17.
The lenses may be available with an ultraviolet absorber (not in all colors and materials). The ultraviolet absorber, Uvinul D-49, has been integrated as an additive within the polymer matrix, blocking up to 97% of light below 380 nm. The UV absorber is 2.2'-dihydroxy-4,4'-dimethoxybenzophenone.
The lens designs have a posterior surface consisting of a base curve and a series of up to four annular spherical or aspherical curves peripheral to the base curve.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.