K070596 · Sharn, Inc. · CCT · Jun 21, 2007 · Anesthesiology
Device Facts
Record ID
K070596
Device Name
TOPICAL APPLICATOR
Applicant
Sharn, Inc.
Product Code
CCT · Anesthesiology
Decision Date
Jun 21, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5170
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Intended for atomizing topical anesthetics to the oropharynx and upper airway regions
Device Story
Topical applicator delivers topical anesthetic solutions to oropharynx, upper airway, nasal passages, or endotracheal tubes. Device consists of a syringe, flexible semi-rigid tube, and atomizer tip/nozzle. Clinician attaches syringe to tube; manual depression of piston generates fine particle spray mist. Used in hospitals and sub-acute settings. Device allows user-controlled dosage. Absence of internal malleable wire (found in some predicates) is compensated by K-resin tubing material, allowing clinician to shape tube for anatomical fit. Benefits include targeted delivery of anesthetic to specific sites while minimizing inhalation risk to alveolar regions due to larger particle size compared to nebulizers.
Clinical Evidence
Bench testing only. Comparative performance testing against predicate using cascade impactor measured total dose, residual mass, simulated clinical dose percentage, particle size, and MMAD. Plume geometry evaluated via digital imaging and photographic grid measuring spray distance, velocity, and dispersion angle over time.
Technological Characteristics
Materials: Acetal/Delrin, K-resin, Polyethylene, Polyvinylchloride (PVC). Biocompatibility: ISO 10993-1. Sensing/Actuation: Manual piston syringe. Form factor: Flexible, semi-rigid tube with atomizer tip. Connectivity: None. Sterilization: Not stated (disposable).
Indications for Use
Indicated for atomizing topical anesthetics to the oropharynx and upper airway regions in individuals requiring topical anesthesia at the direction and discretion of a clinician.
Regulatory Classification
Identification
A laryngotracheal topical anesthesia applicator is a device used to apply topical anesthetics to a patient's laryngotracheal area.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the text "K070596" in a handwritten style. Below this, the words "Topical Applicator" are printed in a smaller, more formal font. The text appears to be part of a label or identifier, possibly related to a product or item.
#### Non-Confidential Summary of Safety and Effectiveness Page 1 of 2
25-May-07
| JUN 2 | 1 | 2007 |
|-------|---|------|
|-------|---|------|
Sharn, Inc. 4517 George Rd. # 200 Tel - 813-889-9614 Fax - 813-886-2701 Tampa, FL 33634 Official Contact: Bruce Tomlinson - President Proprietary or Trade Name: Topical applicator Common/Usual Name: Atomizer Classification Name: Laryngo-Tracheal topical applicator Classification Code: CCT - 21 CFR 868.5170 Device: Topical Applicator Predicate Devices: Wolfe Tory - MADgic - K002255 Sherwood Monoject syringe - K852544 Device Description: The Topical applicator incorporates a small tube with atomizing tip which can connect to a syringe and when placed in the patient's mouth or throat and / or inside an endotracheal tube or in the nasal passages it then delivers topical anesthetic solutions. Indications for Use: Intended for atomizing topical anesthetics to the oropharynx and upper airway regions Environment of Use: Hospitals, sub-acute settings where topical solutions are needed. Patient Population: Individuals requiring topical anesthetic at the direction and discretion of the clinician.
#### Differences Between Other Legally Marketed Predicate Devices
The Sharn Topical Applicator when compared in performance testing for particle size and plume geometry to the predicate, Wolfe Tory MADgic - K002255 was found to be substantially equivalent.
There are no significant differences that affect the safety or effectiveness of the intended device as compared to the predicate devices.
Page 6
pg 1 of 4
{1}------------------------------------------------
:
K070596
Topical Applicator
# Non-Confidential Summary of Safety and Effectiveness
25-May-07
| Features | Predicate<br>Wolfe Tory MADgic K002255 | Proposed Device |
|-----------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use | Intended for the application of<br>topical anesthetics to the<br>oropharynx and upper airway region | Intended for the application of topical<br>anesthetics to the oropharynx and<br>upper airway region |
| Types of medications used | Topical solutions, i.e.,<br>Topical anesthetics<br>Vasoconstrictors | Topical solutions, i.e.,<br>Topical anesthetics<br>Vasoconstrictors |
| Environment of Use /<br>Patient population | Not disclosed | Locations where topical anesthetics<br>solutions are used.<br>Individuals requiring topical<br>anesthetic at the direction and<br>discretion of the clinician. |
| Disposable, non-sterile | Yes | Yes |
| Components | Syringe<br>Flexible, malleable, semi-rigid tube<br>Atomizer tip / nozzle | Syringe (k852544)<br>Flexible, semi-rigid tube<br>Atomizer tip / nozzle |
| Places inserted | Nasal, Orotrachael | Nasal. Orotrachael |
| Dosage Amount | User controlled | User controlled |
| Delivery form | Fine particle spray mist | Fine particle spray mist |
| Spray generated by | Piston syringe | Piston syringe |
| Basic dimensions | Tip diameter - 0.157" OD<br>Tube length - 8.25" overall | Tip diameter - 0.31" OD (7.9 mm)<br>Tube length - 8.25" overall |
| Materials | Polycarbonate<br>Polyvinylchloride (PVC) | Acetal/ Delrin, K-resin<br>Polyethylene, Polyvinylchloride |
| Biocompatibility | Unknown | Tested to ISO 10993-1 |
| Performance testing | Comparative using Cascade impactor to measure the following listed<br>criteria | |
| Total Dose | 2.88 cc - Lidocaine | 2.75 cc - Lidocaine |
| Delivered max 3 cc | 2.87 cc - Saline | 2.78 cc - Saline |
| Residual Mass (cc) Ave. | 0.12 cc | 0.25 cc |
| Simulated Clinical Dose % | 94.7% - Lidocaine<br>95.0% - Saline | 92.3% - Lidocaine<br>94.0% - Saline |
| Particle Size | 470 um - Lidocaine<br>470 um - Saline | 835 um - Lidocaine<br>803 um - Saline |
| MMAD (um) | 1.87 um - Lidocaine<br>1.90 um - Saline | 1.93 um - Lidocaine<br>2.00 um - Saline |
| GSD (um) | Similar | Similar |
| Plume Geometry<br>Evaluated using digital<br>image and photographic grid<br>measuring spray distance,<br>velocity, dispersion angle all<br>vs. time | | |
Page 7
pg 2 of 4
{2}------------------------------------------------
K070596
Topical Applicator
# Discussion and Table of the Comparison and Differences
. I
| Features | Predicate | Proposed Device |
|-----------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Wolfe Tory MADgic K002255 | |
| Indications for use | Intended for the application of<br>topical anesthetics to the | Intended for the application of topical<br>anesthetics to the oropharynx and |
| | oropharynx and upper airway region | upper airway region |
| Types of medications used | Topical solutions, i.e.,<br>Topical anesthetics,<br>Vasoconstrictors | Topical solutions, i.e.,<br>Topical anesthetics<br>Vasoconstrictors |
| Environment of Use /<br>Patient population | Not disclosed | Locations where topical anesthetics<br>solutions are used.<br>Individuals requiring topical<br>anesthetic at the direction and<br>discretion of the clinician. |
| Disposable, non-sterile | Yes | Yes |
| Components | Syringe<br>Flexible, malleable, semi-rigid tube<br>Atomizer tip / nozzle | Syringe (k852544)<br>Flexible, semi-rigid tube<br>Atomizer tip / nozzle |
| Places inserted | Nasal, Orotrachael | Nasal, Orotrachael |
| Dosage Amount | User controlled | User controlled |
| Delivery form | Fine particle spray mist | Fine particle spray mist |
| Spray generated by | Piston syringe | Piston syringe |
| Device dead space (ml) | 0.12 ml device only | 0.13 ml device only |
| Basic dimensions | Tip diameter - 0.157" OD<br>Tube length - 8.25" overall | Tip diameter – 0.31" OD (7.9 mm)<br>Tube length - 8.25" overall |
| Materials | Polycarbonate<br>Polyvinylchloride (PVC) | Acetal / Delrin, K-resin<br>Polyethylene<br>Polyvinylchloride (PVC) |
| Biocompatibility | Unknown | Tested to ISO 10993-1 |
| Performance testing | Comparative using Cascade impactor to measure the following listed<br>criteria | |
| Total Dose | 2.88 cc - Lidocaine | 2.75 cc - Lidocaine |
| Delivered max 3 cc | 2.87 cc - Saline | 2.78 cc - Saline |
| Residual Mass (cc) Ave. | 0.12 cc | 0.25 cc |
| Simulated Clinical Dose % | 94.7% - Lidocaine<br>95.0% - Saline | 92.3% - Lidocaine<br>94.0% - Saline |
| Particle Size | 470 um - Lidocaine | 835 um - Lidocaine |
| MMAD (um) | 470 um - Saline | 803 um - Saline |
| GSD (um) | 1.87 um - Lidocaine<br>1.90 um - Saline | 1.93 um - Lidocaine<br>2.00 um - Saline |
| Plume Geometry | Similar see Section 18 | Similar see Section 18 |
| Evaluated using digital<br>image and photographic<br>grid measuring spray<br>distance, velocity,<br>dispersion angle all vs. time | | |
Page 8
pg 3 of 4
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a sequence of characters that appear to be handwritten. The characters are 'K070596'. The characters are written in a simple, slightly irregular style, giving them a casual appearance. The image is in black and white.
Premarket Notification 510(k) Section 5 - 510(k) Summary
Topical Applicator
# Discussion:
# Particle Size:
While the difference represents a measurable difference, it would not be classified as "substantially different." The intent of the device is to deliver a topical medication to a specific sight in close proximity to the tip. Therefore bigger particles are important, otherwise the medication may be inhaled to another location that is not intended (alveolar). Both particle sizes of the respective devices are sufficiently large to insure this.
The simulated clinical dose for both devices were equal, and almost exactly equal to the "filled dose" minus the "residual", is further verification of this position (i.e., all medication was accounted for at the sight of delivery or remaining in the device. none was inhaled).
Therefore both devices equally performed their function and are "substantially equivalent."
### Tube malleable wire:
The proposed device does not include a malleable wire in the tubing as this is a patented feature. We believe that the intent of the malleable wire was to allow the tube to hold a curve or shape for insertion into the patient's throat. However this feature is not a requirement for any atomizer to be functional.
The lack of a malleable wire does not introduce or create any new safety or performance issues in fact one might consider our design as "safer" as the tube is less stiff. This is not a claim which is proposed to be used or demonstrated.
The use of K-resin for our tubing allows the clinician to bend the tube and for it to "hold" some curve or shape, anatomical, during the intended use.
Therefore there are no significant differences between the proposed device and the predicate.
It is our view that there are no significant differences that affect the safety or effectiveness of the intended device as compared to the predicate device.
Page 9
pq4 of 4
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized eagle-like bird with outstretched wings, symbolizing protection and care. The bird is positioned to the right of a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The overall design is simple and conveys a sense of authority and service.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Sharn, Incorporated C/O Mr. Paul Dryden President ProMedic, Incorporated 3460 Pointe Creek Court # 102 Bonita Springs, Florida 34134-2015
JUN 2 1 2007
Re: K070596
Trade/Device Name: Topical Applicator Regulation Number: 21 CFR 868.5170 Regulation Name: Laryngotracheal Topical Anesthesia Applicator Regulatory Class: II Product Code: CCT Dated: June 11, 2007 Received: June 12, 2007
Dear Mr. Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Dryden
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### Indications for Use Statement
Page 1 of 1
510(k) Number: K070596 Device Name: Topical applicator Indications for Use: Intended for atomizing topical anesthetics to the oropharynx and upper airway regions
Prescription Use XX (Part 21 CFR 801 Subpart D)
or
Over-the-counter use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
signature
(k) Number: .................................................................................................................................................................
...
ാന Sign-Off) sion of Anesthesiology, General Hospital, i Control, Dental Devices
Ko70596
Page 4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.