BLOOD PRESSURE MONITOR, MODELS ZSBP-OXX AND ZSBP-1XX
K070473 · Zen Strong Medical Technology Co., Ltd. · DXN · May 23, 2007 · Cardiovascular
Device Facts
Record ID
K070473
Device Name
BLOOD PRESSURE MONITOR, MODELS ZSBP-OXX AND ZSBP-1XX
Applicant
Zen Strong Medical Technology Co., Ltd.
Product Code
DXN · Cardiovascular
Decision Date
May 23, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
ZSBP-001 and ZSBP-002 blood pressure monitor measures automatically human being systolic, diastolic blood pressure and heart beat rate from wrist by using the oscillometric method for the patient of the age over 18 years old. ZSBP-101 series of blood pressure monitor measures automatically human being systolic, diastolic blood pressure and heat beat rate from upper arm by using the Oscillometric method for the patient of the age over 18 years old. All measurement values can be read out and keep memory on the LCD panel for home care use.
Device Story
Non-invasive electronic blood pressure monitors (wrist and upper arm models) for home use. Device consists of pneumatic cuff, inflation/deflation pump, pressure sensor, and processing unit. Operation: cuff inflates to set pressure, then automatically deflates; pressure changes over time recorded as database; software calculates systolic/diastolic blood pressure and heart rate using oscillometric method. Results displayed on integrated LCD panel; device includes memory function for result storage. Operated by patient without physician involvement.
Clinical Evidence
Bench testing only. Compliance with ANSI/AAMI SP 10-1992, EN 60601-1, and EN 60601-1-2 standards. Software verification performed per FDA guidance.
Technological Characteristics
Non-invasive electronic oscillometric blood pressure monitor. Components: pneumatic pump, pressure sensor, cuff, LCD, and measuring software. Standards: ANSI/AAMI SP 10-1992, EN 60601-1, EN 60601-1-2. Standalone device for home use.
Indications for Use
Indicated for patients over 18 years old for the automatic measurement of systolic and diastolic blood pressure and heart rate via oscillometric method at the wrist (ZSBP-001, ZSBP-002) or upper arm (ZSBP-101). Intended for home care use.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{0}------------------------------------------------
K070473/51
p1/2
## 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is:
1. Submitter's Identifications: Zen Strong Medical Technology Co., Ltd. 6F No. 88, Ning Jin St., Keelung, Taiwan, R.O.C. Contact: Cheng Roei-Sheng Date of Summary Preparation: February 15, 2007 MAY 2 3 2007
- 2. Name of the Device: Blood Pressure Monitor, models ZSBP-001 and ZSBP-002 for wrist type and ZSBP-101 for upper arm type.
- 3. Classification information:
| Regulation Number | 870.1130 |
|-------------------|-----------|
| Medical Specialty | Neurology |
| Product Code | DXN |
| Device Class | II |
| Tier | II |
## 4. Device Description:
Basically two different type of blood pressure monitors are to be included in this 510(K) submission, the wrist type and upper arm type. The main intended use for these two tvpe of blood pressure monitor is as the following description:
1> ZSBP-001 and ZSBP-002 blood pressure monitor. This series of device measures automatically the human being systolic, diastolic blood pressure and heart beat rate from wrist by using the oscillometric method for the patient of ages above 18 years old.
- 2> ZSBP-101 blood pressure monitor measures automatically human being systolic, diastolic blood pressure and heart beat rate from upper arm by using the Oscillometric method for the patient of the age above 18 tears old.
For all of these three blood pressure monitors, the measurement values can be read out and keep memory on the LCD panel for home care use( without the involvement of professional physician).
From the construction point of view, the non-invasive electronic blood pressure monitor was composite of blood pressure measuring circuit via Oscillometric method, pressure sensor, measuring cuff at upper arm or wrist, pneumatic pumping inflation and deflation system, housing, display LCD, and measuring software, and memories ....... etc.
The main operation for the non-invasive electronic blood pressure monitor is carried out in such a way that the measuring cuff at wrist or upper arm is inflated to the set pressure level, then deflated to zero automatically. During the inflation and deflation, the pressure change with respective to time were recorded as the database of measurement. Then the following measuring results will be calculated against the measurement database, and displayed on the LCD of device:
{1}------------------------------------------------
-
- Blood pressure information including systolic and diastolic pressure (calculated via -Oscillometric method)
- Heart beat rate. ।
In addition to the main blood pressure and heart beat rate measuring function, the ZSBP series of blood pressure monitor provides also the memory function for user to store the result of measurement.
- 5. Intended Use:
ZSBP-001 and ZSBP-002 blood pressure monitor measures automatically human being systolic, diastolic blood pressure and heart beat rate from wrist by using the oscillometric method for the patient of the age over 18 years old.
ZSBP-101 series of blood pressure monitor measures automatically human being systolic, diastolic blood pressure and heat beat rate from upper arm by using the Oscillometric method for the patient of the age over 18 years old.
All measurement values can be read out and keep memory on the LCD panel for home care use.
- 6. Comparison to the 510(k) Cleared Device (Predicate Device):
- 1> Health & Life model HL-168 (K050249).
- 2> Micro life model BP-3BT0-AP (K041411).
- 7. Discussion of Non-Clinical Tests Performed Determination of Substantial Equivalence are as follows:
Compliance to applicable voluntary standards includes ANSI/AAMI, SP 10-1992, as well as EN 60601-1, and EN 60601-1-2 requirement. For the conformity of standards, the representative models of ZSBP-001 and ZSBP-101 were chosen as testing samples.
In addition to the compliance of voluntary standards, the software verification has been carried out according to the FDA software guidance.
- 8. Conclusions
The Zen Strong mode! ZSBP-001 and ZSBP-002 blood pressure monitor (measurement at wrist) has the same intended use and technical characteristics as the cleared model HL-168 (K050249), and ZSBP-101 series of blood pressure monitor (measured at upper arm) has the same intended use and technical characteristics as the cleared model BP-3BT0-AP (K041411).
Moreover, verification and validation tests contained in this submission demonstrate that the difference in the submitted demonstrate that the difference in the submitted models could maintain the same safety and effectiveness as that of cleared devices.
In the other words, those engineering difference do not: (1) affect the intended use or (2) alter the fundamental scientific technology of the device, and the new models as mentioned on this submission are considered substantial equivalent to the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS). The logo features a stylized caduceus, a symbol often associated with medicine and healthcare, with three parallel lines that curve and converge. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" is arranged in a circular fashion around the caduceus.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 2 3 2007
Zen Strong Medical Technologies Co. c/o Cheng- Roei-Sheng 6F, No. 88 Ning Jing St. Keelung China (Taiwan)
Re: K070473
Trade/Device Name: Blood Pressure Monitor, models ZSBP-001 and ZSBP-002 for wrist type and ZSBP-101 for upper arm type. Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive blood pressure measurement system Regulatory Class: Class II Product Code: DXN Dated: April 16, 2007 Received: April 16, 2007
Dear Mr. Roei-Sheng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Roei-Sheng
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Blyminuma for
Bram B. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
070473/s1
## Indications For Use
510(k) Number (if known):
Device Name: Blood Pressure Monitor, models ZSBP-001and ZSBP-002 for wrist type, and ZSBP-101 for upper arm type.
Indications For Use:
ZSBP-001 and ZSBP-002 blood pressure monitor measures automatically human being systolic, diastolic blood pressure and heart beat rate from wrist by using the Oscillometric method for the patient of the age over 18 years old.
ZSBP-101 series of blood pressure monitor measures automatically human being systolic, diastolic blood pressure and heart beat rate from upper arm by using the Oscillometric method for the patient of the age over 18 years old.
All measurement values can be read out and keep memory on the LCD panel for home care use.
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use V (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
Bllminson
Page 1 of 1