← Product Code [KNF](/productcode/KNF) · K070315

# LINA LOOP, MODELS: EL-160-4, EL-160-8, EL-200-4, EL-200-8 (K070315)

_Lina Medical Aps · KNF · Oct 30, 2007 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K070315

## Device Facts

- **Applicant:** Lina Medical Aps
- **Product Code:** [KNF](/productcode/KNF.md)
- **Decision Date:** Oct 30, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.4160
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology
- **Attributes:** Therapeutic

## Indications for Use

The LiNA Loop is a 5mm monopolar electrosurgical device intended for cutting and removal of soft tissue in gynecological procedures involving endoscopic hysterectorny. It is used with standard electrosurgical generators.

## Device Story

LiNA Loop is a 5mm monopolar electrosurgical instrument for soft tissue cutting/removal during endoscopic hysterectomy. Device connects to standard electrosurgical generators; utilizes monopolar energy to cut tissue via a loop electrode. Available in two sizes (160mm x 80mm and 200mm x 100mm); features 30mm total monopolar cutting length (15mm uninsulated on each side). Used by surgeons in clinical/OR settings. Provides mechanical loop for tissue capture and electrical path for coagulation/cutting. Single-use, sterile device.

## Clinical Evidence

No clinical data. Bench testing only, specifically electrical testing to confirm safety.

## Technological Characteristics

5mm monopolar electrosurgical instrument; loop electrode design; two sizes (160x80mm, 200x100mm); 30mm total uninsulated cutting length; compatible with standard monopolar electrosurgical generators; sterile, single-use.

## Regulatory Identification

A unipolar endoscopic coagulator-cutter is a device designed to destroy tissue with high temperatures by directing a high frequency electrical current through the tissue between an energized probe and a grounding plate. It is used in female sterilization and in other operative procedures under endoscopic observation. This generic type of device may include the following accessories: an electrical generator, probes and electrical cables, and a patient grounding plate. This generic type of device does not include devices used to perform female sterilization under hysteroscopic observation.

## Predicate Devices

- Omnitech Resectoscope Cutting Loop Electrode ([K981464](/device/K981464.md))
- Cook Ireland Sonnet Polypectomy Snare ([K050294](/device/K050294.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K070315

OCT 30 2003

# 5. 510(k) Summary

[As required by 21 CFR 807.92]

| Submitted by        | LiNA Medical ApS<br>Formervangen 5<br>2600 Glostrup<br>Denmark<br>Phone: +45 4329 6666<br>Fax: +45 4329 6699 |
|---------------------|--------------------------------------------------------------------------------------------------------------|
| Contact person      | Jane Kudsk, QA Manager                                                                                       |
| Date Prepared       | January 25, 2007                                                                                             |
| Device trade name   | LiNA Loop                                                                                                    |
| Common name         | Monopolar Endoscopic Loop                                                                                    |
| Classification name | Coagulator-Cutter, Endoscopic, Unipolar (And accessories)<br>21 CFR 884.4160, Product code KNF<br>Class II   |

## Intended Use

The LiNA Loop is a 5mm monopolar electrosurgical device intended for cutting and removal of soft tissue in gynecological procedures involving endoscopic hysterectorny. It is used with standard electrosurgical generators.

## Description of Device

The LiNA Loop is a 5 mm laparoscopic instrument. It is available with two different loop dimensions; 160mm x 80mm and 200mm x 100mm. The outer 15mm on each side of the loop are not insulated i.e. the monopolar cutting area length is totally 30mm. The device is sterile for single use and is compatible with all standard electrosurgical generators that have a monopolar outlet.

## Substantial Equivalence

The LiNA Loop is substantial equivalent to the Omnitech Resectoscope Cutting Loop Electrode (K981464) and the Cook Ireland Sonnet Polypectomy Snare (K050294) with respect to technical and design features. The submitted devices pose the same type of questions about safety or effectiveness as the existing devices.

## Test Data

Test data include electrical testing which deems the LiNA Loop to be safe under those conditions. No clinical tests have been performed.

{1}------------------------------------------------

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an eagle or bird in flight, composed of three curved lines.

#### Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

# OCT 3 0 2007

LiNA Medical ApS % Mr. Walt Brittle Vice President FDA Compliance Help Desk, Inc. 1289 N. Fordham Blvd., Suite A-128 CHAPEL HILL NC 27517

Re: K070315

Trade Name: LiNA Loop Regulation Number: 21 CFR 884.4160 Regulation Name: Unipolar endoscopic coagulator-cutter and accessories Regulatory Class: II Product Code: KNF Dated: October 20, 2007 Received: October 22, 2007

#### Dear Mr. Brittle:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Orug, and Cosmetic act (Act) that do not require approval of a premarket approval application (MAA), You may, therefore, market the device, subject to the general controls provisions of the (ct. The general controls provisions of the Act include requirements for annual registration, l'isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your it it wire can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.

| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology)                      | 240-276-0120 |
| Other           |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Nancy Brogdon

Nancv C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# 4. Indications for Use Statement

Indications for Use

510(k) Number (if known): K070315

Device Name: LiNA Loop

Indications for Use:

The LiNA Loop is a 5mm monopolar electrosurgical device intended for cutting and removal of soft tissue in gynecological procedures involving endoscopic hysterectorny. It is used with standard electrosurgical generators.

x Prescription Use (Part 21 CFR 801 Subpart D)

t

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

|                                                              | <img alt="Signature" src="signature.png"/> |
|--------------------------------------------------------------|--------------------------------------------|
| (Division Sign-Off)                                          |                                            |
| Division of Reproductive, Abdominal and Radiological Devices |                                            |
| 510(k) Number                                                | K070315                                    |
|                                                              | Page 1 of                                  |

Jan. 25, 2007 LiNA Loop Submission to FDA

Page 4-1

---

**Source:** [https://fda.innolitics.com/device/K070315](https://fda.innolitics.com/device/K070315)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
