← Product Code [NTM](/productcode/NTM) · K070310

# B R A H M S PCT SENSITIVE KRYPTOR TEST SYSTEM (K070310)

_Brahms Aktiengesellschaft · NTM · Mar 31, 2008 · Microbiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K070310

## Device Facts

- **Applicant:** Brahms Aktiengesellschaft
- **Product Code:** [NTM](/productcode/NTM.md)
- **Decision Date:** Mar 31, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.3210
- **Device Class:** Class 2
- **Review Panel:** Microbiology
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K070310 · Mar 31, 2008 | B R A H M S PCT SENSITIVE KRYPTOR TEST SYSTEM | Brahms Aktiengesellschaft | Patient samples (clinical cohort) | A method comparison study was conducted using 184 patient samples to demonstrate substantial equivalence between the B·R·A·H·M·S PCT sensitive KRYPTOR® assay and the predicate B·R·A·H·M·S PCT LIA assay. | Method comparison; Patient samples; Substantial equivalence |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Method comparison study per CLSI EP9-A | Patients requiring PCT measurement; Sample Size: 184; Number of Sites: 3 | B·R·A·H·M·S PCT LIA assay | Correlation of PCT measurements between the new device and the predicate device |

## Indications for Use

The B·R·A·H·M·S PCT sensitive KRYPTOR® is designed for automated detection of PCT (procalcitonin) in human serum or plasma (EDTA, heparin) samples by the immunofluorescent B·R·A·H·M·S PCT sensitive KRYPTOR® assay. The B·R·A·H·M·S PCT sensitive KRYPTOR® is intended for use in conjunction with other laboratory findings and clinical assessments to aid in the risk assessment of critically ill patients on their first day of ICU admission for progression to severe sepsis and septic shock.

## Device Story

The BRAHMS PCT sensitive Kryptor is an automated immunofluorescent assay system for measuring procalcitonin (PCT) in human serum or plasma. It utilizes Time-Resolved Amplified Cryptate Emission (TRACE) technology, where a nitrogen laser excites a cryptate-labeled antibody-antigen-XL665-labeled antibody complex. The resulting delayed fluorescence signal at 665 nm is proportional to the PCT concentration. The system is a closed, fully automated analyzer operated by laboratory personnel. It automatically handles sample dilution for concentrations exceeding 50 ng/ml. The analyzer compares measured signals against a stored standard curve to output quantitative PCT results. These results assist clinicians in assessing the risk of progression to severe sepsis or septic shock in ICU patients. The device provides rapid, automated quantification, reducing manual processing compared to non-automated predicate methods.

## Clinical Evidence

No clinical prospective or retrospective studies were performed. Evidence is based on analytical bench testing, including precision (within-lab and repeatability), linearity (EP6-A), limit of detection (0.0227 ng/ml), limit of quantitation (0.075 ng/ml), and analytical specificity. A method comparison study (n=184) against the predicate device demonstrated a nearly perfect correlation using Passing-Bablock analysis.

## Technological Characteristics

Immunofluorescent assay using TRACE technology. Reagents include cryptate-labeled polyclonal sheep anti-PCT and XL665-labeled monoclonal mouse anti-PCT. Energy source: nitrogen laser (337 nm excitation). Connectivity: closed system, automated analyzer. Dimensions: Model E and Compact variants. Sterilization: N/A (reagents). Software: embedded control system for assay management and calculation.

## Regulatory Identification

An endotoxin assay is a device that uses serological techniques in whole blood. The device is intended for use in conjunction with other laboratory findings and clinical assessment of the patient to aid in the risk assessment of critically ill patients for progression to severe sepsis.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance entitled “Class II Special Controls Guidance Document: Endotoxin Assay.” See § 866.1(e) for the availability of this guidance document.

## Predicate Devices

- B·R·A·H·M·S PCT LIA ([K040887](/device/K040887.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE

A. 510(k) Number:
K070310

B. Purpose for Submission:
To obtain Substantial equivalence for the BRAHMS PCT assay on the KRYPTOR instrument.

C. Measurand:
Procalcitonin

D. Type of Test:
Immunofluorescent assay

E. Applicant:
B·R·A·H·M·S Aktiengesellschaft

F. Proprietary and Established Names:
B·R·A·H·M·S PCT sensitive KRYPTOR Test System

G. Regulatory Information:
1. Regulation section:
21 CFR 866.3610, Endotoxin activity
2. Classification:
Class II
3. Product code:
NTM – Antigen, inflammatory response marker, sepsis
4. Panel:

{1}

Microbiology (83)

## H. Intended Use:

1. Intended use:

The BRAHMS PCT sensitive Kryptor is an immunofluorescent assay used to determine the concentration of PCT (procalcitonin) in human serum and plasma. The BRAHMS PCT sensitive Kryptor is intended for use in conjunction with other laboratory findings and clinical assessments to aid in the risk assessment of critically ill patients on their first day of ICU admission for progression to severe sepsis and septic shock

2. Indication(s) for use:

The BRAHMS PCT sensitive Kryptor is an immunofluorescent assay used to determine the concentration of PCT (procalcitonin) in human serum and plasma. The BRAHMS PCT sensitive Kryptor is intended for use in conjunction with other laboratory findings and clinical assessments to aid in the risk assessment of critically ill patients on their first day of ICU admission for progression to severe sepsis and septic shock

3. Special conditions for use statement(s):

For prescription use

4. Special instrument requirements:

The BRAHMS KRYPTOR analyzer

## I. Device Description:

The contents of the B·R·A·H·M·S PCT sensitive KRYPTOR® assay kit are:

|  Reagent | Quantity for 50 determinations | Content  |
| --- | --- | --- |
|  Cryptate Conjugate | 1 bottle lyophilized | Cryptate conjugate, cryptate labeled, anti-PCT antibody (polyclonal, sheep), 3.2 ml after reconstitution with KRYPTOR® Solution 1 and KRYPTOR® Solution 2  |
|  XL665 Conjugate | 1 bottle lyophilized | XL665 conjugate, XL665 labeled, anti-PCT antibody (monoclonal, mouse), 3.95 ml after reconstitution with KRYPTOR® Solution 1 and KRYPTOR® Solution 2  |
|  Diluent | 1 bottle | Defibrinated human plasma, for diluting samples above 50 ng/ml, ready to use  |

{2}

Controls, Calibrator, and Consumables are provided separate from the reagent kit.

The calibrator is packed as 6 vials of lyophilized recombinant PCT in defibrinated human plasma. It is reconstituted with 0.75 ml osmosed water [range 22.5-27.5 ng/ml].

Controls 1 &amp; 2 are packed as 3 vials each. PCT Control 1 is lyophilized recombinant PCT in defibrinated human plasma. It is reconstituted with 2 ml osmosed water [range: 0.2-0.4 ng/ml]. PCT Control 2 is lyophilized recombinant PCT in defibrinated human plasma. It is reconstituted with 2ml osmosed water [range 8-12 ng/ml]

Consumables consist of 4 x 200 ml vials of solutions 1, 2, 3 and 4 and 5 sachets of a phosphate buffer saline.

Solution 1 is ProClin 150 solution; Solution 2 is KF solution; Solution 3 is active chlorine and sodium hydroxide solution and Solution 4 is sodium hydroxide solution.

## J. Substantial Equivalence Information:

1. Predicate device name(s): BRAHMS PCT LIA
2. Predicate 510(k) number(s): K040887
3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Intended Use | Determination of PCT in human serum or plasma | Same  |
|  Analyte | Procalcitonin | Same  |
|  |   |   |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Assay principle | Immunofluorescent assay | Immunoluminometric assay – non automated  |
|  Detection method | Measuring principle based on TRACE technology which measures the signal emitted from an immunocomplex with time delay | Luminescence signal measured via Luminometer.  |

{3}

|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Instrument requirements | Kryptor analyzer | Luminometer  |
|  Sample volume | 50μl | 20μl  |
|  Measurement range | 0.02-5000ng/ml | 0.3-500ng/ml  |

# K. Standard/Guidance Documents Referenced:

CLSI Guideline EP 5A2 – Evaluation of Precision Performance of Quantitative Measurements and Methods. Approved Guideline  $2^{\text{nd}}$  ed.

CLSI Guideline EP 6 - A - Evaluation of the linearity of quantitative measurement procedures. A statistical approach. Approved guideline Vol 23, No. 16

CLSI Guideline EP 9-A - Method Comparison and Bias Estimation using Patient Samples. Approved Guideline  $2^{\mathrm{nd}}$  ed.

CLSI Guideline EP 17A - Protocols for Determination of Limits of Detection and Limits of Quantitation. Vol. 24 No.34

# L. Test Principle:

The B·R·A·H·M·S KRYPTOR® analyzer is a fully automated system. The B·R·A·H·M·S KRYPTOR® analyzer is a closed system and can only operate utilizing special reagents provided by B·R·A·H·M·S Aktiengesellschaft. The measuring principle is based on Time-Resolved Amplified Cryptate Emission (TRACE®) technology, which measures the signal that is emitted from an immunocomplex with time delay.

The basis of the TRACE® technology is a non-radiative energy transfer from a donor [a cage-like structure with a europium ion in the center (cryptate)] to an acceptor (XL 665). The proximity of donor (cryptate) and acceptor (XL 665) in a formed immunocomplex and the spectral overlap between donor emission and acceptor absorption spectra on the one hand intensifies the fluorescent signal and on the other hand extends the life span of the acceptor signal, allowing for the measurement of temporally delayed fluorescence.

After the sample to be measured has been excited with a nitrogen laser at  $337~\mathrm{nm}$ , the donor (cryptate) emits a long-life fluorescent signal in the milli-second range at  $620~\mathrm{nm}$ , while the acceptor (XL 665) generates a short-life signal in the range of nano-seconds at  $665~\mathrm{nm}$ . When both components are bound in an immunocomplex, both the signal amplification and the prolonged life span of the acceptor signal occur at  $665~\mathrm{nm}$ , and the life is in the microsecond range. This delayed acceptor signal is proportional to the concentration of the analyte to be measured.

The specific fluorescence which is proportional to the antigen concentration is obtained through a double selection: spectral (separation depending on wave-length) and temporal

{4}

(time resolved measurement). This enables an exclusive measurement of the signal emitted by the immunological complex and the ratio between the two wave-lengths (665/620) allows a real-time correction of the variations in optic transmission from the medium.

## M. Performance Characteristics:

### 1. Analytical performance:

#### a. Precision/Reproducibility:

**Internal Precision Study**

Aliquots of 19 samples distributed over the measuring range were assayed in duplicate using 4 different BRAHMS KRYPTOR analyzers with 3 reagent lots. Data is acceptable and is summarized below:

|   | Mean (ng/ml) | Within lab Precision CV% | Repeatability CV%  |
| --- | --- | --- | --- |
|  Pool 1 | 0.056 | 24.2 | 29.9  |
|  Pool 2 | 0.054 | 21.1 | 30.1  |
|  Pool 305 | 0.093 | 13.4 | 13.6  |
|  Pool 3 | 0.288 | 4.9 | 7.4  |
|  Pool 4 | 0.184 | 8.0 | 9.5  |
|  Pool 5 | 0.53 | 4.3 | 3.8  |
|  Pool 6 | 0.59 | 6.0 | 3.8  |
|  Pool 7 | 0.65 | 4.0 | 3.1  |
|  Pool 8 | 1.05 | 3.5 | 1.9  |
|  Pool 9 | 1.53 | 3.4 | 1.7  |
|  Pool 10 | 1.98 | 3.3 | 1.3  |
|  Pool 11 | 4.18 | 3.2 | 1.1  |
|  Pool 12 | 6.13 | 3.3 | 1.0  |
|  Pool 13 | 12.1 | 3.7 | 1.1  |
|  Pool 14 | 20.7 | 5.1 | 1.2  |
|  Pool 15 | 34.8 | 6.1 | 1.2  |
|  HG 50 | 65.8 | 4.5 | 1.0  |
|  HG 1000 | 931 | 4.3 | 1.5  |
|  HG 5000 | 4758 | 4.1 | 1.7  |

**External Precision Study**

Within Lab precision estimates were calculated according to the CLSI guideline EP5-A2.

{5}

Results are acceptable and summarized as follows:

|  Site 1 |   |   |   | Site 2 |   |   |   | Site 3  |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Mean of concentration (ng/ml) | CV (%) | Within Lab SD (ng/ml) | Number of runs (Data points) | Mean of concentration (ng/ml) | CV (%) | Within Lab SD (ng/ml) | Number of runs (Data points) | Mean of concentration (ng/ml) | CV (%) | Within Lab SD (ng/ml) | Number of runs (Data points)  |
|  0.090 | 13.5 | 0.012 | 10 (20) | 0.068 | 19.6 | 0.013 | 10 (20) | 0.089 | 27.7 | 0.025 | 10 (20)  |
|  0.156 | 6.9 | 0.010 | 10 (20) | 0.120 | 12.5 | 0.015 | 10 (20) | 0.156 | 9.6 | 0.015 | 10 (20)  |
|  0.443 | 3.6 | 0.016 | 10 (20) | 0.407 | 5.3 | 0.022 | 10 (20) | 0.445 | 4.0 | 0.018 | 10 (20)  |
|  1.88 | 1.8 | 0.034 | 10 (20) | 1.75 | 4.0 | 0.070 | 10 (20) | 1.89 | 2.5 | 0.047 | 10 (20)  |
|  18.13 | 1.7 | 0.316 | 10 (20) | 14.87 | 5.2 | 0.770 | 10 (20) | 16.54 | 4.6 | 0.761 | 10 (20)  |
|  67.30 | 2.2 | 1.445 | 10 (20) | 57.46 | 3.8 | 2.157 | 10 (20) | 61.08 | 2.7 | 1.650 | 10 (20)  |

In addition total precision ranged from 3.2 – 13.4% CV and within run precision ranged from 1.0 – 13.6% CV

## b. Linearity/assay reportable range:

Nine to ten samples with different levels of PCT were diluted with kit diluents by the KRYPTOR. Three batches of reagents and three analyzers were used in the study. Measurements at each dilution level were done in duplicate. For each sample, at least 4 dilution levels plus the undiluted sample were analyzed. The study was done according to EP6-A. Deviation of the individual samples was calculated in percent deviation of the undiluted sample. Measured concentrations were multiplied with the dilution factor and a linear regression analysis was provided in a graph for each data run. The linearity of diluted samples was acceptable over the whole concentration range.

High concentration samples (&gt;50ng/ml) are detected by the analyzer in the first few seconds of incubation and may be diluted by the appropriate dilution factor, then re-assayed automatically. The system can detect samples greater than 50ng/ml up to 5000ng/ml.

## c. Traceability, Stability, Expected values (controls, calibrators, or methods):

Stability testing was performed by storing the reagent unit, calibrator and controls at 2-8°C up to their expiry date namely 13 months for the reagent and 2 yrs for

{6}

calibrator and controls. Testing was also done under thermo stress conditions namely 2 days at  $18 - 25^{\circ}\mathrm{C}$ , then at  $37^{\circ}\mathrm{C}$ , then frozen and tested after thawing. The stability of the conjugates in the reconstituted reagent was also assessed namely 29 days for the reagent on board the analyzer, 4 hrs at  $18 - 25^{\circ}\mathrm{C}$  for the calibrator and 24 hours at  $2-8^{\circ}\mathrm{C}$  or up to one month at  $&lt; -16^{\circ}\mathrm{C}$ . Results of stability were acceptable.

# d. Detection limit:

The analytical sensitivity for the B·R·A·H·M·S PCT sensitive KRYPTOR® is  $0.02\mathrm{ng / ml}$ . The analytical sensitivity is calculated following CLSI EP17-A guideline as follows:

20 determination of a sample with no PCT are performed. This sample with no PCT was prepared either by storage 1 day at  $37^{\circ}\mathrm{C}$  or by addition of A179 anti-PCT polyclonal antibody. At the end, this "free PCT sample" was different in each run.

20 determination of RS002, a recombinant PCT preparation with a very low amount of PCT are performed. This sample is the same for each run.

This protocol is reproduced three time; 3 different KRYPTOR PCT reagent unit batches and 2 different KRYPTOR were used for these 3 runs.

Then, analysis of measurement is done as follows:

Firstly, Limit of Blank (LoB) value is determined by the 95th percentile of concentrations for a sample that does not contain PCT.

Then, Limit of Detection (LoD = Analytical detection limit) is equal to  $\mathrm{LoB} + \mathrm{C}\beta \mathrm{xSDs}$ , where  $\mathrm{C}\beta$  is derived from the standard Gaussian distribution ( $\mathrm{C}\beta = 1.645 / (1 - 1 / (4\mathrm{xf}))$ , where  $\mathrm{f}$  is the degrees of freedom for the estimated standard deviation SDs) and SDs is the standard deviation of concentration for sample RS002.

The analytical detection limit determination is based on 3 runs utilizing three (3) different lots of B·R·A·H·M·S PCT sensitive KRYPTOR® and two (2) KRYPTOR® instruments.

# Calculation of LoD:

Results of the 60 measurements of sample PCT_RS002

|  replicate | run | value (ng/ml) | run | value (ng/ml) | run | value (ng/ml)  |
| --- | --- | --- | --- | --- | --- | --- |
|  1 | run1 | 0.0353 | run2 | 0.0315 | run3 | 0.0988  |
|  2 | run1 | 0.0569 | run2 | 0.0417 | run3 | 0.0527  |
|  3 | run1 | 0.0568 | run2 | 0.0227 | run3 | 0.0626  |
|  4 | run1 | 0.0185 | run2 | 0.0372 | run3 | 0.0558  |
|  5 | run1 | 0.0373 | run2 | 0.0635 | run3 | 0.0669  |
|  6 | run1 | 0.0438 | run2 | 0.0426 | run3 | 0.0558  |
|  7 | run1 | 0.0432 | run2 | 0.0474 | run3 | 0.0550  |
|  8 | run1 | 0.0148 | run2 | 0.0503 | run3 | 0.0495  |
|  9 | run1 | 0.0556 | run2 | 0.0506 | run3 | 0.0433  |
|  10 | run1 | 0.0495 | run2 | 0.0515 | run3 | 0.0651  |
|  11 | run1 | 0.0491 | run2 | 0.0644 | run3 | 0.0587  |
|  12 | run1 | 0.0497 | run2 | 0.0318 | run3 | 0.0561  |

{7}

Standard deviation of RS002 is calculated based on the pooled 60 determinations.  $\mathrm{SD}_{\mathrm{s}} = 0.01374$

Then, LoD is calculated as follows:

$\mathrm{df} = 60 - 1 =$  59

$\mathrm{C}_{\beta} = 1.6449 / (1 - 1 / 236) =$  1.65190

$\mathrm{SD_s}$  0.01374

$\mathrm{LoD} = \mathrm{LoB} + \mathrm{C}_{\beta}\left(\mathrm{SD}_{\mathrm{s}}\right) =$  0.0227 ng/ml

Limit of Detection for the B·R·A·H·M·S PCT sensitive KRYPTOR® is 0.0227 ng/ml.

The Limit of Quantitation (LOQ) was determined following EP17-P as follows:

. Samples at different targets (from  $0.06\mathrm{ng / ml}$  to  $0.075\mathrm{ng / ml}$ ) were prepared with master calibrators for which actual concentrations were determined independently.

. These samples were run in 5 runs, with 10 replicates per run, thus a total of 50 replicates per sample.

. For the 5 runs, 3 different KRYPTOR instruments and 2 different batches of reagents were used.

. For each sample, the total standard deviation  $(\mathrm{SD}_{\mathrm{s}})$  was calculated as well as the difference between the mean of all replicates and the reference value of the sample (bias) and imprecision with  $95\%$  probability  $(2\times \mathrm{SD}_{\mathrm{s}})$ .

The LOQ determined as the lowest reported concentration level with total error (imprecision + bias)  $\leq 30\%$  was determined at  $0.075\mathrm{ng / ml}$

Results for replicates and  $\mathrm{SD}_{\mathrm{s}}$ , imprecision and bias calculation are shown below:

|  Replicates | Run 1 | Run 2 | Run 3 | Run 4 | Run 5  |
| --- | --- | --- | --- | --- | --- |
|  1 | 0.0868 | 0.0809 | 0.0623 | 0.0687 | 0.0726  |
|  2 | 0.0611 | 0.0761 | 0.0708 | 0.0644 | 0.0861  |
|  3 | 0.0841 | 0.0754 | 0.0622 | 0.0663 | 0.0780  |
|  4 | 0.0525 | 0.1016 | 0.0759 | 0.0887 | 0.0588  |
|  5 | 0.0774 | 0.0751 | 0.0735 | 0.0901 | 0.0782  |
|  6 | 0.0578 | 0.0790 | 0.0569 | 0.0619 | 0.0686  |
|  7 | 0.0735 | 0.0682 | 0.0779 | 0.0658 | 0.0765  |

{8}

|  8 | 0.0872 | 0.0767 | 0.0855 | 0.0798 | 0.0807  |
| --- | --- | --- | --- | --- | --- |
|  9 | 0.0895 | 0.0755 | 0.0871 | 0.0715 | 0.0819  |
|  10 | 0.0604 | 0.0810 | 0.0753 | 0.0693 | 0.0785  |

Average value (ng/ml): 0.0747 ng/ml

Target (ng/ml): 0.0750 ng/ml

Bias (ng/ml): -0.0003 ng/ml

$\mathrm{SD}_{\mathrm{s}}$  (ng/ml): 0.0102 ng/ml

Imprecision (ng/ml): 0.0204 ng/ml

Total Error (ng/ml): 0.0207 ng/ml

Total Error/Target:  $27.6\% &lt; 30\%$

Therefore, it is concluded that LOQ for the B·R·A·H·M·S PCT sensitive KRYPTOR® is 0.075ng/ml

e. Analytical specificity:

The following substances were evaluated with the BRAHMS PCT sensitive Kryptor assay and were found not to affect the test performance at clinically relevant concentrations: Substances tested were bilirubin, hemoglobin, triglycerides, albumin, imipenem, cefotaxim, vancomycin, dopamine, noradrenaline, dobutamine, heparin, furosemide and drugs used in asthmatic and COPD patients. In addition calcitonin, katacalcin, a-CGRP,  $\beta$ -CGRP, Calcitonin salmon and Calcitonin Eel showed no interference when tested with the assay.

f. Assay cut-off:

The assay cut off is  $0.02\mathrm{ng / ml}$

2. Comparison studies:

a. Method comparison with predicate device:

A correlation study was performed between the BRAHMS PCT sensitive KRYPTOR assay and the BRAHMS PCT LIA assay. There were 184 samples from three sites, which had B·R·A·H·M·S PCT LIA measurements of  $0.3\mathrm{ng / ml}$  (the functional assay sensitivity of B·R·A·H·M·S PCT LIA) or higher and/or B·R·A·H·M·S PCT sensitive KRYPTOR® measurements of  $0.06\mathrm{ng / ml}$  (the functional assay sensitivity of B·R·A·H·M·S PCT sensitive KRYPTOR®) or higher. Passing-Bablock analysis shows a nearly perfect correlation of the B·R·A·H·M·S PCT sensitive KRYPTOR® assay and B·R·A·H·M·S PCT LIA assay, as demonstrated in the correlation graph below.

{9}

![img-0.jpeg](img-0.jpeg)

b. Matrix comparison:

A study was conducted comparing serum and plasma samples using 10 spiked patient specimens. The samples were tested in triplicate using serum tubes, heparin plasma tubes, EDTA plasma tubes and citrate plasma tubes. Results demonstrated that both serum and plasma tubes can be used for the assay but citrate plasma tubes are not recommended because concentrations of PCT decreased in those tubes.

3. Clinical studies:

a. Clinical Sensitivity:

N/A

b. Clinical specificity:

N/A

c. Other clinical supportive data (when a. and b. are not applicable):

4. Clinical cut-off:

{10}

Data support the following interpretative risk assessment criteria:

**PCT &gt; 2 ng/ml**

PCT levels above 2.0 ng/ml on the first day of ICU admission represent a high risk for progression to severe sepsis and/or septic shock.

**PCT &lt; 0.5 ng/ml**

PCT levels below 0.5 ng/ml on the first day of ICU admission represent a low risk for progression to severe sepsis and/or septic shock.

**Note:** PCT levels below 0.5 ng/ml do not exclude an infection, because localized infections (without systemic signs) may also be associated with such low levels. If the PCT measurement is done very early after the systemic infection process has started (usually &lt; 6 hours), these values may still be low.

As various non-infectious conditions are known to induce PCT as well, PCT levels between 0.5 ng/ml and 2.0 ng/ml should be reviewed carefully to take into account the specific clinical background and condition(s) of the individual patient.

5. **Expected values/Reference range:**

A study was conducted with the BRAHMS KRYPTOR to determine the prevalence of PCT in a normal population in the US. Data showed that the PCT concentrations for 146 of the 151 samples tested were &lt; 0.1 ng/ml. Five samples were greater than 0.1 ng/ml.

**N. Instrument Name:**

BRAHMS KRYPTOR analyzer

**O. System Descriptions:**

1. **Modes of Operation:**

The analyzer is provided in two models, the Model E and the Compact. They differ in physical size and consequently in the number of wells in the reaction and dilution plates and the number of tests that can be performed. Both models have the same features including the same laser, methodology, incubation time and temperature and calculations/algorithms as well as utilizing the same reagents, calibrators and controls. Documentation provided for Model E is applicable to both models. The analyzer is a fully automated system, able to process multiple samples each day in random access mode. The analyzer is a closed system and can only operate utilizing especially made reagent kits from BRAHMS. The system is based on TRACE (Time Resolved Amplified Cryptate Emission) technology. The analyzer has a built in computer to handle the required controls to store reagents, dispense reagents, patient samples, calibrators and controls, and incubate the reagent-sample mixture and measure sample concentration. The external computer will program the analyzer for the required tests and calculate

11

{11}

patient results.

2. Software:

FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:

Yes ☐ X or No

3. Specimen Identification:

The assay procedure includes registering and/or loading the samples, reagent kit, calibrator and controls as applicable

4. Specimen Sampling and Handling:

A sample volume of 50μl is needed for each test. Initially a worklist for the day is created and the test is started. The sample probe of the analyzer pipettes and dispenses the conjugates from the reagent kit and the sample into the wells. The probe is heated to incubate the reagent-sample mixture so it is at 37°C prior to dispensing and mixing in the reaction well. The analyzer makes periodic measurement of the signal emitted. Samples &gt;50ng/ml are detected, diluted and re-assayed automatically. After measurement of the fluorescent signal, the program compares each result obtained with the stored standard curve.

5. Calibration:

A standard curve does not need to be established for the BRAHMS PCT sensitive KRYPTOR assay on the analyzer. A standard curve is included with the bar code information from the calibration card and is stored in the analyzer. A calibration is carried out before the first use of a reagent batch, and then repeated on a regular basis (i.e. first use and every 15 days thereafter). Calibrations are performed using a disposable calibrator vial in order to readjust the standard curve.

6. Quality Control:

The KRYPTOR QC is provided for use as quality control on board the instrument for the assay of procalcitonin. The kit contains 2 series of 3 lyophilized vials, a bar code card, bar code stick-on labels and the summary of concentration ranges by level. The bar code card contains information related to the control batch, particularly the target concentrations, the standard deviations obtained and the concentration acceptance ranges. Controls are flagged above or below normal (±2 SD) according to the defined control range. The BRAHMS KRYPTOR can automatically check the quality of assays at intervals by statistical analysis on Levy Jennings graphs. National quality assurance guidelines for medical lab tests must be followed e.g. test accuracy and precision can be monitored by means of lab in-house and/or commercially available control materials

12

{12}

P. Other Supportive Instrument Performance Characteristics Data Not Covered In the "Performance Characteristics" Section above:

N/A

Q. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

R. Conclusion:

1. The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

13

---

**Source:** [https://fda.innolitics.com/device/K070310](https://fda.innolitics.com/device/K070310)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
