A method comparison study was conducted using 184 patient samples to demonstrate substantial equivalence between the B·R·A·H·M·S PCT sensitive KRYPTOR® assay and the predicate B·R·A·H·M·S PCT LIA assay.
Patients requiring PCT measurement; Sample Size: 184; Number of Sites: 3
B·R·A·H·M·S PCT LIA assay
Correlation of PCT measurements between the new device and the predicate device
Indications for Use
The B·R·A·H·M·S PCT sensitive KRYPTOR® is designed for automated detection of PCT (procalcitonin) in human serum or plasma (EDTA, heparin) samples by the immunofluorescent B·R·A·H·M·S PCT sensitive KRYPTOR® assay. The B·R·A·H·M·S PCT sensitive KRYPTOR® is intended for use in conjunction with other laboratory findings and clinical assessments to aid in the risk assessment of critically ill patients on their first day of ICU admission for progression to severe sepsis and septic shock.
Device Story
The BRAHMS PCT sensitive Kryptor is an automated immunofluorescent assay system for measuring procalcitonin (PCT) in human serum or plasma. It utilizes Time-Resolved Amplified Cryptate Emission (TRACE) technology, where a nitrogen laser excites a cryptate-labeled antibody-antigen-XL665-labeled antibody complex. The resulting delayed fluorescence signal at 665 nm is proportional to the PCT concentration. The system is a closed, fully automated analyzer operated by laboratory personnel. It automatically handles sample dilution for concentrations exceeding 50 ng/ml. The analyzer compares measured signals against a stored standard curve to output quantitative PCT results. These results assist clinicians in assessing the risk of progression to severe sepsis or septic shock in ICU patients. The device provides rapid, automated quantification, reducing manual processing compared to non-automated predicate methods.
Clinical Evidence
No clinical prospective or retrospective studies were performed. Evidence is based on analytical bench testing, including precision (within-lab and repeatability), linearity (EP6-A), limit of detection (0.0227 ng/ml), limit of quantitation (0.075 ng/ml), and analytical specificity. A method comparison study (n=184) against the predicate device demonstrated a nearly perfect correlation using Passing-Bablock analysis.
Technological Characteristics
Immunofluorescent assay using TRACE technology. Reagents include cryptate-labeled polyclonal sheep anti-PCT and XL665-labeled monoclonal mouse anti-PCT. Energy source: nitrogen laser (337 nm excitation). Connectivity: closed system, automated analyzer. Dimensions: Model E and Compact variants. Sterilization: N/A (reagents). Software: embedded control system for assay management and calculation.
Indications for Use
Indicated for critically ill patients on their first day of ICU admission to aid in risk assessment for progression to severe sepsis and septic shock. Uses human serum or plasma (EDTA or heparin).
Regulatory Classification
Identification
An endotoxin assay is a device that uses serological techniques in whole blood. The device is intended for use in conjunction with other laboratory findings and clinical assessment of the patient to aid in the risk assessment of critically ill patients for progression to severe sepsis.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance entitled “Class II Special Controls Guidance Document: Endotoxin Assay.” See § 866.1(e) for the availability of this guidance document.
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
K070310
B. Purpose for Submission:
To obtain Substantial equivalence for the BRAHMS PCT assay on the KRYPTOR instrument.
C. Measurand:
Procalcitonin
D. Type of Test:
Immunofluorescent assay
E. Applicant:
B·R·A·H·M·S Aktiengesellschaft
F. Proprietary and Established Names:
B·R·A·H·M·S PCT sensitive KRYPTOR Test System
G. Regulatory Information:
1. Regulation section:
21 CFR 866.3610, Endotoxin activity
2. Classification:
Class II
3. Product code:
NTM – Antigen, inflammatory response marker, sepsis
4. Panel:
{1}
Microbiology (83)
## H. Intended Use:
1. Intended use:
The BRAHMS PCT sensitive Kryptor is an immunofluorescent assay used to determine the concentration of PCT (procalcitonin) in human serum and plasma. The BRAHMS PCT sensitive Kryptor is intended for use in conjunction with other laboratory findings and clinical assessments to aid in the risk assessment of critically ill patients on their first day of ICU admission for progression to severe sepsis and septic shock
2. Indication(s) for use:
The BRAHMS PCT sensitive Kryptor is an immunofluorescent assay used to determine the concentration of PCT (procalcitonin) in human serum and plasma. The BRAHMS PCT sensitive Kryptor is intended for use in conjunction with other laboratory findings and clinical assessments to aid in the risk assessment of critically ill patients on their first day of ICU admission for progression to severe sepsis and septic shock
3. Special conditions for use statement(s):
For prescription use
4. Special instrument requirements:
The BRAHMS KRYPTOR analyzer
## I. Device Description:
The contents of the B·R·A·H·M·S PCT sensitive KRYPTOR® assay kit are:
| Reagent | Quantity for 50 determinations | Content |
| --- | --- | --- |
| Cryptate Conjugate | 1 bottle lyophilized | Cryptate conjugate, cryptate labeled, anti-PCT antibody (polyclonal, sheep), 3.2 ml after reconstitution with KRYPTOR® Solution 1 and KRYPTOR® Solution 2 |
| XL665 Conjugate | 1 bottle lyophilized | XL665 conjugate, XL665 labeled, anti-PCT antibody (monoclonal, mouse), 3.95 ml after reconstitution with KRYPTOR® Solution 1 and KRYPTOR® Solution 2 |
| Diluent | 1 bottle | Defibrinated human plasma, for diluting samples above 50 ng/ml, ready to use |
{2}
Controls, Calibrator, and Consumables are provided separate from the reagent kit.
The calibrator is packed as 6 vials of lyophilized recombinant PCT in defibrinated human plasma. It is reconstituted with 0.75 ml osmosed water [range 22.5-27.5 ng/ml].
Controls 1 & 2 are packed as 3 vials each. PCT Control 1 is lyophilized recombinant PCT in defibrinated human plasma. It is reconstituted with 2 ml osmosed water [range: 0.2-0.4 ng/ml]. PCT Control 2 is lyophilized recombinant PCT in defibrinated human plasma. It is reconstituted with 2ml osmosed water [range 8-12 ng/ml]
Consumables consist of 4 x 200 ml vials of solutions 1, 2, 3 and 4 and 5 sachets of a phosphate buffer saline.
Solution 1 is ProClin 150 solution; Solution 2 is KF solution; Solution 3 is active chlorine and sodium hydroxide solution and Solution 4 is sodium hydroxide solution.
## J. Substantial Equivalence Information:
1. Predicate device name(s): BRAHMS PCT LIA
2. Predicate 510(k) number(s): K040887
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | Determination of PCT in human serum or plasma | Same |
| Analyte | Procalcitonin | Same |
| | | |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Assay principle | Immunofluorescent assay | Immunoluminometric assay – non automated |
| Detection method | Measuring principle based on TRACE technology which measures the signal emitted from an immunocomplex with time delay | Luminescence signal measured via Luminometer. |
{3}
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Instrument requirements | Kryptor analyzer | Luminometer |
| Sample volume | 50μl | 20μl |
| Measurement range | 0.02-5000ng/ml | 0.3-500ng/ml |
# K. Standard/Guidance Documents Referenced:
CLSI Guideline EP 5A2 – Evaluation of Precision Performance of Quantitative Measurements and Methods. Approved Guideline $2^{\text{nd}}$ ed.
CLSI Guideline EP 6 - A - Evaluation of the linearity of quantitative measurement procedures. A statistical approach. Approved guideline Vol 23, No. 16
CLSI Guideline EP 9-A - Method Comparison and Bias Estimation using Patient Samples. Approved Guideline $2^{\mathrm{nd}}$ ed.
CLSI Guideline EP 17A - Protocols for Determination of Limits of Detection and Limits of Quantitation. Vol. 24 No.34
# L. Test Principle:
The B·R·A·H·M·S KRYPTOR® analyzer is a fully automated system. The B·R·A·H·M·S KRYPTOR® analyzer is a closed system and can only operate utilizing special reagents provided by B·R·A·H·M·S Aktiengesellschaft. The measuring principle is based on Time-Resolved Amplified Cryptate Emission (TRACE®) technology, which measures the signal that is emitted from an immunocomplex with time delay.
The basis of the TRACE® technology is a non-radiative energy transfer from a donor [a cage-like structure with a europium ion in the center (cryptate)] to an acceptor (XL 665). The proximity of donor (cryptate) and acceptor (XL 665) in a formed immunocomplex and the spectral overlap between donor emission and acceptor absorption spectra on the one hand intensifies the fluorescent signal and on the other hand extends the life span of the acceptor signal, allowing for the measurement of temporally delayed fluorescence.
After the sample to be measured has been excited with a nitrogen laser at $337~\mathrm{nm}$ , the donor (cryptate) emits a long-life fluorescent signal in the milli-second range at $620~\mathrm{nm}$ , while the acceptor (XL 665) generates a short-life signal in the range of nano-seconds at $665~\mathrm{nm}$ . When both components are bound in an immunocomplex, both the signal amplification and the prolonged life span of the acceptor signal occur at $665~\mathrm{nm}$ , and the life is in the microsecond range. This delayed acceptor signal is proportional to the concentration of the analyte to be measured.
The specific fluorescence which is proportional to the antigen concentration is obtained through a double selection: spectral (separation depending on wave-length) and temporal
{4}
(time resolved measurement). This enables an exclusive measurement of the signal emitted by the immunological complex and the ratio between the two wave-lengths (665/620) allows a real-time correction of the variations in optic transmission from the medium.
## M. Performance Characteristics:
### 1. Analytical performance:
#### a. Precision/Reproducibility:
**Internal Precision Study**
Aliquots of 19 samples distributed over the measuring range were assayed in duplicate using 4 different BRAHMS KRYPTOR analyzers with 3 reagent lots. Data is acceptable and is summarized below:
| | Mean (ng/ml) | Within lab Precision CV% | Repeatability CV% |
| --- | --- | --- | --- |
| Pool 1 | 0.056 | 24.2 | 29.9 |
| Pool 2 | 0.054 | 21.1 | 30.1 |
| Pool 305 | 0.093 | 13.4 | 13.6 |
| Pool 3 | 0.288 | 4.9 | 7.4 |
| Pool 4 | 0.184 | 8.0 | 9.5 |
| Pool 5 | 0.53 | 4.3 | 3.8 |
| Pool 6 | 0.59 | 6.0 | 3.8 |
| Pool 7 | 0.65 | 4.0 | 3.1 |
| Pool 8 | 1.05 | 3.5 | 1.9 |
| Pool 9 | 1.53 | 3.4 | 1.7 |
| Pool 10 | 1.98 | 3.3 | 1.3 |
| Pool 11 | 4.18 | 3.2 | 1.1 |
| Pool 12 | 6.13 | 3.3 | 1.0 |
| Pool 13 | 12.1 | 3.7 | 1.1 |
| Pool 14 | 20.7 | 5.1 | 1.2 |
| Pool 15 | 34.8 | 6.1 | 1.2 |
| HG 50 | 65.8 | 4.5 | 1.0 |
| HG 1000 | 931 | 4.3 | 1.5 |
| HG 5000 | 4758 | 4.1 | 1.7 |
**External Precision Study**
Within Lab precision estimates were calculated according to the CLSI guideline EP5-A2.
{5}
Results are acceptable and summarized as follows:
| Site 1 | | | | Site 2 | | | | Site 3 | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Mean of concentration (ng/ml) | CV (%) | Within Lab SD (ng/ml) | Number of runs (Data points) | Mean of concentration (ng/ml) | CV (%) | Within Lab SD (ng/ml) | Number of runs (Data points) | Mean of concentration (ng/ml) | CV (%) | Within Lab SD (ng/ml) | Number of runs (Data points) |
| 0.090 | 13.5 | 0.012 | 10 (20) | 0.068 | 19.6 | 0.013 | 10 (20) | 0.089 | 27.7 | 0.025 | 10 (20) |
| 0.156 | 6.9 | 0.010 | 10 (20) | 0.120 | 12.5 | 0.015 | 10 (20) | 0.156 | 9.6 | 0.015 | 10 (20) |
| 0.443 | 3.6 | 0.016 | 10 (20) | 0.407 | 5.3 | 0.022 | 10 (20) | 0.445 | 4.0 | 0.018 | 10 (20) |
| 1.88 | 1.8 | 0.034 | 10 (20) | 1.75 | 4.0 | 0.070 | 10 (20) | 1.89 | 2.5 | 0.047 | 10 (20) |
| 18.13 | 1.7 | 0.316 | 10 (20) | 14.87 | 5.2 | 0.770 | 10 (20) | 16.54 | 4.6 | 0.761 | 10 (20) |
| 67.30 | 2.2 | 1.445 | 10 (20) | 57.46 | 3.8 | 2.157 | 10 (20) | 61.08 | 2.7 | 1.650 | 10 (20) |
In addition total precision ranged from 3.2 – 13.4% CV and within run precision ranged from 1.0 – 13.6% CV
## b. Linearity/assay reportable range:
Nine to ten samples with different levels of PCT were diluted with kit diluents by the KRYPTOR. Three batches of reagents and three analyzers were used in the study. Measurements at each dilution level were done in duplicate. For each sample, at least 4 dilution levels plus the undiluted sample were analyzed. The study was done according to EP6-A. Deviation of the individual samples was calculated in percent deviation of the undiluted sample. Measured concentrations were multiplied with the dilution factor and a linear regression analysis was provided in a graph for each data run. The linearity of diluted samples was acceptable over the whole concentration range.
High concentration samples (>50ng/ml) are detected by the analyzer in the first few seconds of incubation and may be diluted by the appropriate dilution factor, then re-assayed automatically. The system can detect samples greater than 50ng/ml up to 5000ng/ml.
## c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Stability testing was performed by storing the reagent unit, calibrator and controls at 2-8°C up to their expiry date namely 13 months for the reagent and 2 yrs for
{6}
calibrator and controls. Testing was also done under thermo stress conditions namely 2 days at $18 - 25^{\circ}\mathrm{C}$ , then at $37^{\circ}\mathrm{C}$ , then frozen and tested after thawing. The stability of the conjugates in the reconstituted reagent was also assessed namely 29 days for the reagent on board the analyzer, 4 hrs at $18 - 25^{\circ}\mathrm{C}$ for the calibrator and 24 hours at $2-8^{\circ}\mathrm{C}$ or up to one month at $< -16^{\circ}\mathrm{C}$ . Results of stability were acceptable.
# d. Detection limit:
The analytical sensitivity for the B·R·A·H·M·S PCT sensitive KRYPTOR® is $0.02\mathrm{ng / ml}$ . The analytical sensitivity is calculated following CLSI EP17-A guideline as follows:
20 determination of a sample with no PCT are performed. This sample with no PCT was prepared either by storage 1 day at $37^{\circ}\mathrm{C}$ or by addition of A179 anti-PCT polyclonal antibody. At the end, this "free PCT sample" was different in each run.
20 determination of RS002, a recombinant PCT preparation with a very low amount of PCT are performed. This sample is the same for each run.
This protocol is reproduced three time; 3 different KRYPTOR PCT reagent unit batches and 2 different KRYPTOR were used for these 3 runs.
Then, analysis of measurement is done as follows:
Firstly, Limit of Blank (LoB) value is determined by the 95th percentile of concentrations for a sample that does not contain PCT.
Then, Limit of Detection (LoD = Analytical detection limit) is equal to $\mathrm{LoB} + \mathrm{C}\beta \mathrm{xSDs}$ , where $\mathrm{C}\beta$ is derived from the standard Gaussian distribution ( $\mathrm{C}\beta = 1.645 / (1 - 1 / (4\mathrm{xf}))$ , where $\mathrm{f}$ is the degrees of freedom for the estimated standard deviation SDs) and SDs is the standard deviation of concentration for sample RS002.
The analytical detection limit determination is based on 3 runs utilizing three (3) different lots of B·R·A·H·M·S PCT sensitive KRYPTOR® and two (2) KRYPTOR® instruments.
# Calculation of LoD:
Results of the 60 measurements of sample PCT_RS002
| replicate | run | value (ng/ml) | run | value (ng/ml) | run | value (ng/ml) |
| --- | --- | --- | --- | --- | --- | --- |
| 1 | run1 | 0.0353 | run2 | 0.0315 | run3 | 0.0988 |
| 2 | run1 | 0.0569 | run2 | 0.0417 | run3 | 0.0527 |
| 3 | run1 | 0.0568 | run2 | 0.0227 | run3 | 0.0626 |
| 4 | run1 | 0.0185 | run2 | 0.0372 | run3 | 0.0558 |
| 5 | run1 | 0.0373 | run2 | 0.0635 | run3 | 0.0669 |
| 6 | run1 | 0.0438 | run2 | 0.0426 | run3 | 0.0558 |
| 7 | run1 | 0.0432 | run2 | 0.0474 | run3 | 0.0550 |
| 8 | run1 | 0.0148 | run2 | 0.0503 | run3 | 0.0495 |
| 9 | run1 | 0.0556 | run2 | 0.0506 | run3 | 0.0433 |
| 10 | run1 | 0.0495 | run2 | 0.0515 | run3 | 0.0651 |
| 11 | run1 | 0.0491 | run2 | 0.0644 | run3 | 0.0587 |
| 12 | run1 | 0.0497 | run2 | 0.0318 | run3 | 0.0561 |
{7}
Standard deviation of RS002 is calculated based on the pooled 60 determinations. $\mathrm{SD}_{\mathrm{s}} = 0.01374$
Then, LoD is calculated as follows:
$\mathrm{df} = 60 - 1 =$ 59
$\mathrm{C}_{\beta} = 1.6449 / (1 - 1 / 236) =$ 1.65190
$\mathrm{SD_s}$ 0.01374
$\mathrm{LoD} = \mathrm{LoB} + \mathrm{C}_{\beta}\left(\mathrm{SD}_{\mathrm{s}}\right) =$ 0.0227 ng/ml
Limit of Detection for the B·R·A·H·M·S PCT sensitive KRYPTOR® is 0.0227 ng/ml.
The Limit of Quantitation (LOQ) was determined following EP17-P as follows:
. Samples at different targets (from $0.06\mathrm{ng / ml}$ to $0.075\mathrm{ng / ml}$ ) were prepared with master calibrators for which actual concentrations were determined independently.
. These samples were run in 5 runs, with 10 replicates per run, thus a total of 50 replicates per sample.
. For the 5 runs, 3 different KRYPTOR instruments and 2 different batches of reagents were used.
. For each sample, the total standard deviation $(\mathrm{SD}_{\mathrm{s}})$ was calculated as well as the difference between the mean of all replicates and the reference value of the sample (bias) and imprecision with $95\%$ probability $(2\times \mathrm{SD}_{\mathrm{s}})$ .
The LOQ determined as the lowest reported concentration level with total error (imprecision + bias) $\leq 30\%$ was determined at $0.075\mathrm{ng / ml}$
Results for replicates and $\mathrm{SD}_{\mathrm{s}}$ , imprecision and bias calculation are shown below:
| Replicates | Run 1 | Run 2 | Run 3 | Run 4 | Run 5 |
| --- | --- | --- | --- | --- | --- |
| 1 | 0.0868 | 0.0809 | 0.0623 | 0.0687 | 0.0726 |
| 2 | 0.0611 | 0.0761 | 0.0708 | 0.0644 | 0.0861 |
| 3 | 0.0841 | 0.0754 | 0.0622 | 0.0663 | 0.0780 |
| 4 | 0.0525 | 0.1016 | 0.0759 | 0.0887 | 0.0588 |
| 5 | 0.0774 | 0.0751 | 0.0735 | 0.0901 | 0.0782 |
| 6 | 0.0578 | 0.0790 | 0.0569 | 0.0619 | 0.0686 |
| 7 | 0.0735 | 0.0682 | 0.0779 | 0.0658 | 0.0765 |
{8}
| 8 | 0.0872 | 0.0767 | 0.0855 | 0.0798 | 0.0807 |
| --- | --- | --- | --- | --- | --- |
| 9 | 0.0895 | 0.0755 | 0.0871 | 0.0715 | 0.0819 |
| 10 | 0.0604 | 0.0810 | 0.0753 | 0.0693 | 0.0785 |
Average value (ng/ml): 0.0747 ng/ml
Target (ng/ml): 0.0750 ng/ml
Bias (ng/ml): -0.0003 ng/ml
$\mathrm{SD}_{\mathrm{s}}$ (ng/ml): 0.0102 ng/ml
Imprecision (ng/ml): 0.0204 ng/ml
Total Error (ng/ml): 0.0207 ng/ml
Total Error/Target: $27.6\% < 30\%$
Therefore, it is concluded that LOQ for the B·R·A·H·M·S PCT sensitive KRYPTOR® is 0.075ng/ml
e. Analytical specificity:
The following substances were evaluated with the BRAHMS PCT sensitive Kryptor assay and were found not to affect the test performance at clinically relevant concentrations: Substances tested were bilirubin, hemoglobin, triglycerides, albumin, imipenem, cefotaxim, vancomycin, dopamine, noradrenaline, dobutamine, heparin, furosemide and drugs used in asthmatic and COPD patients. In addition calcitonin, katacalcin, a-CGRP, $\beta$ -CGRP, Calcitonin salmon and Calcitonin Eel showed no interference when tested with the assay.
f. Assay cut-off:
The assay cut off is $0.02\mathrm{ng / ml}$
2. Comparison studies:
a. Method comparison with predicate device:
A correlation study was performed between the BRAHMS PCT sensitive KRYPTOR assay and the BRAHMS PCT LIA assay. There were 184 samples from three sites, which had B·R·A·H·M·S PCT LIA measurements of $0.3\mathrm{ng / ml}$ (the functional assay sensitivity of B·R·A·H·M·S PCT LIA) or higher and/or B·R·A·H·M·S PCT sensitive KRYPTOR® measurements of $0.06\mathrm{ng / ml}$ (the functional assay sensitivity of B·R·A·H·M·S PCT sensitive KRYPTOR®) or higher. Passing-Bablock analysis shows a nearly perfect correlation of the B·R·A·H·M·S PCT sensitive KRYPTOR® assay and B·R·A·H·M·S PCT LIA assay, as demonstrated in the correlation graph below.
{9}

b. Matrix comparison:
A study was conducted comparing serum and plasma samples using 10 spiked patient specimens. The samples were tested in triplicate using serum tubes, heparin plasma tubes, EDTA plasma tubes and citrate plasma tubes. Results demonstrated that both serum and plasma tubes can be used for the assay but citrate plasma tubes are not recommended because concentrations of PCT decreased in those tubes.
3. Clinical studies:
a. Clinical Sensitivity:
N/A
b. Clinical specificity:
N/A
c. Other clinical supportive data (when a. and b. are not applicable):
4. Clinical cut-off:
{10}
Data support the following interpretative risk assessment criteria:
**PCT > 2 ng/ml**
PCT levels above 2.0 ng/ml on the first day of ICU admission represent a high risk for progression to severe sepsis and/or septic shock.
**PCT < 0.5 ng/ml**
PCT levels below 0.5 ng/ml on the first day of ICU admission represent a low risk for progression to severe sepsis and/or septic shock.
**Note:** PCT levels below 0.5 ng/ml do not exclude an infection, because localized infections (without systemic signs) may also be associated with such low levels. If the PCT measurement is done very early after the systemic infection process has started (usually < 6 hours), these values may still be low.
As various non-infectious conditions are known to induce PCT as well, PCT levels between 0.5 ng/ml and 2.0 ng/ml should be reviewed carefully to take into account the specific clinical background and condition(s) of the individual patient.
5. **Expected values/Reference range:**
A study was conducted with the BRAHMS KRYPTOR to determine the prevalence of PCT in a normal population in the US. Data showed that the PCT concentrations for 146 of the 151 samples tested were < 0.1 ng/ml. Five samples were greater than 0.1 ng/ml.
**N. Instrument Name:**
BRAHMS KRYPTOR analyzer
**O. System Descriptions:**
1. **Modes of Operation:**
The analyzer is provided in two models, the Model E and the Compact. They differ in physical size and consequently in the number of wells in the reaction and dilution plates and the number of tests that can be performed. Both models have the same features including the same laser, methodology, incubation time and temperature and calculations/algorithms as well as utilizing the same reagents, calibrators and controls. Documentation provided for Model E is applicable to both models. The analyzer is a fully automated system, able to process multiple samples each day in random access mode. The analyzer is a closed system and can only operate utilizing especially made reagent kits from BRAHMS. The system is based on TRACE (Time Resolved Amplified Cryptate Emission) technology. The analyzer has a built in computer to handle the required controls to store reagents, dispense reagents, patient samples, calibrators and controls, and incubate the reagent-sample mixture and measure sample concentration. The external computer will program the analyzer for the required tests and calculate
11
{11}
patient results.
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ X or No
3. Specimen Identification:
The assay procedure includes registering and/or loading the samples, reagent kit, calibrator and controls as applicable
4. Specimen Sampling and Handling:
A sample volume of 50μl is needed for each test. Initially a worklist for the day is created and the test is started. The sample probe of the analyzer pipettes and dispenses the conjugates from the reagent kit and the sample into the wells. The probe is heated to incubate the reagent-sample mixture so it is at 37°C prior to dispensing and mixing in the reaction well. The analyzer makes periodic measurement of the signal emitted. Samples >50ng/ml are detected, diluted and re-assayed automatically. After measurement of the fluorescent signal, the program compares each result obtained with the stored standard curve.
5. Calibration:
A standard curve does not need to be established for the BRAHMS PCT sensitive KRYPTOR assay on the analyzer. A standard curve is included with the bar code information from the calibration card and is stored in the analyzer. A calibration is carried out before the first use of a reagent batch, and then repeated on a regular basis (i.e. first use and every 15 days thereafter). Calibrations are performed using a disposable calibrator vial in order to readjust the standard curve.
6. Quality Control:
The KRYPTOR QC is provided for use as quality control on board the instrument for the assay of procalcitonin. The kit contains 2 series of 3 lyophilized vials, a bar code card, bar code stick-on labels and the summary of concentration ranges by level. The bar code card contains information related to the control batch, particularly the target concentrations, the standard deviations obtained and the concentration acceptance ranges. Controls are flagged above or below normal (±2 SD) according to the defined control range. The BRAHMS KRYPTOR can automatically check the quality of assays at intervals by statistical analysis on Levy Jennings graphs. National quality assurance guidelines for medical lab tests must be followed e.g. test accuracy and precision can be monitored by means of lab in-house and/or commercially available control materials
12
{12}
P. Other Supportive Instrument Performance Characteristics Data Not Covered In the "Performance Characteristics" Section above:
N/A
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
1. The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
13
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.