PAINT- ON POLISH AGENT

K070306 · Dentsply International, Inc. · EBD · Mar 22, 2007 · Dental

Device Facts

Record IDK070306
Device NamePAINT- ON POLISH AGENT
ApplicantDentsply International, Inc.
Product CodeEBD · Dental
Decision DateMar 22, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3310
Device ClassClass 2

Indications for Use

Surface and enamel margin treatment of Direct, light-cured composite restorations. 1. Surface treatment of Indirect restorations; composite resin or methacrylate veneers, 2. inlays, onlays, crown and bridge retainers.

Device Story

Paint-on Polish Agent is a liquid dental coating applied to permanent and temporary restorations to eliminate manual polishing and serve as a sealant to reduce micro-leakage at the tooth-composite interface. The formulation contains fillers for wear resistance and is cured using LED or halogen light. It provides stain resistance, strong adhesion, and blue fluorescence. Used in dental clinics by practitioners; applied directly to restoration surfaces. An oxygen-inhibited layer forms post-cure, removable with ethanol or isopropanol. Benefits include improved surface finish, reduced micro-leakage, and enhanced aesthetic properties.

Clinical Evidence

Bench testing only. Biocompatibility testing performed for cytotoxicity, genotoxicity, irritation, and sensitization. No clinical data provided.

Technological Characteristics

Liquid coating material for resin fillings. Contains fillers for wear resistance. Curing mechanism: light-cured (LED or halogen). Biocompatible materials. Class II device (21 CFR 872.3310).

Indications for Use

Indicated for surface and enamel margin treatment of direct light-cured composite restorations and surface treatment of indirect restorations including composite resin or methacrylate veneers, inlays, onlays, crown and bridge retainers.

Regulatory Classification

Identification

A coating material for resin fillings is a device intended to be applied to the surface of a restorative resin dental filling to attain a smooth, glaze-like finish on the surface of the filling.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K670306 # SECTION 5: 510(k) SUMMARY MAR 2 2 2007 DENTSPLY International Susquehanna Commerce Center West 221 West Philadelphia Street, Suite 60 York, PA 17405-0872 | CONTACT: | Helen Lewis | |----------------------------|--------------------------------------------------------------------| | DATE PREPARED: | January 30, 2007 | | TRADE OR PROPRIETARY NAME: | Paint-on Polish Agent | | CLASSIFICATION NAME: | Coating Material for Resin Fillings,<br>Class II per 21CFR872.3310 | | PREDICATE DEVICES: | Biscover LV. K043168. | ### DEVICE DESCRIPTION: The Paint-on Polish Agent is liquid Polish agent designed eliminate the Polishing procedure by painting the formulation on the surface of restorations directly. It is indicated for use on both permanent and temporary restorations. Paint-on Polish can also be used as a sealant to decrease the micro-leakage at tooth-composite restoration interface. Due to the unique chemistry, Paint-on Polish formulation can be cured by both LED and halogen light. The addition of fillers provides excellent wear resistance for the formulation. The cured coating surface has a strong adhesion to restoration and excellent stain resistance to food and drinks. It also emits blue fluorescence close to enamel surface. Paint-on Polish formulation has a thin oxygen inhibited layer after curing, but it can be easily removed by ethanol or isopropanol without reducing the glossiness of coating. ### INTENDED USE: - Surface and enamel margin treatment of Direct, light-cured composite restorations. 1. - Surface treatment of Indirect restorations; composite resin or methacrylate veneers, 2. inlays, onlays, crown and bridge retainers. ## TECHNOLOGICAL CHARACTERISTICS: All of the components found in Paint-on Polish Agent have been used in legally marketed devices and were found safe for dental use. Paint-on Polish has been evaluated and passed biocompatibility testing for cytoxicity, genotoxicity, irritation and sensitization. We believe that the prior use of the components of Paint-on Polish in legally marketed devices, the performance data provided, and the biocompatibility data provided support the safety and effectiveness of Paint-on Polish for the indicated uses. 000010 {2}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/1/Picture/1 description: The image shows the logo for the United States Department of Health and Human Services. The logo consists of a stylized eagle with three overlapping wings, representing the department's mission to protect the health of all Americans and provide essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle symbol. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Helen Lewis Director of Corporate Compliance and Regulatory Affairs DENTSPLY International Susquehanna Commerce Center West 221 West Philadelphia Street, Suite 60 York, Pennsylvania 17405-0872 MAR 2 & 2007 Re: K070306 Trade/Device Name: Paint-on Polish Agent Regulation Number: 21 CFR 872.3310 Regulation Name: Coating Material for Resin Fillings Regulatory Class: II Product Code: EBD Dated: January 30, 2007 Received: February 09, 2007 Dear Ms. Lewis: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ ### Page 2 - Ms. Helen Lewis Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510/k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Chiu S. Lin, PhD Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # SECTION 4: INDICATIONS FOR USE K07 0306 510(k) Number (if known): Device Name: Paint-on Polish Agent Indications for Use: Paint-on Polish Agent is indicated for - 1. Surface and enamel margin treatment of Direct, light-cured composite restorations. - 2. Surface treatment of Indirect restorations; composite resin or methacrylate veneers, inlays, onlays, crown and bridge retainers. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Sver Qumne ...tion Cont... 70304 00009 Paint-on Polish Agent
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...