OPUS MAGNUM PI KNOTLESS FIXATION DEVICE

K070227 · Arthrocare Corp. · MBI · Apr 16, 2007 · Orthopedic

Device Facts

Record IDK070227
Device NameOPUS MAGNUM PI KNOTLESS FIXATION DEVICE
ApplicantArthrocare Corp.
Product CodeMBI · Orthopedic
Decision DateApr 16, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Opus® Magnum™ PI bone anchor with inserter is indicated for use in fixation of soft tissue to bone. Examples of such procedures include: Shoulder: Bankart Repair, SLAP lesion repair, acromio-clavicular separation, rotator cuff repair, capsule shift/capsule-labral reconstruction, biceps tenodesis and deltoid repair Ankle: Lateral instability, medial instability, Achilles tendon repair/reconstruction and midfoot reconstruction Foot: Hallux valgus reconstruction Elbow: Tennis elbow repair, biceps tendon attachment Knee: Extra-capsular repairs; reattachment of: medial collateral ligament, posterior oblique ligament or joint capsule closure to anterior proximal tibia; extra capsular reconstruction, ITB tenodesis; patellar ligament and tendon avulsions

Device Story

Bone anchor with inserter handle; used in arthroscopic and orthopedic procedures; facilitates soft tissue-to-bone fixation. Device functions as mechanical fastener; replaces stainless steel version with PEEK material. Operated by surgeons in clinical/OR settings. Provides secure attachment of tendons/ligaments to bone; aids in structural repair of joints; improves patient recovery outcomes.

Clinical Evidence

No clinical data; bench testing only.

Technological Characteristics

Bone anchor with inserter handle; material: PEEK (polyether-etherketone); mechanical fixation principle; non-degradable; class II device.

Indications for Use

Indicated for soft tissue to bone fixation in shoulder, ankle, foot, elbow, and knee orthopedic procedures, including repairs for instability, tendon/ligament reconstruction, and avulsions.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Ko70227 pge 'f² # 510(K) SUMMARY ### ARTHROCARE CORPORATION OPUS MAGNUM PI KNOTLESS FIXATION DEVICE APR 16 2007 ### General Information | Submitter Name/Address: | ArthroCare Corporation<br>680 Vaqueros Avenue<br>Sunnyvale, CA 94085-3523 | |---------------------------------|-----------------------------------------------------------------------------------------------------| | Establishment Registration No.: | 2951580 | | Contact Person: | Laura N. Kasperowicz<br>Sr. Manager, Regulatory Affairs | | Date Prepared: | January 23, 2007 | | Device Description | | | Trade Name: | Opus® Magnum™ PI | | Generic/Common Name: | Bone Anchor, Fastener, Fixation, Soft Tissue | | Classification Name: | Fastener, Fixation, Nondegradeable, Soft Tissue<br>(Class II per 21 CFR 888.3040, Product code: MB) | | Predicate Devices | | | Opus Magnum | K012125 (Cleared 09/17/01) | | Opus LabraLock P | K061349 (Cleared 07/14/06) | ### Product Description The Opus Magnum™ PI device is a bone anchor with inserter handle designed for specific indications in arthroscopic and orthopedic procedures. ### Indications For Use The Opus® Magnum™ PI bone anchor with inserter is indicated for use in fixation of soft tissue to bone. Examples of such procedures include: Shoulder: Bankart Repair, SLAP lesion repair, acromio-clavicular separation, rotator cuff repair, capsule shift/capsule-labral reconstruction, biceps tenodesis and deltoid repair Ankle: Lateral instability, medial instability, Achilles tendon repair/reconstruction and midfoot reconstruction Foot: Hallux valgus reconstruction Elbow: Tennis elbow repair, biceps tendon attachment Knee: Extra-capsular repairs; reattachment of: medial collateral ligament, posterior oblique ligament or joint capsule closure to anterior proximal tibia; extra capsular reconstruction, ITB tenodesis; patellar ligament and tendon avulsions {2}------------------------------------------------ pge yf2 ## 510(K) SUMMARY ### Substantial Equivalence By definition, substantial equivalence means that a device has the same intended use and technical characteristics as the predicate device, or has the same intended use and different technological characteristics, but can be demonstrated to be as safe and effective as the predicate device. The Opus® Magnum™ PI design and technology is substantially equivalent to the existing Opus® Magnum™ Knotless Fixation Device cleared by the Food & Drug Administration [K012125, and subsequently amended by K020172, K031083 and K042914]. The Opus" Magnum™ PI material is substantially equivalent to the existing Opus" LabraLock™ P Knotless Fixation Device cleared by the Food & Drug Administration [K061349]. The differences between the Opus® Magnum™ PI and the predicate devices do not raise any questions regarding the safety and effectiveness of the implant. Furthermore, the materials are well characterized and have been used in predicate devices with similar indications. The device, as designed, is as safe and effective as predicate devices. ### Summary and Reason for 510k Notification The purpose of this 510k is to notify the Food and Drug Administration of a new product, the Opus Magnum™ PI Knotless Fixation Device. This new product is substantially equivalent to the Opus® Magnum™ Knotless Fixation Device originally cleared under K012125 [and subsequently amended by K020172. K031083 and K042914], but is manufactured from PEEK (polyether-etherketone) as opposed to stainless steel. {3}------------------------------------------------ Image /page/2/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of three faces in profile, stacked on top of each other. Re: Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 ArthroCare Corporation % Ms. Laura N. Kasperowicz Senior Manager, Regulatory Affairs 680 Vaqueros Avenue Sunnyvale, California 94085-3523 APR 1 6 2007 K070227 Trade/Device Name: Opus® Magnum™ PI Knotless Fixation Device Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI, HWC Dated: January 23, 2007 Received: January 24, 2007 Dear Ms. Kasperowicz: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ Page 2 - Ms. Laura N. Kasperowicz This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or 240-276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Darbar Bonchum Mark N. Melk Difector . Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ page 1 of 1 ### INDICATIONS FOR USE STATEMENT 510(k) Number: Device Name: Opus™ Magnum™ PI Knotless Fixation Device Indications for Use: The Opus® Magnum™ PI bone anchor with inserter is indicated for use in fixation of soft tissue to bone. Examples of such procedures include: Shoulder; Bankart Repair, SLAP lesion repair, acromio-clavicular separation, rotator cuff repair, capsule shift/capsule-labral reconstruction, biceps tenodesis and deltoid repair Ankle: Lateral instability, medial instability, Achilles tendon repair/reconstruction and midfoot reconstruction Foot: Hallux valgus reconstruction Elbow: Tennis elbow repair, biceps tendon attachment KO 7022 Knee: Extra-capsular repairs; reattachment of: medial collateral ligament, posterior oblique ligament or joint capsule closure to anterior proximal tibia; extra capsular reconstruction, ITB tenodesis; patellar ligament and tendon avulsions X Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) NO (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Barbara Buchus Division of Ger neral, Restorative. and Neurological Devices **510(k) Number** K070227
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...