SUNIRAY II DIGITAL RADIOGRAPHY SYSTEM

K070219 · Suni Medical Imaging, Inc. · MUH · Feb 7, 2007 · Dental

Device Facts

Record IDK070219
Device NameSUNIRAY II DIGITAL RADIOGRAPHY SYSTEM
ApplicantSuni Medical Imaging, Inc.
Product CodeMUH · Dental
Decision DateFeb 7, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.1800
Device ClassClass 2
Attributes3rd-Party Reviewed

Indications for Use

The SuniRay II Digital Radiography System is used to collect dental x-ray photons and convert them into electronic impulses that may be stored, viewed, and manipulated for diagnostic use by dentists.

Device Story

SuniRay II Digital Radiography System captures dental X-ray photons using CMOS-based sensors (#1 and #2 sizes) with high-resolution scintillator screens. Sensors connect via USB electronics box to a computer. System converts X-ray photons into electronic signals for digital storage, viewing, and manipulation. Used in dental clinics by dentists. Software includes device drivers and a Graphical User Interface (GUI) for image acquisition, enhancement (zoom, pan, colorize, contrast/brightness), comparative analysis, and storage. System reduces X-ray dosage compared to traditional film. Output viewed on computer monitor; supports printing or email. Benefits include real-time digital imaging and improved diagnostic workflow.

Clinical Evidence

Bench testing only. The device meets EN 60601 requirements for safety and electromagnetic compatibility. No clinical data provided.

Technological Characteristics

CMOS-based integrated circuit sensor with high-resolution scintillator screen. USB-connected electronics box for data acquisition and firmware control. Operates on Windows XP. Connectivity via USB port. Complies with EN 60601 safety and EMC standards.

Indications for Use

Indicated for dental patients requiring intra-oral radiographic imaging for diagnostic purposes. Used by dentists.

Regulatory Classification

Identification

An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Suni Medical Imaging, Inc. - Pre-Market Notification for the SuniRay II Digital Radiography System - 1/5/2007 # 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS K070219 #### Device Name Trade Name Classification Name Common Name .... SuniRay II Digital Radiography System Digital Extraoral Source X-Ray System Digital X-Ray 7 2007 EEB #### Classification Device Panel Procode Reference Class II Radiology MUH 21 CFR 872.1800 #### Registration Number 3003952803 #### Sponsor Name and Address Suni Medical Imaging, Inc. 6840 Via Del Oro San Jose, CA 95119 #### Manufacturer Name and Address Suni Medical Imaging, Inc. 6840 Via Del Oro San Jose, CA 95119 #### Reason for Pre-Market Notification New design #### Equivalent Legally Marketed Device K021718 Suni Intraoral Imaging System #### Indications for Use: The SuniRay II Digital Radiography System is used to collect dental x-ray photons and convert them into electronic impulses that may be stored, viewed, and manipulated for diagnostic use by dentists. {2}------------------------------------------------ # 510(k) Summary of Safety and Effectiveness The Suni SuniRay II Digital Radiography System ("System") will produce real time digital intra-oral images with a reduction in X-ray dosages as compared to film. The System accomplishes this by replacing X-ray film with an electronic sensor that captures the X-ray photons and converts the photons to an electronic signal which in turn is captured in a computer for viewing, manipulating, storing, and outputting (printing or Email). The System consists of two sensor versions (designated as #1 & #2 sizes; equivalent to dental film sizes), an attached USB electronics box that controls the sensor and interfaces to the computer USB port. The System also includes software drivers that control the USB electronics box, and a Graphical User Interface (GUI) that allows the user access to the data and control functions of the System. The GUI used with the device was designed by Apteryx, Inc. who holds 510(k) K983111 for their product. Apteryx is responsible for both product validation and GMP compliance. The SuniRay II System software utilizes the Windows XP environment. The System sensors are in two formats. The #2 size is a larger sensor typical of the European format and the # 1 size is an intermediate size. The sensors consist of a CMOS type integrated circuit, and a high resolution scintillator screen that converts the photons from the X-rays into visible light, which is then acquired by the CMOS imaging integrated circuit. The sensors are encapsulated with a three-foot cable that is connected to the USB electronics box. The USB electronics box plugs into a computer USB port via a supplied USB cable. The USB electronics box contains the support and control circuitry for the sensor and allows for data communications with the computer. The USB electronics box contains all necessary circuits for sensor data acquisition as well as memory for firmware control of the CMOS X-ray sensor and USB. The USB electronics box communicates with the computer under control of a specific device driver that is active with the GUI. The System software functions on three levels: (1) The computer operating system (Microsoft Windows) controls the computer, user interface, and file structure; (2) Primary control of the sensor and bus functions is achieved by proprietary software and is either embedded firmware or in non user accessible drivers; and (3) A Graphical User Interface (GUI) allows the user to control the x-ray function, control of the sensor data acquisition, and image viewing, manipulation and output. Examples of the GUI include image capture, enhanced viewing features (zoom, pan, colorize, contrast/brightness, comparative analysis, etc) image organization, and storage. The System meets the requirements of EN 60601 for safety and electromagnetic compatibility. For this Pre-Market Notification, similarities and changes made to this System from the predicate are summarized in Section 8. {3}------------------------------------------------ Image /page/2/Picture/0 description: The image shows a circular logo with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES" written around the edge. Inside the circle is a stylized graphic of three human figures connected at the hands, forming a chain-like shape. The figures are abstract and represented by curved lines, giving the impression of movement and connection. The logo is black and white. ### DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850 Suni Medical Imaging, Inc. % Mr. Morten Simon Christensen Staff Engineer & FDA Office Coordinator Underwriters Laboratories, Inc. 455 E. Trimble Road SAN JOSE CA 95131 7 2007 Re: K070219 Trade/Device Name: SuniRay II Digital Radiography System Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulation Name: Extraoral source x-ray system Regulatory Class: II Product Code: MUH Dated: January 19, 2007 Received: January 23, 2007 Dear Mr. Christensen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device American in microade of the have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Image /page/2/Picture/12 description: The image shows a circular logo with the letters "FDA" in the center. Above the letters, the numbers "1906-1996" are displayed. Below the letters, the word "Centennial" is written in a cursive font. There are three stars below the word "Centennial". Protecting and Promoting Public Health {4}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter: | 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 | |----------------|----------------------------------|--------------| | 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 | | 21 CFR 894.xxx | (Radiology) | 240-276-0120 | | Other | | 240-276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html. Sincerely yours. Nancy C Hodgon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Suni Medical Imaging, Inc. – Pre-Market Notification for the SuniRay II Digital Radiography System – 1/5/2007 # Indications for Use 510(k) Number (if known): Hot assigned K070217 Device Name: SuniRay II Digital Radiography System ... Indications for Use: The SuniRay II Digital Radiography System is used to collect dental x-ray photons and convert them into electronic impulses that may be stored, viewed, and manipulated for diagnostic use by dentists. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) David A. Szymanski --- (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 070219 510(k) Number Page 1-2
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...