K070204 · Welch Allyn Protocol, Inc. · MSX · Mar 23, 2007 · Cardiovascular
Device Facts
Record ID
K070204
Device Name
LINK TO ACUITY CLINICIAN NOTIFIER
Applicant
Welch Allyn Protocol, Inc.
Product Code
MSX · Cardiovascular
Decision Date
Mar 23, 2007
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.2300
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Acuity® Central Monitoring System, in all configurations including the hardwired and wireless Acuity LT and Mobile Acuity LT System configurations, is intended to be used by clinicians for the central monitoring of neonatal, pediatric and adult patients in healthcare facilities. In addition to the central monitoring of patient data, waveforms, and alarms and alerts, Acuity System software can include optional modules to provide extended recording and analysis of patient data. Acuity System software can include the following optional modules: The Full Disclosure module stores patient data for up to 96 hours (24 hours with Acuity LT and . Mobile Acuity LT). The Arrhythmia Analysis module provides real-time monitoring and alarms for specific changes in cardiac rhythms. The clinician is responsible for determining the clinical significance of each detected arrhythmia event or alarm. The Arrhythmia Analysis module is not intended for use with neonatal patients. The ST Analysis module provides real-time monitoring and alarms for ST segment deviations for patients with suspected heart disease and anomalies. The clinician is responsible for determining the clinical significance of each selected ST segment deviation or alarm. The ST Analysis option is not intended for use with neonatal patients. The Welch Allyn Connectivity Server (WACS) module consists of a server platform on which one or more of the following software options are installed: Link To Acuity (Acuity System patient information delivered to mobile devices running Clinician Notifier software), Web Server (Acuty System patient printout files available on Web browsers) and HL7 options.
Device Story
Link To Acuity Clinician Notifier is a mobile alarm management extension for the Acuity Central Monitoring System. It receives patient alarm data, real-time waveforms, and historical alarm details from the Acuity system via the Welch Allyn Connectivity Server (WACS). Clinicians use non-proprietary mobile devices to view patient status, respond to alarms, and track clinician-patient assignments. The system supports barcode scanning for patient ID entry. It is intended for use by authorized healthcare professionals (nurses, physicians, technicians) as a supplemental notification tool, not as a primary alarm device. It facilitates remote monitoring, allowing clinicians to respond to patient needs from outside the central station.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including simulated clinical engineering test protocols and pre-validation clinical entry testing to verify software performance and alarm management functionality.
Technological Characteristics
Networked software system; operates on WACS server platform and mobile devices. Connectivity via wireless/hardwired network. Software-based alarm management and data visualization. Quality system conforms to ISO 13485:2003 and 21 CFR 820.
Indications for Use
Indicated for central monitoring of neonatal, pediatric, and adult patients in healthcare facilities. Arrhythmia and ST analysis modules are contraindicated for neonatal patients.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Welch Allyn. The text is in a bold, sans-serif font. The logo is black and the background is white.
Special 510(k) Premarket Notification Link To Acuity® Clinician Notifier Option (Modification to Acuity® Central Monitoring System)
# Premarket Notification [510(k)} Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92. Information supporting claims of substantial equivalence, as defined under the Federal Food, Drug and Cosmetic Act, respecting safety and effectiveness, is outlined below.
| Date Prepared: | March 5, 2007 |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name of Submitter: | Welch Allyn Protocol, Inc.<br>8500 S.W. Creekside Place<br>Beaverton, OR 97008-7107 USA<br><br>Contact:<br>Jeffrey W. Wagner, Director, Regulatory Affairs<br>Phone: (503) 530-7909<br>Fax: (503) 526-4901 |
| Device Proprietary Name: | Link To Acuity® Clinician Notifier<br>(Modification to the Acuity® Central Monitoring System) |
| Classification Name: | System, network and communication, physiological monitors (MSX)<br>Detector and Alarm, Arrhythmia (DSI) |
| Common/Usual Names: | Central monitoring system |
## Predicate Devices:
| K052160 | Acuity® Central Monitoring Station | Welch Allyn Protocol, Inc. |
|---------|------------------------------------|----------------------------|
| K052975 | S/5 Pocket Viewer | Datex-Ohmeda |
# Device Description:
The Acuity® Central Monitoring System, in all configurations including the hardwired and wireless Acuity LT and Mobile Acuity LT System configurations, is intended to be used by clinicians for the central monitoring of neonatal, pediatric and adult patients in healthcare facilities.
In addition to the central monitoring of patient data, waveforms, and alarms and alerts, Acuity System software can include optional modules to provide extended recording and analysis of patient data. Acuity System software can include the following optional modules:
- The Full Disclosure module stores patient data for up to 96 hours (24 hours with Acuity LT and . Mobile Acuity LT).
MAR 2 3 2007
K070204/51
{1}------------------------------------------------
# WelchAllyn
Special 510(k) Premarket Notification Link To Acuity® Clinician Notifier Option (Modification to Acuity® Central Monitoring System)
C7C20451 72/3
- . The Arrhythmia Analysis module provides real-time monitoring and alarms for specific changes in cardiac rhythms. The clinician is responsible for determining the clinical significance of each detected arrhythmia event or alarm. The Arrhythmia Analysis module is not intended for use with neonatal patients.
- . The ST Analysis module provides real-time monitoring and alarms for ST segment deviations for patients with suspected heart disease and anomalies. The clinician is responsible for determining the clinical significance of each selected ST segment deviation or alarm. The ST Analysis option is not intended for use with neonatal patients.
- . The Welch Allyn Connectivity Server (WACS) module consists of a server platform on which one or more of the following software options are installed: Link To Acuity (Acuity System patient information delivered to mobile devices running Clinician Notifier software), Web Server (Acuty System patient printout files available on Web browsers) and HL7 options.
The Link To Acuity option is Welch Allyn's mobile alarm management solution. It consists of Clinician Notifier software for non-proprietary mobile devices and administrative software for the Welch Allyn Connectivity Server (WACS).
Mobile devices running the Clinician Notifier software deliver patient alarm information and real-time waveforms gathered from patient monitors connected to the Acuity Central Monitoring System. The devices are not intended for use as primary alarm notification devices. Devices running Clinician Notifier software are not directly connected to patients.
The Link To Acuity option is designed to extend the patient monitoring functions of the Acuity Central Monitoring System. The software enables administrators to track the status of clinician-patient assignments, and it enables clinicians to track, respond to and view Acuity System patient alarms, view historical alarm details and waveforms, and view real-time patient waveforms. The mobile devices can be used for barcode scanning to enter patient ID and room number.
The Link To Acuity option is to be used by authorized health care professionals using standard institutional procedures and good clinical practice guidelines for patient monitoring. Staff training in the operation of the Link To Acuity option is essential for optimal use. Users should be skilled at the level of a technician, nurse, physician, health care provider or medical specialist, with the knowledge and experience to acquire and interpret patients' vital signs data.
Individuals using the Link To Acuity option should be familiar with its operation as described in the directions for use manual, and they should understand all warnings and cautions in the directions for use manual.
# Technological Characteristics:
The Welch Allyn Link To Acuity Clinician Notifier option allows for alarm management remote from the central monitoring station. Through a network of wireless and/or hardwired devices connected to the Welch Allyn Acuity Central Monitoring System, alarms, real time data, and other information can be viewed from a generic mobile computer configured with Welch Allyn Link To Acuity Clinician Notifier option software. The Welch Allyn Link To Acuity Clinician Notifier is to be used as a supplement to other patient monitoring and not as a primary alarm notification device.
{2}------------------------------------------------
# WelchAllyn
Special 510(k) Premarket Notification Link To Acuity® Clinician Notifier Option (Modification to Acuity® Central Monitoring System)
<07020451 \$ 3/3
# Link To Acuity® Clinician Notifier Features
# View Patient Info
Status of all patients in unit History of alarms Real-time waveform List of current alarms Details of an alarm
# Respond to Alarms and Alerts
Patient alarms delivered to Clinician Notifier devices Two possible alarm delivery modes - Escalation and Broadcast Alarm notifications Respond to patient alarms
#### Security
Login Logout
# Patient Assignment
Assign patients to clinicians Enter and confirm a patient's ID
# Performance Data:
The subject device has been tested in accordance with the following Welch Allyn documents using production equivalent units prior to market release.
| 830-1423-00 | Test Plan, Acuity V7.00 |
|-------------|--------------------------------------------------------------------------------|
| 831-1285-00 | Pre-Validation Clinical Entry Test Report Clinician Notifier 2.35/WACS 2.10.00 |
| 830-1472-00 | Welch Allyn Link to Acuity® - Simulated Clinical Engineering Test Protocol |
| 831-1248-00 | Link To Acuity® Simulated Clinical Testing Report |
A risk analysis identifying potential hazards and documenting mitigation of the hazards has been developed and applied as part of Welch Allyn Protocol's product development procedure. Welch Allyn Protocol's Quality System conforms to 21 CFR 820 and is certified to ISO 13485:2003.
# Conclusion:
Based on the information contained in this submission, Welch Allyn Protocol, Inc. believes that the subject device is substantially equivalent to the predicate devices.
{3}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 2 3 2007
Welch Allyn Protocol, Inc. c/o Mr. Jeffrey Wagner Director, Regulatory Affairs 8500 SW Creekside Place Beaverton, OR 97008
Re: K070204
Trade/Device Name: Link to Acuity Clinical Notifier Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac monitor (including cardiotachometer and rate alarm) Regulatory Class: Class II (two) Product Code: MSX Dated: March 12, 2007 Received: March 13, 2007
Dear Mr. Wagner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Wagner
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Blyminima for
Bran D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K070204/S1
Special 510(k) Premarket Notification Link To Acuity® Clinician Notifier Option (Modification to Acuity® Central Monitoring System)
### Statement of Indications for Use
The Acuity® Central Monitoring System, in all configurations including the hardwired and wireless Acuity LT and Mobile Acuity LT System configurations, is intended to be used by clinicians for the central monitoring of neonatal, pediatric and adult patients in healthcare facilities.
In addition to the central monitoring of patient data, waveforms, and alarms and alerts, Acuity System software can include optional modules to provide extended recording and analysis of patient data. Acuity System software can include the following optional modules:
- . The Full Disclosure module stores patient data for up to 96 hours with Acuity LT and Mobile Acuity LT).
- . The Arrhythmia Analysis module provides real-time monitoring and alarms for specific changes in cardiac rhythms. The clinician is responsible for determining the clinical significance of each detected arrhythmia event or alarm. The Arrhythmia Analysis module is not intended for use with neonatal patients.
- . The ST Analysis module provides real-time monitoring and alarms for ST segment deviations for patients with suspected heart disease and anomalies. The clinician is responsible for determining the clinical significance of each selected ST segment deviation or alarm. The ST Analysis option is not intended for use with neonatal patients.
- . The Welch Allyn Connectivity Server (WACS) module consists of a server platform on which one or more of the following software options are installed: Link To Acuity System patient information delivered to mobile devices running Clinician Notifier software), Web Server (Acuity System patient printout files available on Web browsers) and HL7 options.
### Applicant:
Welch Allyn Protocol, Inc. 8500 S.W. Creekside Place Beaverton, OR 97008-7107 USA
510(k) Number: K070204
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (Part 21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
| Office of Device Evaluation (ODE) |
|-----------------------------------|
|-----------------------------------|
| (Division Sign-Off) | |
|------------------------------------|---------|
| Division of Cardiovascular Devices | |
| 510(k) Number | K070204 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.