INSCOPE TISSUE ANCHORS, KNOTTING ELEMENTS AND TISSUE ANCHOR AND KNOTTING ELEMENT APPLIER

K070151 · Ethicon Endo-Surgery, LLC · OCW · Aug 15, 2007 · Gastroenterology, Urology

Device Facts

Record IDK070151
Device NameINSCOPE TISSUE ANCHORS, KNOTTING ELEMENTS AND TISSUE ANCHOR AND KNOTTING ELEMENT APPLIER
ApplicantEthicon Endo-Surgery, LLC
Product CodeOCW · Gastroenterology, Urology
Decision DateAug 15, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Tissue Apposition System (TAS) is indicated for endoscopic placement of suture(s) and approximation of soft tissue.

Device Story

InScope Tissue Apposition System is a sterile, single-use disposable device for endoscopic soft tissue approximation; used with endoscopes having ≥2.8 mm working channel. System components include Tissue Anchor Applier, Tissue Anchor with suture, Knotting Element Applier, and Knotting Element. Operation: Tissue Anchor Applier places metal anchors in GI tract tissue; Knotting Element Applier is advanced over suture ends to tension tissue; Knotting Element is deployed to secure suture and cut trailing ends. Used by physicians in clinical settings to treat GI defects like ulcers and perforations. Benefits include minimally invasive tissue closure.

Clinical Evidence

Bench and animal testing only. Bench testing confirmed holding strength of the Knotting Element is equivalent to the Bard EndoCinch suture anchor and meets USP requirements. Porcine animal model testing demonstrated successful closure of colonic and gastric perforations with normal healing results. Biocompatibility established per ISO 10993-1.

Technological Characteristics

Sterile, single-use disposable system. Components: Tissue Anchor Applier, Knotting Element Applier, anchors, and knotting elements. Mechanical actuation via endoscope working channel (≥2.8 mm). Biocompatibility per ISO 10993-1. No electronic or software components.

Indications for Use

Indicated for endoscopic placement of sutures and approximation of soft tissue in the gastrointestinal tract for patients requiring endoscopic defect closure, such as ulcers and perforations.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K070151 p. 1 of 2 ## 510(k) Summary ## AUG 15 2007 Ethicon Endo-Surgery, LLC Company 475 Calle C Guaynabo, Puerto Rico 00969 Contact Kimberly Shoemaker, RAC Group Manager, Regulatory Affairs Ethicon Endo-Surgery, Inc. 4545 Creek Road Cincinnati, OH 45242 Telephone: (513) 337-8123 Fax: (513) 337-2123 Email: kshoemal@eesus.jnj.com Date Prepared January 15, 2007 Device Name Trade Name: InScope™ Tissue Apposition System Common or Usual Name: Tissue Apposition System Classification Name: Endoscope and Accessories[21CFR 876.1500 (OCW, MXW, GAW)] Bard® EndoCinch™ Suturing System (K003956) Predicate Device LSI Solutions Flexible Suture Placement Device and Acc (K011016) Ethicon Endo-Surgery" Endoscopic Suturing System (K061770) EndoGastric Solutions StomaphyX™ endoluminal fastener and delivery system (K062875) ### Device Description The InScope™ Tissue Apposition System is a sterile, single patient use disposable suture system for approximating and securing soft tissue within the gastrointestinal tract. It is intended to perform suturing in conjunction with endoscopes having a working channel of 2.8 mm or larger. There are four essential devices, and two secondary components: - Tissue Anchor Applier 1. - 2. Tissue Anchor with suture a. Tissue Anchor Loader - Knotting Element Applier 3. - 4. Knotting Element - a. Knotting Element Loader {1}------------------------------------------------ The InScope™ Tissue Anchor Applier is used to place and anchor sutures in the tissue of the GI tract. Suture position is maintained via a metal tissue anchor structure attached to the distal end of each suture strand. Following suture placement, the Knotting Element Applier is advanced over the trailing ends of the anchored sutures and advanced distally. After the sutures are properly tensioned to appose tissue, the Knotting Element is deployed to secure the suture and cut the trailing ends. The Tissue Apposition System can be used to treat a variety of defects endoscopically, including ulcers and perforations. Indications for Use The InScope™ Tissue Apposition System is indicated for endoscopic placement of suture(s) and approximation of soft tissue. Technological Characteristics The InScope™ Tissue Apposition System is similar in design to the Bard® EndoCinch™ and Ethicon Endo-Surgery® Endoscopic Suturing System predicate devices, as both devices provide an endoscopic means of suture termination. The new device is different from these predicates devices as it does not attach externally to an endoscope, but is advanced down the working channel of the endoscope. Performance Data Bench and animal testing was performed to demonstrate the new device performs as intended. Bench testing demonstrated the holding strength of the InScope™ Knotting Element component is substantially equivalent to the Bard® EndoCinch™ suture anchor, and that the anchor attachment force meets the USP requirement. Testing in the porcine animal model demonstrated successful closure of colonic and gastric perforations, with normal results for the healing process. Each material in the InScope™ Tissue Apposition System was assessed for biocompatibility using ISO 10993-1: "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing" and FDA General Program Memorandum #G95-1 : Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices - Part 1 : Evaluation and Testing." and each was found to be biocompatible. Biocompatibility for both limited patient contacting and permanent patient contacting/implant materials has been established through history of use in other marketed Ethicon Endo-Surgery medical devices and biocompatibility test results. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an image of an eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 AUG 15 2007 Ethicon Endo-Surgery, LLC % Ethicon Endo-Surgery, Inc. Ms. Kimberly Shoemaker, RAC Group Manager, Regulatory Affairs 4545 Creek Road Cincinnati, Ohio 45242 Re: K070151 Trade/Device Name: InScope™ Tissue Apposition System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: OCW, MXW, GAW Dated: July 10, 2007 Received: July 11, 2007 Dear Ms. Shoemaker: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 - Ms. Kimberly Shoemaker, RAC This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Your may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ K070151 p. 1 of 1 # Indications for Use 510(k) Number (if known): __K070151 Device Name: InScope™ Tissue Apposition System Indications for Use: The Tissue Apposition System (TAS) is indicated for endoscopic placement of suture(s) and approximation of soft tissue. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1 (Posted November 13, 2003) (Division Sign-Off) Division of General, Restorative, and Neurological Devices 510(k) Number
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...