← Product Code [DPS](/productcode/DPS) · K070136

# AUDICOR 200 SYSTEM, MODELS 200D AND 200S (K070136)

_Inovise Medical, Inc. · DPS · Apr 13, 2007 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K070136

## Device Facts

- **Applicant:** Inovise Medical, Inc.
- **Product Code:** [DPS](/productcode/DPS.md)
- **Decision Date:** Apr 13, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2340
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Audicor 200 System, when used with AUDICOR Sensors on the chest wall, is intended for use in acquiring, analyzing and reporting ECG and heart sound data and to provide interpretation of the data in an integrated report for consideration by physicians. In addition the Audicor 200 System can be used with an add-on laptop computer that allows data to be presented in the format of real-time and trended display of cardiac indices derived from simultaneous ECG data and heart sounds data. The interpretations of ECG and heart sound data offered by the device are only significant when used in conjunction with physician over read as well as consideration of other relevant patient data. The device is intended for use only under the direct supervision of a physician, and is for use on adults (> 18 years).

## Device Story

Device captures 10-second ECG and heart sound data via chest wall sensors; processes inputs to derive cardiac indices and interpretive statements; outputs integrated reports or real-time/trended displays via connected laptop. Used in clinical settings for patients suspected of heart failure or acute coronary syndrome; operated under physician supervision. System calculates parameters including LVST, LVDT, PADT, AAFT, EMAT, QS2, R-R interval, S3/S4 strength, and S1/S2 intensity ratio. Modification adds Systolic Dysfunction Index (SDI) and specific interpretive statements for patients >40 years. Physician reviews output to support clinical decision-making regarding therapeutic treatment or CRT studies.

## Clinical Evidence

No clinical data provided; bench testing only. Device demonstrated compliance with AAMI EC-11, EN 60601-1, EN 60601-1-2, and IEC 60601-2-25 standards.

## Technological Characteristics

System utilizes chest wall sensors for simultaneous ECG and acoustic heart sound acquisition. Processes data to derive temporal and intensity-based cardiac parameters. Connectivity via laptop for trended display. Complies with AAMI EC-11, EN 60601-1, and IEC 60601-2-25. Software performs automated analysis of cardiac intervals and heart sound intensity.

## Regulatory Identification

An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.

## Predicate Devices

- AUDICOR 200 system ([K043074](/device/K043074.md))
- Audicor Advanced Parameters ([K051450](/device/K051450.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K070136

## Summary of Safety and Effectiveness

As required by 21 CFR, part 807.92

.....

:

111

| Submitted By:                 | Inovise Medical, Inc.<br>10565 SW Nimbus Ave, Suite 100<br>Portland, OR 97233-4311<br>Phone 503-431-3837<br>Fax 503-431-3801                                                                                                                                                                                                                                                                                                                                                                                                 |  | APR 13 20 |  |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|-----------|--|
| Contact:                      | Steve Hesler<br>Director, Quality and Regulatory                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |  |           |  |
| Date Prepared:                | January 15, 2007                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |  |           |  |
| Proprietary<br>Name:          | SDI Modification to AUDICOR 200 System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |  |           |  |
| Common/ Usual<br>Name:        | Electrocardiograph/Acoustic Cardiograph                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |  |           |  |
| Classification:               | 870.2340, DPS class II, Electrocardiograph                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |  |           |  |
|                               | 870.1875, DQD, class II, Electronic Stethoscope                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |  |           |  |
| Performance<br>Standards:     | AAMI EC 11                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |  |           |  |
| Intended Use:                 | The Audicor 200 System, when used with AUDICOR Sensors on the chest wall, is<br>intended for use in acquiring, analyzing and reporting ECG and heart sound data<br>and to provide interpretation of the data in an integrated report for consideration by<br>physicians. In addition the Audicor 200 System can be used with an add-on laptop<br>computer that allows data to be presented in the format of real-time and trended<br>display of cardiac indices derived from simultaneous ECG data and heart sounds<br>data. |  |           |  |
|                               | The interpretations of ECG and heart sound data offered by the device are only<br>significant when used in conjunction with physician over read as well as<br>consideration of other relevant patient data.                                                                                                                                                                                                                                                                                                                  |  |           |  |
|                               | The device is intended for use only under the direct supervision of a physician, and<br>is for use on adults (> 18 years).                                                                                                                                                                                                                                                                                                                                                                                                   |  |           |  |
| Device<br>Description:        | The Audicor 200 is a stand-alone device that can be used to capture 10-second<br>evaluations of ECG and heart sounds in patients suspected of heart failure or<br>acute coronary syndrome. In addition the system can be connected to a<br>compatible laptop computer and used to display and analyze patient data over<br>time in a trended format. The trending system can be used in monitoring patient<br>changes during therapeutic treatment or during CRT studies.                                                    |  |           |  |
| Test Summary &<br>Conclusion: | The Audicor 200 System has been tested to the applicable requirements of the<br>following standards, and shown to comply.                                                                                                                                                                                                                                                                                                                                                                                                    |  |           |  |
|                               | • AAMI EC-11 Standard for Diagnostic Electrocardiographic Devices<br>• EN 60601-1 (UL 60601-1) Standard for Medical Electrical Equipment: General<br>Requirements for Safety                                                                                                                                                                                                                                                                                                                                                 |  |           |  |
|                               | • EN 60601-1-2 Standard for Medical Electrical Equipment Part 1-2: General<br>Requirements for Safety - Collateral Standard: Electromagnetic Compatibility –<br>Requirements and Tests                                                                                                                                                                                                                                                                                                                                       |  |           |  |
|                               | • IEC 60601-2-25 Medical Electrical Equipment Part 2-25: Particular requirement<br>for the safety of electrocardiographs                                                                                                                                                                                                                                                                                                                                                                                                     |  |           |  |

APR 1 3 2007

:

{1}------------------------------------------------

- AAMI EC 11 Diagnostic Electrocardiographic Devices .
Substantial Equivalence: The Inovise Audicor 200 System with added interpretive statements is substantially equivalent to the AUDICOR® 200 system (K043074, cleared as Liberty) with the incorporation of Audicor Advanced Parameters (K051450, cleared as Extended Measurements)

Technological Characteristics:

The SDI modification to the Audicor 200 System and the AUDICOR 200 predicate device are technologically equivalent in that both acquire ECG and heart sounds data from adult patients then present the data in the AUDICOR report format which can include graphic display of MI and LVH conditions along with detection and display of S3 and S4 heart sounds. Both systems can analyze and display "Advanced Parameters" derived from ECG and heart sounds information to include:

- . Left Ventricular Systolic Time (LVST)
- Left Ventricular Diastolic Time (LVDT) ●
- Pre-atrial Diastolic Filling Time (PADT)
- Accelerated Atrial Filling Time (AAFT) ●
- QS1 (EMAT) .
- QS2 .
- . R-R Interval
- S3 Strength .
- S4 Strength .
- S1/S2 Intensity Ratio �

The SDI modification to the Audicor 200 System includes addition of a new parameter, the Systolic Dysfunction Index (SDI). The system will also display added interpretive statements when analysis of patient data warrants. The added statements, appropriate only for patients over 40 years of age, are:

- Prolonged EMAT, cconsider LV systolic dysfunction .
- Consider LV systolic dysfunction [reason block] .
- Consider severe LV systolic dysfunction with diastolic dysfunction freason . block]
- � Consider elevated LV filling pressure
- S3 detected -consider acute heart failure in presence of dyspnea

{2}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR 1 3 2007

Inovise Medical, Inc. c/o Mr. Steve C. Hesler Director, Quality and Regulatory 10565 SW Nimbus Avenue, Suite 100 Portland, Oregon 97223-4311

Re: K070136

Trade Name: SDI Modification to AUDICOR® 200 System Regulation Number: 21 CFR 870.2340 and 870.1875 Regulation Name: Electrocardiograph and Stethoscope Regulatory Class: Class II Product Code: DPS and DQD Dated: January 15, 2007 Received: January 16, 2007

Dear Mr. Hessler:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act 

{3}------------------------------------------------

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Bzemmema for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K070136

## Indications for Use

510(k) Number (if known):

SDI Modification to AUDICOR ® 200 System Device Name:

Indications For Use:

The Audicor 200 System, when used with AUDICOR Sensors on the chest wall, is intended for use in acquiring, analyzing and reporting ECG and heart sound data and to provide interpretation of the data in an integrated report for consideration by physicians. In addition, the Audicor 200 System can be used with an add-on laptop computer that allows data to be presented in the format of real-time and trended display of advanced parameters derived from simultaneous ECG data and heart sounds data.

The interpretations of ECG and heart sound data offered by the device are only significant when used in conjunction with physician over read as well as consideration of other relevant patient data.

The device is intended for use only under the direct supervision of a physician, and is for use on adults (≥ 18 years).

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

B. Summa

ardiovi

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/device/K070136](https://fda.innolitics.com/device/K070136)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
