QUICKSCREEN BARBITURATE 300 SCREENING TEST, MODELS 9019, 9018; MULTI DRUG SCREENING TEST, MODEL 9317T; CUP MULTI DRUG SC
K070098 · Phamatech, Inc. · DIS · Dec 7, 2007 · Clinical Toxicology
Device Facts
Record ID
K070098
Device Name
QUICKSCREEN BARBITURATE 300 SCREENING TEST, MODELS 9019, 9018; MULTI DRUG SCREENING TEST, MODEL 9317T; CUP MULTI DRUG SC
Applicant
Phamatech, Inc.
Product Code
DIS · Clinical Toxicology
Decision Date
Dec 7, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3150
Device Class
Class 2
Indications for Use
The QuickScreen Barbiturates Test is a qualitative in-vitro diagnostic screen that provides a preliminary result for the detection/presence of Barbiturates in urine. Tests for barbiturates cannot distinguish between abused drugs and certain prescribed medications. The cut-off concentration will be 300 ng/ml (secobarbital). It is intended for professional use only. This test provides only a preliminary test result. A more specific alternate testing method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Other confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are observed.
Device Story
QuickScreen devices are rapid, qualitative, one-step lateral flow immunoassays for detecting drug metabolites in urine. Input is a urine sample; device utilizes visual color sandwich immunochemical technology where specific antibodies bind to target analytes. Output is a visual qualitative result (presence/absence of drug) read by a professional user. Used in clinical or professional settings to provide preliminary screening results. Results require confirmation via GC/MS or other analytical methods. Assists healthcare providers in identifying potential drug abuse; clinical judgment required for interpretation.
Clinical Evidence
Clinical evidence includes clinical sample correlation studies and blind labeled spiked studies. Results demonstrated >97.9% correlation for barbiturates and >98% correlation for multi-drug panels when compared to predicate devices and GC/MS methodology. Overall accuracy reported as >97% in professional use settings. No clinical data beyond correlation/spiked studies provided.
Technological Characteristics
Lateral flow, one-step visual color sandwich immunoassay. Detects target analytes via antibody-antigen-antibody complexes. Form factors include test strips and drug cups. Standalone, non-electronic, visual readout. No external energy source required.
Indications for Use
Indicated for professional use in the qualitative detection of amphetamine, cocaine, methamphetamine, opiates, PCP, barbiturates, benzodiazepines, methadone, and THC in human urine to assist in the diagnosis and treatment of drug abuse. Cannot distinguish between abused drugs and certain prescribed medications.
Regulatory Classification
Identification
A barbiturate test system is a device intended to measure barbiturates, a class of hypnotic and sedative drugs, in serum, urine, and gastric contents. Measurements obtained by this device are used in the diagnosis and treatment of barbiturate use or overdose and in monitoring levels of barbiturate to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). A barbiturate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Predicate Devices
Phamatech QuickScreen Model 9026 and 9025 (k82152)
Phamatech QuickScreen Pro Drug Cup Model 9195X and 9153 T (k001397, k043167)
Submission Summary (Full Text)
{0}
1
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k070098
B. Purpose for Submission:
Change in the barbiturate cutoff concentration from 200 ng/mL to 300 ng/mL
C. Measurand:
Barbiturates
D. Type of Test:
Qualitative
E. Applicant:
Phamatech, Inc.
F. Proprietary and Established Names:
Phamatech QuickScreen™ Barbiturates Test Models 9019, 9018
Phamatech QuickScreen™ Pro Multi Drug Screening Test model 9317T and 9187Z
G. Regulatory Information:
1. Regulation section:
21 CFR 862.3150 Barbiturate test system
2. Classification:
II
3. Product code:
DIS
4. Panel:
{1}
91 (Toxicology)
## H. Intended Use:
1. Intended use(s):
See indications for use statement below.
2. Indication(s) for use:
**Phamatech QuickScreen™ Barbiturates Test Models 9019, 9018**
The QuickScreen Barbiturates Test is a qualitative in-vitro diagnostic screen that provides a preliminary result for the detection/presence of barbiturates in urine. Tests for barbiturates cannot distinguish between abused drugs and certain prescribed medications. The cut-off concentration will be 300 ng/ml (secobarbital). It is intended for professional use only.
**Phamatech QuickScreen™ Pro Multi Drug Screening Test Model 9317T**
QuickScreen™ Pro Multi Drug Screening Test is an in vitro diagnostic test for the qualitative detection of amphetamine, cocaine, methamphetamine, opiates, PCP, barbiturates, benzodiazepines, methadone and THC in urine. Tests for prescription drugs cannot distinguish between abused drugs and certain prescribed medications. Measurements obtained by this device are used in the diagnosis and treatment of drug abuse. It is intended for professional use only.
| Amphetamine | 1000 ng/ml |
| --- | --- |
| Cocaine | 300 ng/ml |
| Methamphetamine | 1000 ng/ml |
| Opiates | 2000 ng/ml |
| PCP | 25 ng/ml |
| Barbiturates (Secobarbital) | 300 ng/ml |
| Benzodiazepines | 200 ng/ml |
| Methadone | 300 ng/ml |
| THC | 50 ng/ml |
**Phamatech QuickScreen™ Pro Drug Cup Model 9187Z**
QuickScreen™ Pro Drug Cup is an in vitro diagnostic test for the qualitative detection of amphetamine, cocaine, methamphetamine, opiates, PCP, Barbiturates, benzodiazepines, methadone and THC in urine. Tests for prescription drugs cannot distinguish between abused drugs and certain prescribed medications. Measurements obtained by this device are used in the diagnosis and treatment of drug abuse. It is intended for professional use only.
{2}
3
Amphetamine 1000 ng/ml
Cocaine 300 ng/ml
Methamphetamine 1000 ng/ml
Opiates 2000 ng/ml
PCP 25 ng/ml
Barbiturates (Secobarbital) 300 ng/ml
Benzodiazepines 200 ng/ml
Methadone 300 ng/ml
THC 50 ng/ml
3. Special conditions for use statement(s):
This test provides only a preliminary test result. A more specific alternate testing method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Other confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are observed.
4. Special instrument requirements:
Not applicable.
I. Device Description:
The single barbiturate test and multi drug test device employs lateral flow immunochromatographic technology and is based on the principle of competitive binding. The device is available in the cassette, dipstick and cup formats. Drugs, if present in concentrations below the cutoff level, will not saturate the binding sites of the antibody coated particles on the drug specific test strips. The goat-anti-rabbit IgG antibody-coated particles will then be captured by immobilized drug-specific conjugate. If the level of drug in the urine specimen is below the cutoff concentration, the T line appears as a visible burgundy line. If the level of drug in the urine specimen is above the cutoff, no T line develops. The control line (C line) serves as an internal quality control of the system. It should always appear as a burgundy-colored band regardless of the presence of the drug.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Phamatech QuickScreen Model 9026 and 9025 (Cassette and Dipstick)
Phamatech QuickScreen Pro Drug Cup Model 9195X and 9153 T (Integrated Cup and Dip Card).
2. Predicate K number(s):
k82152, k001397 and k043167
{3}
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Cup and Card | | |
| Specimen | Urine | Urine |
| Methodology | Lateral flow immunoassay | Lateral flow immunoassay |
| Type | Qualitative | Qualitative |
| Analytes | Secobarbital | Secobarbital |
| User | Professional use | Professional Use |
| Cup and Dip card | | |
| Methodology | Lateral flow immunoassay | Lateral flow immunoassay |
| Formats | Cup and Dip card | Cup and Dip card |
| Specimen | Urine | Urine |
| Type | Qualitative | Qualitative |
| Analytes | cocaine (benzoylecgonine)
THC,
opiates,
amphetamine,
methamphetamine,
benzodiazepines,
barbiturates,
methadone
PCP | cocaine (benzoylecgonine)
THC,
opiates,
amphetamine,
methamphetamine,
benzodiazepines,
barbiturates,
methadone
PCP |
| | | |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Cutoff | Secobarbital 300 mg/mL | Secobarbital 200 mg/mL |
K. Standard/Guidance Document Referenced (if applicable):
None were referenced.
L. Test Principle:
The QuickScreen™ Barbiturate tests are chromatographic absorbent devices in which drug metabolites in a sample compete with drug conjugate, immobilized on a porous support membrane, for a limited number of goat anti-rabbit IgG antibody binding sites. When a urine sample is applied to a test device, the barbiturates, if present in the test sample at a concentration of less than 300 ng/mL, will not block the binding sites of the immobilized antigen conjugate in the test region. The dye conjugate will migrate to the test region with the urine. There the dye conjugate will bind to barbiturate derivative immobilized in the test region of the membrane. As a result, a visible rose pink band will form in the test region, indicating a negative
{4}
result. Conversely, where barbiturate concentrations are 300 ng/mL or more the antibody - dye conjugate binds to free drug forming a complex that competes with the immobilized antigen conjugate in the test region. This prevents the development of a rose pink band, indicating a positive sample.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
Eight normal urine pools were spiked with secobarbital to concentrations of 0, 122, 162, 274, 300, 400, 490 and 800 ng/mL and confirmed by GC/MS. One replicate per level was assayed five times per day for 10 days with the cassette format. The results are shown in the table below.
| Phenobarbital Conc. (ng/mL) | % Cutoff Concentration | % Correct |
| --- | --- | --- |
| 0 | <25% of cutoff | 100 |
| 122 | 26% to 50% of cutoff | 100 |
| 162 | 51% to 75% of cutoff | 100 |
| 274 | 76% to cutoff | 100 |
| 300 | Cutoff to 125% of cutoff | 100 |
| 400 | 126% to 150% of cutoff | 100 |
| 490 | 151% to 175% of cutoff | 100 |
| 800 | >176% of cutoff | 100 |
The sponsor conducted a bridging study to assess the equivalency of the 4 format types (dipstick, cassette and dip-card and cup). Fifty GC/MS confirmed urine samples that ranged from 50 to 2000 ng/mL were obtained from clinical sites and were tested in duplicates. The sponsor's results showed 100% agreement between formats.
#### b. Linearity/assay reportable range:
Not applicable. This assay is intended for qualitative use.
#### c. Traceability, Stability, Expected values (controls, calibrators, or methods):
This device has an internal process control. A line appearing in the control region confirms that sufficient sample volume has been applied and
{5}
that the sample has migrated correctly on the test strip. Users are informed not to interpret the test if a line does not form in the control region. External controls are not supplied with this device but the manufacturer recommends the use of external controls in the labeling.
d. Detection limit:
See assay cutoff section below.
e. Analytical specificity:
To test the consistency of the test line and control line development, the sponsor tested four urine control pools and five patient samples. The pools and samples were analyzed in replicates of three using three lots of the QuickScreen Barbiturate 300 dipstick test. The results indicate that the development times were under the ten minutes given in the package insert.
The sponsor conducted a prozone study to test the effects of high barbiturate concentrations on the device. A GC/MS confirmed positive urine sample (>3000 ng/mL) was tested with three lots of the QuickScreen Barbiturate 300 test. The sponsor reported no interference with high barbiturate concentrations.
The sponsor conducted a prozone effect (hook-effect) study to evaluate the devices ability to detect high barbiturate concentrations. A GC/MS confirmed positive for barbiturate (concentration greater than 3000 mg/mL) was tested in three lots of the QuickScreen™ Test. All results for the barbiturates were positive and sponsor reports that there was no prozone effect observed.
The sponsor conducted a pH study to test for pH effects on the device. Three urine samples were aliquoted at barbiturate concentrations of 0, 835, and 3000 ng/mL. Each sample was then again aliquoted into five individual samples and each was adjusted to pH 4.5 to 8.5 (increments of 1 pH). The sponsor reports no adverse effect with varying pH levels.
The sponsor conducted a test strip flexibility study to determine the dip time for the strip test. Five urine samples were selected at random for this study. The samples had barbiturate concentrations of 0, 0, 449, 531 and 1146 ng/mL. The sample dip times were varied from 10 seconds to 600 seconds (10 minutes). The samples were run in three assays in triplicates. The samples produced 100% of the expected results.
The sponsor studied structurally related compounds to test for cross-reactivity in the QuickScreen Barbiturates 300 test. The compounds were prepared in normal human urine and the results (below) were expressed as the amount of compound capable of giving a result equivalent to 300 ng/mL secobarbital.
6
{6}
7
| Alphenal 0.400 μg/mL | Butabarbital >0.800 μg/mL | Pentobarbital 0.250 μg/mL | Amobarbital 0.500 μg/mL |
| --- | --- | --- | --- |
| Butabital 0.125 μg/mL | Phenobarbital 0.050 μg/mL | Secobarbital 0.300 μg/mL | |
A wide variety of compounds (n=167 see package insert for a complete list) were tested to 100,000 ng/mL and were found not to interfere or cross react. Additionally, the following compounds were tested to the stated concentration and were found not to cross react.
| Alprazolam -25 μg/mL | Fentanyl- 10 μg/mL | LSD 2.5 μg/mL |
| --- | --- | --- |
| Barbituric Acid 1000 μg/mL | Heroin – 10 μg/mL | Ethylmorphine- 10 μg/mL |
| 11-Hydroxy Δ⁹ HC- 5 μg/mL | | |
f. Assay cut-off:
The sponsor conducted a cutoff/sensitivity study on 80 assay urine samples. The results are presented below and support the sponsors claimed cutoff of 300 ng/mL.
| % of the cutoff concentration | Result | N |
| --- | --- | --- |
| 0-20% | 100% negative | 10/10 |
| 26-50% | 100% negative | 10/10 |
| 51-75% | 100% negative | 10/10 |
| 76-100% | 80% negative | 8/10 |
| 101-125% | 100% positive | 10/10 |
| 126-150% | 100% positive | 10/10 |
| 151-175% | 100% positive | 10/10 |
| 176-200% | 100% positive | 10/10 |
2. Comparison studies:
a. Method comparison with predicate device:
The sponsor tested 152 GC/MS confirmed urine samples obtained from two clinical sites with two lots of BAR 300 (cassette and strip). QuickScreen test correctly detected >99% of the positive samples and >96% of the negative samples. The study included members of the barbiturate family that included amobarbital, pentobarbital, secobarbital, butalbital and alphenal, Test results
{7}
below were compared to their know GC/MS results. Samples were considered positive if they were at or higher than the cutoff of 300 ng/mL.
| | | | | Specimen cutoff range by GC/MS results | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Quick Screen GC/MS | | 0-150 ng/mL | 151-225 ng/mL | 226-300 ng/mL | 300-375 ng/mL | 375-450 ng/mL | >450 ng/mL | % Agreement |
| | - | 19 | 42 | 36 | 0 | 0 | 0 | 96% |
| | + | 0 | 0 | 2 | 31 | 6 | 16 | 99% |
b. Matrix comparison:
Not applicable: this device is only for use with urine samples.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Not applicable
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.