K070077 · B-K Medical Aps · IYO · Mar 5, 2007 · Radiology
Device Facts
Record ID
K070077
Device Name
ULTRASOUND SCANNER PROFOCUS, MODEL 2202
Applicant
B-K Medical Aps
Product Code
IYO · Radiology
Decision Date
Mar 5, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1560
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Ultrasound scanner and transducers for B, Tissue and Contrast Harmonic Imaging, M, PWD,CWD, Color Doppler and combined mode imaging. Guidance of biopsy needles, geometrical measurements and calculation of parameters. Non monitoring ECG for superimposing the ultrasound information. An optional 3-D unit can reconstruct a series of 2-D images into a single 3-D volume and display this on the screen.
Device Story
ProFocus 2202 is a diagnostic ultrasound system utilizing pulsed echo and Doppler imaging. It processes acoustic signals from linear, convex, and phased array transducers to generate B-mode, M-mode, PWD, CWD, and Color Flow Mapping images. Features include Tissue/Contrast Harmonic Imaging, needle guidance, and geometric measurements. An optional 3-D unit reconstructs 2-D images into 3-D volumes. An optional ECG signal can be superimposed on ultrasound displays. Operated by clinicians in clinical settings, the system provides real-time visualization and quantitative data to assist in diagnosis and interventional guidance. Patient benefits include non-invasive diagnostic imaging and precise needle placement during biopsies.
Clinical Evidence
Bench testing only. System safety verified per IEC 60601-1. Acoustic output measured and calculated per AIUM 1998 standards. No clinical data presented.
Technological Characteristics
Diagnostic ultrasound system; linear, convex, phased array, and mechanical sector transducers. Patient contact materials comply with ISO 10993-1. Modes: B, M, PWD, CWD, CFM, Tissue/Contrast Harmonic Imaging. Track 3 acoustic output (Ispta ≤ 720 mW/cm², MI ≤ 1.9, TI ≤ 6.0). Safety tested to IEC 60601-1. Connectivity includes optional ECG input.
Indications for Use
Indicated for diagnostic ultrasound imaging or fluid flow analysis in patients requiring abdominal, cardiac, fetal, intraoperative, neurosurgical, obstetric, pediatric, transcranial (adult cephalic), transrectal, small organ, transvaginal, or musculoskeletal examinations.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{1}------------------------------------------------
### Attachment 1
# K070071
### 510(k) Summary:
This summary is provided as part of this Premarket Notification in compliance with 21CRF, Section 807.92.
Submitters name: B-K Medical Address: Mileparken 34, DK2730 Herlev, Denmark Phone: +45 44528100 Fax: +45 44528199 Contact person: Villy Braender, Regulatory Manager Date prepared: 5 January 2007
Trade name: Ultrasound Scanner ProFocus 2202 Common name: Diagnostic Ultrasound System Classification names: Ultrasonic Pulsed Echo Imaging System (90 IYO, CFR 892.1560) Ultrasonic Pulsed Doppler Imaging System (90 IYN, CFR 892.1560) Diagnostic Ultrasonic Transducer (90 ITX, CFR 892.1570)
Identification of predicate, legally marketed device: Siemens Medical Solutions USA Inc: Acuson CV70 Cardiovascular system (K032111)
### Device description:
2202 supports the following scanning modes and combinations thereof:
B-mode, M-mode, CWD-mode, PWD mode and CFM mode. Tissue harmonic imaging, Contrast harmonic imaging.
An optional ECG signal can be superimposed the ultrasound information in all modes and mode combinations.
An optional 3-D unit can reconstruct a series of 2-D images into a single 3-D volume and display this on the screen.
The system can perform simple geometric measurements, and perform calculations in the areas of Vascular, Urology, Cardiology and OB/GYN applications,
The system can guide biopsy- and puncture needles.
### Transducers
Transducers are linear and convex phased arrays and mechanical sector. The patient contact materials comply with ISO10993-1 All transducers used together with 2202 are Track 3 transducers.
#### Acoustic output
The system will assure that the acoustic output always will stay below the pre-amendments upper limits i.e. Ispta ≤ 720 mW/cm² and MI ≤ 1.9 (Track 3, non ophthalmic). The Thermal Index values are maximum 6.0, i.e. TI ≤ 6.0
### Clinical measurement accuracy.
Clinical measurements and calculations are described and accuracies are provided with the User Guide.
### Thermal, mechanical and electrical safety.
The scanner 2202 has been tested by a recognized, certified body according to IEC 60601-1.
I
MAR 0 5 2007
{2}------------------------------------------------
## Attachment 1
### Acoustic Output Reporting
The Acoustic Output Reporting is made according to the standards required by "Information for Manufacturers Seeking Clearance of Diagnostic Ultrasound Systems and Transducers, FDA, CDRH, September 30, 1997"
The acoustic output is measured and calculated according to: "Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment" (AIUM 1998).
### Intended use.
See comparison below
### Technological characteristics compared to the predicate device.
The predicate device has the same major technological characteristics as the subject device, see comparison below.
Comparison with Siemens Medical Solutions USA Inc: Acuson CV70 Cardiovascular system (K032111)
| | ProFocus 2202 in this application | K032111 |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended uses | Abdominal, Cardiac, Fetal,<br>Intraoperative, Neurosurgery,<br>Obstetrics, Pediatrics,<br>Transrectal, Small organs,<br>Transvaginal, Musculoskeletal<br>(superficial, conventional),<br>Peripheral Vascular.<br>Transcranial (Adult cephalic) | Abdominal. Intraoperative,<br>Small Parts, Transcranial,<br>OB/GYN, Cardiac,<br>Transesophageal. Pelvic,<br>Neonatal/Adult Cephalic,<br>Vascular, Musculoskeletal,<br>Superficial Musculoskeletal,<br>and Peripheral Vascular<br>applications. |
| General device description | Measurements and calculations<br>Track 3 (Index display).<br>Scanning modes and mode<br>combinations: B, B+M,<br>B+CWD, B+D, B+C, B+D+C.<br>(CWD is continuous wave<br>Doppler, D is PWD, C is Color<br>Flow mapping Doppler including<br>Velocity+Variance display,<br>Amplitude (power) Doppler.<br>B mode includes Tissue<br>Harmonic Imaging (THI) and<br>Contrast Harmonic Imaging<br>(CHI).<br>3D imaging | Measurements and analysis<br>Track 3 (Index display)<br>B-Mode, M-Mode, Pulsed<br>(PW) Doppler Mode,<br>Continuous (CW) Doppler<br>Mode, Color Doppler Mode,<br>Amplitude Doppler Mode, a<br>combination of modes, or<br>Harmonic Imaging, or<br>3D imaging |
| Acoustic output | Ispta ≤ 720 mW/cm² and MI ≤<br>1.9 (Track 3, non ophthalmic).<br>TI ≤ 6.0 | Not in 510(k)summary, except<br>that it has index display<br>according to Display standard. |
| General safety and<br>effectiveness | AIUM/NEMA UD-3<br>AIUM/NEMA UD-2 | UL 2601-1<br>CSA C22.2 No. 601-1<br>AIUM/NEMA UD-3<br>AIUM/NEMA UD-2 |
| | 93/42/EEC Medical Devices<br>Directive,<br>EN/IEC 60601-1<br>EN/IEC 60601-1-1<br>EN/IEC 60601-1-2<br>EN/IEC 60601-2-37<br>EN/ISO 10993-1 | 93/42/EEC Medical Devices<br>Directive<br>EN/IEC 60601-1<br>EN/IEC 60601-1-1<br>EN/IEC 60601-1-2<br>IEC 1157<br>ISO 10993 |
| Labeling | Please refer to section 4.8 | Not in 510(k) summary) |
{3}------------------------------------------------
### Attachment 1
Conclusion: The device ProFocus 2202 in this application has similar intended uses, and in particular the subject for the application Transcranial (Adult cephalic), is the same.
B-K Medical ApS therefore believes, that 2202 is substantially equivalent to K032111.
: 100 : 100 : 100 : 100 : 100
{4}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Villy Brænder Official Correspondent B-K Medical Mileparken 34, DK-2730 Herlev DENMARK
MAR C 5 2007
Re: K070077
Trade Name: Ultrasound Scanner Profocus 2202 Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulation Number: 21 CFR 892.1560 Regulations Name: Ultrasound pulsed echo imaging system Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasound transducer Regulatory Class: II Product Code: IYO, IYN, and ITX Dated: January 5, 2007 Received: January 10, 2007
Dear Mr. Brænder:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the Ultrasound Scanner Profocus 2202, as described in your premarket notification:
Transducer Model Number
8827
{5}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:
> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
{6}------------------------------------------------
If you have any questions regarding the content of this letter, please contact Andrew Kang, M.D. at (240) 276-3666.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known):
K070077
Device Name: Ultrasound Scanner Profocus 2202
Indications For Use:
Ultrasound scanner and transducers for B, Tissue and Contrast Harmonic Imaging, M, PWD,CWD, Color Doppler and combined mode imaging. Guidance of biopsy needles, geometrical measurements and calculation of parameters. Non monitoring ECG for superimposing the ultrasound information. An optional 3-D unit can reconstruct a series of 2-D images into a single 3-D volume
and display this on the screen.
Clinical applications: Abdominal, Cardiac, Fetal, Intraoperative, Neurosurgery, Obstetrics, Pediatrics, Transcranial (=Adult cephalic), Transrectal, Small organs, Transvaginal, Musculoskeletal.
Details on specific Indication for Use forms
Prescription Use / AND/OR Over-The-Counter Use **__**
(Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of
Nancy C. Brogdon
(Division Sign Off) Division of Reproductive. Abdominal, and Radiological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________
{8}------------------------------------------------
#### Diagnostic Ultrasound Indications for Use Form
System: 2202
K070077
# Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|---------------------------------------------------|----------------------|----------------------|------------------------------|-------------------------|--------------------|
| Clinical Application | A | B | M | PWD | Tissue-<br>and<br>contrast<br>harmonic<br>imaging | Color<br>Doppl<br>er | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify 1) | Continous<br>Wave) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | P | P | P | P | P | P | | P | |
| Abdominal | | P | P | P | P | P | P | | P | |
| Intraoperative (specify) | | P | P | P | P | P | P | | P | |
| Intraoperative Neurological | | P | P | P | P | P | P | | P | |
| Pediatric | | P | P | P | P | P | P | | P | |
| Small Organ (specify) | | P | P | P | P | P | P | | P | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | N | N | N | N | N | N | | N | N |
| Cardiac | | P | P | P | P | P | P | | P | N |
| Transesophageal | | | | | | | | | | |
| Transrectal | | P | P | P | P | P | P | | P | |
| Transvaginal | | P | P | P | P | P | P | | P | |
| Transurethral | | P | P | P | P | P | P | | P | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | P | P | P | P | P | P | | P | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | P | P | P | P | P | P | | P | |
| Musculo-skeletal Superficial | | P | P | P | P | P | P | | P | |
| Other (specify) | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
1) B+M, B+D, B+C, B+D+C. B mode includes Tissue-and Contrast Harmonic Imaging. Additional Comments:_ D is PWD, C is Color Doppler. Fetal is often called Obstetrics
> (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Nancy C. Swedon
li ingeinn Ovision of Reproductive, Abdominal, and Radiological Devi 570(k) Number _
{9}------------------------------------------------
#### Diagnostic Ultrasound Indications for Use Form
System: 2202
Transducer 8827
K070077
Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|---------------------------------------------------|----------------------|----------------------|------------------------------|-------------------------|-------------------------|
| Clinical Application | A | B | M | PWD | Tissue-<br>and<br>contrast<br>harmonic<br>imaging | Color<br>Doppl<br>er | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify 1) | Continous<br>Wave<br>CW |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | N | N | N | N | N | N | | N | N |
| Cardiac | | E | E | E | E | E | E | | E | N |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
1) B+M, B+D, B+CW, B+C, B+C, B mode includes Tissue-and Contrast Harmonic Imaging. Additional Comments: D is PWD, C is Color Doppler. Fetal is often called Obstetrics
> (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Exaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Invision Sidn-Off) of Reproductive, Abdomin olomcal Devices
anda
KD702
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.