QWIKCHECK TEST STRIPS- REAGENT STRIPS FOR SEMEN ANALYSIS

K063864 · Medical Electronic Systems , Ltd. · POV · Mar 26, 2007 · Hematology

Device Facts

Record IDK063864
Device NameQWIKCHECK TEST STRIPS- REAGENT STRIPS FOR SEMEN ANALYSIS
ApplicantMedical Electronic Systems , Ltd.
Product CodePOV · Hematology
Decision DateMar 26, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.5220
Device ClassClass 2

Indications for Use

QwikCheck™ Test Strips are for in vitro diagnostic use for the determination of pH and leukocytes (WBCs) in semen. Test results are determined by comparing the color of the test patches to the color chart provided on the bottle label. The test is for professional use only.

Device Story

QwikCheck™ Test Strips consist of reagent-impregnated test patches for semen analysis. Device inputs are semen samples applied to test strips; principle of operation is colorimetric chemical reaction. User (professional clinician) compares resulting patch color to provided bottle label color chart to determine pH and leukocyte levels. Output is semi-quantitative visual assessment of semen pH and WBC concentration. Used in clinical settings to assist in diagnostic evaluation of semen quality.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Two-parameter dipstick. Reagent pads contain methyl red and bromothymol blue (pH) and 3-indoly-phenol ester and benzendiazonium salt (leukocytes). Visual colorimetric readout. Shelf life 18 months; storage 15-30°C.

Indications for Use

Indicated for professional in vitro diagnostic determination of pH and leukocytes (WBCs) in human semen samples.

Regulatory Classification

Identification

An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”

Submission Summary (Full Text)

{1} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: K063864 B. Purpose for Submission: Traditional 510(k) for a new indication for use of a previously cleared device. C. Measurand: Leukocytes (WBCs) and pH in semen D. Type of Test: Two-parameter test strip for determination of pH and non-quantitative measurement of leukocytes in semen. E. Applicant: Medical Electronic Systems, Ltd. F. Proprietary and Established Names: QwickCheck™ Test Strips-Reagent Strips for Semen Analysis G. Regulatory Information: 1. Regulation section: There are no regulations for semen analysis devices. 2. Classification: Class II 3. Product code: MNA {2} 4. Panel: 85 (Ob/Gyn) H. Intended Use: 1. Intended use(s): QwickCheck Test Strips are for in vitro diagnostic use for the determination of pH and leukocytes (WBCs) in semen. Test results are determined by comparing the color of the test patches to the color chart provided on the bottle label. The test is for professional use only. 2. Indication(s) for use: N/A 3. Special conditions for use statement(s): N/A 4. Special instrument requirements: N/A I. Device Description: The QwickCheck Test Strip is a two parameter test strip (dipstick) for the determination of pH and leukocytes (WBCs) in semen. Test results are determined by comparing the color of the test patches to the color chart provided on the bottle label. J. Substantial Equivalence Information: 1. Predicate device name(s): International Newtech Development Inc. Urinalysis Reagent Strips 2. Predicate 510(k) number(s): K993850 {3} 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Material Composition on Reagent Pads | pH: methyl red and bromothymol blue | Same | | | Leukocytes: 3-indoly-phenol ester and benzendiazonium salt | | | Procedural Steps | • Fresh undiluted sample is used | Same | | | • Sample applied to test pads by dropper. | | | | • Read result at appropriate time intervals (1 min for pH, 2 min for WBCs; color of test patch is compared to color scale on bottle. | | | Shelf Life | 18 months | Same | | Storage Conditions | Room temperature (15-30°C). Keep away from direct sunlight and moisture. | Same | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Matrix | Semen | Urine | K. Standard/Guidance Document Referenced (if applicable): N/A {4} L. Test Principle: For pH: Acid/base color reaction between two indicators (methyl red and bromothymol blue. For Leukocytes: The color of the leukocyte reagent pad reacts to the presence of esterases found in granulocytes. M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: A field study was designed to test the ability of typical lab and clinic users to accurately interpret the QwickCheck test strips for pH and WBCs based on the color reaction. The study was performed at three sites, one in-house at MES and two external sites with a minimum of three operators per site. The data indicated that correct interpretations were achieved 100% of the time. The study included 132 pH results and 132 WBC results (12 samples x 11 testers). For pH, the exact level was recorded in 126/132 cases (95%), and there were six cases where the pH was interpreted as 0.5 units (one color block) higher than the assigned value: the protocol allowed a variance of plus/minus one color block. For WBCs, the correct interpretation was achieved in 132/132 cases (100%). b. Linearity/assay reportable range: N/A c. Traceability, Stability, Expected values (controls, calibrators, or methods): Shelf life is 18 months. d. Detection limit: N/A e. Analytical specificity: f. Assay cut-off: N/A {5} 2. Comparison studies: a. Method comparison with predicate device: N/A b. Matrix comparison: N/A 3. Clinical studies: a. Clinical Sensitivity: N/A b. Clinical specificity: N/A c. Other clinical supportive data (when a. and b. are not applicable): N/A 4. Clinical cut-off: N/A 5. Expected values/Reference range: N/A N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 5
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...