LIBERTY NORMAL GLUCOSE CONTROL SOLUTION, MODEL 2120045/ HIGH GLUCOSE CONTROL SOLUTION, MODEL 2120047
K063855 · Liberty Healthcare Group, Inc. · JJX · Feb 1, 2007 · Clinical Chemistry
Device Facts
Record ID
K063855
Device Name
LIBERTY NORMAL GLUCOSE CONTROL SOLUTION, MODEL 2120045/ HIGH GLUCOSE CONTROL SOLUTION, MODEL 2120047
Applicant
Liberty Healthcare Group, Inc.
Product Code
JJX · Clinical Chemistry
Decision Date
Feb 1, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1660
Device Class
Class 1
Attributes
Pediatric
Indications for Use
The Liberty Control Solutions are intended for in vitro diagnostic use by healthcare professionals and in the home by people with diabetes mellitus to assess the performance of the Liberty Blood Glucose Monitoring System.
Device Story
Liberty Control Solutions are viscosity-adjusted, aqueous liquid controls containing known quantities of glucose; packaged in plastic dropper-tipped bottles. Used by healthcare professionals and patients with diabetes mellitus in clinical or home settings to verify the accuracy and performance of the Liberty Blood Glucose Monitoring System. The solution contains a red dye to assist users in visually confirming application to test strips. The device is non-hazardous and contains no human or animal-derived materials. Users apply the control solution to a test strip; the glucose monitoring system measures the glucose level; the user compares the result against expected ranges to ensure the system is functioning correctly.
Clinical Evidence
No clinical data. Bench testing only. Stability determined via real-time studies (24-month shelf-life, 90-day open-vial stability). Traceability established via in-house glucose preparation; acceptable ranges assigned via repeat analysis using three lots of test strips.
Technological Characteristics
Buffered aqueous solution containing D-Glucose, viscosity modifier, preservatives, and red dye. Packaged in plastic dropper-tipped bottle. Non-sterile. Evaluated per CLSI EP5-A for precision performance. No human or animal-derived materials.
Indications for Use
Indicated for in vitro diagnostic use by healthcare professionals and individuals with diabetes mellitus to verify the performance of the Liberty Blood Glucose Monitoring System.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
Predicate Devices
AgaMatrix Liberty Control Solutions (Normal and High Levels) (K052762)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k063855
B. Purpose for Submission:
New Device
C. Measurand:
Quality control materials for glucose
D. Type of Test:
N/A
E. Applicant:
Liberty Healthcare Group, Inc.
F. Proprietary and Established Names:
Liberty Control Solutions (Normal Level and High Level)
G. Regulatory Information:
1. Regulation section:
21 CFR § 862.1660 Quality control material (assayed and unassayed)
2. Classification:
Class I
3. Product code:
JJX, single (specified) analyte controls (assayed and unassayed)
4. Panel:
Clinical Chemistry (75)
H. Intended Use:
1. Intended use(s):
For in vitro diagnostic use (i.e. for external use only) by healthcare professional and in the home by people with diabetes mellitus to assess the performance of the Liberty Blood Glucose Monitoring System.
2. Indication(s) for use:
See intended use section above
3. Special conditions for use statement(s):
Over-The-Counter Use
3. Special instrument requirements:
Liberty Blood Glucose Monitoring System
{1}
2
I. Device Description:
The Liberty Control Solutions are liquid glucose control solutions for use with the Liberty Blood Glucose Monitors. Each control consists of a buffered aqueous solution of D-Glucose, a viscosity modifier, preservatives, red dye, and other non-reactive ingredients. The device is non-sterile, over-the counter product and is intended for external use only. It is non-hazardous and contains no human or animal derived materials.
The product is packaged in a plastic dropper tipped bottle for easy application of the control solution to the test strip and contains sufficient volume (3.6 mL) to run 75 tests. A red coloration is included to aid the user to visually confirm application of the control.
J. Substantial Equivalence Information:
1. Predicate device name(s):
AgaMatrix Liberty Control Solution
2. Predicate 510(k) number(s):
k052762
3. Comparison with predicate:
Both devices contain D-Glucose and is non-hazardous and contains no human or animal derived materials.
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Indications for Use | Used to check the performance of Liberty Blood Glucose Monitoring System. | Used with the Liberty Meter and Test Strips as a quality control check to verify the accuracy of blood glucose test results. |
| Target Population | Professional and home use | Professional and home use |
| Matrix | Buffered Aqueous Solution | Buffered Aqueous Solution |
| Container | Plastic bottle with dropper-tip | Plastic bottle with dropper-tip |
| Analytes | Glucose | Glucose |
| Number of Levels | 2 | 2 |
K. Standard/Guidance Document Referenced (if applicable):
CLSI EP5-A, Evaluation of the Precision Performance of Clinical Chemistry Devices.
L. Test Principle:
Not Applicable
{2}
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Not Applicable
b. Linearity/assay reportable range:
Not Applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The D-Glucose used in this control is traceable to an in-house glucose preparation. Values are assigned by repeat analysis using three different lots of test strips. The mean and standard deviation are used to establish the acceptable range for each control level.
Stability characteristics of the Liberty Glucose Control MID were determined using real-time studies. The unopened shelf-life is 24 months and the open vial stability is 90 days at the recommended storage of 36 °F to 86 °F.
d. Detection limit:
Not Applicable
e. Analytical specificity:
Not Applicable
f. Assay cut-off:
Not Applicable
2. Comparison studies:
a. Method comparison with predicate device:
Not Applicable
b. Matrix comparison:
Not Applicable
3. Clinical studies:
a. Clinical Sensitivity:
{3}
Not Applicable
b. Clinical specificity:
Not Applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not Applicable
4. Clinical cut-off:
Not Applicable
5. Expected values/Reference range:
Not Applicable
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
4