The NMR LipoProfile® test, when used with the NMR Profiler, an automated NMR spectrometer, measures lipoprotein particles to quantify LDL particle number (LDL-P), HDL cholesterol (HDL-C), and triglycerides in serum and plasma using nuclear magnetic resonance (NMR) spectroscopy. LDL-P and these NMR-derived concentrations of triglycerides and HDL-C are used in conjunction with other lipid measurements and clinical evaluation to aid in the management of lipoprotein disorders associated with cardiovascular disease. The test is performed and provided as a service by Liposcience Laboratory.
Device Story
The NMR LipoProfile test is a laboratory-developed service provided by Liposcience Laboratories. It utilizes a 400 MHz proton NMR spectrometer to analyze serum or plasma samples. The system measures the composite NMR signal at ~0.8 ppm, which corresponds to methyl group protons of lipids in VLDL, LDL, and HDL subclasses. An analysis server running proprietary software performs deconvolution of this signal to determine the amplitudes of specific lipoprotein subclasses. These amplitudes are converted into concentrations of LDL particle number (LDL-P), HDL cholesterol, and triglycerides. The device is operated by laboratory technicians at the Liposcience facility. Healthcare providers use the resulting lipid profile, alongside other clinical evaluations, to manage cardiovascular disease risk. The test provides a quantitative assessment of lipoprotein particles, offering an alternative to traditional colorimetric enzyme assays.
Clinical Evidence
Clinical performance evaluated via three prospective/nested case-control studies (VA-HIT, EPIC-Norfolk, Women's Health Study). VA-HIT (n=1061) showed baseline LDL-P associated with CHD events (OR 1.31, p=0.004). EPIC-Norfolk (n=2888) showed LDL-P associated with CAD events (multivariable OR 1.37, p=0.02). Women's Health Study (n=27,673) showed LDL-P associated with CVD events (HR 2.51, p<0.001). Method comparison (n=5,362) against predicate chemistry analyzers showed strong correlation (R²=0.897 for HDL-C; R²=0.929 for Triglycerides).
Technological Characteristics
The system uses a 400 MHz proton NMR spectrometer. Reagents include NMR Diluent (Na2EDTA, CaCl2, KCl, Na2HPO4, NaN3), NMR Wash (Triton X-100, Liqui Nox), and NMR Calibrator (Trimethyl Acetate). The system is a centralized laboratory service. Software performs signal deconvolution of the 0.8 ppm methyl group proton signal. Connectivity is via a sample handler interfaced with the NMR spectrometer. Calibration is performed daily using TMA.
Indications for Use
Indicated for use in patients to aid in the management of lipoprotein disorders associated with cardiovascular disease by quantifying LDL particle number, HDL cholesterol, and triglycerides in serum and plasma.
Regulatory Classification
Identification
A lipoprotein test system is a device intended to measure lipoprotein in serum and plasma. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.
{1}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
k063841
B. Purpose for Submission:
New Test System (laboratory service)
C. Measurand:
LDL-particle number, HDL cholesterol, triglycerides
D. Type of Test:
Nuclear magnetic resonance (NMR)
E. Applicant:
Liposcience Inc.
F. Proprietary and Established Names:
NMR Lipoprofile Assay
G. Regulatory Information:
1. Regulation section: Lipoprotein test system 21 CFR 862.1475
Cholesterol test system 21 CFR 862.1175
Triglyceride test system 21 CFR 862.1705
Quality control material 21 CFR 862.1660
Calibrator 21 CFR 862.1150
2. Classification: 862.1475, 862.1175, 862.1705, 862.1660: Class I, subject to limitation of exemptions (21 CFR 862.9(c)(4)); 862.1150, Class II
3. Product code: JJY (control material), CDT (triglyceride), LBS (hdl- cholesterol)
JIT (calibrator), MRR (low density lipoprotein)
4. Panel: Chemistry, 75
{2}
H. Intended Use:
1. Intended use(s):
The NMR LipoProfile® test, when used with the NMR Profiler, an automated NMR spectrometer, measures lipoprotein particles to quantify LDL particle number (LDL-P), HDL cholesterol (HDL-C), and triglycerides in serum and plasma using nuclear magnetic resonance (NMR) spectroscopy. LDL-P and these NMR-derived concentrations of triglycerides and HDL-C are used in conjunction with other lipid measurements and clinical evaluation to aid in the management of lipoprotein disorders associated with cardiovascular disease. The test is performed and provided as a service by Liposcience Laboratory.
2. Indication(s) for use:
See intended use, above.
3. Special conditions for use statement(s):
For prescription use only.
4. Special instrument requirements:
This device is cleared as a service (in-house). Samples are sent to Liposcience Laboratories to be run on the Lipoprofile II NMR, using the system's software.
I. Device Description:
The test system includes a 400 MHz proton NMR spectrometer interfaced with a commercial sample handler, an analysis server containing the software to analyze digitized spectral data, and the following reagents:
NMR Diluent - aqueous solution containing Na₂EDTA (5.0 mM), CaCl₂ (1.0 mM), KCL (120 mM), Na₂HPO₄ (50 mM), NaN₃ (0.02%), pH 7.4; NMR Wash - Triton X-100-0.1% v/v, Liqui Nox 0.1% v/v in deionized water, pH 10.0; NMR Calibrator - aqueous solution of Trimethyl Acetate (TMA) disodium salt (15.0 mM) containing Na₂EDTA (5.0 mM), CaCl₂ (3.0 mM), KCl (120 nM), D₂O 10% v/v; NMR LipoProfile Quality Control material - two levels of pooled human serum-based control material, (Control A and Control B), with pre-determined target ranges.
The control materials contain human source material. Each donor unit is tested by FDA - approved methods and found non-reactive for hepatitis B surface antigen (HBsAg), antibody to hepatitis C, and antibody to HIV-1/2, all products using human source material should be handled as potentially infectious, because no test method can offer complete assurance that infectious agents are absent. Products should be handled according to established good laboratory practices.
J. Substantial Equivalence Information:
1. Predicate device name(s):
{3}
LDL-Cholesterol: k020724, Dimension Automated LDL Cholesterol Flex Reagent
Triglycerides: k971324, Carolina Liquid Chemistries
HDL-Cholesterol: k971126, Genzyme, Liquid N-geneous HDL Cholesterol Kit
Quality controls materials: k993606 Carolina Liquid Chemistries
2. Predicate 510(k) number(s):
k020724, k971324, k971162, k993606
3. Comparison with predicate:
The device is similar in indications to the predicate assays, i.e., to quantify lipoprotein, HDL cholesterol (HDL-C), and triglycerides in serum and plasma to aid in the management of lipoprotein disorders associated with cardiovascular disease. However, the test principles, technology, and one of the measurands differ between the devices. This device measures NMR signals and uses an algorithm developed by the manufacturer, to calculate concentrations of LDL particle number, high density lipoprotein cholesterol, and triglycerides. The predicate devices utilize colorimetric enzyme assay technology on chemistry analyzers. In addition, the predicate device reports LDL-cholesterol, this device reports LDL-particle number. The predicate device is for use by clinical laboratories on clinical chemistry analyzers. Specimens to be tested with Lipoprofile are sent to Liposcience laboratories for measurement and analysis using the manufacturer's system.
K. Standard/Guidance Document Referenced (if applicable):
1. CLSI. Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline. CLSI document EP6-A, 2003.
2. CLSI. Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline – Second Edition. CLSI document EP9-A2, 2002.
L. Test Principle:
The test system involves measurement of the 400 MHz proton NMR spectrum of a plasma or serum sample, deconvolution of the composite signal at ~0.8 ppm to produce the signal amplitudes of the lipoprotein subclasses that contribute to the composite plasma signal, and conversion of these subclass signal amplitudes to lipoprotein subclass concentrations. The 0.8 ppm plasma NMR signal arises from the methyl group protons of the lipids carried in the VLDL, LDL, and HDL subclasses of varying diameter. The NMR signals from the various lipoprotein subclasses have unique and distinctive frequencies and lineshapes, each of which are accounted for in the deconvolution analysis model. LDL subclass particle concentrations, in units of nanomoles of particles per liter (nmol/L), are summed to give the reported total LDL particle concentration (LDL-P). By employing conversion factors that assume that the various lipoprotein subclass particles have cholesterol and triglyceride contents characteristic of normolipidemic individuals, HDL cholesterol and triglyceride concentrations are also derived.
{4}
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
All analytical studies described below were performed on patient samples or pools.
a. Precision/Reproducibility:
Within-run precision:
Two pools of samples were analyzed in 20 replicates on the same run. The observed within-run precision is shown below.
| | Pool 1 - | | | Pool 2 - | | |
| --- | --- | --- | --- | --- | --- | --- |
| | TG (mg/dL) | HDL-C (mg/dL) | LDL-P (nMol/L) | TG (mg/dL) | HDL-C (mg/dL) | LDL-P (nMol/L) |
| Mean | 188.8 | 41.4 | 2221.8 | 74.6 | 57.1 | 1041.8 |
| SD | 2.0 | 0.5 | 49.1 | 1.2 | 0.4 | 47.7 |
| CV% | 1.1 | 1.3 | 2.2 | 1.5 | 0.7 | 4.6 |
Inter-assay precision:
The same pools used to test within-run precision were run on 20 different runs over 4 days. Results are shown below:
| | Pool 1 - | | | Pool 2 - | | |
| --- | --- | --- | --- | --- | --- | --- |
| | TG (mg/dL) | HDL-C (mg/dL) | LDL-P (nMol/L) | TG (mg/dL) | HDL-C (mg/dL) | LDL-P (nMol/L) |
| Mean | 219.1 | 42.4 | 1924.8 | 79.5 | 55.5 | 1052.9 |
| SD | 2.92 | 1.17 | 66.7 | 1.7 | 0.9 | 68.4 |
| CV% | 1.33 | 2.75 | 3.5 | 2.1 | 1.5 | 6.5 |
Precision near the upper limits of the assay:
In addition, 3 serum pools were prepared, each having a concentration at least as high as the upper limit of analytical measurement range for each analyte. Each pool was analyzed in 10 separate runs (conducted on 10 separate NMR analyzers) on a single day. Observed percents CV ranged from 2-4%.
b. Linearity/assay reportable range:
Two pools (one high and one low) were prepared from patient samples. The target values of the high and low pools were derived from the averaged values of 6 different runs. The
{5}
high and low pools were then mixed and diluted in different proportions to produce multiple HDL-C, triglyceride, and LDL-P levels. Preparations were kept at $2 - 8^{\circ}\mathrm{C}$ , and analyzed on 6 different runs. Results are tabulated below.
HDL-C:
| Target Value (mg/dL) | 7.0 | 14.1 | 28.1 | 29.7 | 32.0 | 36.0 | 47.7 | 67.4 | 87.0 | 106.7 | 160.0 |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Observed Average | 7.0 | 13.9 | 28.4 | 29.5 | 32.1 | 36.0 | 47.7 | 66.4 | 86.1 | 106.2 | 161.1 |
| %bias | -0.4 | -1.1 | 1.0 | -0.7 | 0.1 | 0.2 | -0.1 | -1.4 | -1.1 | -0.5 | 0.7 |
Triglycerides:
| Target value (mg/dL) | 5.4 | 10.9 | 21.8 | 43.6 | 79.5 | 133.4 | 223.2 | 492.7 | 941.8 | 1390.9 |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Observed average | 5 | 11.3 | 22.1 | 48.6 | 88 | 143.1 | 232 | 513.7 | 964.9 | 1414 |
| % bias | -8.8 | 3.6 | 1.4 | 11.4 | 10.6 | 7.3 | 3.9 | 4.3 | 2.4 | 1.7 |
LDL-P:
| Target value (nMol/)L | 297.3 | 594.6 | 663.0 | 765.6 | 936.6 | 1450 | 2305 | 3160 | 4015 |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Observed average | 300.6 | 594.4 | 624.0 | 799.6 | 959.0 | 1622 | 2498 | 3306 | 4074 |
| % Bias | 1.11 | -0.03 | -5.87 | 4.44 | 2.39 | 11.91 | 8.37 | 4.63 | 1.48 |
Reportable ranges, based on the lower limit of detection and validation by method comparison are:
LDL-P: 300-3500 nmol/L
HDL-C: 7-140 mg/dL
Triglycerides: 5-1100 mg/dL
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
TMA (Trimethylacetic acid, Sodium salt) is used as the NMR calibrator for the NMR clinical analyzers. TMA is used routinely as a calibrator once daily during instrument startup to establish daily normalization factors. It also serves as a quality assessment tool to ensure quality NMR spectra are produced by the NMR analyzer. New calibrator material is run in parallel with the existing calibrator in five separate runs.
The stability of the TMA calibrator material and storage conditions was studied over an 18-month period. The TMA calibrator was stored at room temperature and at $4\mathrm{degC}$ , in both plastic vials and glass bottles. Aliquots were taken and TMA signal methyl integrals were measured once every two months during the 18 month study period. There were no significant changes in the integral of the stored TMA material during the study period for the different storage conditions. The coefficient of variation over the study period (where
{6}
measurements were made in 6 different NMR analyzers) for the TMA integrals were less than 1%.
Control material consists of human serum at two levels (A and B). To assign values, new controls are run on every available NMR clinical analyzer in-house during a two week period in parallel with the then existing controls. Means, Standard Deviations and $\% \mathrm{CV}$ s are computed and new targets are established.
Stability for control material over 17 months was determined. The percents deviation, of observed mean values relative to the target mean values, for control materials across multiple NMR analyzers were under $5\%$ .
# d. Detection limit:
Serial dilutions of sample pools as well as a blank solution (NMR diluent) were run on 10 different instruments, two times each per instrument to collect 20 data points. Data were compiled and analyzed to determine bias and $\% \mathrm{CV}$ for each dilution.
The lower limits of the reportable range, defined by the limits of detection, are: LDL-P: 300-nmol/L; HDL-C: 7 mg/dL, Triglycerides: 5 mg/dL.
The limits of quantitation, determined to achieve total error $< 20\%$ , are: LDL-P: 300-nmol/L; HDL-C: 10 mg/dL, Triglycerides: 25 mg/dL.
# e. Analytical specificity:
Twenty drugs representing classes of five major categories of medications that are likely to be taken by the target population were selected and evaluated for their possible interference with the NMR assay. Pure drugs were purchased and prepared in solutions which were spiked into 6 different pools of plasma samples to attain levels similar to the highest possible blood concentrations as shown in the following table. Mean percent differences for each substance were $<5\%$ .
| Drugs | Concentration (uMol/L) |
| --- | --- |
| Simvastatin | 30 |
| Fenofibrate | 70 |
| Nicotinic acid Sod salt | 2200 |
| Acetylsalicylic acid | 880 |
| Acetaminophen | 1050 |
| Naproxen Sod | 470- |
| Ibuprofen | 550 |
| Piroxicam | 10 |
| Hydrochlorothiazide | 27 |
| Triamterene | 31 |
| Furosemide | 38 |
| Metoprolol | 23 |
| Nifedipine | 37 |
{7}
7
Enalapril maleate 13
Hydralazine 60
Isosorbide dinitrate 80
Clopidogrel 28
Glipizide 7
2,4-
Thiazolidinedione 61
| Potential Interferent | Concentration |
| --- | --- |
| Bilirubin | 20 mg/dL |
| Creatinine | 20 mg/dL |
| Hemoglobin | 500 mg/dL |
| Urea | 300 mg/dL |
| Uric acid | 40 mg/dL |
f. Assay cut-off:
Not applicable; this is a quantitative assay.
2. Comparison studies:
a. Method comparison with predicate device:
HDL-C and triglycerides were evaluated by split sample comparison to the predicate devices. Samples were from apparently healthy men and women, ages ranging from 44-84 (mean 61). Points on the plot, as well as calculations of correlation coefficients and standard errors are based on singlicate measurements. The HDL-C analysis includes 5362 plasma samples with values ranging from 15 to 140 mg/dL. The triglyceride analysis includes the same 5362 samples with triglyceride values ranging from 21 to 400 mg/dL.
{8}
HDL-C:

Regression Statistics
| | Coefficient | Standard Error | Lower 95% | Upper 95% | R² |
| --- | --- | --- | --- | --- | --- |
| | | | | | 0.897 |
| Intercept | 0.2577 | 0.2430 | -0.2283 | 0.7437 | |
| X Variable | 0.9828 | 0.0045 | 0.9737 | 0.9918 | |
Triglycerides:

{9}
Regression Statistics
| | Coefficient | Standard Error | Lower 95% | Upper 95% | R² |
| --- | --- | --- | --- | --- | --- |
| | | | | | 0.929 |
| Intercept | 10.7278 | 0.5026 | 9.7226 | 11.7329 | |
| X Variable | 0.9524 | 0.0036 | 0.9452 | 0.9596 | |
An additional method comparison study for triglycerides was performed to evaluate samples across the entire assay reportable range. The study evaluated 199 patient specimens. Results are shown below.


Regression Statistics
| | Coefficient | Standard Error | Lower 95% | Upper 95% | R² |
| --- | --- | --- | --- | --- | --- |
| X Variable | 1.01 | 0.02 | 0.97 | 1.05 | 0.924 |
| Intercept | -12.56 | 9.93 | -32.42 | 7.29 | |
b. Matrix comparison:
The various specimen tubes recommended by the manufacturer for use with this assay are serum drawn in gel barrier NMR LipoTubes (Greiner, Inc. Part #456293), or red-top collection tubes; or plasma drawn into EDTA or heparin collection tubes. Three studies were conducted to validate the equivalence of NMR LipoProfile results using these serum and plasma specimens. In all 3 studies, blood was collected in the specified collection tubes, centrifuged within 1 hour, and immediately tested. The study comparing serum (red-top) to EDTA plasma tested individual specimens from 38 volunteers, the study comparing serum (red-top) to serum (NMR LipoTube) tested individual specimens from 20 volunteers, and the study comparing serum (NMR LipoTube) to heparin plasma tested individual specimens from 35 volunteers. Samples tested spanned most of the assay range. Average biases observed were less than 5% for triglycerides and HDL-C, and
{10}
ranged between 4-8% for LDL-P. No concentration-dependent trends were observed.
## 3. Clinical studies:
a. Clinical Sensitivity:
See section c, below for clinical study descriptions.
b. Clinical specificity:
See section c, below for clinical study descriptions.
c. Other clinical supportive data (when a. and b. are not applicable):
Since there is not a predicate device that measures LDL-P, descriptions and results of 3 clinical studies were provided. (Literature references for these studies are listed in the product package insert).
## Study 1
Samples from the nested case-controlled EPIC study (the European prospective Investigation into Cancer and Nutrition) were evaluated. The study population included men and women, inhabitants of Norfolk, UK, aged 45-79 years. Non-fasting serum samples were collected at the beginning of the study (baseline). Participants were followed for 6 years for the development of CAD, defined as CAD death, MI and angina. The samples tested by the Lipoprofile assay were from cases (n=1003) who developed CAD during the 6 year follow-up period and controls (n=1885) matched for age, gender, and enrollment time who did not develop CAD. None of the cases or controls were taking any lipid altering medication.
Conditional logistic regression was used to calculate odds ratios for future (incident) CAD, adjusted for smoking and blood pressure. Odds ratios for development of CAD by quartiles are shown in the table below.
10
{11}
| | Quartiles | | | | |
| --- | --- | --- | --- | --- | --- |
| | 1 | 2 | 3 | 4 | p$ |
| LDL-P | | | | | |
| Range (nmol/L) | <1278 | 1278-1525 | 1526-1812 | >1812 | |
| Univariable Odds Ratio (95% CI)* | 1.00 | 1.23 (0.97-1.56) | 1.48 (1.17-1.87) | 2.00 (1.58-2.59) | <0.0001 |
| Multivariable Odds Ratio (95% CI)** | 1.00 | 1.13 (0.89-1.44) | 1.21 (0.94-1.54) | 1.37 (1.04-1.83) | 0.02 |
*Univariable odds ratios and 95% confidence intervals were calculated by conditional logistic regression, taking into account matching for gender, age, and enrollment time and adjusted additionally for smoking and systolic blood pressure. **Multivariable odds ratios were from a model adjusted additionally for HDL-C and triglycerides. $^{\S}$ p value for linear trend.
## Study 2:
Samples from a randomized, double-blind, placebo-controlled trial were evaluated by the Lipoprofile assay. Participants in the study were apparently-healthy female health-care professionals, aged 45 years and older, who were free of self-reported CVD. Evaluations were conducted on 27,673 baseline non-fasting EDTA plasma samples from participants. During the study participants were followed for incident CVD, which included non-fatal MI, percutaneous coronary intervention, coronary artery bypass grafting, non-fatal ischemic stroke, and cardiovascular death. Over the (mean) follow-up period of 11 years, 1015 CVD events occurred. Odds ratios associated with quartiles for LDL-P observed in this study are shown below.
| | Quintiles | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | 1 | 2 | 3 | 4 | 5 | p$ |
| LDL-P | | | | | | |
| Range (nmol/L) | <963 | 963-1165 | 1166-1387 | 1388-1703 | >1703 | |
| HR (95% CI)* | 1.00 | 1.37
(1.01-1.85) | 1.35
(1.01-1.81) | 1.80
(1.36-2.38) | 2.51
(1.91-3.30) | <0.001 |
*Hazard ratios and 95% confidence intervals were calculated by Cox proportional hazard regression models adjusted for age, randomized treatment assignment, smoking status, menopausal status, postmenopausal hormone use, blood pressure, diabetes, and body mass index. $^{\S}$ p value for linear trend.
## Study 3:
Samples from men enrolled in the prospective case-controlled Veterans Affairs HDL Intervention Trial (VA-HIT) were evaluated using the Lipoprofile assay. Fasting
{12}
plasma samples were collected at baseline and after 7 months from men younger than 74 years (mean age 64.2) with an established diagnosis of CHD and low levels of LDL-C and HDL-C. Case subjects were men who experienced a CHD event (cardiac death or non-fatal MI) during the mean 5.1 year follow-up period. Control subjects were men matched for age who remained free of CHD events during follow-up. There were 354 CVD events and 697 age-matched controls. Samples from these individuals were evaluated as described below.
Logistic regression models were used to evaluate the associations of CHD events with concentrations of LDL-P measured at baseline and during the trial (7 month time point), adjusted for non-lipid covariates. The table below shows odds ratios for a new CHD event associated with a 1-standard deviation (SD) increment of LDL-P in the subjects treated with placebo (n=546).
| | Baseline | On-Trial |
| --- | --- | --- |
| LDL-P | | |
| Odds Ratio (95% CI)* | 1.31 (1.09-1.57) | 1.19 (0.99-1.43) |
| p value | 0.004 | 0.054 |
*Odds ratios and 95% confidence intervals were calculated for a 1-SD increment of LDL-P using logistic regression models adjusted for age, hypertension, smoking, body mass index, and diabetes.
## 4. Clinical cut-off:
See expected values, below.
## 5. Expected values/Reference range:
In order to determine the distribution of LDL-P levels expected in a representative sampling of the general population, plasma samples (n=5,362) were analyzed from apparently healthy men and women (mean age 61, ranging from 44 to 84 years) enrolled in the Multi-Ethnic Study of Atherosclerosis (MESA), a large epidemiologic study sponsored by the National Heart, Lung, and Blood Institute. This reference population consisted of men (n=2,529) and women (2,833) with the following ethnic make-up: 1467 African-American (27.4%); 2039 Caucasian (38.0%); 658 Chinese (12.3%); and 1198 Hispanic (22.3%). The following table provides the concentrations of LDL-P by percentile in this reference population:
Distribution of LDL-P Observed in a Reference Population - Multi-Ethnic Study of Atherosclerosis (MESA)
| Percentile | All (n=5362) | Men (n=2529) | Women (n=2833)LDL-P |
| --- | --- | --- | --- |
| 1 | 1.31 | 1.09 | 0.99 |
| 2 | 1.00 | 0.99 | 0.99 |
| 3 | 0.99 | 0.99 | 0.99 |
| 4 | 0.99 | 0.99 | 0.99 |
{13}
| | LDL-P (nmol/L) | LDL-P (nmol/L) | (nmol/L) |
| --- | --- | --- | --- |
| 5 | 770 | 800 | 760 |
| 10 | 870 | 900 | 850 |
| 20 | 1000 | 1040 | 970 |
| 30 | 1100 | 1150 | 1060 |
| 40 | 1190 | 1250 | 1150 |
| 50 | 1280 | 1330 | 1230 |
| 60 | 1380 | 1430 | 1330 |
| 70 | 1480 | 1530 | 1440 |
| 80 | 1610 | 1640 | 1570 |
| 90 | 1790 | 1820 | 1760 |
| 95 | 1980 | 1990 | 1970 |
## HDL Cholesterol and Triglycerides
The following reference values for patient classification have been recommended by the NCEP for HDL cholesterol and triglycerides for the assessment and management of CVD risk. Each laboratory should verify the validity of these reference values for the population it serves.
| HDL Cholesterol, mg/dL Classification | |
| --- | --- |
| Low | High |
| <40 | ≥60 |
| Triglycerides, mg/dL Classification | | | |
| --- | --- | --- | --- |
| Normal | Borderline-high | High | Very high |
| <150 | 150-199 | 200-499 | ≥500 |
## N. Instrument Name: NMR Profiler
## O. System Descriptions:
1. Modes of Operation:
The NMR Profiler is a 400 MHz proton nuclear magnetic resonance spectrometer.
2. Software:
{14}
The digitized spectrum is stored in computer memory and the Lipoprofile analysis software performs fitting (deconvolution), determines amplitudes and converts these to particle number or concentration units.
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes
3. Specimen Identification:
Bar code of source tube
4. Specimen Sampling and Handling:
The NMR Profiler is interfaced with a sample handler. Samples are manually placed in the sample handler rack, in the same sequence as bar coded source tubes. Source tubes are sequentially scanned by the Liposcience laboratory technician. This information is stored in the database and in a file required as input to the NMR Profiler. Each spectrum is stored in the file system in a manner that preserves the identity of the sample from which the spectrum was collected.
5. Calibration:
The instrument is calibrated with an aqueous solution of Trimethyl Acetate (TMA) a disodium salt (15.0 mM) containing Na₂EDTA (5.0 mM), CaCl₂ (3.0 mM), KCl (120 nM), D₂O 10% v/v.
6. Quality Control:
NMR LipoProfile Quality Control material include two levels of pooled human serum-based control material, with pre-determined target ranges.
P. Other Supportive Instrument Performance Characteristics Data Not Covered In the "Performance Characteristics" Section above:
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
14
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.