Indications for Use

STEAMPlus™ Sterilization Integrators are intended to be used to monitor parameters in steam sterilization cycles in healthcare facilities. The critical parameters for which the integrators will respond are time and temperature (under the presence of saturated steam). The integrators are intended to be used in steam cycles pperating at 121°C - 134°C.

Device Story

SPSmedical STEAMPlus™ Steam Integrator is a wicking-style chemical indicator for steam sterilizers. Device monitors sterilization parameters (time, temperature, saturated steam) within sterilizer chambers. Operation: when defined sterilization parameters are met, an internal pellet melts, causing dye to migrate from a 'non-safe' zone into a 'safe' zone. Used in wrapped packs, trays, pouches, containers, and unwrapped trays. Provides visual confirmation of conditions required for spore death. Intended for use by healthcare personnel in clinical settings to verify sterilization process effectiveness.

Clinical Evidence

Bench testing only. Testing performed in accordance with AAMI/ANSI ST60:1996 (Sterilization of health care products - Chemical indicators - Part 1: General requirements). Results confirmed the device functions as a Class 5 integrator, reliably integrating time and temperature to indicate conditions for spore death.

Technological Characteristics

Wicking-style chemical integrator. Principle: thermal-sensitive pellet melting and dye migration. Dimensions/form factor: strip/indicator. Standards: AAMI/ANSI ST60:1996. Shelf life: 5 years. Storage: 15-30°C.

Indications for Use

Indicated for monitoring steam sterilization cycles (121°C - 134°C) in healthcare facilities to verify time, temperature, and saturated steam presence.

Regulatory Classification

Identification

Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K063799 # 510(k) Summary of Safety and Effectiveness ### Submitter: JAN - 5 2007 - SPSmedical Supply Corp. . 6789 West Henrietta Road Rush, NY 14543 U.S.A. Phone: (585)-359-0130 Fax: (585)-359-0167 Establishment FDA Registration No.: . 1319130 - Date Summary was prepared . October 12th, 2006 - Gary J. Socola Printed name of person submitting for 510(k) - . ov ocon Signature of person submitting for 510(k) - Vice President, Scientific Affairs . Title of person submitting for 510(k) ## Device Name and Classification | Trade Name: | SPSmedical STEAMPlus™ Steam Integrator | |------------------------|---------------------------------------------------------------------| | Classification Name: | Physical/Chemical Sterilization Process Indicator | | Common Name: | Steam Integrator | | Device Classification: | General Hospital - Class II, Regulation Number<br>880.2800 | | Product Code: | 80JOJ | | Predicate Device: | 3M Thermalog steam sterilization integrator<br>(510(k) No. K813202) | {1}------------------------------------------------ #### Device Description The SPSmedical STEAMPlus™ Steam Integrator is designed to monitor sterilization parameters in steam sterilizers. It is a wicking style integrator which provides a visual indication that proper sterilization parameters were present within the sterilizers' chamber. Once the defined sterilization parameters have been reached the integrators pellet will melt and its dye will migrate from the non safe zone into the integrators safe zone. This will indicate to the user that the defined parameters for the integrator have been met. #### Intended Use The SPSmedical STEAMPlus™ Steam Integrator monitors sterilization parameters in both prevacuum and gravity type steam sterilizers. They are a reliable tool used for the monitoring of steam sterilization processes and provide a visual indication that proper sterilization parameters were present within the sterilizers' chamber. Integrators can be used in wrapped packs, trays, pouches, containers and in unwrapped trays. #### Performance Testing Testing was performed in accordance with AAMVANSI ST60:1996 - Sterilization of health care products - Chemical indicators - Part 1: General requirements. #### Recommended Storage Conditions Store in a cool, dry place (15-30℃). #### Shelf life The shelf life shall be 5 years from the date of manufacture, when stored in a cool, dry place (15-30°C). #### Biocompatibility Risk analysis concludes that the manufacturing and subsequent use of this product has a low associated risk and that the product is safe for its intended use and handling. #### Conclusion Data has demonstrated that the SPSmedical STEAMPlus™ Steam Integrator is equivalent to the predicate device. Results of testing show the SPSmedical STEAMPlus Integrator is a highly reliable and reproducible chemical integrator for steam sterilization processes and meets the performance claims for a class 5 integrator according to AAMI/ANSI ST60. It is capable of integrating time and temperature into an easy to interpret "Safe Zone" where the pellet only melts and advances into the safe zone when conditions needed for spore death have been reached. The STEAMPlus™ Steam Integrator raises no issues related to safety or effectiveness and therefore should be allowed for market in the United States. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird-like figure with flowing ribbons or streamers beneath it. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JAN 0 5 2007 SPSmedical Supply Corporation C/O Mr. Daniel W. Lehtonen Responsible Third Party Official Intertek Testing Services 2307 East Aurora Road, Unit B7 Twinsburg, Ohio 44087 Re: K063799 Trade/Device Name: STEAMPlus™ Steam Sterilization Integrator Regulation Number: 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: II Product Code: JOJ Dated: December 21, 2006 Received: December 22, 2006 Dear Mr. Lehtonen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 – Mr. Lehtonen Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours. f Supa Qunn fa Chiu Lin, Ph.D. Chiu Lin. Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # INDICATIONS for USE STATEMENT Applicant: SPSmedical Supply Corp. 510(k) Number (if known): Device Name: STEAMPlus™ Steam Sterilization Integrator Indications For Use: STEAMPlus™ Sterilization Integrators are intended to be used to monitor parameters in steam sterilization cycles in healthcare facilities. The critical parameters for which the integrators will respond are time and temperature (under the presence of saturated steam). The integrators are intended to be used in steam cycles pperating at 121°C - 134°C. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter' Use X (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Sheila F. Murphy, MD Anesthesiology, General Hosp. Control, Dental Devices K063794 Page 1 of 1
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...