← Product Code [MNI](/productcode/MNI) · K063794

# MODIFICATION TO EASYSPINE SYSTEM (K063794)

_Ldr Spine USA · MNI · Jan 24, 2007 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K063794

## Device Facts

- **Applicant:** Ldr Spine USA
- **Product Code:** [MNI](/productcode/MNI.md)
- **Decision Date:** Jan 24, 2007
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Easyspine System is a posterior, noncervical, pedicle system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: spondylolisthesis (grades 3 and 4 at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment), trauma (i.e., fracture or dislocation), spinal stenosis, deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), tumor, pseudoarthrosis, and failed previous fusion.

## Device Story

Easyspine System is a posterior, noncervical, pedicle screw spinal system used for spinal immobilization and stabilization as an adjunct to fusion. The system consists of rods, pedicle screws, set screws, connectors, and washers. Modifications include the addition of 5 mm and 8 mm diameter screws and washers. The device is implanted by a surgeon in a clinical setting to treat acute and chronic spinal instabilities or deformities. The titanium alloy components provide structural support to the spine. The material is biocompatible and allows for artifact-free imaging (X-ray, CT, MRI). The device functions as a mechanical fixation system to stabilize spinal segments, facilitating fusion and potentially reducing patient pain and disability associated with spinal instability.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Pedicle screw spinal system. Materials: Titanium-6Aluminum-4Vanadium ELI alloy (ASTM F 136). Components: rods, pedicle screws, set screws, connectors, washers. Mechanical fixation device. Non-active, non-software, non-electronic.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- Easyspine System ([K043004](/device/K043004.md))
- Easyspine System ([K064017](/device/K064017.md))
- Easyspine System ([K043094](/device/K043094.md))
- Easyspine System ([K061017](/device/K061017.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

510(k) Summary

Modifications to Easyspine® System

K063794
Page 1 of 2

# Special 510(k): Device Modification 510(k) Summary Easyspine® System

| Sponsor Name:              | LDR Spine USA<br>4030 West Braker Lane, Suite 360<br>Austin, Texas 78759<br>Telephone (512) 344-3333<br>Fax (512) 344-3350                                | JAN 2 4 2007 |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Official Contact:          | James Burrows                                                                                                                                             |              |
| Representative/Consultant: | Floyd G. Larson<br>PaxMed International, LLC<br>11234 El Camino Real, Suite. 200<br>San Diego, CA 92130<br>Telephone (858) 792-1235<br>Fax (858) 792-1236 |              |
| DEVICE NAME                |                                                                                                                                                           |              |
| Classification Name:       | Orthosis, Spinal Pedicle Fixation (MNI)<br>Orthosis, Spondylolisthesis Spinal Fixation (MNH)                                                              |              |
| Trade/Proprietary Name:    | Easyspine® System                                                                                                                                         |              |
| Common Name:               | Pedicle Screw Spinal System                                                                                                                               |              |

### ESTABLISHMENT REGISTRATION NUMBER

LDR Spine USA has submitted an Establishment Registration to FDA. The Establishment Registration number is 3004903783. The Owner/Operator number is 9068076.

#### DEVICE CLASSIFICATION

The Easyspine system is a Class II pedicle screw spinal system (21 CFR 888.3070). The product code for Orthosis, Spondylolistines is Spinal Eixation is MNH. The product code for Orthosis, Spinal Pedicle Fixation is MN. These device opinal intellent is MNT. The product code in Crimsels, Species, Species, Branch.

## PREDICATE DEVICE INFORMATION

The predicate device for this modification is the Easyspine System cleared by FDA under
K043004 and K064017 K043094 and K061017.

{1}------------------------------------------------

#### LABELING

Package labels are similar in format and content to previously cleared labels for the unmodified Easyspine System. The system modifications described in this submission do not necessitate changes to the Package Insert.

#### PACKAGING AND STERILIZATION

Packaging and sterilization will not change with the modifications described in this submission.

#### INTENDED USE

The Easyspine System is a posterior, noncervical, pedicle system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: spondylolisthesis (grades 3 and 4 at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment), trauma (i.e., fracture or dislocation), spinal stenosis, deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), tumor, pseudoarthrosis, and failed previous fusion.

#### DEVICE DESCRIPTION

Modifications to the Easyspine System that are the submission are confined solely to the addition of 5 mm diameter screws, 8 mm diameter screws, and washers (also referred to as spacers.) No components of the predicate (unmodified) system have been modified or deleted.

#### Material Composition and Biocompatibility

The rods, pedicle screws and components, set screws, connectors and washers of the Easyspine System are made of titanium alloy conforming to the requirements of ASTM F 136, Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications. The titanium material is biocompatible, corrosion-resistant, and not toxic in a biologic environment allowing for artifact-free x-ray imaging, computed tomography (CT) and Magnetic Resonance Imaging (MRI).

#### EQUIVALENCE TO MARKETED PRODUCT

The modified Easyspine System has the following similarities to the predicate Easyspine System:

- I has the same intended use.
- . uses the same operating principle.
- incorporates the same basic design.
- 트 incorporates the same materials, and
- . is packaged and sterilized using the same materials and processes.

In summary, the modified Easyspine System described in this submission is, in our opinion, substantially equivalent to the predicate devices.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with its wings spread, clutching a caduceus in its talons. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

LDR Spine USA % Mr. Floyd G. Larson President PaxMed International, LLC 11234 El Camino Real, Suite 200 San Diego, California 92130

JAN 2 4 2007

Re: K063794

Trade/Device Name: Easyspine® System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class II Product Code: MNI, MNH Dated: December 21, 2006 Received: December 22, 2007

Dear Mr. Larson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, EDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Mr. Floyd G. Larson

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120 Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours,

for
N. M. M.

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

### Indications for Use

Applicant: LDR Spine USA

510(k) Number (if known):

Device Name: Easyspine System

Indications for Use:

The LDR Easyspine System is a posterior, noncervical, pedicle system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: spondylolisthesis (grades 3 and 4 at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment), trauma (i.e., fracture or dislocation), spinal stenosis, deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), tumor, pseudoarthrosis, and failed previous fusion.

Prescription Use X Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Pep

(Division Sign-Off) Division of General, Restorative, and Neurological Devices

**510(k) Number** k063794

---

**Source:** [https://fda.innolitics.com/device/K063794](https://fda.innolitics.com/device/K063794)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
