REPROCESSED ENDOSCOPIC TROCARS AND CANNULAS

K063788 · Ascent Healthcare Solutions · NLM · Oct 15, 2007 · Gastroenterology, Urology

Device Facts

Record IDK063788
Device NameREPROCESSED ENDOSCOPIC TROCARS AND CANNULAS
ApplicantAscent Healthcare Solutions
Product CodeNLM · Gastroenterology, Urology
Decision DateOct 15, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

Reprocessed Endoscopic Trocars and Cannulas are intended for use during minimally invasive surgery for temporary dilation access to the abdominal and thoracic cavities for passage of diagnostic, therapeutic and operative instruments into the abdominal and thoracic cavities, and for percutaneous access to hollow body organs. Reprocessed Endoscopic Trocars and Cannulas are indicated to establish a port of entry for endoscopic instruments in patients requiring minimally invasive surgical procedures.

Device Story

Reprocessed Endoscopic Trocars and Cannulas provide port of entry for endoscopic instruments during minimally invasive surgery; facilitate temporary dilation access to abdominal/thoracic cavities and hollow body organs. Devices consist of trocars, cannulas, obturators, and seals in various configurations. Reprocessing involves removal of adherent visible soil and decontamination of previously used devices. Functionality verified via component testing prior to packaging. Used by surgeons in clinical settings; provides access for diagnostic, therapeutic, and operative instruments.

Clinical Evidence

No clinical data. Bench and laboratory testing performed, including biocompatibility, validation of reprocessing, sterilization validation, functional testing, and packaging validation.

Technological Characteristics

Mechanical endoscopic access devices. Materials identical to original predicate devices. Includes various configurations of trocars, cannulas, obturators, and seals. Non-powered, manual operation. Sterilization validation performed as part of reprocessing.

Indications for Use

Indicated for patients requiring minimally invasive surgical procedures to establish a port of entry for endoscopic instruments.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx P910+3 # PART B: 510(k) SUMMARY | Submitter: | Ascent Healthcare Solutions<br>10232 South 51st Street<br>Phoenix, Arizona 85044 | OCT 15 2007 | |----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------| | Contact: | Katie Bray<br>RA Specialist/Biomedical Engineer<br>(480) 763-5300 (o)<br>(480) 763-6089 (f)<br>kbray@ascenths.com | | | Date of preparation: | December 13, 2006 | | | Name of device: | Trade/Proprietary Name: Reprocessed Endoscopic Trocars and<br>Cannulas<br>Classification Name: Laparoscope, general & plastic surgery,<br>reprocessed | | | Predicate Device | 510(k) Title | Manufacturer | | K041795 | Modification to Optical Separator | Applied Medical | | K032889 | Optical Separator | Applied Medical | | K012884 | Dilating Tip Obturator | Applied Medical | | K012968 | Trocar Seal | Applied Medical | | Device description: | Reprocessed Endoscopic Trocars and Cannulas are designed<br>to establish a port of entry for endoscopic instruments used<br>during minimally invasive surgery. Endoscopic trocars and<br>cannulas are available in a variety of configurations and<br>materials as well as trocar and cannula sets. Trocar seals vary<br>between single-port and multi-port seals. | | ### Bladed Trocars | Manufacturer | Description | Model | |-----------------|-----------------------------------------------------------------------|-------| | Applied Medical | 11mm x 100mm Shielded Obturator with Universal Seal, non-<br>threaded | C0638 | | Applied Medical | 12mm x 100mm Shielded Obturator with Universal Seal, non-<br>threaded | C0639 | | Applied Medical | 11mm x 100mm Shielded Obturator with Universal Seal, threaded | C0658 | | Applied Medical | 12mm x 100mm Shielded Obturator with Universal Seal, threaded | C0659 | {1}------------------------------------------------ | く<br>0 | 1637 | -88 | | |--------|------|-----|--| | | | | | | Non-Bladed Trocars | | | |--------------------|------------------------------------------------------------------|-------| | Manufacturer | Description | Model | | Applied Medical | 12mm blunt tip trocar | C0718 | | Applied Medical | 11mm blunt tip trocar | C0717 | | Applied Medical | 12mm x 100mm Optical Separator system, handled, non-threaded | C0124 | | Applied Medical | 12mm x 100mm Optical Separator system, handled, threaded | C0128 | | Applied Medical | 12mm x 100mm Optical Separator system, non-handled, non-threaded | C0126 | | Applied Medical | 12mm x 100mm Optical Separator system, non-handled, threaded | C0130 | | Applied Medical | 11mm x 100mm Optical Separator system, handled, non-threaded | C0114 | | Applied Medical | 11mm x 100mm Optical Separator system, handled, threaded | C0150 | | Applied Medical | 11mm x 100mm Optical Separator system, non-handled, non-threaded | C0116 | | Applied Medical | 11mm x 100mm Optical Separator system, non-handled, threaded | C0152 | | Applied Medical | 8mm x 100mm Optical Separator system, non-threaded | C0675 | | Applied Medical | 8mm x 100mm Optical Separator system, threaded | C0676 | | Applied Medical | 15mm x 100mm Separator system with Universal seal, non-threaded | C0604 | | Applied Medical | 15mm x 100mm Separator system with Universal seal, threaded | C0605 | | Applied Medical | 12mm x 100mm Separator system with Universal seal, non-threaded | C0682 | | Applied Medical | 12mm x 100mm Separator system with Universal seal, threaded | C0680 | | Applied Medical | 11mm x 100mm Separator system with Universal seal, non-threaded | C0667 | | Applied Medical | 11mm x 100mm Separator system with Universal seal, threaded | C0665 | | Applied Medical | 12mm x 100mm Optical Separator, obturator, handled | C0120 | | Applied Medical | 12mm x 100mm Optical Separator, obturator, non-handled | C0122 | | Applied Medical | 12mm x 100mm Separator obturator | C0930 | | Applied Medical | 11mm x 100mm Separator obturator | C0915 | | Applied Medical | 5mm x 100mm Separator obturator | C0901 | | Applied Medical | 5mm x 55mm Separator obturator | C0900 | #### Stability Cones and Seals | Manufacturer | Description | Model | |-----------------|-------------------------------------------------------|-------| | Applied Medical | 10/11mm Conductive Stability Cone | C0704 | | Applied Medical | 10/12mm Conductive Stability Cone | C0705 | | Applied Medical | 12mm disposable non-conductive stability cone | C0707 | | Applied Medical | 11mm disposable non-conductive stability cone | C0706 | | Applied Medical | 12mm x 100mm cannula and Universal seal, non-threaded | C0632 | | Applied Medical | 12mm x 100mm cannula and Universal seal, threaded | C0652 | | Applied Medical | 11mm x 100mm cannula and Universal seal, non-threaded | C0631 | | Applied Medical | 11mm x 100mm cannula and Universal seal, threaded | C0651 | | Applied Medical | Universal® Seals | C0600 | Reprocessed Endoscopic Trocars and Cannulas are intended Intended use: for use during minimally invasive surgery for temporary dilation access to the abdominal and thoracic cavities for passage of diagnostic, therapeutic and operative instruments into the abdominal and thoracic cavities, and for percutaneous access to hollow body organs. Reprocessed Endoscopic Trocars and Cannulas are indicated Indications for use to establish a port of entry for endoscopic instruments in statement: patients requiring minimally invasive surgical procedures. Ascent Healthcare Solutions 10 Reprocessed Endoscopic Trocars and Cannulas Traditional 510(k) {2}------------------------------------------------ K063788 Pg 3 of 3 The design, materials, and intended use of Reprocessed Technological Endoscopic Trocars and Cannulas are identical to the predicate characteristics: devices. The mechanism of action of Reprocessed Endoscopic Trocars and Cannulas is identical to the predicate devices in that the same standard mechanical design, materials, and sizes are utilized. There are no changes to the claims, intended use, clinical applications, patient population, performance specifications, or method of operation. In addition, Ascent Healthcare Solutions's reprocessing of Endoscopic Trocars and Cannulas includes removal of adherent visible soil and decontamination. Each individual Endoscopic Trocars and Cannulas is tested for appropriate function of its components prior to packaging and labeling operations. Performance data: Bench and laboratory testing was conducted to demonstrate performance (safety and effectiveness) of Reprocessed Endoscopic Trocars and Cannulas. This included the following tests: - Biocompatibility . - Validation of reprocessing . - Sterilization Validation ● - Function test(s) . - . Packaging Validation Performance testing demonstrates that Reprocessed Endoscopic Trocars and Cannulas perform as originally intended. Ascent Healthcare Solutions concludes that the modified Conclusion: devices (Reprocessed Endoscopic Trocars and Cannulas) are safe, effective, and substantially equivalent to the predicate devices as described herein. > Ascent Healthcare Solutions Reprocessed Endoscopic Trocars and Cannulas Traditional 510(k) {3}------------------------------------------------ Image /page/3/Picture/1 description: The image is a circular seal for the Department of Health & Human Services - USA. The seal features the department's emblem, which is a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the emblem. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 OCT 1 5 2007 % Ms. Katie Bray Regulatory Affairs Engineer 10232 South 51st Street Phoenix, Arizona 85044 Ascent Healthcare Solutions Re: K063788 Trade/Device Name: Reprocessed Endoscopic Trocars and Cannulas Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: NLM Dated: October 3, 2007 Received: October 5 2007 Dear Ms. Bray: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ Page 2 - Ms. Katie Bray This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. Fa Nh Dm Mark N. Melkerson Ar Qr 10 lixdr Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ K063788 ## List of Devices: | Manufacturer | Description | Model | |-----------------|------------------------------------------------------------------|-------| | Applied Medical | 12mm blunt tip trocar | C0718 | | Applied Medical | 11mm blunt tip trocar | C0717 | | Applied Medical | 12mm x 100mm Optical Separator system, handled, non-threaded | C0124 | | Applied Medical | 12mm x 100mm Optical Separator system, handled, threaded | C0128 | | Applied Medical | 12mm x 100mm Optical Separator system, non-handled, non-threaded | C0126 | | Applied Medical | 12mm x 100mm Optical Separator system, non-handled, threaded | C0130 | | Applied Medical | 11mm x 100mm Optical Separator system, handled, non-threaded | C0114 | | Applied Medical | 11mm x 100mm Optical Separator system, handled, threaded | C0150 | | Applied Medical | 11mm x 100mm Optical Separator system, non-handled, non-threaded | C0116 | | Applied Medical | 11mm x 100mm Optical Separator system, non-handled, threaded | C0152 | | Applied Medical | 8mm x 100mm Optical Separator system, non-threaded | C0675 | | Applied Medical | 8mm x 100mm Optical Separator system, threaded | C0676 | | Applied Medical | 15mm x 100mm Separator system with Universal seal, non-threaded | C0604 | | Applied Medical | 15mm x 100mm Separator system with Universal seal, threaded | C0605 | | Applied Medical | 12mm x 100mm Separator system with Universal seal, non-threaded | C0682 | | Applied Medical | 12mm x 100mm Separator system with Universal seal, threaded | C0680 | | Applied Medical | 11mm x 100mm Separator system with Universal seal, non-threaded | C0667 | | Applied Medical | 11mm x 100mm Separator system with Universal seal, threaded | C0665 | | Applied Medical | 12mm x 100mm Optical Separator, obturator, handled | C0120 | | Applied Medical | 12mm x 100mm Optical Separator, obturator, non-handled | C0122 | | Applied Medical | 12mm x 100mm Separator obturator | C0930 | | Applied Medical | 11mm x 100mm Separator obturator | C0915 | | Applied Medical | 5mm x 100mm Separator obturator | C0901 | | Applied Medical | 5mm x 55mm Separator obturator | C0900 | | Manufacturer | Description | Model | |-----------------|-------------------------------------------------------------------|-------| | Applied Medical | 11mm x 100mm Shielded Obturator with Universal Seal, non-threaded | C0638 | | Applied Medical | 12mm x 100mm Shielded Obturator with Universal Seal, non-threaded | C0639 | | Applied Medical | 11mm x 100mm Shielded Obturator with Universal Seal, threaded | C0658 | | Applied Medical | 12mm x 100mm Shielded Obturator with Universal Seal, threaded | C0659 | | Manufacturer | Description | Model | |-----------------|-------------------------------------------------------|-------| | Applied Medical | 10/11mm Conductive Stability Cone | C0704 | | Applied Medical | 10/12mm Conductive Stability Cone | C0705 | | Applied Medical | 12mm disposable non-conductive stability cone | C0707 | | Applied Medical | 11mm disposable non-conductive stability cone | C0706 | | Applied Medical | 12mm x 100mm cannula and Universal seal, non-threaded | C0632 | | Applied Medical | 12mm x 100mm cannula and Universal seal, threaded | C0652 | | Applied Medical | 11mm x 100mm cannula and Universal seal, non-threaded | C0631 | | Applied Medical | 11mm x 100mm cannula and Universal seal, threaded | C0651 | | Applied Medical | Universal® Seals | C0600 | {6}------------------------------------------------ K063788 pg 1 of 1 #### Indications for Use Statement 1. 510(k) Number (if known): Device Name: Reprocessed Endoscopic Trocars and Cannulas Indications for Use: Reprocessed Endoscopic Trocars are indicated for use to establish a port of entry for endoscopic instruments in patients requiring minimally invasive surgical procedures. AND/OR Prescription Use X (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Division of General, Restorative, and Neurological Devices 1106374 510(k) Number
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...